EIC Requirements
Electron Ion Collider

Global EIC System Requirements
Global requirements associated with the Electron Ion Collider.
- NameWBSDescriptionUpdatedStatusTBD
GBL
GBL-BACKGRDS
- 6.0Backgrounds in the detector system must be kept to a very low level to extract the signals of interest05/16/2025In ProcessFALSE
- 6.0The central and far-forward detectors must be integrated into the accelerator and IR lattice, including the vacuum, controls, and beam protection systems, in such a way as to minimize backgrounds from scattered particles, such as beam-gas05/16/2025In ProcessFALSE
- 6.0The detectors must be well protected against background created by the beam. This implies that there must not be particle background from the electron and hadron beams hitting the interaction region vacuum walls. Unavoidable lost particles05/16/2025In ProcessFALSE
- 6.0Residual pressure levels must remain within limits defined by the background level which can be accepted by the detector.05/16/2025In ProcessFALSE
GBL-BEAMPOL
- 6.0Polarization is required for (i) double-spin asymmetries, requiring both electron and hadron beams to be polarized, and (ii) single-spin asymmetries, requiring only the electron or hadron beam to05/16/2025In ProcessFALSE
- 6.0High beam polarizations, ≥70%, are mandatory to reduce the statistical uncertainties.05/16/2025In ProcessFALSE
- 6.0Electron- and proton beams as well as light ion beams such as He-3 need to be spin polarized accordingly in collisions with a polarization degree of 70% on average (over time).05/16/2025In ProcessFALSE
- 6.0The accelerator chain and the injection systems for electron must be able to provide beams with sufficiently high polarization at a sufficiently high injection frequency to maintain high electron polarization.05/16/2025In ProcessFALSE
- 6.0The polarization direction in collisions needs to be adjustable in the longitudinal direction and the transverse direction with respect to the hadron beam direction05/16/2025In ProcessFALSE
- 6.0The polarization direction in collisions needs to be adjustable in the longitudinal direction for the electron beam.05/16/2025In ProcessFALSE
- 6.0The polarimeters are required to be able to measure the beam polarization on a bunch-by bunch level.05/16/2025In ProcessFALSE
GBL-BUNCHPARAMS
- 6.0Good Control of luminosity and polarization is essential for the EIC. The EIC measurements require that the instantaneous luminosity L be measurable to 1%.05/16/2025In ProcessFALSE
- 6.0The precision of double spin asymmetries is dependent on the relative luminosity measurement R = (L++/--)/(L+-/-+), which shall be determined with an accuracy <10-4.05/16/2025In ProcessFALSE
- 6.0The collider shall be constructed and operated such that bunch polarization and polarization orientation in the interaction point can be measured for each electron and proton bunch,05/16/2025In ProcessFALSE
- 6.0The collider shall be constructed and operated such that luminosity per bunch crossing and relative luminosity for the spin different spin direction combinations (++, --, +- and -+)05/16/2025In ProcessFALSE
GBL-CENMASSENG
- 6.0The EIC must cover a large range of center of mass energy's. The collider shall be designed for center of mass energies in the range of 29 GeV to 140 GeV (electrons and protons)05/16/2025In ProcessFALSE
- 6.0The corresponding requirement for the maximum proton beam energy is 275 GeV and for maximum electron energy is 18 GeV.05/16/2025In ProcessFALSE
- 6.0The corresponding crequirment for minimum proton beam energy is 41 GeV and for minimum electron energy is 5 GeV.05/16/2025In ProcessFALSE
- 6.0The center of mass energy range for Electron-Ion collisions shall be (in case of electron gold collisions, a representative case) 29 GeV to 89 GeV (collisions of electrons with gold ions).05/16/2025In ProcessFALSE
- 6.0The corresponding requirment for Au ion energies are 110 GeV/nucleon and 41 GeV per nucleon.05/16/2025In ProcessFALSE
- 6.0The electron storage ring must be designed such that the revolution frequency of the electron beam equals the revolution frequency of a 133 GeV proton beam in the hadron storage ring.05/16/2025In ProcessFALSE
- 6.0For low energy hadron operations, the hadron storage ring lattice must include an option of a bypass such that the path length for protons with an energy of 41 GeV - or any ion species with05/16/2025In ProcessFALSE
GBL-DETACCEPT
- 6.0the scattered electron alone must be precisely measured, including a range of angles within a few milliradians of the beam; semi-inclusive measurements, which require detection of at least one hadron in coincidence with the scattered electron; and exclusive processes, which require the05/16/2025In ProcessFALSE
- 6.0the detector must be “hermetic”, with an acceptance that includes all angles, up to those of particles scattered within a few milliradians of the colliding beam directions.05/16/2025In ProcessFALSE
- 6.0The main detector, comprising tracking, calorimetry, and particle identification for scattered particles with pseudorapidity in the range η = -4 to +4 must fit within a space -4.5 to +5 meters from the collision point.05/16/2025In ProcessFALSE
- 6.0For scattered particles whose energy and momentum are very close to those of the circulating beams, far forward and backward detectors must be integrated with the accelerator components of the IR.05/16/2025In ProcessFALSE
- 6.0The required experimental equipment includes:Very forward detectors to complete the hermetic coverage, such as Roman pots to detect scattered protons that remain inside the beam pipe, and large acceptance zero-degree calorimetry to effectively detect neutrons from the break-up of nuclei05/16/2025In ProcessFALSE
- 6.0Polarized beams require the implementation of electron, proton, and light-ion polarimetry.05/16/2025In ProcessFALSE
- 6.0The EIC detector will have to cope with collision rates up to ~500 kHz at full luminosity.05/16/2025In ProcessFALSE
GBL-EFFICIENCY_REDUNDANCY
- 6.0The facility shall be planned and designed preferably with standardized components that can be used in several hardware systems of the collider.05/16/2025In ProcessFALSE
- 6.0Multipurpose components shall be used wherever they are not compromising performance, cost or schedule.05/16/2025In ProcessFALSE
- 6.0To minimize performance risks, commissioning and collider maturing periods and to reduce initial trouble shooting efforts, existing and proven technology shall be used wherever possible.05/16/2025In ProcessFALSE
- 6.0Use of new technology must be motivated by substantial increase in performance, tolerances, service friendliness, maintainability, manufacturability, availability on the market, cost, and schedule and reasonable research and development effort.05/16/2025In ProcessFALSE
GBL-ENVIRON
- 6.0Ecologically and environmentally sensitive areas such as the Peconic River that crosses the EIC facility must not be affected by EIC construction activities.05/16/2025In ProcessFALSE
- 6.0Precautions shall be taken at locations with expected high beam loss, that activation of oil, ground water, and cooling water is kept within level defined by the document C-AD OPM 9.1.12.05/16/2025In ProcessFALSE
- 6.0The inefficient use of electrical power shall be avoided by appropriate energy conscientious design.05/16/2025In ProcessFALSE
- 6.0Precautions shall be taken at locations with expected high beam loss, that activation of soil, ground water, and cooling water is kept ALARA, and that controls are implemented to minimize environmental impacts and exposure to personnel.05/16/2025In ProcessFALSE
- 6.0In addition, DOE buildings are subject to the requirements for efficiency and sustainability in DOE O 436.1, Departmental Sustainability.05/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the environment, the facility must comply to the National Environmental Policy Act (NEPA)Implementing Procedures, 10 CFR 102105/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Compliance with Floodplain/Wetlands Environmental Review Requirements, 10 CFR 102205/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Endangered Species Act (Title 16 - Conservation: Chapter 35-Endangered Species, 16 USC 153105/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Historic Sites Act of 1935, 16 USC 461-46705/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Archaeological and Historic Preservation Act, 16 USC 46905/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the National Historic Preservation Act, 16 USC 47005/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Archaeological Resources Protection Act of 1979, 16 USC 470aa-470ll05/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Migratory Bird Treaty Act, 16 USC 703-71205/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Protection of Historic Resources, 36 CFR 80005/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Protection of Environment /Protection of Stratospheric Ozone, 40 CFR 8205/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Resource Conservation and Recovery Act (RCRA), 40 CFR 239 – 28205/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the National Environmental Policy Act of 1969, et seq., as amended, 42 USC 4321-434705/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Trees and Plants/Protected Native Plants, 6 NYCRR 193.305/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Environmental Remediation Programs, 6 NYCRR 37505/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Article 15, Title 5 - Protection of Waters, 6 NYCRR 60805/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Article 24 - Freshwater Wetlands, 6 NYCRR 662 & 66305/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Article 15, Title 27 - Wild, Scenic, Recreational River Systems Act, 6 NYCRR 66605/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the New York State Environmental Quality Review Act, 8 NYCRR Part 10105/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Notification Requirements under CERCLA and Title III of the Superfund Amendments and Reauthorization Act of 1986, 40 CFR 302.605/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Transportation/Hazardous Materials Regulations/ Hazardous Materials Table, Special Provisions, Hazardous Materials Communications, Emergency Response Information, and Training Requirements, 49 CFR 17205/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Solid Wastes, 6 NYCRR 360-364.905/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Hazardous Waste Management System: General, 6 NYCRR 37005/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Identification and Listing of Hazardous Wastes, 6 NYCRR 37105/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Standards for the Management of Special Hazardous Wastes and Specific Types of Hazardous Waste Management Facilities, 6 NYCRR 37405/16/2025In ProcessFALSE
- 6.0In order to minimize the impact of the EIC on the Standards for the Management of Used Oil, New York State Department of Environmental Conservation, 6 NYCRR 374-205/16/2025In ProcessFALSE
- 6.0New York State Department of Health, State Sanitary Code, Drinking Water Supplies , 10 NYCRR 505/16/2025In ProcessFALSE
- 6.0The Bald and Golden Eagle Protection Act., 16 USC 668 a-d05/16/2025In ProcessFALSE
- 6.0Discharge of Oil, 40 CFR 110.605/16/2025In ProcessFALSE
- 6.0Protection of Environment/Oil Pollution Prevention, 40 CFR 11205/16/2025In ProcessFALSE
- 6.0Code of Federal Regulations, National Pollutant Discharge Elimination System, 40 CFR 122-131, 13305/16/2025In ProcessFALSE
- 6.0National Primary and Secondary Drinking Water Standards, 40 CFR 141-14305/16/2025In ProcessFALSE
- 6.0Underground Injection Control, 40 CFR 144 - 14805/16/2025In ProcessFALSE
- 6.0November 1978, "Regulations for Implementing the Procedural Provisions of the National Environmental Policy Act," Council on Environmental Quality, U.S. Code of Federal Regulations, 40 CFR 1500-150805/16/2025In ProcessFALSE
- 6.0Resource Conservation and Recovery Act/Standards Applicable to Generators of Hazardous Waste, 40 CFR 262 & 264-26505/16/2025In ProcessFALSE
- 6.0Standards for the Management of Used Oil, 40 CFR 27905/16/2025In ProcessFALSE
- 6.0Protection of the Environment/National Oil and Hazardous Substances Pollution Contingency Plan, 40 CFR 30005/16/2025In ProcessFALSE
- 6.0Standards of Performance for New Stationary Sources (NSPS), 40 CFR 60 - Subpart A05/16/2025In ProcessFALSE
- 6.0Standards of Performance for Stationary Compression Ignition Internal Combustion Engine, 40 CFR 60 Subpart IIII (as amended June 28, 2011)05/16/2025In ProcessFALSE
- 6.0National Emissions Standards for Hazardous Air Pollutants (NESHAPs)- General Provisions, 40 CFR 61 - Subpart A05/16/2025In ProcessFALSE
- 6.0National Emission Standards for Hazardous Air Pollutants (NESHAPs) - National Emission Standards for Emissions of Radionuclides Other Than Radon from Department of Energy Facilities, 40 CFR 61 - Subpart H05/16/2025In ProcessFALSE
- 6.0Mandatory Greenhouse Gas Reporting, 40 CFR 9805/16/2025In ProcessFALSE
- 6.0USC 1996. July 1983, "CEQ Regulations for Implementing the Procedural Provisions of the National Environmental Policy Act," Council on Environmental Quality, Federal Register, 48 FR 3426305/16/2025In ProcessFALSE
- 6.0Endanger and Threatened Wildlife and Plants; Listing the Northern Long-Eared Bat as an Endangered Species - Proposed Rule, 50 CFR 17 Dept. of Interior, Fish and Wild Life Service [Fed. Reg. Vol.78 No.191, Oct 2, 2013]05/16/2025In ProcessFALSE
- 6.0New York State Department of Environmental Conservation/Prevention and Control of Air contamination and Air Pollution, 6 NYCRR 200 - 23405/16/2025In ProcessFALSE
- 6.0New York State Department of Environmental Conservation, "Hazardous Substance Bulk Storage Regulations,", 6 NYCRR 595-59905/16/2025In ProcessFALSE
- 6.0New York State Department of Environmental Conservation, Storage and Handling of Petroleum/Petroleum Clean-up and Removal, 6 NYCRR 611 and 61305/16/2025In ProcessFALSE
- 6.0State Pollutant Discharge Elimination System (SPDES) Permits, 6 NYCRR 75005/16/2025In ProcessFALSE
- 6.0Sustainable Acquisition Program (Oct 2010)(SC Alternate 1)(Sep 2018), BSA Contract No. DE-SC0012704 - Clause I.134 (DEAR 970.5223-7)05/16/2025In ProcessFALSE
- 6.0Pollution Prevention And Right-to-know Information (May 2011) ( Alternate I), BSA Contract No. DE-SC0012704 - Clause I.52 — FAR 52.223-505/16/2025In ProcessFALSE
- 6.0Estimate Of Percentage Of Recovered Material Content For EPA Designated Items (May 2008), BSA Contract No. DE-SC0012704 - Clause I.54 — FAR 52.223-905/16/2025In ProcessFALSE
- 6.0Ozone-Depleting Substances and High Global Warming Potential Hydrofluorocarbons (Jun 2016), BSA Contract No. DE-SC0012704 - Clause I.56 (FAR 52.223-11)05/16/2025In ProcessFALSE
- 6.0Compliance With Environmental Management Systems (May 2011), BSA Contract No. DE-SC0012704 - Clause I.62 — FAR 52.223-1905/16/2025In ProcessFALSE
- 6.0Aerosols (Jun 2016), BSA Contract No. DE-SC0012704 - Clause I.62A - FAR 52.223-2005/16/2025In ProcessFALSE
- 6.0Foams (Jun 2016), BSA Contract No. DE-SC0012704 - Clause I.62B - FAR 52.223-2105/16/2025In ProcessFALSE
- 6.0EO 13990: Climate Crisis; Efforts to Protect Public Health and Environment and Restore Science, January 20, 2021 (EO 13693 was revoked by EO 13990, EO 14008, EO 14057, EO 14082)05/16/2025In ProcessFALSE
- 6.0EO 14008: Tackling the Climate Crisis at Home and Abroad, January 27, 202105/16/2025In ProcessFALSE
- 6.0(EO 13693 was revoked by EO 13990, EO 14008, EO 14057, EO 14082)05/16/2025In ProcessFALSE
- 6.0EO 14057: Catalyzing Clean Energy Industries and Jobs Through Federal Sustainability, December 8, 2021 (EO 13693 was revoked by EO 13990, EO 14008, EO 14057, EO 14082)05/16/2025In ProcessFALSE
- 6.0EO 14082: Implementation of the Energy and Infrastructure Provisions of the Inflation Reduction Act of 2022, September 12, 2022 (EO 13693 was revoked by EO 13990, EO 14008, EO 14057, EO 14082)05/16/2025In ProcessFALSE
- 6.0CRD – Radioactive Waste Management, O 435.1 Chg 2 (AdminChg)05/16/2025In ProcessFALSE
- 6.0Departmental Sustainability, O 436.1 (May 2, 2011)05/16/2025In ProcessFALSE
- 6.0Radiation Protection of the Public and the Environment, O 458.1 Chg 4 (LtdChg)9-15-202005/16/2025In ProcessFALSE
- 6.0Toxic and Hazardous Materials Storage and Handling Controls, Suffolk County Sanitary Code - Article 1205/16/2025In ProcessFALSE
GBL-INTEGRAT
- 6.0The EIC shall be designed such as to seamlessly integrate into the existing RHIC systems05/16/2025In ProcessFALSE
- 6.0Duplication of existing functionality and infrastructure previously used for RHIC must be avoided.05/16/2025In ProcessFALSE
- 6.0RHIC components which are becoming part of the EIC shall remain unaltered wherever possible.05/16/2025In ProcessFALSE
- 6.0The Electron-Ion collider rings shall use the existing RHIC tunnel and major changes of the present RHIC accelerator tunnel and the experimental halls must be avoided05/16/2025In ProcessFALSE
- 6.0Present shielding measures of RHIC, in particular the RHIC berm must stay in place and its integrity as a radiation shielding measure must not be compromised.05/16/2025In ProcessFALSE
- 6.0Existing RHIC buildings (service buildings) must be used wherever possible.05/16/2025In ProcessFALSE
- 6.0New accelerator controls systems must be designed such as to interface to the existing or upgraded existing hadron accelerator control system without major additional05/16/2025In ProcessFALSE
- 6.0The EIC hadron ring must be able to accept beam from the AGS via the AtE (former AtR) transfer line.05/16/2025In ProcessFALSE
GBL-IONSPEC
- 6.0The EIC collider shall include the capacity to produce ion beams of a large range in A from protons to uranium.05/16/2025In ProcessFALSE
- 6.0The EIC injector complex must be configured such that polarized deuteron beams can be added with a minimum of additional hardware (such as a polarized deuteron source and deuteron polarimetry).05/16/2025In ProcessFALSE
GBL-LUMI
- 6.0The EIC shall be designed to achieve peak electron-proton luminosities between 1033cm-2s-1 and 1034 cm-2s-1. Comment: With strong hadron cooling (Lpeak = Lavg), 1033 cm-2s-1 yields an integrated luminosity of 1.5 fb-1 per month.05/16/2025In ProcessFALSE
- 6.0The peak electron-proton luminosity of the EIC shall reach values between one and ten times 1033 cm-2 s-1 in the range 29 to 140 GeV of center of mass energies, Lpeak= (1-10) ∙ 1033 cm-2 s-1 for 29 GeV< Ecm<140 GeV.05/16/2025In ProcessFALSE
- 6.0The design shall aim to maximize the range of center of mass energies where the peak electron-proton luminosity reaches values close to Lpeak=1034 cm-2 s-1.05/16/2025In ProcessFALSE
- 6.0The luminosity averaged between two subsequent injections of hadron beams Lavg shall be close to 90% of the peak luminosity.05/16/2025In ProcessFALSE
- 6.0The collider shall be designed such that these luminosity goals can be achieved within the first five years of operations.05/16/2025In ProcessFALSE
- 6.0Studies of the spatial distributions of quarks and gluons in the proton with polarized beams; Shall require an integrated luminosity of up to 100 fb-105/16/2025In ProcessFALSE
- 6.0The choice of beam species, energies, and spin orientation shall allow multiple measurements simultaneously per operating period.05/16/2025In ProcessFALSE
GBL-OPEREFF
- 6.0The EIC collider design choices must consider high levels of operational efficiency and reliability to maximize the physics outcome.05/16/2025In ProcessFALSE
- 6.0Operating procedures which minimize the time between collision runs which includes the time for beam injection, collision adjustment and tuning shall be required to achieve high levels of operational efficiency and reliability to maximize the physics outcome.05/16/2025In ProcessFALSE
- 6.0Consistently achieving good performance parameters near the anticipated design goals shall be required to achieve high levels of operational efficiency and reliability to maximize the physics outcome.05/16/2025In ProcessFALSE
- 6.0Minimizing unscheduled downtime by technically reliable accelerator hardware (thus large mean time between failure, MTBF) and short repair and replacement times (which implies short times between repairs, MTTR) shall be required to achieve high levels of operational efficiency and reliability to maximize the physics outcome.05/16/2025In ProcessFALSE
- 6.0Switching center of mass energy shall not require changing or major moving accelerator components (rewiring maybe unavoidable but shall be designed such as to minimize tie and effort)05/16/2025In ProcessFALSE
- 6.0All components need to be removeable/exchangeable without modifications to buildings and access to tunnels and service buildings.05/16/2025In ProcessFALSE
- 6.0Reliability is defined as time when beam is available as a fraction of scheduled time with beam. The difference between delivered and scheduled time is failure time. The EIC shall meet or exceeded a target of 80% reliability .05/16/2025In ProcessFALSE
GBL-SAFETY_BUILDINGS&FIRE
- 6.0The new Buildings needed for the EIC will be constructed to meet the building and fire protection code that is outlined in the New York State Uniform Fire Prevention and Building Code (2020 Edition) (Division of Code Enforcement and Administration (ny.gov)).05/16/2025In ProcessFALSE
- 6.0This is compliant with the DOE Order 420.lC "Facility Safety," construction of new facilities and significant modifications of existing facilities shall meet the applicable parts of the latest edition of the International Building Code (IBC, 2018 edition) and remains in compliance with DOE Orders and Standards directions in particular with DOE-STD-1066 section 2.2.4.05/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Building Code of New York State Chapters 2 to 3505/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Existing Building Code of New York State, Chapters 2 to 1605/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Fire Code of New York State, Chapters 2 to 6705/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Fuel Gas Code of New York State, Chapters 2 to 605/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Mechanical Code of New York State, Chapters 2 to 1505/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Plumbing Code of New York State, Chapters 2 to 1505/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Property Maintenance Code of New York State, Chapters 2 to 805/16/2025In ProcessFALSE
- 6.0The amended and updated version of the NYS Code Books incorporates by reference the 2020 Residential Code of New York State, Chapters 2 to 4405/16/2025In ProcessFALSE
- 6.0Rules on fire protection include 29 CFR 1910 Sub Part L Fire Protection05/16/2025In ProcessFALSE
- 6.0Rules on fire protection include ANSI Z 535.1 Safety Color Code05/16/2025In ProcessFALSE
- 6.0Rules on fire protection include DOE-STD-1066 Fire Protection Design Criteria05/16/2025In ProcessFALSE
GBL-SAFETY_CRYO&PRESS
- 6.0The completed collider complex must comply to ASME BPVC Section VIII - Rules for Construction of Pressure Vessels, Division 1‚ Rules for Construction of Pressure Vessels and Division 2‚ Alternative05/16/2025In ProcessFALSE
- 6.0The completed collider complex must comply to ASME B31.3 Process Piping05/16/2025In ProcessFALSE
- 6.0The completed collider complex must comply to BNL-8715-2008-IR Vacuum Systems Consensus Guidelines for DOE Accelerator Laboratories05/16/2025In ProcessFALSE
- 6.0The completed collider complex must comply to Compressed Gas Association (CGA) Standard S-1.3, Pressure Relief Device Standards Part 3 – Stationary Storage Containers for Compressed Gases05/16/2025In ProcessFALSE
- 6.0The completed collider complex must comply to ISO 21013-3 Cryogenic vessels - Pressure-relief accessories for cryogenic service Part 3: Sizing capacity determination05/16/2025In ProcessFALSE
GBL-SAFETY_EGRESS
- 6.0New construction in the collider complex must be constructed such as to avoid confined space in areas that must be entered by personnel for maintenance and repair. GBL SAFETY_EGRESS.02 Shall avoiding confined space in areas that must be entered by personnel not be possible or lead to unreasonable conditions, the collider design must include mitigation of the corresponding hazard.05/16/2025In ProcessFALSE
- 6.0Any work locations in the collider tunnel shall not be further away than 400 ft from the next tunnel exit when the tunnel has sprinkler protection.05/16/2025In ProcessFALSE
- 6.0The exit path from any work location in the collider tunnel to the next exit must be unobstructed by accelerator components and shall not require underpasses with less than 36 inches width and 36 inches height.05/16/2025In ProcessFALSE
- 6.0Markings shall be provided to denote and preserve access to the duck under.05/16/2025In ProcessFALSE
- 6.0In certain regions a ladder or stair shall be used to access a platform on top of the magnets for egress.05/16/2025In ProcessFALSE
- 6.0Emergency lighting and illuminated exit signs will be provided.05/16/2025In ProcessFALSE
GBL-SAFETY_ELEC
- 6.0Electrical equipment purchased for the accelerator will be certified by a Nationally Recognized Testing Laboratory (NRTL) whenever possible. GBL_SAFETY_ELEC All equipment05/16/2025In ProcessFALSE
- 6.0All equipment will adhere to Classification of Flammable Liquids, Gases, or Vapors and of Hazardous (Classified) Locations for Electrical Installations in Chemical Process05/16/2025In ProcessFALSE
- 6.0All equipment will adhere to National Electrical Code, NFPA 7005/16/2025In ProcessFALSE
- 6.0All equipment will adhere to Electrical Standard for Industrial Machinery, NFPA 7905/16/2025In ProcessFALSE
- 6.0All equipment will adhere to Standard for Competency of Third-Party Field Evaluation Bodies, NFPA 79005/16/2025In ProcessFALSE
GBL-SAFETY_GEN
- 6.0The elements EIC collider complex must be designed and built to meet the BNL Subject Based Management System (https://sbms.bnl.gov/).05/16/2025In ProcessFALSE
- 6.0The elements EIC collider complex must be designed and built to meet the DOE 10CFR851Worker Safety and Health Program.05/16/2025In ProcessFALSE
GBL-SAFETY_LASERS
- 6.0Lasers and laser enclosures will be designed to comply with ANSI Z136.1, American National Standard for the Safe Use of Lasers.05/16/2025In ProcessFALSE
GBL-SAFETY_MAGFIELDS
- 6.0Magnetic fields may be present in accelerator components and personnel exposure will not exceed the thresholds set in ACGIH05/16/2025In ProcessFALSE
GBL-SAFETY_ODH
- 6.0Oxygen deficiency hazards must be avoided by providing oxygen monitoring, ventilation, adequate warning systems and corresponding training of operating and maintenance staff.05/16/2025In ProcessFALSE
- 6.0Oxygen deficiency hazards are evaluated using the methodology in SBMS, Oxygen Deficiency Hazards (ODH), System Classification and Controls.05/16/2025In ProcessFALSE
GBL-SAFETY_RAD
- 6.0The radiation safety measures of the EIC shall be compliant with the DOE order 10 CFR 835, Energy/Occupational Radiation Protection and the more stringent requirements set forth in the BNL SBMS.05/16/2025In ProcessFALSE
- 6.0All exposures shall be As Low as Reasonably Achievable.05/16/2025In ProcessFALSE
- 6.0Radiation emitted by the EIC accelerator beams and due particle losses must be shielded such that dose limits outside the accelerator enclosures are 25 mrem annually for an inadvertently exposed person05/16/2025In ProcessFALSE
- 6.0Radiation emitted by the EIC accelerator beams and due particle losses must be shielded such that dose limits outside the accelerator enclosures are 5 mrem annually at the site boundary05/16/2025In ProcessFALSE
- 6.0Radiation emitted by the EIC accelerator beams and due particle losses must be shielded such that dose limits outside the accelerator enclosures are 20 mrem during a fault condition05/16/2025In ProcessFALSE
- 6.0Radiation emitted by the EIC accelerator beams and due particle losses must be shielded such that dose limits outside the accelerator enclosures are 0.5 mrem in 1 hour or 20 mrem in one week for continuously05/16/2025In ProcessFALSE
GBL-SAFETY_RF
- 6.0The RF systems must comply to the ASME BPVC, https://www.asme.org/codes-standards/find-codes-standards/bpvc-complete-code-boiler-pressure-vessel-code-complete-set.05/16/2025In ProcessFALSE
- 6.0The operating EIC has to following the review, inspection and maintenance requirements of the ASME BPVC.05/16/2025In ProcessFALSE
- 6.0The RF exposure to personnel will not exceed the thresholds set in ACGIH Threshold Limit Values – 2016.05/16/2025In ProcessFALSE
GBL-SAFETY_VACUUM
- 6.0Vacuum systems will comply to guidelines set forth in the publication, Vacuum Systems Consensus Guideline for Department of Energy Accelerator Laboratories”, BNL-81715-2008-IR https://intranet.bnl.gov/esh/shsd/seg/refdoc/pressuresafety/vacuum_standard.doc05/16/2025In ProcessFALSE
GBL-SAFETY_WATERCOOL
- 6.0In the design of water-cooling systems, precaution need to be taken to avoid endangering the staff by dangerous bacterial infections. GBL_SAFETY_WATERCOOL.02 The suppression of unwanted effects such as algae in open water circuits must avoid aggressive chemicals.05/16/2025In ProcessFALSE
- 6.0The guideline for design of cooling water systems must obey the ANSI/ASHRAE Standard 188. Legionellosis: Risk Management for Building Water Systems. establishes minimum legionellosis risk management requirements for building water systems. (This is also in the BNL SBMS.)05/16/2025In ProcessFALSE
- 6.0The guideline for design of cooling water systems must obey the State Requirement Protection Against Legionella. 10 NYCRR Part 4 – Protection Against Legionella: Subpart 4-1, Cooling Towers.05/16/2025In ProcessFALSE
GBL-UPGRADES
- 6.0The EIC plans to include previsions to allow future a upgrade that is (but is not limited to) two interaction regions and two colliding beam detectors.05/16/2025In ProcessFALSE
- 6.0The EIC must be planned such that a second interaction region and a second detector can be integrated in the collider with a minimum of cost and effort.05/16/2025In ProcessFALSE
- 6.0Integrating a second detector is achieved by designing the present beam trajectories in the possible area for the next IR around IP8 such that a second IR can be introduced without imposing a difference05/16/2025In ProcessFALSE
- 6.0The design of the electron and hadron beam optics and their higher order correction must offer sufficient margin so that the accelerator performance is not compromised by the 2nd IR.05/16/2025In ProcessFALSE
- 6.0Obsolete detector or equipment in the hall around IP8 shall be removed to avoid impeding the construction of a possible 2nd IR05/16/2025In ProcessFALSE
- 6.0The facility must be upgradable to operation with colliding electron and polarized deuteron beams.05/16/2025In ProcessFALSE

Electron Injector System Requirements
General, functional and performance requirements associated with the Electron Injector Systems of the Electron Ion Collider.
- NameWBSDescriptionUpdatedStatusTBD
EIS : Electron Injector System Performance Requirements
EIS-RCS : Rapid Cycling Synchrotron (WBS 6.03.02)
EIS-RCS-MAG : EIS RCS Magnet (WBS 6.03.02.02)
EIS-RCS-MAG-C0 : EIS RCS Magnet C0 (WBS 6.03.02.01)
EIS-RCS-MAG-Q0 : EIS RCS Magnet Q0 (WBS 6.03.02.01)
EIS-RCS-MAG-Q0_6 : EIS RCS Magnet Q0_6 (WBS 6.03.02.01)
EIS-RCS-MAG-S0 : EIS RCS Magnet S0 (WBS 6.03.02.01)
EIS-RCS-MAG-D0 : EIS RCS Magnet D0 (WBS 6.03.02.02)
EIS-RCS-MAG-CORR : RCS Corrector Magnet
EIS-RCS-MAG-DIP : RCS Dipole Magnet
EIS-RCS-MAG-PS : RCS Magnet Power Supply
EIS-RCS-MAG-QUAD : RCS Quadrupole Magnet
EIS-RCS-MAG-SEXT : RCS Sextupole Magnet
EIS-RCS-PS : EIS RCS Magnet Power Supply (WBS 6.03.02.03)
EIS-RCS-PS-C0 : EIS RCS Magnet Power Supply C0 (WBS 6.03.02.03)
EIS-RCS-PS-CO : EIS RCS Magnet Power Supply CO (WBS 6.03.02.03)
EIS-RCS-PS-D0 : EIS RCS Magnet Power Supply D0 (WBS 6.03.02.03)
EIS-RCS-PS-Q0 : EIS RCS Magnet Power Supply Q0 (WBS 6.03.02.03)
EIS-RCS-PS-Q0_6 : EIS RCS Magnet Power Supply Q0_6 (WBS 6.03.02.03)
EIS-RCS-PS-S0 : EIS RCS Magnet Power Supply S0 (WBS 6.03.02.03)
EIS-RCS-PS-SXT : EIS RCS Magnet Power Supply SXT (WBS 6.03.02.03)
EIS-RCS-VAC : EIS RCS Vacuum (WBS 6.03.02.04)
EIS-RCS-INST : EIS RCS Instrumentation (WBS 6.03.02.05)
EIS-RCS-INST-BPM : EIS RCS Instrumentation Beam Position Monitor (WBS 6.03.02.05)
EIS-RCS-INST-BPM-ELEC : EIS HETL Instrumentation Beam Profile Monitor Electronic (WBS 6.03.02.05.01)
EIS-RCS-INST-BPM-PU : EIS RCS Instrumentation Beam Profile Monitor Pick-Up (WBS 6.03.02.05.01)
EIS-RCS-INST-CM : EIS RCS Instrumentation Current and Charge Monitor (WBS 6.03.02.05)
EIS-RCS-INST-SLM : EIS RCS Instrumentation Synchrotron Light Monitor (WBS 6.03.02.05)
EIS-RCS-INST-TM : EIS RCS Instrumentation Tune Monitor (WBS 6.03.02.05)
EIS-RCS-INST-TPM : EIS RCS Instrumentation Transverse Profile Monitor (WBS 6.03.02.05)
EIS-RCS-INST-BBA : RCS Beam Based Alignment Monitor
EIS-RCS-INST-BC : RCS Bunch Charge Monitor
EIS-RCS-INST-BCM : RCS Beam Current Monitor
EIS-RCS-INST-BLM : RCS Beam Loss Monitor
EIS-RCS-INST-BP : RCS Bunch Profile Monitor
EIS-RCS-INST-TUNE : RCS Tune Monitor
EIS-RCS-PPD : EIS Pulsed Power Device System (WBS 6.03.03.05)
EIS-RCS-PPD-INJ_PULSE_KICK : EIS RCS Injection Pulse Power Kicker System (WBS 6.03.03.05.01)
EIS-RCS-PPD-INJ_PULSE_SEPT : EIS RCS Injection Pulsed Power Septum System (WBS 6.03.03.05.02)
EIS-RCS-PPD-EXT_PULSE_KICK : EIS RCS Extraction Pulse Power Kicker System (WBS 6.03.03.05.03)
EIS-RCS-PPD-EXT_RF_KICK : EIS RCS Extraction Pulse Power System (WBS 6.03.03.05.04)
EIS-RCS-PPD-INJ_RF_KICK : EIS RCS Injection Pulse Power System (WBS 6.03.03.05.04)
EIS-RCS-PPD-EXT_PULSE_SEPT : EIS RCS Extraction Pulsed Power Septum System (WBS 6.03.03.05.05)
EIS-RCS-PPD-BUMP : RCS Extraction System Bump Magnet
EIS-RCS-PPD-FKICK : RCS Extraction System Fast Kicker Magnet
EIS-RCS-PPD-SEPT : RCS Extraction System Septum Magnet
EIS-RCS-CONT : EIS RCS Controls (WBS 6.07.02)
EIS-RCS-CONT-INJ : EIS RCS Controls System Injection Tuning (WBS 6.07.02)
EIS-RCS-RF : EIS RCS RF System (WBS 6.08)
EIS-RCS-RF-ACAV:591M
EIS-HETL : EIS High Energy Transfer Line (HETL) System (WBS 6.03.03)
EIS-HETL-MAG : EIS HETL Magnet (WBS 6.03.03.01)
EIS-HETL-MAG-B1 : EIS HETL Magnet B1 (WBS 6.03.03.01)
EIS-HETL-MAG-B2 : EIS HETL Magnet B2 (WBS 6.03.03.01)
EIS-HETL-MAG-D1 : EIS HETL Magnet D1 (WBS 6.03.03.01)
EIS-HETL-MAG-D2 : EIS HETL Magnet D2 (WBS 6.03.03.01)
EIS-HETL-MAG-D3 : EIS HETL Magnet D3 (WBS 6.03.03.01)
EIS-HETL-MAG-INDSEPT1 : EIS HETL Magnet INDSEPT1 (WBS 6.03.03.01)
EIS-HETL-MAG-KH1 : EIS HETL Magnet KH1 (WBS 6.03.03.01)
EIS-HETL-MAG-KV1 : EIS HETL Magnet KV1 (WBS 6.03.03.01)
EIS-HETL-MAG-Q1 : EIS HETL Magnet Q1 (WBS 6.03.03.01)
EIS-HETL-MAG-Q2 : EIS HETL Magnet Q2 (WBS 6.03.03.01)
EIS-HETL-MAG-SEPT1 : EIS HETL Magnet SEPT1 (WBS 6.03.03.01)
EIS-HETL-MAG-SEPT2 : EIS HETL Magnet SEPT2 (WBS 6.03.03.01)
EIS-HETL-MAG-ESR_INJ_BMP : EIS HETL Injection Bump ESR Magnet (WBS 6.03.03.05.08)
EIS-HETL-MAG-BUMP : High Energy Transfer Line Bump Magnet
EIS-HETL-MAG-SEPT : High Energy Transfer Line Septum Magnets
EIS-HETL-PS : EIS HETL Magnet Power Supply (WBS 6.03.03.02)
EIS-HETL-PS-B1 : EIS HETL Magnet Power Supply B1 (WBS 6.03.03.02)
EIS-HETL-PS-B2 : EIS HETL Magnet Power Supply B2 (WBS 6.03.03.02)
EIS-HETL-PS-D1 : EIS HETL Magnet Power Supply D1 (WBS 6.03.03.02)
EIS-HETL-PS-D2 : EIS HETL Magnet Power Supply D2 (WBS 6.03.03.02)
EIS-HETL-PS-D3 : EIS HETL Magnet Power Supply D3 (WBS 6.03.03.02)
EIS-HETL-PS-INDSEPT1 : EIS HETL Magnet Power Supply INDSEPT1 (WBS 6.03.03.02)
EIS-HETL-PS-KH1 : EIS HETL Magnet Power Supply KH1 (WBS 6.03.03.02)
EIS-HETL-PS-KV1 : EIS HETL Magnet Power Supply KV1 (WBS 6.03.03.02)
EIS-HETL-PS-Q1 : EIS HETL Magnet Power Supply Q1 (WBS 6.03.03.02)
EIS-HETL-PS-Q2 : EIS HETL Magnet Power Supply Q2 (WBS 6.03.03.02)
EIS-HETL-PS-SEPT1 : EIS HETL Magnet Power Supply SEPT1 (WBS 6.03.03.02)
EIS-HETL-PS-SEPT2 : EIS HETL Magnet Power Supply SEPT2 (WBS 6.03.03.02)
EIS-HETL-PS-ESR_INJ_BMP : EIS HETL Injection Bump ESR Power Supply (WBS 6.03.03.05)
EIS-HETL-VAC : EIS HETL Vacuum (WBS 6.03.03.03)
EIS-HETL-INST : EIS HETL Instrumentation System (WBS 6.03.03.04)
EIS-HETL-INST-BPM : EIS HETL Instrumentation Beam Profile Monitor (WBS 6.03.03.04)
EIS-HETL-INST-BPM-ELEC : EIS HETL Instrumentation Beam Profile Monitor Electronic (WBS 6.03.03.04.01)
EIS-HETL-INST-BPM-PU : EIS HETL Instrumentation Beam Profile Monitor Pick-Up (WBS 6.03.03.04.01)
EIS-HETL-INST-CM : EIS HETL Instrumentation Charge Monitor (WBS 6.03.03.04)
EIS-HETL-INST-PM : EIS HETL Instrumentation Profile Monitor (WBS 6.03.03.04)
EIS-HETL-INST-IC : High Energy Transfer Line Integrating Current Monitor
EIS-HETL-INST-POL : High Energy Transfer Line Polarization Measurement System
EIS-HETL-ESR : High Energy Transfer Line ESR
EIS-HETL-ESR-INJ : High Energy Transfer Line ESR Injection Section
EIS-HETL-SLINE : High Energy Transfer Stripline
EIS-HETL-SLINE-FKICK : High Energy Transfer Fast Stripline Kicker
EIS-METL : EIS Medium Energy Transfer Line (METL) (WBS 6.03.03)
EIS-METL-MAG : EIS METL Magnet (WBS 6.03.03.01)
EIS-METL-MAG-BMP1 : EIS METL Magnet BMP1 (WBS 6.03.03.01)
EIS-METL-MAG-DH1 : EIS METL Magnet DH1 (WBS 6.03.03.01)
EIS-METL-MAG-DH2 : EIS METL Magnet DH2 (WBS 6.03.03.01)
EIS-METL-MAG-DH3 : EIS METL Magnet DH3 (WBS 6.03.03.01)
EIS-METL-MAG-DV1 : EIS METL Magnet DV1 (WBS 6.03.03.01)
EIS-METL-MAG-KH1 : EIS METL Magnet KH1 (WBS 6.03.03.01)
EIS-METL-MAG-KV1 : EIS METL Magnet KV1 (WBS 6.03.03.01)
EIS-METL-MAG-Q1 : EIS METL Magnet Q1 (WBS 6.03.03.01)
EIS-METL-MAG-Q_BATES : EIS METL Magnet BATES (WBS 6.03.03.01)
EIS-METL-MAG-SEPTUM : EIS METL Magnet SEP1 (WBS 6.03.03.01)
EIS-METL-MAG-SPINSOL : EIS METL Magnet SOL1 (WBS 6.03.03.01)
EIS-METL-MAG-DIP : Medium Energy Transfer Line Dipole Magnet
EIS-METL-MAG-QUAD : Medium Energy Transfer Line Quadrupole Magnet
EIS-METL-PS : EIS METL Magnet Power Supply (WBS 6.03.03.02)
EIS-METL-PS-BMP1 : EIS METL Magnet Power Supply BMP1 (WBS 6.03.03.02)
EIS-METL-PS-DH1 : EIS METL Magnet Power Supply DH1 (WBS 6.03.03.02)
EIS-METL-PS-DH2 : EIS METL Magnet Power Supply DH2 (WBS 6.03.03.02)
EIS-METL-PS-DH3 : EIS METL Magnet Power Supply DH3 (WBS 6.03.03.02)
EIS-METL-PS-DV1 : EIS METL Magnet Power Supply DV1 (WBS 6.03.03.02)
EIS-METL-PS-KH1 : EIS METL Magnet Power Supply KH1 (WBS 6.03.03.02)
EIS-METL-PS-KV1 : EIS METL Magnet Power Supply KV1 (WBS 6.03.03.02)
EIS-METL-PS-Q1 : EIS METL Magnet Power Supply Q1 (WBS 6.03.03.02)
EIS-METL-PS-Q2 : EIS METL Magnet Power Supply Q2 (WBS 6.03.03.02)
EIS-METL-PS-Q3 : EIS METL Magnet Power Supply Q3 (WBS 6.03.03.02)
EIS-METL-PS-Q4 : EIS METL Magnet Power Supply Q4 (WBS 6.03.03.02)
EIS-METL-PS-Q_BATES : EIS METL Magnet Power Supply BATES (WBS 6.03.03.02)
EIS-METL-PS-SEPTUM : EIS METL Magnet Power Supply SEP1 (WBS 6.03.03.02)
EIS-METL-PS-SPINSOL : EIS METL Magnet Power Supply SOL1 (WBS 6.03.03.02)
EIS-METL-VAC : EIS METL Vacuum (WBS 6.03.03.03)
EIS-METL-INST : EIS METL Instrumentation (WBS 6.03.03.04)
EIS-METL-INST-BPM : EIS METL Instrumentation Beam Profile Monitor (WBS 6.03.03.04)
EIS-METL-INST-BPM-ELEC : EIS METL Instrumentation Beam Profile Monitor Electronic (WBS 6.03.03.04.01)
EIS-METL-INST-BPM-PU : EIS METL Instrumentation Beam Profile Monitor Pick-Up (WBS 6.03.03.04.01)
EIS-METL-INST-CM : EIS METL Instrumentation Charge Monitor (WBS 6.03.03.04)
EIS-METL-INST-PM : EIS METL Instrumentation Profile Monitor (WBS 6.03.03.04)
EIS-METL-INJ : Medium Energy Transfer Line Electron Injector Injection Section
EIS-METL-INJ-MAG : Medium Energy Transfer Line Electron Injector Injection Section Magnet
EIS-METL-INJ-MAG-FKICK : Medium Energy Transfer Line Electron Injector Injection Section Fast Kicker Magnet
EIS-METL-INJ-MAG-SEP : Medium Energy Transfer Line Electron Injector Injection Section Septum Magnet
EIS-METL-RCS : Medium Energy Transfer Line RCS
EIS-METL-RCS-INJ : Medium Energy Transfer Line RCS Injection Section
EIS-METL-RCS-INJ-MAG : Medium Energy Transfer Line RCS Injection Section Magnet
EIS-METL-RCS-INJ-MAG-BUMP : Medium Energy Transfer Line RCS Injection Section Bump Magnet
EIS-METL-RF : Medium Energy Transfer Line RF
EIS-METL-RF-FKICK : Medium Energy Transfer Line Fast RF Kicker
EIS-GUN : EIS Electron Source Gun (WBS 6.03.04.01)
EIS-GUN-TIMING : EIS Electron Source Gun Timing (WBS 6.03.04.01)
EIS-GUN-VAC : EIS Electron Source Gun Vacuum (WBS 6.03.04.01.02)
EIS-GUN-ESOURCE : EIS Electron Source Gun Electron Beam (WBS 6.03.04.02)
EIS-GUN-LASER : EIS Electron Source Gun Laser (WBS 6.03.04.02)
EIS-GUN-INST : EIS Electron Source Gun Instrumentation (WBS 6.03.04.02.06)
EIS-GUN-INST-EM : EIS Electron Source Gun Instrumentation Emittance Monitor (WBS 6.03.04.02.06)
EIS-GUN-INST-F_Cup : EIS Electron Source Gun Instrumentation Faraday Cup (WBS 6.03.04.02.06)
EIS-GUN-INST-MOT : EIS Electron Source Gun Instrumentation Mott Polarimeter (WBS 6.03.04.02.06)
EIS-GUN-INST-SR : EIS Electron Source Gun Instrumentation Spin Rotator (WBS 6.03.04.02.06)
EIS-GUN-CONT : EIS Electron Source Gun Controls System (WBS 6.07.02)
EIS-GUN-MAG : Gun Diagnostic Line Magnet
EIS-LINAC : EIS Electron Source LINAC (WBS 6.03.04.01)
EIS-LINAC-ACCEL : EIS Electron Source LINAC Accelerating System (WBS 6.03.04.01)
EIS-LINAC-BEAM : EIS Electron Source LINAC Beamline Electron Bunch (WBS 6.03.04.01)
EIS-LINAC-BUNCH_118MHZ : EIS Electron Source LINAC Buncher 118MHz (WBS 6.03.04.01)
EIS-LINAC-BUNCH_2856MHZ : EIS Electron Source LINAC Buncher 2856MHz (WBS 6.03.04.01)
EIS-LINAC-BUNCH_560MHZ : EIS Electron Source LINAC Buncher 560MHz (WBS 6.03.04.01)
EIS-LINAC-CONT : EIS Electron Source LINAC Controls (WBS 6.03.04.01)
EIS-LINAC-DECHIRP : EIS Electron Source LINAC De Chirp System (WBS 6.03.04.01)
EIS-LINAC-INST : EIS Electron Source LINAC Instrumentation (WBS 6.03.04.01)
EIS-LINAC-INST-BLM : EIS Electron Source LINAC Bunch Length Monitor (WBS 6.03.04.01)
EIS-LINAC-INST-BPM : EIS Electron Source LINAC Beam Position Monitor (WBS 6.03.04.01)
EIS-LINAC-INST-CM : EIS Electron Source LINAC Bunch Charge Monitors (ICT, WCM, FC) (WBS 6.03.04.01)
EIS-LINAC-INST-PM : EIS Electron Source LINAC Transverse Profile Monitor (PM) (WBS 6.03.04.01)
EIS-LINAC-VAC : EIS Electron Source LINAC Vacuum (WBS 6.03.04.01)
EIS-LINAC-ZIGZAG : EIS Electron Source LINAC Zigzag (WBS 6.03.04.01)
EIS-LINAC-ZIGZAG-MAG : Zigzag Section Magnet
EIS-LINAC-ZIGZAG-MAG-CHIRP : Zigzag Section De-Chirp Cavity
EIS-LINAC-ZIGZAG-MAG-DIP : Zigzag Section Dipole Magnet
EIS-LINAC-ZIGZAG-MAG-QUAD : Zigzag Section Quadrupole Magnet
EIS-LINAC-BNCHR : Buncher Section
EIS-LINAC-BNCHR-INST : Buncher Section Instrumentation
EIS-LINAC-BNCHR-INST-BLM : Buncher Section Bunch Length Monitor
EIS-LINAC-BNCHR-INST-BPM : Buncher Section Beam Position Monitor
EIS-LINAC-BNCHR-INST-EM : Buncher Section Emittance Monitor
EIS-LINAC-BNCHR-INST-ICT : Buncher Section Integrating Current Transformer Monitor
EIS-LINAC-BNCHR-INST-LM : Buncher Section Beam Loss Monitor
EIS-LINAC-BNCHR-INST-TPM : Buncher Section Transverse Profile Monitor
EIS-LINAC-BNCHR-MAG : Buncher Section Magnet
EIS-LINAC-BNCHR-MAG-CORR : Buncher Section Corrector Magnet
EIS-LINAC-BNCHR-MAG-HELM : Buncher Section Helmholtz Magnet
EIS-LINAC-BNCHR-MAG-SOL : Buncher Section Solenoid Magnet
EIS-LINAC-BNCHR-RF : Buncher Section RF System
EIS-LINAC-BNCHR-VAC : Buncher Section Vacuum
EIS-LINAC-CAV : LINAC RF Cavity
EIS-LINAC-DUMP : LINAC Beam Dump
EIS-LINAC-DUMP-INST : LINAC Dump Line
EIS-LINAC-MAG : LINAC Magnet
EIS-LINAC-MAG-QUAD : LINAC Quadrupole Magnet
EIS-LINAC-MAG-WF : LINAC Window Frame Steering Magnet
EIS-LETL : Gun Low Energy Transfer Line
EIS-MPS : Protection System
EIS-MPS-ABORT : Protection System Beam Abort
EIS-MPS-COLLIMATION : Protection System Collimator
EIS-MPS-CONTROLS
EIS-MPS-GENERAL
EIS-MPS-RADSHIELD : Protection System Beam Dump
EIS-TLINE
Electron Storage Ring Requirements
General, functional and performance requirements associated with the Electron Storage Ring of the Electron Ion Collider.
- NameWBSDescriptionUpdatedStatusTBD
ESR : Electron Storage Ring (WBS 6.02.02)
- 6.02.02The ESR shall be able to accept single bunches of spin-polarized electrons as injected from the RCS at the energies specified in the Master Parameter Table (MPT). [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.02All ESR quadrupoles shall be designed to facilitate beam based alignment.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02The ESR main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02The ESR alignment requirements are established by dynamic aperture and polarization tracking. The ESR RMS alignment tolerances shall be such that all the beam parameter listed in the MPT can be satisfied. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice shall provide a minimum dynamic aperture of 10 sigma with respect to Gaussian electron beam distribution in all three dimensions (horizontal, vertical, and longitudinal) having a vertical emittance of half the horizontal design emittance.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The field harmonic measurements shall be measured at the reference radius of 25mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet reference radius for field homogeneity shall be 17 (mm).03/02/2026ApprovedFALSE
- 6.02.02.06The maximum beam excursion orbit shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet reference radius for homogeneity shall be 17 (mm).02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR Lattice shall contain provisions for correctors such as skew quadrupoles, Dipole correctors etc. as needed.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall have a depolarization time long enough to ensure an average polarization of 70%, assuming an injected polarization level of 80% and a bunch replacement rate of up to two bunches per second.02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber shall provide sufficient horizontal and vertical aperture to accommodate; a +/-15 sigma beam, where the vertical RMS beam size is based on the emittance of a fully coupled beam, plus an additional 10 mm horizontal and 5 mm vertical margin to account for expected orbit errors.02/09/2026ApprovedFALSE
- 6.02.02.06The typical (standard) vacuum chamber aperture shall be 80 x 36 mm.02/09/2026ApprovedFALSE
- 6.02.02.06Special aperture requirements and/or aperture file shall be provided by or approved by physics.02/09/2026ApprovedFALSE
- 6.02.02The dynamic pressure around the ESR shall be consistent with a beam gas lifetime of >10[hrs] with the design currents after an integrated beam current of 1000 [A.h].02/09/2026ApprovedFALSE
- 6.02.02.06There shall be no upper pressure limit as long as the average pressure is maintained.02/09/2026ApprovedFALSE
- 6.02.02.06The average vacuum level in the ESR Arc sections after conditioning (for 1000Ahrs) shall be <5x10-9 Torr.02/09/2026ApprovedFALSE
- 6.02.02.06On 15 m on each side (or one vacuum sector) of the SRF cavities shall be processed to class ISO 5.02/09/2026ApprovedFALSE
- 6.02.02There shall be no pressure bumps in the ESR exceeding (TBD)[Torr]02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber and all its components shall be designed to withstand a total synchrotron radiation load of 10 MW, considering the uneven linear load particularly related to the super-bends.02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber material shall be chosen such that the SR power can be intercepted by the arc chambers and in addition good radiation shielding will be provided to prevent damage to other components.02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.04.05The impedance of the entire ESR vacuum system, including the interaction regions in IR06 and IR08, shall allow for the bunch intensities, beam currents, and bunch numbers contained in the Master Parameter Table (MPT). [Document#:EIC-SEG-RSI-005]05/16/2025ApprovedFALSE
- 6.02.02.06The vacuum system global impedance shall be less than the impedance budget as provided by accelerator physics.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall deliver a spin polarized electron beam with time-averaged polarization of at least 70 percent.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
- 6.02.02The polarization lifetime of the beam in the ESR shall be maximized to maximize time-averaged polarization based on the given replacement frequency and polarization degree of the bunches provided by the injector.02/09/2026ApprovedFALSE
- 6.02.02The polarization lifetime maximum optimization in the ESR shall be accomplished by proper spin matching which minimizes the spin diffusion.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
- 6.02.02The ESR shall include a system to absorb the energy of bunches which are periodically and continually ejected from the ESR to accommodate freshly polarized bunches.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall be capable of having all polarized electron bunches in the ring continually replaced while the electron beam is in collision with a hadron beam, thus allowing for arbitrary spin patterns for the collider experiments.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
- 6.02.02The ESR shall be capable of delivering bunches with longitudinal spins to the IP.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall provide electron bunches having the bunch parameters specified in the MPT. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall contain an array of regular FODO cells02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall accommodate slightly different average arc radii in the individual arcs by adjusting the drift spaces between individual elements in each FODO cell.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR super-bends shall generate additional synchrotron radiation damping to support a large beam-beam parameter of 0.1 and to create the required horizontal design emittance in the Master Parameter Table (MPT) when the ESR is operated at energies below 10 GeV. [Document: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The polarity of the ESR center bending magnet shall be capable of being wired in reverse to control the beam emittance and to damp the beam. The polarity will be dictated by the beam energy.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The FODO cell shall operate with a horizontal and vertical betatron phase advance of 60 degrees per arc section at beam energies of 10 GeV and below.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02The ESR Sextupole wiring scheme shall create the required sextupole families needed per arc to maximize dynamic aperture at the 60 degrees per FODO cell phase advance at < 10 GeV.02/09/2026ApprovedFALSE
- 6.02.02The FODO cell shall operate with a horizontal and vertical betatron phase advance of 90 degrees per arc section to maintain the required horizontal beam emittance defined in the MPT at 18 GeV. [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.02The vertical emittance shall be controlled by appropriate beam orbit manipulations and horizontal-vertical cross coupling.02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include dual-plane Beam Position Monitors (BPMs) adjacent to each vertically focusing quadrupole. Provisions shall be made in the vacuum chamber design to install additional dual-plane BPMs at the horizontally focusing quadrupoles, if needed.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up shall provide dual plane (horizontal and vertical) beam positional measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The roll angle error of the ESR BPM pickup, given as the horizontal measurement plane defined by the BPM fiducials witrh respect to. the horizontal plane of the ESR ring shall be less than +/- 20 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall fulfill resolution and accuracy requirements over the following two defined transverse beam position ranges with respect vacuum chamber center, referenced to the mechanical fiducials on the BPM pickups.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The beam position monitor pick-up assembly shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02The ESR BPMs shall have turn-by-turn orbit measurement capability based on a single, remotely selectable bunch out of the fully filled bunch train to enable injection optimization.02/09/2026ApprovedFALSE
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following capabilities defined for the low intensity pilot injection energies and high intensity collision energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following time resolutions for data refresh defined for the beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following time resolutions for data logging defined for the beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following measurement resolutions defined for beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate reliability with capability to withstand a lifetime radiation dose of TBD MGy.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate in an ambient temperature degree from X (C) to X (C).02/09/2026ReviewedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up shall provide dual plane (horizontal and vertical) beam positional measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The roll angle error of the ESR BPM pickup, given as the horizontal measurement plane defined by the BPM fiducials witrh respect to. the horizontal plane of the ESR ring shall be less than +/- 20 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall fulfill resolution and accuracy requirements over the following two defined transverse beam position ranges with respect vacuum chamber center, referenced to the mechanical fiducials on the BPM pickups.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The beam position monitor pick-up assembly shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a beam current monitor to measure average beam current.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall have the ability to measure the average beam current over a range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall provide an average current measurement with a resolution of less than 5 (uA /√Hz).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measurement drift tolerance shall be less than 10 (uA) over 1 (hr).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy of better than +/- 1 (%) at 250 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy between the range of 250 (mA) to 2.5 (A) at +/- 0.5 (%).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer Calibration system shall be capable of providing an equivalent ESR DC current over the full operating range within 0.25% over the beam current range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall have a remote controlled self calibration system02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measured average current shall be archived at a rate of 1 Hz02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure individual bunch charges and bunch pattern.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring witness bunch for 1/e for a fixed gain over the beam lifetime.02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure transverse beam profiles.02/09/2026ApprovedFALSE
- 6.02.02.05.05.01The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure longitudinal beam profiles.02/09/2026ApprovedFALSE
- 6.02.02.05.05.02The requirements for longitudinal feedback are ??? TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.02The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02The ESR longitudinal bunch profile monitor needs turn-by-turn capability based on a single bunch in the fully filled bunch train to allow timing and energy adjustment for injection optimization.02/09/2026ApprovedFALSE
- 6.02.02.05.05.02The requirements for longitudinal feedback are ??? TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.02The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include system to measure H & V betatron tunes.02/09/2026ApprovedFALSE
- 6.02.02.05.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the horizontal kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the vertical kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The location of the tune meter kicker striplines in the ESR shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The kicker waveform (risetime and shape) requirements shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The impedance of the kicker beamline device shall be approved by beam Physics.02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall facilitate all required feedback systems (slow transverse, longitudinal and transverse bunch-by-bunch)02/09/2026ApprovedFALSE
- 6.02.04.02The slow orbit feedback correction output rate shall be 10 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The slow orbit feedback BPM data averaging period shall be tbd -02/09/2026In ProcessFALSE
- 6.02.02.05.05.01The transverse slow feedback system bandwidth shall bs 10 Hz02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include beam loss monitor system with detectors located only at select regions of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.05.02BLM shall be needed to needed to protect sensitive equipment.02/09/2026In ProcessFALSE
- 6.02.02.05.02The number of BLM installed in the ESR shall be TBD ea02/09/2026In ProcessFALSE
- 6.02.02.05.02BLM shall be installed at the following locations in the ESR TBD02/09/2026In ProcessFALSE
- 6.02.02.05.02The sensitivity of the BLM detectors shall be TBD units?02/09/2026In ProcessFALSE
- 6.02.02.05.02Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/09/2026In ProcessFALSE
- 6.02.02.05.02The response time from loss detection to abort shall be TBD us02/09/2026In ProcessFALSE
- 6.02.02The ESR magnets shall meet the requirements defined by the physics lattice.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall have the required field quality to meet the operational needs.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The “super-bends” in the ESR ARC sections shall consist of two long dipoles on either end of a short dipole.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The good field region of the ESR dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02All ESR quadrupoles shall be designed to facilitate beam based alignment.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02The maximum integrated field strength of the ESR FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.02The aperture of all ESR magnets shall be large enough to accommodate the ESR vacuum chamber.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 52 and 150 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet pole tip radius shall be 40 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 150 and 52 mm.02/09/2026ApprovedFALSE
- 6.02.02The ESR Shall have a conventional orbit corrector scheme, with single-plane correctors located at the respective quadrupoles. The strength of these correctors needs to be chosen to correct for any source of orbit distortion and should have enough margin for beam based diagnostic purposes, Harmonic Spin bumps and emittance generating bumps.02/09/2026ApprovedFALSE
- 6.02.02All dipoles in the ESR, including those in IRs 6 and 8, shall be connected in series to a single main power supply.02/09/2026ApprovedFALSE
- 6.02.02The ESR main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02The ESR quadrupoles in the straight sections IR02, IR04, IR10 and IR12 and in the transition from the arc to the straight section structure shall be wired to provide the optimized betatron phase advance across each straight section, as required for dynamic aperture optimization.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR quadrupoles shall have provisions to vary individual strengths by approximately 1% for beam-based alignment purposes.02/09/2026ApprovedFALSE
- 6.02.02The ESR sextupole power supply scheme shall be laid out such that the sextupole family structure can be configured for both the 60 and the 90 degree lattice along with a small number of individually powered sextupoles in the transition regions between arcs and straight sections with minimal effort, cost and minimizing any risk of error.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnet power supplies shall be capable of providing the stability the ESR needs to operate02/09/2026ApprovedFALSE
- 6.02.02The ESR RF Systems shall be designed to fulfill all necessary parameters as set by the Master Parameter Table (MPT). [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF System shall utilize superconductivity.02/09/2026ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum designed operational temperature shall be 2 K.05/16/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum designed operational pressure shall be 30 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath designed operational pressure stability shall be ±0.1 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cavity helium jacket shall have a minimum helium bath vapor surface area of 0.049 m^2.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed helium supply operational temperature shall be 5.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed helium supply operational pressure shall be 3 to 3.5 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return temperature shall be 20 to 100 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return pressure shall be 2.4 to 2.6 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return temperature shall be 4.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return pressure shall be 30 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity maximum Niobium temperature shall be 5 K during operation.05/16/2025ApprovedFALSE
- 6.08.04.01The warm beamline maximum vacuum shall be 5.0e-7 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The cold beamline maximum vacuum shall be 1.0e-9 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The beamline vacuum maximum leak rate shall be 5e-10 mbar L/s.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall conform to the ESR lattice.02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.08.04.01The SRF cryomodule cavity beam axis to the tunnel floor shall be vertically alignable to 1381.09 ± 20 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in X shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Y shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Z shall be ± 5 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ± 0.04 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall be installed in the straight sections of the ESR lattice within the existing RHIC tunnel in IR10.02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.08.04.01The SRF Cryomodule maximum length shall be 7.2 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule maximum width shall be 2.15 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule maximum height shall be 1.7 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cryomodule cavity beam axis to the tunnel floor shall be vertically alignable to 1381.09 ± 20 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in X shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Y shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Z shall be ± 5 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ± 0.04 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum height shall be 2.1m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum length (not including vacuum jacketed lines) shall be 1.5 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum width shall be 1.0 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic valve box minimum vertical stay clear height above the cryomodule shall be 0.92 m.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall conform to the EIC Code of Record.02/09/2026ApprovedFALSE
- 6.08.04.01All cryomodule surfaces accessible to workers shall be within the temperature range of 283 to 333 K.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME B31.3.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME BPVC.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASTM C1055.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 70.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 70E.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by API 520 & API 521.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by AWS.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by CGA S1.3.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards as directed by the DOE Vacuum Vessel Consensus Standards.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to meet or exceed the maximum working pressures defined by the EIC pressure document (Document No. TBD).05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems within the tunnel shall operate within its yearly radiation exposure budget.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall have a minimum operating lifetime of 20 years02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall operate through a minimum of 200 thermal cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner 1% range tuning cycles shall be 100,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner full range tuning cycles shall be 1,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF System shall be designed to minimize unscheduled downtime, maintenance time and repair time to achieve ESR operational availability.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 300K to 150K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve steady state temperature with the cavity bath at 4K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve a full warm-up cycle from 4K to 295K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall operate through a minimum of 200 thermal cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner 1% range tuning cycles shall be 100,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner full range tuning cycles shall be 1,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The manufactured SRF Cryomodule Cavity shall produce no field emission at 4 MV.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.08.04.01All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.05/16/2025ApprovedFALSE
- 6.08.04.01The active SRF cavity tuning mechanism components (motor/gearbox/drive mechanism) shall be replaceable and maintainable in-situ.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity slow tuner tuning rate shall be 800 Hz/s.05/16/2025ApprovedFALSE
- 6.02.02The ESR Storage RF System shall be designed to accelerate electrons.02/09/2026ApprovedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the cryomodule shall be 30 kW for all EIC design energies and currents.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E11.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- 6.02.02The phase advance of each straight section shall be tunable in order to optimize the dynamic aperture of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The field harmonic measurements shall be measured at the reference radius of 25mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The reference integrated field for the measurement shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR straight sections IR02, IR04, IR10 and IR12 shall be based on FODO cells.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The reference integrated field for the measurement shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02Ther ESR shall have matching sections at the ends of each of the straight sections to compensate for the different FODO cell lengths with respect to the arc FODO cells imposed by geometric constraints.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The maximum physical magnet length shall be 0.88 m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The maximum magnet axial dimensions shall be X = 45 cm and Y = 45 cm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber shall provide sufficient horizontal and vertical aperture to accommodate; a +/-15 sigma beam, where the vertical RMS beam size is based on the emittance of a fully coupled beam, plus an additional 10 mm horizontal and 5 mm vertical margin to account for expected orbit errors.02/09/2026ApprovedFALSE
- 6.02.02.06The typical (standard) vacuum chamber aperture shall be 80 x 36 mm.02/09/2026ApprovedFALSE
- 6.02.02.06Special aperture requirements and/or aperture file shall be provided by or approved by physics.02/09/2026ApprovedFALSE
- 6.02.02The dynamic pressure around the ESR shall be consistent with a beam gas lifetime of >10[hrs] with the design currents after an integrated beam current of 1000 [A.h].02/09/2026ApprovedFALSE
- 6.02.02.06There shall be no upper pressure limit as long as the average pressure is maintained.02/09/2026ApprovedFALSE
- 6.02.02.06The average vacuum level in the ESR Arc sections after conditioning (for 1000Ahrs) shall be <5x10-9 Torr.02/09/2026ApprovedFALSE
- 6.02.02.06On 15 m on each side (or one vacuum sector) of the SRF cavities shall be processed to class ISO 5.02/09/2026ApprovedFALSE
- 6.02.02There shall be no pressure bumps in the ESR exceeding (TBD)[Torr]02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber and all its components shall be designed to withstand a total synchrotron radiation load of 10 MW, considering the uneven linear load particularly related to the super-bends.02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber material shall be chosen such that the SR power can be intercepted by the arc chambers and in addition good radiation shielding will be provided to prevent damage to other components.02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.04.05The impedance of the entire ESR vacuum system, including the interaction regions in IR06 and IR08, shall allow for the bunch intensities, beam currents, and bunch numbers contained in the Master Parameter Table (MPT). [Document#:EIC-SEG-RSI-005]05/16/2025ApprovedFALSE
- 6.02.02.06The vacuum system global impedance shall be less than the impedance budget as provided by accelerator physics.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice shall provide a minimum dynamic aperture of 10 sigma with respect to Gaussian electron beam distribution in all three dimensions (horizontal, vertical, and longitudinal) having a vertical emittance of half the horizontal design emittance.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The field harmonic measurements shall be measured at the reference radius of 25mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet reference radius for field homogeneity shall be 17 (mm).03/02/2026ApprovedFALSE
- 6.02.02.06The maximum beam excursion orbit shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet reference radius for homogeneity shall be 17 (mm).02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR alignment requirements are established by dynamic aperture and polarization tracking. The ESR RMS alignment tolerances shall be such that all the beam parameter listed in the MPT can be satisfied. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02The minimum dynamic aperture shall be achieved in two optics configurations (60 and 90 degrees betatron phase advance per FODO cell) at all operational beam energies in the Master Parameter Table (MPT), and with one and with two low-beta insertions. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.06The maximum beam excursion orbit shall be TBD02/09/2026In ProcessFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR Lattice shall contain provisions for correctors such as skew quadrupoles, Dipole correctors etc. as needed.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall have an average arc bending radius of approximately 380 meters02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall contain an array of regular FODO cells02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall accommodate slightly different average arc radii in the individual arcs by adjusting the drift spaces between individual elements in each FODO cell.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall meet the requirements defined by the physics lattice.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall have the required field quality to meet the operational needs.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The “super-bends” in the ESR ARC sections shall consist of two long dipoles on either end of a short dipole.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The good field region of the ESR dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The aperture of all ESR magnets shall be large enough to accommodate the ESR vacuum chamber.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 52 and 150 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet pole tip radius shall be 40 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 150 and 52 mm.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02All dipoles in the ESR, including those in IRs 6 and 8, shall be connected in series to a single main power supply.02/09/2026ApprovedFALSE
- 6.02.02The ESR Lattice shall contain provisions for correctors such as skew quadrupoles, Dipole correctors etc. as needed.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR electron beam shall orbit in the clockwise direction as seen from above.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall contain an array of regular FODO cells02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall accommodate slightly different average arc radii in the individual arcs by adjusting the drift spaces between individual elements in each FODO cell.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall meet the requirements defined by the physics lattice.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall have the required field quality to meet the operational needs.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The “super-bends” in the ESR ARC sections shall consist of two long dipoles on either end of a short dipole.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The good field region of the ESR dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02All dipoles in the ESR, including those in IRs 6 and 8, shall be connected in series to a single main power supply.02/09/2026ApprovedFALSE
- 6.02.02The aperture of all ESR magnets shall be large enough to accommodate the ESR vacuum chamber.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 52 and 150 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet pole tip radius shall be 40 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 150 and 52 mm.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02The ESR Lattice shall be tilted with respect to the plane of the HSR along the chord between IP6 and IP8 by 200 microradians to avoid interference with other components and such that the ESR is above the HSR at IP12.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall have a circumference such that the revolution frequency of the stored electron beam matches the revolution frequency of 133 GeV protons stored in the HSR, with the proton beam orbit centered in the HSR beampipe.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall contain an array of regular FODO cells02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall accommodate slightly different average arc radii in the individual arcs by adjusting the drift spaces between individual elements in each FODO cell.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall meet the requirements defined by the physics lattice.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall have the required field quality to meet the operational needs.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The “super-bends” in the ESR ARC sections shall consist of two long dipoles on either end of a short dipole.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The good field region of the ESR dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02All dipoles in the ESR, including those in IRs 6 and 8, shall be connected in series to a single main power supply.02/09/2026ApprovedFALSE
- 6.02.02The aperture of all ESR magnets shall be large enough to accommodate the ESR vacuum chamber.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 52 and 150 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet pole tip radius shall be 40 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 150 and 52 mm.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall use the “inner” aisle (closest to the inner tunnel wall) of the tunnel from IR4 to IR6 and from IR8 to IR12, and the “outer” aisle (closest to the outer tunnel wall) from IR12 to IR4 and IR6 to IR8.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall contain an array of regular FODO cells02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall accommodate slightly different average arc radii in the individual arcs by adjusting the drift spaces between individual elements in each FODO cell.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR Sextupole wiring scheme shall create the required sextupole families needed per arc to maximize dynamic aperture at the 90 degrees per FODO cell phase advance at 18 GeV.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall meet the requirements defined by the physics lattice.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall have the required field quality to meet the operational needs.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The “super-bends” in the ESR ARC sections shall consist of two long dipoles on either end of a short dipole.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The good field region of the ESR dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02All dipoles in the ESR, including those in IRs 6 and 8, shall be connected in series to a single main power supply.02/09/2026ApprovedFALSE
- 6.02.02The aperture of all ESR magnets shall be large enough to accommodate the ESR vacuum chamber.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 52 and 150 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet pole tip radius shall be 40 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 150 and 52 mm.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall support collisions with the HSR in IR6, to accommodate colliding beam interaction regions and detectors for nuclear physics experiments.02/09/2026ApprovedFALSE
- 6.02.02The ESR main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02The maximum integrated field strength of the ESR FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.02The ESR shall support two low-beta insertions (colliding beam interaction regions) at IRs 06 and 08.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.06The maximum beam excursion orbit shall be TBD02/09/2026In ProcessFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR beam dynamics shall incorporate the need for collision points with the HSR in IR6 and IR8.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall have provisions made at IR8 to accommodate a future 2nd colliding beam interaction region with low-beta section, spin rotators and crab cavities.02/09/2026ApprovedFALSE
- 6.02.02The ESR main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02The maximum integrated field strength of the ESR FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.02The ESR shall support two low-beta insertions (colliding beam interaction regions) at IRs 06 and 08.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.06The maximum beam excursion orbit shall be TBD02/09/2026In ProcessFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall have provisions made at IR12 to accommodate beam polarimetry measurements, damper systems, special instrumentation.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall have provisions made to accommodate electron beam injection and extraction elements in IR4.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR fast abort system shall be located in the IR2 straight section.02/09/2026ApprovedFALSE
- 6.02.02The ESR shall be designed to minimize unscheduled downtime, maintenance time and repair time to achieve EIC operational availability.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR shall contain an Abort system to dump the beam.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The materials shall be C / Al / Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The energy deposited during abort shall be 320 kJ02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The frequency of thermal cycle shall be 1 hour02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window thickness shall be tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window material shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Length shall be 50 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Internal diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Temperature sensors shall be yes02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The BPMs shall be 402/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Correctors shall be 402/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Corrector PS shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Cooling / pumping shall be yes02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The deflection shall be 2 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 1.2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply shall be 1600 A02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Y-chamber aperture shall be 36 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02May need to add additional window requirements for other leg of Lambertson magnet TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The gradient shall be 17 T/m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply shall be 1600 A02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 70 cm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The aperture radius shall be 50 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The number of kickers shall be 602/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Rise time shall be 900 ns02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Fall time shall be NA sec02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flat top time shall be 13 us02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The waveshape shall be trap02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The painting shall be vertical02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum field shall be 0.12 T02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The total deflection shall be 16 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum current shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum voltage shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The inductance with cable shall be TBD (uH)02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Max rep rate shall be 100 kV/pC02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flat top repeatability shall be NA Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The beam abort kicker shall be tbd %02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flatness of flat top/pulse form shall be 1 %02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The cooling type shall be w (W,A)02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The ESR Abort system shall contain a beam dump to safely absorb the energy of the stored beam in a controlled fashion.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The materials shall be C / Al / Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The energy deposited during abort shall be 320 kJ02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window thickness shall be tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The frequency of thermal cycle shall be 1 hour02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window material shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR shall have a collimation system capable of ensuring a sufficiently low background at the detector.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be vertical.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR injection absorber has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The thermal duty cycle shall be 2 Hz. 2 Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR detector absorbers shall be placed at Sector 5.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The detector absorbers have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be tbd kW. tbd kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 21 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 560-720 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator shall be horizontal.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be range shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary vertical collimator shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary horizontal collimator shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W on the tip of the jaw. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary vertical collimators shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary horizontal collimators shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR shall have a collimation system capable protecting all machine elements in case of failure.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be vertical.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR injection absorber has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The thermal duty cycle shall be 2 Hz. 2 Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR detector absorbers shall be placed at Sector 5.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The detector absorbers have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be tbd kW. tbd kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 21 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 560-720 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator shall be horizontal.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be range shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary vertical collimator shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary horizontal collimator shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W on the tip of the jaw. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary vertical collimators shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary horizontal collimators shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
- 6.02.02The ESR shall reach an availability consistent with the overall availability of the entire EIC as specified in the GRD. [Document#:EIC-SEG-RSI-010]02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.04.03.01.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).07/29/2025ApprovedFALSE
- 6.02.02.03.04.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).03/02/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed with components capable to withstand a lifetime radiation dose of 1 MGy.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).02/09/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet shall be designed with components capable to withstand a lifetime radiation dose of 1 MGy.02/09/2026ApprovedFALSE
- 6.02.02All ESR components and systems shall be designed and installed in line with all relevant regulatory codes and in full compliance with BNL SBMS.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR shall have provisions made to accommodate a control system which can operate the ESR consistent will the overall control of the EIC and other EIC systems and to ensure the ESR meets all the Physics requirements needed to deliver the physics goals of the EIC.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR control system shall facilitate all ESR global control requirements.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR control system shall facilitate all network, relational database and data archiving required.02/09/2026ApprovedFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.02The ESR RF Systems within the tunnel shall operate within its yearly radiation exposure budget.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.02.04.02The control system shall facilitate all machine protection systems required02/09/2026ApprovedFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR control system shall facilitate all EIC machine timing required.02/09/2026ApprovedFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR control system shall facilitate fast orbit feedback integration systems as required.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR control system shall facilitate all physics application support required.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include dual-plane Beam Position Monitors (BPMs) adjacent to each vertically focusing quadrupole. Provisions shall be made in the vacuum chamber design to install additional dual-plane BPMs at the horizontally focusing quadrupoles, if needed.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up shall provide dual plane (horizontal and vertical) beam positional measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The roll angle error of the ESR BPM pickup, given as the horizontal measurement plane defined by the BPM fiducials witrh respect to. the horizontal plane of the ESR ring shall be less than +/- 20 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall fulfill resolution and accuracy requirements over the following two defined transverse beam position ranges with respect vacuum chamber center, referenced to the mechanical fiducials on the BPM pickups.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The beam position monitor pick-up assembly shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02The ESR BPMs shall have turn-by-turn orbit measurement capability based on a single, remotely selectable bunch out of the fully filled bunch train to enable injection optimization.02/09/2026ApprovedFALSE
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following capabilities defined for the low intensity pilot injection energies and high intensity collision energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following time resolutions for data refresh defined for the beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following time resolutions for data logging defined for the beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following measurement resolutions defined for beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate reliability with capability to withstand a lifetime radiation dose of TBD MGy.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate in an ambient temperature degree from X (C) to X (C).02/09/2026ReviewedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up shall provide dual plane (horizontal and vertical) beam positional measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The roll angle error of the ESR BPM pickup, given as the horizontal measurement plane defined by the BPM fiducials witrh respect to. the horizontal plane of the ESR ring shall be less than +/- 20 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall fulfill resolution and accuracy requirements over the following two defined transverse beam position ranges with respect vacuum chamber center, referenced to the mechanical fiducials on the BPM pickups.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The beam position monitor pick-up assembly shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a beam current monitor to measure average beam current.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall have the ability to measure the average beam current over a range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall provide an average current measurement with a resolution of less than 5 (uA /√Hz).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measurement drift tolerance shall be less than 10 (uA) over 1 (hr).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy of better than +/- 1 (%) at 250 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy between the range of 250 (mA) to 2.5 (A) at +/- 0.5 (%).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer Calibration system shall be capable of providing an equivalent ESR DC current over the full operating range within 0.25% over the beam current range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall have a remote controlled self calibration system02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measured average current shall be archived at a rate of 1 Hz02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure individual bunch charges and bunch pattern.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring witness bunch for 1/e for a fixed gain over the beam lifetime.02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure transverse beam profiles.02/09/2026ApprovedFALSE
- 6.02.02.05.05.01The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure longitudinal beam profiles.02/09/2026ApprovedFALSE
- 6.02.02.05.05.02The requirements for longitudinal feedback are ??? TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.02The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02The ESR longitudinal bunch profile monitor needs turn-by-turn capability based on a single bunch in the fully filled bunch train to allow timing and energy adjustment for injection optimization.02/09/2026ApprovedFALSE
- 6.02.02.05.05.02The requirements for longitudinal feedback are ??? TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.02The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include system to measure H & V betatron tunes.02/09/2026ApprovedFALSE
- 6.02.02.05.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the horizontal kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the vertical kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The location of the tune meter kicker striplines in the ESR shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The kicker waveform (risetime and shape) requirements shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The impedance of the kicker beamline device shall be approved by beam Physics.02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall facilitate all required feedback systems (slow transverse, longitudinal and transverse bunch-by-bunch)02/09/2026ApprovedFALSE
- 6.02.04.02The slow orbit feedback correction output rate shall be 10 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The slow orbit feedback BPM data averaging period shall be tbd -02/09/2026In ProcessFALSE
- 6.02.02.05.05.01The transverse slow feedback system bandwidth shall bs 10 Hz02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include beam loss monitor system with detectors located only at select regions of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.05.02BLM shall be needed to needed to protect sensitive equipment.02/09/2026In ProcessFALSE
- 6.02.02.05.02The number of BLM installed in the ESR shall be TBD ea02/09/2026In ProcessFALSE
- 6.02.02.05.02BLM shall be installed at the following locations in the ESR TBD02/09/2026In ProcessFALSE
- 6.02.02.05.02The sensitivity of the BLM detectors shall be TBD units?02/09/2026In ProcessFALSE
- 6.02.02.05.02Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/09/2026In ProcessFALSE
- 6.02.02.05.02The response time from loss detection to abort shall be TBD us02/09/2026In ProcessFALSE
- 6.02.02The ESR RF System shall be designed to minimize unscheduled downtime, maintenance time and repair time to achieve ESR operational availability.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 300K to 150K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve steady state temperature with the cavity bath at 4K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve a full warm-up cycle from 4K to 295K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall operate through a minimum of 200 thermal cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner 1% range tuning cycles shall be 100,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner full range tuning cycles shall be 1,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The manufactured SRF Cryomodule Cavity shall produce no field emission at 4 MV.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.08.04.01All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.05/16/2025ApprovedFALSE
- 6.08.04.01The active SRF cavity tuning mechanism components (motor/gearbox/drive mechanism) shall be replaceable and maintainable in-situ.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity slow tuner tuning rate shall be 800 Hz/s.05/16/2025ApprovedFALSE
- 6.02.04.04The ESR shall have provisions made to accommodate a cryogenic system to cool and operate all elements which need cryogenic cooling .02/09/2026ApprovedFALSE
- 6.02.02The ESR shall have provisions made to accommodate a RF system capable of operating at parameters defined in MPT. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
ESR-ARC : Electron Storage Ring Arc Section (WBS 6.02.02)
- ESR-ARC EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02The ESR lattice arc magnet structure shall contain an array of regular FODO cells02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR lattice arc magnet structure shall accommodate slightly different average arc radii in the individual arcs by adjusting the drift spaces between individual elements in each FODO cell.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR super-bends shall generate additional synchrotron radiation damping to support a large beam-beam parameter of 0.1 and to create the required horizontal design emittance in the Master Parameter Table (MPT) when the ESR is operated at energies below 10 GeV. [Document: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The polarity of the ESR center bending magnet shall be capable of being wired in reverse to control the beam emittance and to damp the beam. The polarity will be dictated by the beam energy.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The FODO cell shall operate with a horizontal and vertical betatron phase advance of 60 degrees per arc section at beam energies of 10 GeV and below.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02The ESR Sextupole wiring scheme shall create the required sextupole families needed per arc to maximize dynamic aperture at the 60 degrees per FODO cell phase advance at < 10 GeV.02/09/2026ApprovedFALSE
- 6.02.02The FODO cell shall operate with a horizontal and vertical betatron phase advance of 90 degrees per arc section to maintain the required horizontal beam emittance defined in the MPT at 18 GeV. [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.02The ESR Sextupole wiring scheme shall create the required sextupole families needed per arc to maximize dynamic aperture at the 90 degrees per FODO cell phase advance at 18 GeV.02/09/2026ApprovedFALSE
- 6.02.02The vertical emittance shall be controlled by appropriate beam orbit manipulations and horizontal-vertical cross coupling.02/09/2026ApprovedFALSE
ESR-INJ : Electron Storage Ring Injection System (WBS 6.02.02)
- 6.02.02The ESR injection system shall inject a single bunch onto the closed orbit.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system shall be installed in IR4.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection point shall be upstream of IP4, where the ESR is installed along the outer tunnel wall.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system shall be capable of injecting one bunch per second.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system kickers shall deflect the incoming beam in the horizontal direction.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection kicker pulse form shall be such that fewer than 10 stored bunches receive a kick that leads to betatron oscillations of these bunches that correspond to less than 0.1 RMS transverse beam sizes.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system shall upon injection of a new bunch into a given bucket extract the spent stored bunch by the same kicker pulse.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system shall deflect replaced (extracted) bunches towards a dedicated beam dump.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system beam dump for replaced bunches (replacement dump) shall be internal to the beam pipe.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system replacement dump shall be capable of absorbing one bunch per second, with the bunch parameters listed in the Master Parameter Table. [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system shall be designed with a 50 percent safety margin for absorption of spent bunches into the replacement dump.02/09/2026ApprovedFALSE
- 6.02.02The ESR injection system shall have radiation shielding provided around the replacement dump.02/09/2026ApprovedFALSE
- 6.02.02There shall be no slow injection/extraction bump in the ESR.02/09/2026ApprovedFALSE
- 6.02.02Incoming bunches from the RCS shall have their transverse emittances matched to the equilibrium emittances in the ESR within 10 percent.02/09/2026ApprovedFALSE
- 6.02.02The energy, bunch length, and momentum spread of the incoming bunches from the RCS shall be matched to the ESR bucket.02/09/2026ApprovedFALSE
- 6.02.02The layout of the ESR injection and extraction beam lines shall minimize the required kick angle by taking advantage of the split yokes of the APS quadrupoles in the area.02/09/2026ApprovedFALSE
ESR-INJ-DUMP : Electron Storage Ring Injection Abort Beam Dump (WBS 6.04.04.03.01.02)
- 6.04.04.03.01.02The ESR replacement beam dump shall be located in IR4.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR replacement beam dump shall be located downstream of the ESR injection kicker.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR replacement beam dump shall be internal.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR replacement beam dump shall be based on the ESR injection kicker, which extracts a single bunch towards the replacement dump as it injects a fresh bunch into the ESR.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The spent ESR bunch shall be extracted in the horizontal plane.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR horizontal half aperture for the circulating beam at the location of the replacement beam dump shall correspond to at least 15 horizontal RMS beam sizes, based on the emittances defined in the Master Parameter Table (MPT), plus an additional 10 mm. [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR vertical half aperture for the circulating beam at the location of the replacement beam dump shall correspond to at least 15 vertical RMS beam sizes, assuming a fully coupled beam based on the emittances defined in the MPT, plus an additional 5 mm. [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR replacement beam dump shall be capable of safely absorbing one bunch per second, with the bunch parameters defined in the MPT. [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The material(s) of the ESR replacement beam dump shall be consistent with these requirements.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The geometry of the ESR replacement dump shall be consistent with the impedance budget requirements of the ESR.02/09/2026ApprovedFALSE
ESR-STRAIGHT : Electron Storage Ring Straight Section (WBS 6.02.02)
- ESR-STRAIGHT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02The phase advance of each straight section shall be tunable in order to optimize the dynamic aperture of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The field harmonic measurements shall be measured at the reference radius of 25mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The reference integrated field for the measurement shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR straight sections IR02, IR04, IR10 and IR12 shall be based on FODO cells.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The reference integrated field for the measurement shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02Ther ESR shall have matching sections at the ends of each of the straight sections to compensate for the different FODO cell lengths with respect to the arc FODO cells imposed by geometric constraints.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The maximum physical magnet length shall be 0.88 m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The maximum magnet axial dimensions shall be X = 45 cm and Y = 45 cm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
ESR-MAG : ESR Magnet (WBS 6.02.02.03)
- ESR-MAG EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02The ESR magnets shall meet the requirements defined by the physics lattice.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The ESR magnets shall have the required field quality to meet the operational needs.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02The “super-bends” in the ESR ARC sections shall consist of two long dipoles on either end of a short dipole.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The good field region of the ESR dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02All ESR quadrupoles shall be designed to facilitate beam based alignment.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02The maximum integrated field strength of the ESR FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.02The aperture of all ESR magnets shall be large enough to accommodate the ESR vacuum chamber.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 52 and 150 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet pole tip radius shall be 40 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 150 and 52 mm.02/09/2026ApprovedFALSE
- 6.02.02The ESR Shall have a conventional orbit corrector scheme, with single-plane correctors located at the respective quadrupoles. The strength of these correctors needs to be chosen to correct for any source of orbit distortion and should have enough margin for beam based diagnostic purposes, Harmonic Spin bumps and emittance generating bumps.02/09/2026ApprovedFALSE
- 6.02.02All dipoles in the ESR, including those in IRs 6 and 8, shall be connected in series to a single main power supply.02/09/2026ApprovedFALSE
- 6.02.02The ESR main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02The ESR quadrupoles in the straight sections IR02, IR04, IR10 and IR12 and in the transition from the arc to the straight section structure shall be wired to provide the optimized betatron phase advance across each straight section, as required for dynamic aperture optimization.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02The ESR quadrupoles shall have provisions to vary individual strengths by approximately 1% for beam-based alignment purposes.02/09/2026ApprovedFALSE
- 6.02.02The ESR sextupole power supply scheme shall be laid out such that the sextupole family structure can be configured for both the 60 and the 90 degree lattice along with a small number of individually powered sextupoles in the transition regions between arcs and straight sections with minimal effort, cost and minimizing any risk of error.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
ESR-MAG-Q:50 : ESR Quadrupole Magnet (Q50) (WBS 6.01.02.02)
- 6.01.02.02The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.02/09/2026ApprovedFALSE
- 6.01.02.02The physical magnet length shall be <0.5 m.02/09/2026ApprovedFALSE
- 6.01.02.02The effective magnet length shall be 0.5 m.02/09/2026ApprovedFALSE
- 6.01.02.02The pole tip radius of the magnet shall be 40 mm.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet integrated grad field G shall be 9.4 T.02/09/2026ApprovedFALSE
- 6.01.02.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.01.02.02The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).02/09/2026ApprovedFALSE
- 6.01.02.02The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet bore field shall require the following multipole content:02/09/2026ApprovedFALSE
- 6.01.02.02b2 = 10000 , a2 = N/A02/09/2026ApprovedFALSE
- 6.01.02.02b3 = HV +/- 2.2 , a3 = +/- 202/09/2026ApprovedFALSE
- 6.01.02.02b4 = HV +/- 2.4 , a4 = +/- 0.702/09/2026ApprovedFALSE
- 6.01.02.02b5 = HV +/- 1.0 , a5 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b6 = HV +/- 1.0 , a6 = +/- 0.202/09/2026ApprovedFALSE
- 6.01.02.02b7 = HV +/- 1.0 , a7 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b8 = HV +/- 1.0 , a8 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b9 = HV +/- 1.0 , a9 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b10 = HV +/- 1.0 , a10 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b11 = HV +/- 1.0 , a11 = +/- 0.402/09/2026ApprovedFALSE
- 6.01.02.02b12 = HV +/- 1.0 , a12 = +/- 0.302/09/2026ApprovedFALSE
- 6.01.02.02b13 = HV +/- 1.0 , a13 = +/- 0.202/09/2026ApprovedFALSE
- 6.01.02.02b14 = HV +/- 1.0 , a14 = +/- 0.1502/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).02/09/2026In ProcessFALSE
- 6.01.02.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.02/09/2026In ProcessFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-Q:50 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
ESR-MAG-Q:60 : ESR Quadrupole Magnet (Q60) (WBS 6.01.02.02)
- 6.01.02.02The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.02/09/2026ApprovedFALSE
- 6.01.02.02The physical magnet length shall be <0.6 m.02/09/2026ApprovedFALSE
- 6.01.02.02The effective magnet length shall be 0.6 m.02/09/2026ApprovedFALSE
- 6.01.02.02The pole tip radius of the magnet shall be 40 mm.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet integrated grad field G shall be 11.0 T.02/09/2026ApprovedFALSE
- 6.01.02.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.01.02.02The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).02/09/2026ApprovedFALSE
- 6.01.02.02The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet bore field shall require the following multipole content:02/09/2026ApprovedFALSE
- 6.01.02.02b2 = 10000 , a2 = N/A02/09/2026ApprovedFALSE
- 6.01.02.02b3 = HV +/- 2.2 , a3 = +/- 202/09/2026ApprovedFALSE
- 6.01.02.02b4 = HV +/- 2.4 , a4 = +/- 0.702/09/2026ApprovedFALSE
- 6.01.02.02b5 = HV +/- 1.0 , a5 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b6 = HV +/- 1.0 , a6 = +/- 0.202/09/2026ApprovedFALSE
- 6.01.02.02b7 = HV +/- 1.0 , a7 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b8 = HV +/- 1.0 , a8 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b9 = HV +/- 1.0 , a9 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b10 = HV +/- 1.0 , a10 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b11 = HV +/- 1.0 , a11 = +/- 0.402/09/2026ApprovedFALSE
- 6.01.02.02b12 = HV +/- 1.0 , a12 = +/- 0.302/09/2026ApprovedFALSE
- 6.01.02.02b13 = HV +/- 1.0 , a13 = +/- 0.202/09/2026ApprovedFALSE
- 6.01.02.02b14 = HV +/- 1.0 , a14 = +/- 0.1502/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).02/09/2026In ProcessFALSE
- 6.01.02.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.02/09/2026In ProcessFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-Q:60 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
ESR-MAG-Q:80 : ESR Quadrupole Magnet (Q80) (WBS 6.01.02.02)
- 6.01.02.02The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.02/09/2026ApprovedFALSE
- 6.01.02.02The physical magnet length shall be <0.8 m.02/09/2026ApprovedFALSE
- 6.01.02.02The effective magnet length shall be 0.8 m.02/09/2026ApprovedFALSE
- 6.01.02.02The pole tip radius of the magnet shall be 40 mm.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet integrated grad field G shall be 15.1 T.02/09/2026ApprovedFALSE
- 6.01.02.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.01.02.02The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).02/09/2026ApprovedFALSE
- 6.01.02.02The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet bore field shall require the following multipole content:02/09/2026ApprovedFALSE
- 6.01.02.02b2 = 10000 , a2 = N/A02/09/2026ApprovedFALSE
- 6.01.02.02b3 = HV +/- 2.2 , a3 = +/- 202/09/2026ApprovedFALSE
- 6.01.02.02b4 = HV +/- 2.4 , a4 = +/- 0.702/09/2026ApprovedFALSE
- 6.01.02.02b5 = HV +/- 1.0 , a5 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b6 = HV +/- 1.0 , a6 = +/- 0.202/09/2026ApprovedFALSE
- 6.01.02.02b7 = HV +/- 1.0 , a7 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b8 = HV +/- 1.0 , a8 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b9 = HV +/- 1.0 , a9 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b10 = HV +/- 1.0 , a10 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b11 = HV +/- 1.0 , a11 = +/- 0.402/09/2026ApprovedFALSE
- 6.01.02.02b12 = HV +/- 1.0 , a12 = +/- 0.302/09/2026ApprovedFALSE
- 6.01.02.02b13 = HV +/- 1.0 , a13 = +/- 0.202/09/2026ApprovedFALSE
- 6.01.02.02b14 = HV +/- 1.0 , a14 = +/- 0.1502/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).02/09/2026In ProcessFALSE
- 6.01.02.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.02/09/2026In ProcessFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-Q:80 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
ESR-MAG-SX:24 : ESR Sextupole Magnet (SX) (WBS 6.01.02.02)
- 6.01.02.02The magnet shall be a single function sextupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to have a splitable pole toaccommodate the Vacuum beam pipe installation.02/09/2026ApprovedFALSE
- 6.01.02.02The physical magnet length shall be <0.32 m.02/09/2026ApprovedFALSE
- 6.01.02.02The effective magnet length shall be 0.24 m02/09/2026ApprovedFALSE
- 6.01.02.02The magnet install center and install alignment must be within atranslational value of +/-150(um) and a rotational alignment valueof +/-0.5 (mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet integrated grad field G shall be 97.2 T/m.02/09/2026ApprovedFALSE
- 6.01.02.02The magnetic field, center and alignment, within the magnet mustbe known to within a translational value of +/-50 (um) and arotational alignment value of +/-0.5 (mrad).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.01.02.02The harmonic reference radius and current shall be 25 (mm) and 161 (A).02/09/2026ApprovedFALSE
- 6.01.02.02The Field at the reference radius and current shall be 405 (T/m^2).02/09/2026ApprovedFALSE
- 6.01.02.02The magnet bore field shall require the following multipole content:02/09/2026ApprovedFALSE
- 6.01.02.02b3 = 10000 a3 = N/A02/09/2026ApprovedFALSE
- 6.01.02.02b4 = HV +/- 5.3 a4 = HV +/- 7.102/09/2026ApprovedFALSE
- 6.01.02.02b5 = HV +/- 1.6 a5 = HV +/- 1.902/09/2026ApprovedFALSE
- 6.01.02.02b6 = HV +/- 1.0 a6 = +/- 2.002/09/2026ApprovedFALSE
- 6.01.02.02b7 = HV +/- 1.0 a7 = +/- 0.802/09/2026ApprovedFALSE
- 6.01.02.02b8 = HV +/- 1.0 a8 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b9 = HV +/- 1.0 a9 = +/- 0.502/09/2026ApprovedFALSE
- 6.01.02.02b10 = HV +/- 1.0 a10 = +/- 0.302/09/2026ApprovedFALSE
- 6.01.02.02b11 = HV +/- 1.0 a11 = +/- 0.2802/09/2026ApprovedFALSE
- 6.01.02.02b12 = HV +/- 1.0 a12 = +/- 0.2602/09/2026ApprovedFALSE
- 6.01.02.02b13 = HV +/- 1.0 a13 = +/- 0.2402/09/2026ApprovedFALSE
- 6.01.02.02b14 = HV +/- 1.0 a14 = +/- 0.2202/09/2026ApprovedFALSE
- 6.01.02.02b15 = HV +/- 1.0 a15 = +/- 0.2102/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).02/09/2026In ProcessFALSE
- 6.01.02.02The magnet design and verification process shall ensure the finalmagnet will meet the reliability needs of the EIC over it plannedoperational life of >20 Years.02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to operate reliably given thecumulative radiation dose of TBD Rads it will experience over thelifetime of the EIC of >20 Years.02/09/2026In ProcessFALSE
- 6.02.02.03.02Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the LCW connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.02The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.02The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.02The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-TD:273 : ESR Dipole Magnet (D13) (WBS 6.02.02.03.01.01)
- 6.02.02.03.01.01nan02/09/2026ApprovedFALSE
- 6.02.02.03.01.01nan02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.01.01nan02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be measured using the following multipole homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints:02/09/2026ApprovedFALSE
- 6.02.02.03.01.01nan02/09/2026ApprovedFALSE
- 6.02.02.03.01Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall purchase the cables that go from the power supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide trim coils.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.01/08/2026Not ApplicableFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/08/2026Not ApplicableFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.01The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders.03/02/2026ReviewedFALSE
- 6.02.02.03.01The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-TD:273 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
ESR-MAG-TD:89 : ESR Dipole Magnet (D2) (WBS 6.02.02.03.01.02)
- 6.02.02.03.01.02nan02/09/2026ApprovedFALSE
- 6.02.02.03.01.02nan02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.01.02nan02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be measured using the following multipole homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints:02/09/2026ApprovedFALSE
- 6.02.02.03.01.02nan02/09/2026ApprovedFALSE
- 6.02.02.03.01Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall purchase the cables that go from the power supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide trim coils.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.01/08/2026Not ApplicableFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/08/2026Not ApplicableFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.01The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders.03/02/2026ReviewedFALSE
- 6.02.02.03.01The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-TD:89 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
ESR-MAG-TD:380 : ESR Spin Rotator Dipole Magnet (DROT) (WBS 6.02.02.03.01.03)
- 6.02.02.03.01.03The magnet shall be a single function dipole with a vertical field.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet shall require trim coils capable of trimming the field within +/- TBD (%) of the Peak field.02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet shall require current taps for operation TBD ( Y or N)02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The physical magnet length shall be <3.8 m.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet model length shall be 3.8 m.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet bore gap and Width shall be 52 mm.02/09/2026ReviewedFALSE
- 6.02.02.03.01.03The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.02.02.03.01.03Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet integrated dipole field (B) shall be 1.2 T.m.02/09/2026ReviewedFALSE
- 6.02.02.03.01.03The magnet good field aperture dAx required shall be 32.59 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet good field aperture dAy required shall be 11.46 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet to magnet field variability between magnets shall be 1 %.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The field shall be measured at two locations as follows: Harmonic Measurements region 1; Rref1=TBD mm centered at (-TBD,TBD) mm, Harmonic Measurements region 2; Rref2=TBD mm centered at (TBD,TBD) mm.02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The reference field for the different measurements shall be: Measurement 1; Bref1=TBD (T) in Region TBD, Region TBD , Measurement 2; Bref2=TBD (T) in Region TBD, Region TBD.02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet bore field shall have a field homogeneity in region TBD, of better than dB/B <10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1 & 2.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: b1 = 10000, Region 2: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region1: -4<b2<4, Region2: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.02.02.03.01.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall purchase the cables that go from the power supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide trim coils.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.01/08/2026In ProcessFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/08/2026In ProcessFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.01The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders.03/02/2026ReviewedFALSE
- 6.02.02.03.01The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-TD:550 : ESR Dipole Magnet (B2ER) (WBS 6.02.02.03.01.03)
- 6.02.02.03.01.03The magnet shall be a single function magnet with a vertical dipole field along the beam axis.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026On HoldFALSE
- 6.02.02.03.01.03The physical magnet length shall be <TBD m.02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet bore gap and Width shall be 48 mm.02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet shall be designed to fit within the following envelope:02/09/2026On HoldFALSE
- 6.02.02.03.01.03Magnet installation volume tolerances TBD02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet integrated dipole field (B) shall be TBD T.m.02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet good field aperture dAx required shall be TBD mm.02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet good field aperture dAy required shall be TBD mm.02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet to magnet field variability between magnets shall be 5 %.02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026On HoldFALSE
- 6.02.02.03.01.03The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/09/2026On HoldFALSE
- 6.02.02.03.01.03The Field at the reference radius at the design energy of 18 GeV shall be TBD (T) .02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet bore field shall require the following multipole content:02/09/2026On HoldFALSE
- 6.02.02.03.01.03b1 = 10000 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b2 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b3 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b4 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b5 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b6 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b7 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b8 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b9 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b10 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b11 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b12 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b13 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b14 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b15 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03b16 < 100 (10^-4)02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026On HoldFALSE
- 6.02.02.03.01.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall purchase the cables that go from the power supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide trim coils.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide several current taps to facilitate running at different operational modes.01/08/2026Not ApplicableFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/08/2026Not ApplicableFALSE
- 6.02.02.03.01The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.01The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.01The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.01The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders.03/02/2026ReviewedFALSE
- 6.02.02.03.01The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.01ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-SXT:57 : ESR Long Sextupole Magnet (SXL) (WBS 6.02.02.03.02)
- 6.02.02.03.02The magnet shall be a single function sextupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.02The maximum physical magnet length shall be 0.63 m.02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet pole tip radius shall be 49 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet shall be designed to fit within the following envelope.02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet maximum axial dimensions shall be X = 71.1 cm and Y = 92.3 cm.02/09/2026In ProcessFALSE
- 6.02.02.03.02The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet integrated gradient field, G, shall be 230.9 T/m.02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet-to-magnet field variability between magnets shall be less than 0.1%.02/09/2026ReviewedFALSE
- 6.02.02.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet reference radius and current shall be 25 mm and 161 A.02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnetic field at the reference radius and current shall be 405 T/m^2.02/09/2026ApprovedFALSE
- 6.02.02.03.02The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.02b3 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.02-5.00 < b4 < 5.00 -7.00 < a4 < 7.0002/09/2026ApprovedFALSE
- 6.02.02.03.02-1.40 < b5 < 1.40, -1.70 < a5 < 1.7002/09/2026ApprovedFALSE
- 6.02.02.03.02-1.00 < b6 < 1.00, -1.00 < a6 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.02-0.75 < b7 < 0.75, -0.40 < a7 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.02-0.50 < b8 < 0.50, -0.30 < a8 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.02-5.00 < b9 < 5.00, -0.30 < a9 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.02-0.35 < b10 < 0.35, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.02-0.30 < b11 < 0.30, -0.15 < a11 < 0.1502/09/2026ApprovedFALSE
- 6.02.02.03.02-0.30 < b12 < 0.30, -0.15 < a12 < 0.1502/09/2026ApprovedFALSE
- 6.02.02.03.02-0.25 < b13 < 0.25, -0.10 < a13 < 0.1002/09/2026ApprovedFALSE
- 6.02.02.03.02-0.20 < b14 < 0.20, -0.10 < a14 < 0.1002/09/2026ApprovedFALSE
- 6.02.02.03.02-1.00 < b15 < 1.00, -0.10 < a15 < 0.1002/09/2026ApprovedFALSE
- 6.02.02.03.02-0.20 < b16 < 0.20, -0.10 < a16 < 0.1002/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).02/09/2026In ProcessFALSE
- 6.02.02.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/12/2026In ProcessFALSE
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) to be utilized by the magnets.01/12/2026In ProcessFALSE
- The magnet group shall provide all distribution design of the piping (or hoses) for Low Conductivity Water (LCW) from the tunnel header to the Cryomodules to be utilized by the magnets.01/12/2026In ProcessFALSE
- The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/12/2026In ProcessFALSE
- The Accelerator Installation WBS CAM shall schedule and fund cable installation from the power supply to the magnets by the appropriate technical support group.01/12/2026In ProcessFALSE
- The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/12/2026In ProcessFALSE
- The Power Supply Group shall make connections of the main winding magnet power cables at the magnet terminals and power supply ends.01/12/2026In ProcessFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.01/12/2026In ProcessFALSE
- The Power Supply Group shall provide a power supply to be utilized by the magnets.01/12/2026In ProcessFALSE
- The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM.01/12/2026In ProcessFALSE
- The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/12/2026In ProcessFALSE
- The Accelerator Installation WBS CAM shall schedule and fund installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support group.01/12/2026In ProcessFALSE
- The Power Supply Group shall purchase the cables that go from the thermal switch on the magnet to the Power Supply.01/12/2026In ProcessFALSE
- The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends.01/12/2026In ProcessFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s).01/12/2026In ProcessFALSE
- The Magnet Group shall provide the terminal block with labeling to facilitate installation of the connection for thermal protection.01/12/2026In ProcessFALSE
- The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/12/2026In ProcessFALSE
- The Vacuum Group shall design the beampipe integration to the magnets.01/12/2026In ProcessFALSE
- The Accelerator Installation WBS CAM shall schedule and fund installing the beampipe through the magnets and installing the magnets by the appropriate technical support group.01/12/2026In ProcessFALSE
- The Vacuum Group will provide temporary thermal couples to the magnets for beamline bakeout. The magnet and coil temperature shall not exceed 150 degrees F during beamline bakeout.01/12/2026In ProcessFALSE
- The Magnet Group shall provide analysis of potential thermally conductive paths between the magnet and beampipe.01/12/2026In ProcessFALSE
- The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe.01/12/2026In ProcessFALSE
- The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/12/2026In ProcessFALSE
- The Survey Group shall fiducialize the magnet based on magnetic center after they have been installed onto the girder.01/12/2026In ProcessFALSE
- The Magnet Group shall provide magnet design CAD models to the Mechanical Design Group to facilitate the design of the magnet girder.01/12/2026In ProcessFALSE
- The Accelerator Installation WBS CAM shall schedule and fund installing the final assembled magnet girder to the tunnel by the appropriate technical support group.01/12/2026In ProcessFALSE
- The Mechanical Design Group shall model the tunnel to define the required spacial location of the magnets and girders.01/12/2026In ProcessFALSE
- The Survey Group will provide the arrangement location for the magnets and girders.01/12/2026In ProcessFALSE
- The Magnet Group shall provide analysis and approval of the magnet installations and potential interferences to the Mechanical Design Group.01/12/2026In ProcessFALSE
ESR-MAG-QN:80 : ESR Narrow Quadrupole Magnet (QN) (WBS 6.02.02.03.04.01)
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-QN:80 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.03.04.01The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to have a splitable yoke.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The maximum physical magnet length shall be 0.88 m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet pole tip radius shall be 40 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The maximum magnet axial dimensions shall be X = 45 cm and Y = 45 cm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet install center and install alignment must be within a translational value of +/- 150 (um) and a rotational alignment value of +/- 0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet-to-magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 50 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The field harmonic measurements shall be measured at the reference radius of 25mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The reference integrated field for the measurement shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The integrated field over the length of the magnet at the reference radius shall meet the following harmonic multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.01b2=10000, a2 = N/A02/09/2026ApprovedFALSE
- 6.02.02.03.04.01-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-0.30 < b14 < 0.30, -0.02 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.01-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.04.03.01.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).07/29/2025ApprovedFALSE
- 6.02.02.03.04.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
ESR-MAG-QLA:135 : ESR Large Aperture Quadrupole Magnet (Q_LA) (WBS 6.02.02.03.04.02)
- 6.02.02.03.04.02The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet shall be require shunt resistors for beam-based alignment, TBD A at TBD GeV.02/09/2026Not ApplicableFALSE
- 6.02.02.03.04.02The magnet shall be designed to have a splitable yoke to accommodate the Vacuum beam pipe installation.02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The maximum physical magnet length shall be 1.35(m).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet pole tip radius shall be 104 (mm).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The maximum magnet axial dimensions shall be X = 90 (cm) and Y = 80 (cm).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet integrated grad field G shall be 9.7 (T).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet to magnet field variability between magnets shall be less than 0.25%.02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The field harmonic measurements shall be measured at the reference radius of 25mm.02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The reference field for the measurement shall be 9.7 (T).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.02.02.03.04.02b2=1000002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-2.00 < b3 < 2.00, -1.00 < a3 < 1.0002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.75 < b4 < 1.75, -0.47 < a4 < 0.4602/09/2026ApprovedFALSE
- 6.02.02.03.04.02-0.50 < b5 < 0.50, -0.30 < a5 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.75 < b6 < 1.75, -0.20 < a6 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b7 < 1.00, -0.50 < a7 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b8 < 1.00, -0.50 < a8 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b9 < 1.00, -0.50 < a9 < 0.5002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b10 < 1.00, -0.20 < a10 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b11 < 1.00, -0.40 < a11 < 0.4002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b12 < 1.00, -0.30 < a12 < 0.3002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b13 < 1.00, -0.20 < a13 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-0.30 < b14 < 0.30, -0.20 < a14 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b15 < 1.00, -0.20 < a15 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.02-1.00 < b16 < 1.00, -0.20 < a16 < 0.2002/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet coils shall have a maximum current leak of 10 uA during Hi-Pot test at nominal operating conditions corresponding to 1 (kV).02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04.02The magnet shall be designed with components capable to withstand a lifetime radiation dose of 1 MGy.02/09/2026ApprovedFALSE
- 6.04.03.02Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) to be utilized by the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The magnet group shall provide all distribution design of the piping (or hoses) for Low Conductivity Water (LCW) from the tunnel header to the Cryomodules to be utilized by the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund cable installation from the power supply to the magnets by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall make connections of the main winding magnet power cables at the magnet terminals and power supply ends.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall provide a power supply to be utilized by the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall purchase the cables that go from the thermal switch on the magnet to the Power Supply.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s).01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide the terminal block with labeling to facilitate installation of the connection for thermal protection.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Vacuum Group shall design the beampipe integration to the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing the beampipe through the magnets and installing the magnets by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Vacuum Group will provide temporary thermal couples to the magnets for beamline bakeout. The magnet and coil temperature shall not exceed 150 degrees F during beamline bakeout.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide analysis of potential thermally conductive paths between the magnet and beampipe.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe.01/08/2026Not ApplicableFALSE
- 6.04.03.02The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Survey Group shall fiducialize the magnet based on magnetic center after they have been installed onto the girder.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide magnet design CAD models to the Mechanical Design Group to facilitate the design of the magnet girder.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing the final assembled magnet girder to the tunnel by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Mechanical Design Group shall model the tunnel to define the required spacial location of the magnets and girders.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Survey Group will provide the arrangement location for the magnets and girders.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide analysis and approval of the magnet installations and potential interferences to the Mechanical Design Group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Mechanical Design Group shall provide the spacial limitation between the beampipe and the magnet pole after all radiation shielding has been added.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Accelerator Installation WBS CAM01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Accelerator Installation WBS CAM01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/08/2026Not ApplicableFALSE
ESR-MAG-TH:20 : ESR Horizontal Corrector Magnet (TH) (WBS 6.02.02.03.05.01)
- 6.02.02.03.05.01The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet axial dimensions shall be X = 43.2 cm and Y = 43.2 cm.03/02/2026ApprovedFALSE
- 6.02.02.03.05.0102/09/2026Not ApplicableFALSE
- 6.02.02.03.05.01All skew multipole terms (a1 through a16) shall be within ± 1.03/02/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet should not be designed to limit CrossTalk requirements.02/09/2026Not ApplicableFALSE
- 6.02.02.03.05.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet axis.02/09/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet coils shall pass a Hi-Pot test at nominal operating conditions corresponding to TBD (V).02/09/2026In ProcessFALSE
- 6.02.02.03.05The Magnet Group shall provide the requirements to the Infrastructure Group that defines the building and utilities along with associated heatload analyses for the ESR Correctors.03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location with sufficiently cooled space within the existing RHIC tunnel that satisfies the requirements of the ESR corrector magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.05The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.05The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-TH:20 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 52 and 150 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet reference radius for field homogeneity shall be 17 (mm).03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet integrated field strength variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within reference radius at half and full field excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.007 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.01b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b2 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-15 < b3 < 1503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b5 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b7 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b9 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b11 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b13 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b15 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < b16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).03/02/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed with components capable to withstand a lifetime radiation dose of 1 MGy.02/09/2026ApprovedFALSE
ESR-MAG-TV:20 : ESR Vertical Corrector Magnet (TV) (WBS 6.02.02.03.05.01)
- 6.02.02.03.05.01The magnet shall be designed to fit within the following envelope:02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet axial dimensions shall be X = 61 cm and Y = 43.2 cm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.0102/09/2026Not ApplicableFALSE
- 6.02.02.03.05.01The magnet should not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet axis.02/09/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet coils shall pass a Hi-Pot test at nominal operating conditions corresponding to TBD (V).02/09/2026In ProcessFALSE
- 6.02.02.03.05The Magnet Group shall provide the requirements to the Infrastructure Group that defines the building and utilities along with associated heatload analyses for the ESR Correctors.03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location with sufficiently cooled space within the existing RHIC tunnel that satisfies the requirements of the ESR corrector magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.05The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.05The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
- ESR-MAG-TV:20 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.03.05.01The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum physical magnet length shall be 0.2 m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The minimal magnet bore gap and width shall be 150 and 52 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet reference radius for homogeneity shall be 17 (mm).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The maximum magnet-to-magnet variability shall be 0.5%.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 500 (um) and a rotational alignment value of +/-0.3 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet field quality shall be measured within the reference radius at half and full excitation.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01a1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a2 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a16 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a3 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a4 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a5 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a6 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a7 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a8 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a9 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a10 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a11 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a12 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a13 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-5 < a14 < 502/09/2026ApprovedFALSE
- 6.02.02.03.05.01-25 < a15 < 2502/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).02/09/2026In ProcessFALSE
- 6.02.02.03.05.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet shall be designed with components capable to withstand a lifetime radiation dose of 1 MGy.02/09/2026ApprovedFALSE
ESR-MAG-THLA:50 : ESR IR10 Horizontal Corrector Magnet (TH_LA) (WBS 6.02.02.03.05.02)
- 6.02.02.03.05.02The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The maximum physical magnet length shall be 0.6 m.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The minimal magnet bore gap and width shall be 208 and 208 (mm).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed to fit within the following envelope:03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The maximum magnet axial dimensions shall be X = 45 (cm) and Y = 63 (cm).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet install center and install alignment must be within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet integrated dipole field (B) shall be 12 (mT-m).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet reference radius for field homogeneity shall be 17 (mm).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The maximum magnet-to-magnet variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-500 (um) and a rotational alignment value of +/-0.3 (mrad).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet field quality shall be measured within the reference radius at half and full field excitation.03/02/2026ApprovedFALSE
- 6.02.02.03.05.0203/02/2026Not ApplicableFALSE
- 6.02.02.03.05.02The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02b1 = 1000003/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b2 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < b3 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b4 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < b5 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b6 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < b7 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b8 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < b9 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b10 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < b11 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b12 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < b13 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b14 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < b15 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < b16 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed with components capable to withstand a lifetime radiation dose of 1 MGy.03/02/2026ApprovedFALSE
- 6.02.02.03.05The Magnet Group shall provide the requirements to the Infrastructure Group that defines the building and utilities along with associated heatload analyses for the ESR Correctors.03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location with sufficiently cooled space within the existing RHIC tunnel that satisfies the requirements of the ESR corrector magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.05The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.05The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-TVLA:50 : ESR IR10 Vertical Corrector Magnet (TV_LA) (WBS 6.02.02.03.05.02)
- 6.02.02.03.05.02The magnet shall operate with a single functions with a Dipole field type and Horizontal field direction.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The maximum physical magnet length shall be 0.6 m.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The minimal magnet bore gap and width shall be 208 and 208 (mm).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed to fit within the following envelope:03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The maximum magnet axial dimensions shall be X = 63 (cm) and Y = 45 (cm).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet install center and install alignment must be within a translational value of +/- 500(um) and a rotational alignment value of +/-0.3 (mrad).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet integrated dipole field (B) shall be 12 (mT-m).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet reference radius for homoegeneity shall be 17 (mm).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The maximum magnet-to-magnet variability shall be 0.5%.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- 500 (um) and a rotational alignment value of +/-0.3 (mrad).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet field quality shall be measured within the reference radius at half and full excitation.03/02/2026ApprovedFALSE
- 6.02.02.03.05.0203/02/2026Not ApplicableFALSE
- 6.02.02.03.05.02The integrated field over the length of the magnet shall have a variation with respect to transverse offset of less than |dB|/B = 0.003 and shall meet the following multipole content requirements.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02a1 = 1000003/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a2 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < a3 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a4 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < a5 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a6 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < a7 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a8 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < a9 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a10 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < a11 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a12 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < a13 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a14 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-25 < a125 < 2503/02/2026ApprovedFALSE
- 6.02.02.03.05.02-5 < a16 < 503/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +15 (C) to +50 (C).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet axis.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.03/02/2026ApprovedFALSE
- 6.02.02.03.05.02The magnet shall be designed with components capable to withstand a lifetime radiation dose of 1 MGy.03/02/2026ApprovedFALSE
- 6.02.02.03.05The Magnet Group shall provide the requirements to the Infrastructure Group that defines the building and utilities along with associated heatload analyses for the ESR Correctors.03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location with sufficiently cooled space within the existing RHIC tunnel that satisfies the requirements of the ESR corrector magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.05The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.05The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.05The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.05The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.05ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-QS:25 : ESR Skew Quadrupole Magnet (QS) (WBS 6.04.03.03.02.01)
- 6.04.03.03.02.01The magnet shall be a single function quadrupole with a skew field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The physical magnet length shall be <0.25 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.01The magnet model length shall be 0.25 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.01The magnet pole tip radius shall be 67.9 mm.02/09/2026ReviewedFALSE
- 6.04.03.03.02.01The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.01Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The magnet integrated grad field G shall be 1.5 T/m.02/09/2026ReviewedFALSE
- 6.04.03.03.02.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The field at the reference radius at the design energy of 18 GeV shall be TBD (T/m) .02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The magnet bore field shall require the following multipole content:02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b1 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b2 = 10000 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b3 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b4 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b5 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b6 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b7 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b8 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b9 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b10 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b11 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b12 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b13 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b14 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b15 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01b16 < 1 (10^-4)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04Magnet protection is required to be provided from a power supply in the tunnel to be utilized to safely operate the normal conducting QS:25 magnet within the ESR.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-Q:120 : ESR Quadrupole Magnet (Q120) (WBS 6.04.03.03.02.02)
- 6.04.03.03.02.02The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation.02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The physical magnet length shall be <1.2 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.02The magnet pole tip radius shall be 40 mm.02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.02Magnet installation tolerances, Max X = 42 cm May Y = 42 cm02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet integrated grad field G shall be 18.9 T/m.02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The field harmonic measurements shall be measured at Rref=25mm.02/09/2026ApprovedFALSE
- 6.04.03.03.02.02The reference field for the measurement shall be TBD (T/m).02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02b2=1000002/09/2026ApprovedFALSE
- 6.04.03.03.02.02-2.71<b3<0.72, -2.13<a3<2.1102/09/2026ApprovedFALSE
- 6.04.03.03.02.02-3.58<b4<1.2, -0.47<a4<0.4602/09/2026ApprovedFALSE
- 6.04.03.03.02.02-0.53<b5<0.23, -0.29<a5<0.3502/09/2026ApprovedFALSE
- 6.04.03.03.02.02-1.75<b6<-1.12, -0.07<a6<0.0802/09/2026ApprovedFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02-1.06<b10<-0.66, 0<a10<002/09/2026ApprovedFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02-0.12<b14<-0.08, 0<a14<002/09/2026ApprovedFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet shall be designed to meet the following fringe field requirements The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet center.02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet shall be designed to be cooled and sustained at an operational temperature range of +X (C) to +X (C).02/09/2026In ProcessFALSE
- 6.04.03.03.02.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
ESR-MAG-QROT:114 : ESR Quadrupole Magnet (QSS3) (WBS 6.04.03.03.02.03)
- 6.04.03.03.02.03The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The physical magnet length shall be <1.14 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet pole tip radius shall be 67.9 mm.02/09/2026ReviewedFALSE
- 6.04.03.03.02.03The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet integrated grad field G shall be 24 T/m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The field harmonic measurements shall be measured at Rref=25mm.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The reference field for the measurement shall be TBD (T/m).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03b2=1000002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-2.71<b3<0.72, -2.13<a3<2.1102/09/2026ApprovedFALSE
- 6.04.03.03.02.03-3.58<b4<1.2, -0.47<a4<0.4602/09/2026ApprovedFALSE
- 6.04.03.03.02.03-0.53<b5<0.23, -0.29<a5<0.3502/09/2026ApprovedFALSE
- 6.04.03.03.02.03-1.75<b6<-1.12, -0.07<a6<0.0802/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-1.06<b10<-0.66, 0<a10<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-0.12<b14<-0.08, 0<a14<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-QROT:122 : ESR Quadrupole Magnet (QSS1) (WBS 6.04.03.03.02.03)
- 6.04.03.03.02.03The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The physical magnet length shall be <1.22 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet pole tip radius shall be 67.9 mm.02/09/2026ReviewedFALSE
- 6.04.03.03.02.03The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet integrated grad field G shall be 24 T/m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The field harmonic measurements shall be measured at Rref=25mm.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The reference field for the measurement shall be TBD (T/m).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03b2=1000002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-2.71<b3<0.72, -2.13<a3<2.1102/09/2026ApprovedFALSE
- 6.04.03.03.02.03-3.58<b4<1.2, -0.47<a4<0.4602/09/2026ApprovedFALSE
- 6.04.03.03.02.03-0.53<b5<0.23, -0.29<a5<0.3502/09/2026ApprovedFALSE
- 6.04.03.03.02.03-1.75<b6<-1.12, -0.07<a6<0.0802/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-1.06<b10<-0.66, 0<a10<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-0.12<b14<-0.08, 0<a14<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-QROT:138 : ESR Quadrupole Magnet (QSS4) (WBS 6.04.03.03.02.03)
- 6.04.03.03.02.03The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The physical magnet length shall be <1.38 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet pole tip radius shall be 67.9 mm.02/09/2026ReviewedFALSE
- 6.04.03.03.02.03The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet integrated grad field G shall be 24 T/m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The field harmonic measurements shall be measured at Rref=25mm.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The reference field for the measurement shall be TBD (T/m).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03b2=1000002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-2.71<b3<0.72, -2.13<a3<2.1102/09/2026ApprovedFALSE
- 6.04.03.03.02.03-3.58<b4<1.2, -0.47<a4<0.4602/09/2026ApprovedFALSE
- 6.04.03.03.02.03-0.53<b5<0.23, -0.29<a5<0.3502/09/2026ApprovedFALSE
- 6.04.03.03.02.03-1.75<b6<-1.12, -0.07<a6<0.0802/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-1.06<b10<-0.66, 0<a10<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-0.12<b14<-0.08, 0<a14<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-QROT:188 : ESR Quadrupole Magnet (QSS5) (WBS 6.04.03.03.02.03)
- 6.04.03.03.02.03The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The physical magnet length shall be <1.88 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet pole tip radius shall be 67.9 mm.02/09/2026ReviewedFALSE
- 6.04.03.03.02.03The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet integrated grad field G shall be 24 T/m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The field harmonic measurements shall be measured at Rref=25mm.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The reference field for the measurement shall be TBD (T/m).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03b2=1000002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-2.71<b3<0.72, -2.13<a3<2.1102/09/2026ApprovedFALSE
- 6.04.03.03.02.03-3.58<b4<1.2, -0.47<a4<0.4602/09/2026ApprovedFALSE
- 6.04.03.03.02.03-0.53<b5<0.23, -0.29<a5<0.3502/09/2026ApprovedFALSE
- 6.04.03.03.02.03-1.75<b6<-1.12, -0.07<a6<0.0802/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-1.06<b10<-0.66, 0<a10<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-0.12<b14<-0.08, 0<a14<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-QROT:197 : ESR Quadrupole Magnet (QSS2) (WBS 6.04.03.03.02.03)
- 6.04.03.03.02.03The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The physical magnet length shall be <1.97 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet pole tip radius shall be 67.9 mm.02/09/2026ReviewedFALSE
- 6.04.03.03.02.03The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet integrated grad field G shall be 24 T/m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The field harmonic measurements shall be measured at Rref=25mm.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The reference field for the measurement shall be TBD (T/m).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03b2=1000002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-2.71<b3<0.72, -2.13<a3<2.1102/09/2026ApprovedFALSE
- 6.04.03.03.02.03-3.58<b4<1.2, -0.47<a4<0.4602/09/2026ApprovedFALSE
- 6.04.03.03.02.03-0.53<b5<0.23, -0.29<a5<0.3502/09/2026ApprovedFALSE
- 6.04.03.03.02.03-1.75<b6<-1.12, -0.07<a6<0.0802/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-1.06<b10<-0.66, 0<a10<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-0.12<b14<-0.08, 0<a14<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-QROT:87 : ESR Quadrupole Magnet (QLS3) (WBS 6.04.03.03.02.03)
- 6.04.03.03.02.03The magnet shall be single function quadrupole with a normal field rotation.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The physical magnet length shall be <0.87 m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet pole tip radius shall be 67.9 mm.02/09/2026ReviewedFALSE
- 6.04.03.03.02.03The magnet shall be designed to fit within the following envelope. TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03Magnet installation tolerances TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet integrated grad field G shall be 24 T/m.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet to magnet field variability between magnets shall be less than 0.1%.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The field harmonic measurements shall be measured at Rref=25mm.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The reference field for the measurement shall be TBD (T/m).02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03b2=1000002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-2.71<b3<0.72, -2.13<a3<2.1102/09/2026ApprovedFALSE
- 6.04.03.03.02.03-3.58<b4<1.2, -0.47<a4<0.4602/09/2026ApprovedFALSE
- 6.04.03.03.02.03-0.53<b5<0.23, -0.29<a5<0.3502/09/2026ApprovedFALSE
- 6.04.03.03.02.03-1.75<b6<-1.12, -0.07<a6<0.0802/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-1.06<b10<-0.66, 0<a10<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-0.12<b14<-0.08, 0<a14<002/09/2026ApprovedFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03-, -02/09/2026Not ApplicableFALSE
- 6.04.03.03.02.03The magnet shall not be designed to limit CrossTalk requirements.02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet shall be designed to meet the following fringe field requirements TBD02/09/2026In ProcessFALSE
- 6.04.03.03.02.03The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.04.03.03.02.03The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-GIRDER
- 6.02.02.03.04Reference related Vacuum interface document for the interface details. (I-ESR-VAC-XXX.XXX)03/02/2026ApprovedFALSE
- 6.02.02.03.06Reference related Vacuum interface document for the interface details. (I-ESR-VAC-XXX.XXX)03/02/2026ApprovedFALSE
- 6.02.02.03.08Reference related Magnet interface document for the interface details. (I-ESR-MAG-XXX.XXX))03/02/2026ApprovedFALSE
- 6.02.02.03.10Reference related Power Supply interface document for the interface details. (I-ESR-PS-XXX.XXX)03/02/2026ApprovedFALSE
- 6.02.02.03.11Reference related Magnet interface document for the interface details. (I-ESR-MAG-XXX.XXX)03/02/2026ApprovedFALSE
- 6.02.02.03.13Reference related Magnet interface document for the interface details. (I-ESR-MAG-XXX.XXX)03/02/2026ApprovedFALSE
- 6.02.02.03.15Reference related Beam Instrumentation interface document for the interface details. (I-ESR-INST-XXX.XXX)03/02/2026ApprovedFALSE
- 6.02.02.03.17Reference related Beam Instrumentation interface document for the interface details. (I-ESR-INST-XXX.XXX)03/02/2026ApprovedFALSE
- 6.02.02.03.19Machine Protection Group shall provide the design for the Synchrotron Radiation Shielding including mounting provisions.03/02/2026ApprovedFALSE
- 6.02.02.03.20Magnet Girder Group shall incorporate mounting provisions into the integrated girder design to accommodate the synchrotron radiation shielding.03/02/2026ApprovedFALSE
- 6.02.02.03.21ASR System Installation and Final Integration shall provide funding and scheduling for the installation of synchrotron radiation shielding on the integrated girder.03/02/2026ApprovedFALSE
- 6.02.02.03.23Magnet Girder Group shall provide the integrated girder design with tunnel mounting provisions.03/02/2026ApprovedFALSE
- 6.02.02.03.25The Physics Group shall provide the allocated spatial location based on the lattice for the integrated girder.03/02/2026ApprovedFALSE
- 6.02.02.03.26The Mechanical Design Group shall provide the volume within the tunnel to be occupied by the integrated girder and CAD model to ensure the design does not interfere with any components in the tunnel.03/02/2026ApprovedFALSE
- 6.02.02.03.27The Magnet Girder Group shall have fiducial marks added to the integrated girder allowing the center of the girder in respect to the beamlines to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ApprovedFALSE
- 6.02.02.03.28ASR System Installation and Final Integration shall provide funding and scheduling to provide tunnel survey details to ensure the integrated girder can be aligned in the tunnel.03/02/2026ApprovedFALSE
- 6.02.02.03.29ASR System Installation and Final Integration shall provide funding and scheduling to align and install the integrated girder such that the fiducial marks are located with respect to the install locations provided by the survey group to ensure the integrated girder is aligned.03/02/2026ApprovedFALSE
ESR-MAG-Q120
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-QLA:120
- 6.02.02.03.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall provide a power supply to be utilized by the magnets03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Systems Integration and Final Installation Group03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.04The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.04The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.04The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.04The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.04ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-SX:57
- 6.02.02.03.02Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW).03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the LCW connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall provide a power supply to be utilized by the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Systems Integration and Final Installation Group.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the electrical connections to the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection03/02/2026ReviewedFALSE
- 6.02.02.03.02The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe03/02/2026ReviewedFALSE
- 6.02.02.03.02The mechanical design group shall have fiducial marks added to the magnet allowing the center of beamline to be located. It should identify the location of the magnet center and its alignment with respect to the beamline.03/02/2026ReviewedFALSE
- 6.02.02.03.02The BNL Magnet Group shall procure the girders for the Systems Integration and Final Installation Group to install.03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the magnets on the girders by the appropriate technical support groups such that the fiducial marks on the magnet are located with respect to the install locations provided by the survey group to ensure the magnet is aligned..03/02/2026ReviewedFALSE
- 6.02.02.03.02The Vacuum Group shall design the beampipe integration to the magnets.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Vacuum Group will provide temporary thermal couples to the magnets and incorporate the defined thermal design limits for the magnet into its bakeout procedure for the ESR beamline which satisfies the Magnet Group design.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Magnet Group will provide supports for beam pipes.03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.03/02/2026ReviewedFALSE
- 6.02.02.03.02The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.02The ESR Physics Group will provide the arrangement location for the magnets and girders03/02/2026ReviewedFALSE
- 6.02.02.03.02ASR System Installation and Final Integration shall provide funding and scheduling for surveying and installing the magnet fiducials by the appropriate technical support groups.03/02/2026ReviewedFALSE
ESR-MAG-THLA:60a
- 6.04.03.02The Magnet Group shall provide the requirements to the Infrastructure Group that defines the building and utilities along with associated heatload analyses for the ESR Correctors.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Infrastructure group shall provide a sufficiently cooled space within the existing RHIC tunnel that satisfies the requirements of the ESR corrector magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund cable installation from the power supply to the magnets by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall make connections of the main winding magnet power cables at the magnet terminals and power supply ends.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall provide a power supply to be utilized by the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall purchase the cables that go from the thermal switch on the magnet to the Power Supply.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s).01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide the terminal block with labeling to facilitate installation of the connection for thermal protection.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Vacuum Group shall design the beampipe integration to the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing the beampipe through the magnets and installing the magnets by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Vacuum Group will provide temporary thermal couples to the magnets for beamline bakeout. The magnet and coil temperature shall not exceed 150 degrees F during beamline bakeout.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide analysis of potential thermally conductive paths between the magnet and beampipe.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe.01/08/2026Not ApplicableFALSE
- 6.04.03.02The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Survey Group shall fiducialize the magnet based on magnetic center after they have been installed onto the girder.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide magnet design CAD models to the Mechanical Design Group to facilitate the design of the magnet girder.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing the final assembled magnet girder to the tunnel by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Mechanical Design Group shall model the tunnel to define the required spacial location of the magnets and girders.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Survey Group will provide the arrangement location for the magnets and girders.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide analysis and approval of the magnet installations and potential interferences to the Mechanical Design Group.01/08/2026Not ApplicableFALSE
ESR-MAG-TVLA:60a
- 6.04.03.02The Magnet Group shall provide the requirements to the Infrastructure Group that defines the building and utilities along with associated heatload analyses for the ESR Correctors.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Infrastructure group shall provide a sufficiently cooled space within the existing RHIC tunnel that satisfies the requirements of the ESR corrector magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund cable installation from the power supply to the magnets by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall make connections of the main winding magnet power cables at the magnet terminals and power supply ends.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall provide a power supply to be utilized by the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a terminal block with standard polarity labels to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall purchase the cables that go from the thermal switch on the magnet to the Power Supply.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s).01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide the terminal block with labeling to facilitate installation of the connection for thermal protection.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide a electrical and thermal parameters to the Power Supply Group to be utilized in selecting necessary power supply needs.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Vacuum Group shall design the beampipe integration to the magnets.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing the beampipe through the magnets and installing the magnets by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Vacuum Group will provide temporary thermal couples to the magnets for beamline bakeout. The magnet and coil temperature shall not exceed 150 degrees F during beamline bakeout.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide analysis of potential thermally conductive paths between the magnet and beampipe.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe.01/08/2026Not ApplicableFALSE
- 6.04.03.02The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Survey Group shall fiducialize the magnet based on magnetic center after they have been installed onto the girder.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide magnet design CAD models to the Mechanical Design Group to facilitate the design of the magnet girder.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Accelerator Installation WBS CAM shall schedule and fund installing the final assembled magnet girder to the tunnel by the appropriate technical support group.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Mechanical Design Group shall model the tunnel to define the required spacial location of the magnets and girders.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Survey Group will provide the arrangement location for the magnets and girders.01/08/2026Not ApplicableFALSE
- 6.04.03.02The Magnet Group shall provide analysis and approval of the magnet installations and potential interferences to the Mechanical Design Group.01/08/2026Not ApplicableFALSE
ESR-PS : ESR Magnet Power Supply (WBS 6.02.02.04)
- ESR-PS EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02.03.04.01The magnet integrated gradient field, G, shall be 11.3 T.02/09/2026ApprovedFALSE
- 6.02.02.03.05.01The magnet integrated dipole field (B) shall be 12 milli T-m.02/09/2026ApprovedFALSE
- 6.02.02The ESR magnet power supplies shall be capable of providing the stability the ESR needs to operate02/09/2026ApprovedFALSE
ESR-PS-D : ESR Main Dipole Magnet (D) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-td:380, esr-td:273, esr-td:89 and esr-td:550.02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 1600 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply during a fault condition shall limit the maximum voltage of the magnet-to-ground to less than 2500 (V) for a period of 1 minute. The magnet should have sufficient design margin for this level of voltage:02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply synchronization timing of synchronization shall be 100 (us).02/09/2026Not ApplicableFALSE
- 6.02.02.04The main dipole power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply shall limit current ripple (RMS) to 10 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply shall limit ripple to 1000 (ppm) in the 1 k(Hz)–8 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply shall limit ripple in the 8 k(Hz)–40 k(Hz) range per the specified by figure in P-ESR-PS-main dipole.09.04.01.02/09/2026ApprovedFALSE
- 6.02.02.04nan02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole power supply shall have a tunable switching frequency over a tuning range of +/- 500 (Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The main dipole rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TD:89.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TD:273.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TD:380.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TD:550.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-Q:120 : ESR Quadrupole Magnet (Q:120) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-q:120.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QL:120 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:120 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-Q:120.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-Q:50 : ESR Quadrupole Magnet (Q:50) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-q:50.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:500 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:50 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-Q:50.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-Q:60 : ESR Quadrupole Magnet (Q:60) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-q:60.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:60 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-Q:60.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-Q:80 : ESR Quadrupole Magnet (Q:80) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-q:80.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-Q:80.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QLA:120 : ESR Quadrupole Magnet (QLA:120) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qla:120.02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QLA:120 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QLA:120.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QN:80 : ESR Narrow Quadrupole Magnet (QN:80) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-q:80.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QN:80 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QN:80 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QN:80 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QN:80 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QN:80 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QN:80 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The Q:80 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QN:80.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QROT:114 : ESR Quadrupole Spin Rotator Magnet (QROT:114) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qrot:114.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:114 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QROT:114.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QROT:122 : ESR Quadrupole Spin Rotator Magnet (QROT:122) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qrot:122.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:122 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QROT:122.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QROT:138 : ESR Quadrupole Spin Rotator Magnet (QROT:138) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qrot:138.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:138 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QROT:138.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QROT:188 : ESR Quadrupole Spin Rotator Magnet (QROT:188) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qrot:188.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:188 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QROT:188.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QROT:197 : ESR Quadrupole Spin Rotator Magnet (QROT:197) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qrot:197.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:197 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QROT:197.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QROT:87 : ESR Quadrupole Spin Rotator Magnet (QROT:87) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qrot:87.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QROT:87 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QROT:87.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-QS:25 : ESR Quadrupole Magnet (QS:25) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-qs:25.02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply for all series connected Quadrupoles BBA will be achieved by a circuit connected across each quad magnet that will divert a percentage of the operational current defined in the technical magnet documentation. [BNL-228833-2025-TECH and EIC-ADD-TN-134]02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply for individually powered quadrupoles shall be able to transition from 95% of its operational current to its operational current at 1 (Hz) without exceeding the operational setpoint current for beam-based alignment.02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 10 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply shall limit current ripple (RMS) to 15 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 power supply shall limit current ripple (RMS) to 500 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The QS:25 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-QS:25.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-SX:24 : ESR Sextupole Magnet (SX:57) Power Supply (WBS 6.02.02.04)
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including allocated space for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel . (I-ESR-MAG-SX:24.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-SX:57 : ESR Sextupole Magnet (SX:114) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-sx:57.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 100 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply shall provide a minimal current setpoint resolution of 16 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply shall limit current ripple (RMS) to 40 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 power supply shall limit current ripple (RMS) to 10000 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:57 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 1.7 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-SX:57.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-TD:273 : ESR Dipole Magnet (TD:273) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-td:273.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 1600 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 50 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:237 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply shall limit current ripple (RMS) to 10 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply shall limit ripple to 1000 (ppm) in the 1 k(Hz)–8 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:273 power supply shall limit ripple in the 8 k(Hz)–40 k(Hz) range per the specified by figure in P-ESR-PS-TD:273.09.04.01.02/09/2026ApprovedFALSE
- 6.02.02.04nan02/09/2026ApprovedFALSE
- 6.02.02.04The TD:237 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TD:273.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-TD:380 : ESR Dipole Magnet (TD:380) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-td:380.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 1600 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 50 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply shall limit current ripple (RMS) to 10 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply shall limit ripple to 1000 (ppm) in the 1 k(Hz)–8 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 power supply shall limit ripple in the 8 k(Hz)–40 k(Hz) range per the specified by figure in P-ESR-PS-TD:380.09.04.01.02/09/2026ApprovedFALSE
- 6.02.02.04nan02/09/2026ApprovedFALSE
- 6.02.02.04The TD:380 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for Electrical Power interface details inside of tunnel. (I-ESR-MAG-TD:380.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-TD:550 : ESR Dipole Magnet (TD:550) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-td:550.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply during operation shall include a ground fault protection system to limit the maximum voltage of the magnet-to-ground to less than 1600 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 50 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply shall limit current ripple (RMS) to 10 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply shall limit ripple to 1000 (ppm) in the 1 k(Hz)–8 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 power supply shall limit ripple in the 8 k(Hz)–40 k(Hz) range per the specified by figure in P-ESR-PS-TD:550.09.04.01.02/09/2026ApprovedFALSE
- 6.02.02.04nan02/09/2026ApprovedFALSE
- 6.02.02.04The TD:550 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for Electrical Power interface details inside of tunnel. (I-ESR-MAG-TD:550.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-TD:89 : ESR Dipole Magnet (TD:89) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-td:89.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 1600 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 50 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply shall provide a minimal current setpoint resolution of 18 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply shall limit current ripple (RMS) to 10 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply shall limit ripple to 1000 (ppm) in the 1 k(Hz)–8 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 power supply shall limit ripple in the 8 k(Hz)–40 k(Hz) range per the specified by figure in P-ESR-PS-TD:89.09.04.01.02/09/2026ApprovedFALSE
- 6.02.02.04nan02/09/2026ApprovedFALSE
- 6.02.02.04The TD:89 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 0.3 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TD:89.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-TH:20 : ESR Corrector Magnet (TH:20) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-th:20.02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply shall be capable of compensating slow orbit motion with a 1/f spectrum, the required magnet current to be equivalent to +/- 5mA at 10 Hz02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 100 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply shall provide a minimal current setpoint resolution of 16 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply shall limit current ripple (RMS) to 100 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 power supply shall limit current ripple (RMS) to 10000 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The TH:20 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 1.7 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TH:20.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-THLA:50 : SR Large Aperture Corrector Magnet (THLA:50) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-thla:60.03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply shall provide a DC current.03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply shall be capable of compensating slow orbit motion with a 1/f spectrum, the required magnet current to be equivalent to +/- 5mA at 10 Hz03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 100 (ppm).03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply shall provide a minimal current setpoint resolution of 16 (Bit).03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply synchronization timing of synchronization shall be 100 (us).03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply shall limit current ripple (RMS) to 100 (ppm) of full-scale current in the 0–1 k(Hz) range.03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 power supply shall limit current ripple (RMS) to 10000 (ppm) of full scale current greater than 1k(Hz).03/02/2026ApprovedFALSE
- 6.02.02.04The THLA:60 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 1.7 (C).03/02/2026ApprovedFALSE
ESR-PS-TV:20 : ESR Corrector Magnet (TV:20) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-tv:20.02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply shall be capable of compensating slow orbit motion with a 1/f spectrum, the required magnet current to be equivalent to +/- 5mA at 10 Hz02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 100 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply shall provide a minimal current setpoint resolution of 16 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply shall limit current ripple (RMS) to 100 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 power supply shall limit current ripple (RMS) to 10000 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The TV:20 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 1.7 (C).02/09/2026ApprovedFALSE
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TV:20.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-TVLA:50 : ESR Large Aperture Corrector Magnet (TVLA:50) Power Supply (WBS 6.02.02.04)
- 6.02.02.04The magnet model being powered is esr-tvla:60.03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply shall provide a DC current.03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply shall be capable of compensating slow orbit motion with a 1/f spectrum, the required magnet current to be equivalent to +/- 5mA at 10 Hz03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 100 (ppm).03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply shall provide a minimal current setpoint resolution of 16 (Bit).03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply synchronization timing of synchronization shall be 100 (us).03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply shall limit current ripple (RMS) to 100 (ppm) of full-scale current in the 0–1 k(Hz) range.03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 power supply shall limit current ripple (RMS) to 10000 (ppm) of full scale current greater than 1k(Hz).03/02/2026ApprovedFALSE
- 6.02.02.04The TVLA:60 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 1.7 (C).03/02/2026ApprovedFALSE
ESR-PS-SX24
- 6.02.02.04The magnet model being powered is esr-sx:24.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply shall meet all requirements to deliver the magnet operational parameters defined in the technical magnet documentation. [Document#: EIC-SEG-RSI-110]02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply during operation shall limit the maximum voltage of the magnet-to-ground to less than 300 (V).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply shall provide a DC current.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply long-term stability (1 second to 10 hours) at maximum operating current shall be 100 (ppm).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply shall provide a minimal current setpoint resolution of 16 (Bit).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply synchronization timing of synchronization shall be 100 (us).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply setpoint and all of the PS analog readbacks shall be synchronized to the line to reduce noise.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply shall limit current ripple (RMS) to 40 (ppm) of full-scale current in the 0–1 k(Hz) range.02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 power supply shall limit current ripple (RMS) to 10000 (ppm) of full scale current greater than 1k(Hz).02/09/2026ApprovedFALSE
- 6.02.02.04The SX:24 rack mounted power supply and its controls rack shall be designed to be cooled and sustained at an operational temperature range of +23.9 (C) +/- 1.7 (C).02/09/2026ApprovedFALSE
ESR-PS-THLA:60
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-THLA:60.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-PS-TVLA:60
- 6.02.02.04Power Supply Group shall provide design details that defines the building, rack layouts and power and low conductivity water utility requirements for power supplies and its associated subsystems.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the power supplies and its subcomponents which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure Group shall provide indoor enviromental control for the power supply buildings which satisfies the power supply equipment operating parameters.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the power supply rack and its subsystems design, including the spatial location, thermal and weight details.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the plan and funding for the procurement of the racks to house power supplies and its sub components which satisfies power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling to move and install power supply racks by the appropriate technical support group and its subsystems which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of power supplies into the racks by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each power supply rack and the freestanding equipment.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each power supply rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Infustructure mechanical cooling shall provide the LCW piping design and procurement of materials from the wall distribution to connections to the power supply DC Busses which satisfies the power supply group design, including waterflow switches and one contact per switch.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each power supply DC Buss by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design and funding for procurement for each power supply UPS, including AC cable distribution from the wall mount distribution to each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable distribution by the appropriate technical support group which satisfies the power supply group design into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall shall provide funding and scheduling for the installation of UPS by the appropriate technical support group into each power supply rack.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the design details defining the DC power cable routing from the power supply and its associated subsystems into the tunnels.01/08/2026ReviewedFALSE
- 6.02.02.04Infrastructure group shall provide tunnel penetrations for DC cable and bus work which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04ASR System Installation and Final Integration shall provide funding and scheduling for installation of the AC cable from the each power supply in the building to its corresponding magnet, including termination by the appropriate technical support group which satisfies the power supply group design.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related magnet interface document for magnet electrical power interface details related to the connection inside of tunnel. (I-ESR-MAG-TVLA:60.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide an "ON" status output through the MPS interface which satisfies MPS design.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall define the design and location of the magnet interlock PLCs for normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for the Machine protection interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide personal protection system interlock interface and connection pinouts.01/08/2026ReviewedFALSE
- 6.02.02.04Reference related Personnel Protection System (PSS) interface document for the PSS interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the zPSCs for all EIC Power Supplies.01/08/2026ReviewedFALSE
- 6.02.02.04Power Supply Group shall provide the magnet interlock PLCs and Modbus for all EIC normal conducting magnet(s).01/08/2026ReviewedFALSE
- 6.02.02.04Reference related controls interface document for Controls interface details. (I-ESR-CNTRL-XXX.XXX)01/08/2026ReviewedFALSE
ESR-INST : ESR Instrumentation System (WBS 6.02.02.05)
- ESR-INST EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02The ESR instrumentation system shall include dual-plane Beam Position Monitors (BPMs) adjacent to each vertically focusing quadrupole. Provisions shall be made in the vacuum chamber design to install additional dual-plane BPMs at the horizontally focusing quadrupoles, if needed.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up shall provide dual plane (horizontal and vertical) beam positional measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The roll angle error of the ESR BPM pickup, given as the horizontal measurement plane defined by the BPM fiducials witrh respect to. the horizontal plane of the ESR ring shall be less than +/- 20 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall fulfill resolution and accuracy requirements over the following two defined transverse beam position ranges with respect vacuum chamber center, referenced to the mechanical fiducials on the BPM pickups.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The beam position monitor pick-up assembly shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02The ESR BPMs shall have turn-by-turn orbit measurement capability based on a single, remotely selectable bunch out of the fully filled bunch train to enable injection optimization.02/09/2026ApprovedFALSE
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following capabilities defined for the low intensity pilot injection energies and high intensity collision energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following time resolutions for data refresh defined for the beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following time resolutions for data logging defined for the beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following measurement resolutions defined for beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate reliability with capability to withstand a lifetime radiation dose of TBD MGy.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate in an ambient temperature degree from X (C) to X (C).02/09/2026ReviewedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up shall provide dual plane (horizontal and vertical) beam positional measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The roll angle error of the ESR BPM pickup, given as the horizontal measurement plane defined by the BPM fiducials witrh respect to. the horizontal plane of the ESR ring shall be less than +/- 20 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall fulfill resolution and accuracy requirements over the following two defined transverse beam position ranges with respect vacuum chamber center, referenced to the mechanical fiducials on the BPM pickups.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The beam position monitor pick-up assembly shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a beam current monitor to measure average beam current.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall have the ability to measure the average beam current over a range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall provide an average current measurement with a resolution of less than 5 (uA /√Hz).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measurement drift tolerance shall be less than 10 (uA) over 1 (hr).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy of better than +/- 1 (%) at 250 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy between the range of 250 (mA) to 2.5 (A) at +/- 0.5 (%).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer Calibration system shall be capable of providing an equivalent ESR DC current over the full operating range within 0.25% over the beam current range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall have a remote controlled self calibration system02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measured average current shall be archived at a rate of 1 Hz02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure individual bunch charges and bunch pattern.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring witness bunch for 1/e for a fixed gain over the beam lifetime.02/09/2026ApprovedFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure transverse beam profiles.02/09/2026ApprovedFALSE
- 6.02.02.05.05.01The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include a system to measure longitudinal beam profiles.02/09/2026ApprovedFALSE
- 6.02.02.05.05.02The requirements for longitudinal feedback are ??? TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.02The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02The ESR longitudinal bunch profile monitor needs turn-by-turn capability based on a single bunch in the fully filled bunch train to allow timing and energy adjustment for injection optimization.02/09/2026ApprovedFALSE
- 6.02.02.05.05.02The requirements for longitudinal feedback are ??? TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.02The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include system to measure H & V betatron tunes.02/09/2026ApprovedFALSE
- 6.02.02.05.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the horizontal kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the vertical kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The location of the tune meter kicker striplines in the ESR shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The kicker waveform (risetime and shape) requirements shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The impedance of the kicker beamline device shall be approved by beam Physics.02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall facilitate all required feedback systems (slow transverse, longitudinal and transverse bunch-by-bunch)02/09/2026ApprovedFALSE
- 6.02.04.02The slow orbit feedback correction output rate shall be 10 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The slow orbit feedback BPM data averaging period shall be tbd -02/09/2026In ProcessFALSE
- 6.02.02.05.05.01The transverse slow feedback system bandwidth shall bs 10 Hz02/09/2026In ProcessFALSE
- 6.02.02The ESR instrumentation system shall include beam loss monitor system with detectors located only at select regions of the ESR.02/09/2026ApprovedFALSE
- 6.02.02.05.02BLM shall be needed to needed to protect sensitive equipment.02/09/2026In ProcessFALSE
- 6.02.02.05.02The number of BLM installed in the ESR shall be TBD ea02/09/2026In ProcessFALSE
- 6.02.02.05.02BLM shall be installed at the following locations in the ESR TBD02/09/2026In ProcessFALSE
- 6.02.02.05.02The sensitivity of the BLM detectors shall be TBD units?02/09/2026In ProcessFALSE
- 6.02.02.05.02Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/09/2026In ProcessFALSE
- 6.02.02.05.02The response time from loss detection to abort shall be TBD us02/09/2026In ProcessFALSE
ESR-INST-BPM : ESR Instrumentation Beam Postion Monitor System (WBS 6.02.02.05.01)
ESR-INST-BPM-ELEC : ESR Instrumentation Beam Postion Monitor (WBS 6.02.02.05.01.01)
- 6.02.02.05.01.01The ESR BPM Electronics shall provide a turn-by-turn measurement which is defined as a range of 1 (us) to the revolution period of bunches combined.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics for the first 20 BPMs after injection shall be capable of measuring individual bunch positions to minimize betatron oscillations of the newly injected bunches.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of delivering an array of at least 1024 consecutive single-turn position measurements at a continuous rate of 1 (Hz).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of delivering average beam orbit measurements at a continuous rate of 1 (Hz).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of delivering bunch-by-bunch beam orbit measurements at a continuous rate of 1 (Hz).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of bunch lifetime (1/e) measurements for the 2 non-colliding bunches during the 2.5 minute bunch duration.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of delivering fast orbit feedback measurements at a continuous rate of 1 (kHz).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of delivering slow orbit feedback measurements at a continuous rate of 1 (Hz).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of logging an array of at least TBD consecutive single-turn position measurements at a continuous rate of TBD Hz.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of logging average beam orbit measurements at a continuous rate of TBD Hz.02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of logging bunch-by-bunch beam orbit measurements at a continuous rate of TBD Hz.02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of logging bunch lifetime (1/e) measurements during the XXX minute bunch duration.02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of logging fast orbit feedback measurements at a continuous rate of TBD Hz.02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The ESR BPM electronics shall be capable of logging slow orbit feedback measurements at a continuous rate of TBD Hz.02/09/2026In ProcessFALSE
- 6.02.02.05.01.01For pilot bunches the ESR BPM electronics shall have the following measurement resolutions defined for beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a measurement range grater than or equal to 2 (nC).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a single turn measurement horizontal and vertical resolution greater than or equal to (100um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an average orbit measurement resolution over 1 second greater than or equal to (30um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an reproducibility from run to run greater than or equal to (30um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM electronics measurement drift due to thermal variations (0.5hrs) shall be no greater than 50 (µm).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01For newly injected low charge refill bunches the ESR BPM electronics shall have the following measurement resolutions defined for beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a measurement range grater than or equal to 2 (nC).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a single turn measurement horizontal and vertical resolution resolution greater than or equal to (H=50 V=10 um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an average orbit measurement resolution over 1 second greater than or equal to TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an reproducibility from run to run greater than or equal to TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM electronics measurement drift due to thermal variations (0.5hrs) shall be no greater than 10 (µm).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01For newly injected high charge refill bunches the ESR BPM electronics shall have the following measurement resolutions defined for beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a measurement range between 7 to 28 (nC).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a single turn measurement horizontal and vertical resolution resolution greater than or equal to (H=50 V=10 um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an average orbit measurement resolution over 1 second greater than or equal to TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an reproducibility from run to run greater than or equal to TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM electronics measurement drift due to thermal variations (0.5hrs) shall be no greater than 5 (µm).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01For stored beam low charge refill bunches the ESR BPM electronics shall have the following measurement resolutions defined for beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a measurement range grater than or equal to 2 (nC).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a single turn measurement horizontal and vertical resolution resolution greater than or equal to (30um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an average orbit measurement resolution over 1 second greater than or equal to TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an reproducibility from run to run greater than or equal to TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM electronics measurement drift due to thermal variations (0.5hrs) shall be no greater than 30 (µm).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01For stored beam high charge refill bunches the ESR BPM electronics shall have the following measurement resolutions defined for beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a measurement range between 7 to 28 (nC).02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The BPM Electronics shall have a single turn measurement horizontal and vertical resolution resolution greater than or equal to (10 um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an average orbit measurement resolution over 1 second greater than or equal to (5um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM Electronics shall have an reproducibility from run to run greater than or equal to (5um) TBD (RMS).02/09/2026In ProcessFALSE
- 6.02.02.05.01.01The BPM electronics measurement drift due to thermal variations (0.5hrs) shall be no greater than 10 (µm).02/09/2026ReviewedFALSE
- 6.04.06.01Instrumentation Group shall provide design details that defines the building, rack layouts and power and LCW utilities for beam position monitor electronics and its associated subsystems.09/24/2025In ProcessFALSE
- 6.04.06.01Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the beam position monitor electronics and its subcomponents which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.01Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the beam position monitor electronics and its subcomponents which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.01Infrastructure Group shall provide funding and scheduling for the installation of the LCW utilities for distribution to the beam position monitor electronics and its subcomponents which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.01Infrastructure Group shall provide indoor enviromental control for the beam position monitor electronic buildings which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.01Instrumentation Group shall provide the beam position monitor electronic rack and its subsystems design, including the spacial location, thermal and weight details.09/24/2025In ProcessFALSE
- 6.04.06.01Instrumentation Group shall provide the plan and funding for the procurement of the racks to house beam position monitor electronics and its sub components which satisfies the design.09/24/2025In ProcessFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling to move and install beam position monitor electronic racks by the appropriate technical support group and its subsystems which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling for the installation of beam position monitor electronics into the racks by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.01Instrumentation Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each beam position monitor electronic rack.09/24/2025In ProcessFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling for the design and AC cable tray to contain the AC cable distribution from the wall mount distribution to each beam position monitor electronic rack.01/08/2026In ProcessFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable and AC cable tray from the wall mount distribution to each beam position monitor electronic rack by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.01Instrumentation Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.09/24/2025In ProcessFALSE
- 6.04.06.01Infrastructure Group shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling for the LCW piping design and procurement of materials from the wall distribution to connections to the beam position monitor electronic cooling loops which satisfies the Instrumentation Group design, including waterflow switches and one contact per switch.01/08/2026In ProcessFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling for the installation of LCW piping from the wall mount distribution to each beam position monitor electronic DC Buss by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.01Instrumentation Group shall provide the design and funding for procurement for each beam position monitor electronic UPS, including AC cable distribution from the wall mount distribution to each beam position monitor electronic rack.09/24/2025In ProcessFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling for the installation of UPS and AC cable distribution cabling by the appropriate technical support group which satisfies the Instrumentation Group design into each beam position monitor electronic rack.01/08/2026In ProcessFALSE
- 6.04.06.01Instrumentation Group shall provide the design details defining the signal cable routing from the beam position monitor electronic and its associated subsystems into the tunnels.09/24/2025In ProcessFALSE
- 6.04.06.01Infrastructure group shall provide tunnel penetrations for signal cables which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.01Reference related beam position monitor interface document for signal cable termination iterface details inside of tunnel. (I-ESR-INST-BPM-PU.XX)09/24/2025In ProcessFALSE
- 6.04.06.01Machine Protection System (MPS) Group shall define the design details including input connections and data required to monitor beam position monitor electronic status.09/24/2025In ProcessFALSE
- 6.04.06.01Instrumentation Group shall provide output from the MPS interface the data which satisfies MPS design.09/24/2025In ProcessFALSE
- 6.04.06.01Reference related controls interface document for Controls interface signal details. (I-ESR-CNTRL-XXX.XX)09/24/2025In ProcessFALSE
- ESR-INST-BPM-ELEC EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following capabilities defined for the low intensity pilot injection energies and high intensity collision energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR Beam Position Monitor (BPM) Electronics shall have the following time resolutions for data refresh defined for the beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following time resolutions for data logging defined for the beam energies:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall have the following measurement resolutions defined for beam energies as defined in the Master Parameter Table: [EIC Document: EIC-SEG-RSI-005]:02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate reliability with capability to withstand a lifetime radiation dose of TBD MGy.02/09/2026ReviewedFALSE
- 6.02.02.05.01.01The ESR BPM Electronics shall be designed to operate in an ambient temperature degree from X (C) to X (C).02/09/2026ReviewedFALSE
ESR-INST-BPM-PU : ESR Instrumentation Beam Postion Monitor Pick-up (WBS 6.02.02.05.01.02)
- 6.02.02.05.01.02The first ESR 20 BPMs downstream of injection shall have the capability to measure the position of injected bunches, in the presence of already circulating bunches that are spaced ~10.15 (ns) apart.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall have single-turn / turn-by-turn trajectory (TbT) measurements at a rate of 12 (us) measurement (averaging) time with turns separated based on the 1 (μs) abort gap.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall have Closed Orbit (CO) measurements for beam orbit feedback and for day-to-day orbit correction purposes, polarization optimization, trouble shooting, etc.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall have Closed Orbit (CO) measurements averaging over n= TBD1 Turns or TBD1 (s).02/09/2026In ProcessFALSE
- 6.02.02.05.01.02The ESR BPMs shall have +/- 3 (mm) horizontal and vertical beam position range in the 80 (mm) ×36 (mm) chamber for position measurements in luminosity (LUMI) and fill (FILL) operations, and for Beam-Based Alignment (BBA) measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPMs shall have +/- 10 (mm) horizontal and vertical beam position range in the 80 (mm) ×36 (mm) chamber for position measurements in machine commissioning (COMM) operations.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The resolution, RMS, short term (ns, µs time scales) of the ESR BPMs shall be ≤ 100 (μm) for position measurements in single-bunch / bunch-by-bunch (BbB) and single-turn / turn-by-turn trajectory (TbT) modes during machine commissioning (COMM) operation.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The resolution, RMS, short term (ns, µs time scales) of the first 20 ESR BPMs after injection shall be ≤ 15 (μm) for position measurements in single-bunch / bunch-by-bunch (BbB) during ESR luminosity (LUMI) operation.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The resolution, RMS, short term (ns, µs time scales) of the ESR BPMs shall be ≤ 30 (μm) for position measurements in single-turn / turn-by-turn trajectory (TbT) modes in ESR luminosity (LUMI) operation.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR resolution, RMS, short term (ns, µs time scales) of the ESR BPMs shall be ≤ 5 (μm) for position measurements in Closed Orbit (CO) mode in ESR luminosity (LUMI) operation.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The resolution, RMS, short term (ns, µs time scales) of the ESR BPMs shall be ≤ 2 (μm) for position measurements in Closed Orbit (CO) mode when performing ESR Beam Based Alignment (BBA) measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The allowable measurement drift of the ESR BPMs shall be ≤ +/-1 (μm) in Closed Orbit (CO) mode in ESR luminosity (LUMI) operation with a time window of 1 (s).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The allowable measurement drift of the ESR BPMs shall be ≤ +/-30 (μm) in Closed Orbit (CO) mode in ESR fill (FILL) operation with a time window of ~30 (min).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The allowable measurement drift of the ESR BPMs shall be ≤ +/-5 (μm) in Closed Orbit (CO) mode in ESR luminosity (LUMI) operation within the store ~8 (hours).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The scaling accuracy of the ESR BPMs shall be ≤ +/- 100 (μm) in the +/-3 (mm) measurement range defined for fill (FILL), luminosity (LUMI), and Beam Based Alignment (BBA) operations.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The scaling accuracy of the ESR BPMs shall be ≤ +/- 300 (μm) in the +/-10 (mm) range defined for the beam commissioning (COMM) operation.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The mechanical center of the ESR Beam Position Monitor Pickups, i.e., the horizontal and vertical position alignment tolerances with respect to its fiducials shall be known to a certainty within +/- 200 (um).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up mechanical center x-y offset alignment tolerances with respect to its fiducials to the nearby quad magnetic center shall be ≤ TBD1 (um).02/09/2026In ProcessFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-up mechanical roll alignment tolerances with respect to its fiducials to the nearby quad magnetic center shall be ≤ TBD1 (mrad).02/09/2026In ProcessFALSE
- 6.02.02.05.01.02Wakefields and beam impedance of the ESR Beam Position Monitor Pickup design shall be meet the beam physics needs, with the geometric loss faction being below 2e-4 (V/pC) for 7 (\mm) bunch length and any strong eigen-resonances above approximately 15 (GHz).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR beam position monitor pick-ups assembly shall be a radiation hardened device.02/09/2026ApprovedFALSE
- 6.04.06.01Instrumentation Group shall provide design details that defines the mechanical connection for beam position monitor button to the beam position monitor pick-up assembly.01/08/2026ReviewedFALSE
- 6.04.06.01Instrumentation Group shall provide thermal design limits for the beam position monitor pick-up assembly deformation due to thermal exposure.01/08/2026ReviewedFALSE
- 6.04.06.01Mechanical engineering group shall define the design details for beam position monitor pickup assembly to satisfy the instrumentation thermal regulation design.01/08/2026ReviewedFALSE
- 6.04.06.01Vacuum group shall provide provisions to incorporate the beam position monitor pick-up assembly which satisfies the instrumentation and mechanical Group design into the beamline.01/08/2026ReviewedFALSE
- 6.04.06.01Instrumentation Group shall provide the BPM buttons to be installed onto the beam position monitor pick-up assembly01/08/2026ReviewedFALSE
- 6.04.06.01The physics group shall indicate the location of the beam position monitor pick-up assembly in the lattice designated by a marker.01/08/2026ReviewedFALSE
- 6.04.06.01Vacuum group shall define the installation details for the mechanical connection of beam position monitor pick-up assembly to the beamline.01/08/2026ReviewedFALSE
- 6.04.06.01Vacuum group shall install the beam position monitor pick-up assembly into the vacuum chamber assembly.01/08/2026ReviewedFALSE
- 6.04.06.01Physics Group shall provide the expected radiation dose in its installation location and if required the physics design of the radiation shielding for the beam position monitor pick-up assembly and its associated subsystems to mitigate exposure.01/08/2026ReviewedFALSE
- 6.04.06.01Instrumentation Group shall provide radiation design limits for the beam position monitor and cable radiation hardness and/or sheilding required to mitigate damage from the expected radiation dose.01/08/2026ReviewedFALSE
- 6.04.06.01Instrumentation Group shall provide design details that defines the signal cable connection for beam position monitor pick-up assembly and its associated subsystems.01/08/2026ReviewedFALSE
- 6.04.06.01ASR System Installation and Final Integration shall provide funding and scheduling to connect the beam position monitor required signal cable by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.01Reference related beam position monitor electronics interface document for signal cable routing and installation interface details inside of tunnel. (I-ESR-INST-BPM-ELEC.XX)01/08/2026ReviewedFALSE
- 6.04.06.01Instrumentation Group shall provide thermal design details for beam position monitor buttons installed into the beam position monitor pick-up assembly.01/08/2026ReviewedFALSE
- 6.04.06.01Vacuum group shall incorporate the defined thermal design limits for the beam position monitor pick-up assembly into its bakeout procedure for the ESR beamline which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.01Vacuum group shall define the cleaniness and installation details for the mechanical connection of the beam position monitor pick-up assembly to the beamline.01/08/2026ReviewedFALSE
- 6.04.06.01Instrumentation Group shall provide funding to insure cleanliness installation of the beam position monitor pick-up design by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- ESR-INST-BPM-PU EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.01.02The ESR beam position monitor pick-up shall provide dual plane (horizontal and vertical) beam positional measurements.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The ESR BPM system shall fulfill resolution and accuracy requirements over the following two defined transverse beam position ranges with respect vacuum chamber center, referenced to the mechanical fiducials on the BPM pickups.02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The roll angle error of the ESR BPM pickup, given as the horizontal measurement plane defined by the BPM fiducials witrh respect to. the horizontal plane of the ESR ring shall be less than +/- 20 (mrad).02/09/2026ApprovedFALSE
- 6.02.02.05.01.02The beam position monitor pick-up assembly shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
ESR-INST-BLM : ESR Instrumentation Beam Loss Monitor (WBS 6.02.02.05.02)
- ESR-INST-BLM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.02BLM shall be needed to needed to protect sensitive equipment.02/09/2026In ProcessFALSE
- 6.02.02.05.02The number of BLM installed in the ESR shall be TBD ea02/09/2026In ProcessFALSE
- 6.02.02.05.02BLM shall be installed at the following locations in the ESR TBD02/09/2026In ProcessFALSE
- 6.02.02.05.02The sensitivity of the BLM detectors shall be TBD units?02/09/2026In ProcessFALSE
- 6.02.02.05.02Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/09/2026In ProcessFALSE
- 6.02.02.05.02The response time from loss detection to abort shall be TBD us02/09/2026In ProcessFALSE
ESR-INST-DCCT : ESR Instrumentation Current and Charge Monitor (WBS 6.02.02.05.03)
- 6.02.02.05.03The ESR DC Current Transformer beamline device impedance shall be approved by beam physics.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer digitizer rate shall be 720 (Hz) and stored in an array of 1 (s) duration for post mortum use.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall be a radiation hardened device.02/09/2026ApprovedFALSE
- ESR-INST-DCCT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.03The ESR DC Current Transformer shall have the ability to measure the average beam current over a range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer shall provide an average current measurement with a resolution of less than 5 (uA /√Hz).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measurement drift tolerance shall be less than 10 (uA) over 1 (hr).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy of better than +/- 1 (%) at 250 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system average beam current measurement shall have an absolute accuracy between the range of 250 (mA) to 2.5 (A) at +/- 0.5 (%).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer system shall have a remote controlled self calibration system02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer Calibration system shall be capable of providing an equivalent ESR DC current over the full operating range within 0.25% over the beam current range of 0.15 (mA) to 2500 (mA).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR DC Current Transformer measured average current shall be archived at a rate of 1 Hz02/09/2026ApprovedFALSE
ESR-INST-DCCT-CM
- 6.04.06.02Instrumentation Group shall provide design details that defines the location limitations and the mechanical connection for the DC current transformer assembly and its associated subsystems.01/08/2026ReviewedFALSE
- 6.04.06.02The physics group shall indicate the location of the DC current transformer in the lattice designated by a marker that satisfies the instrumentation groups design limits01/08/2026ReviewedFALSE
- 6.04.06.02Vacuum group shall define the installation details for the mechanical connection for DC current transformer assembly and its associated subsystems to the vacuum beamline.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to install DC current transformer by the appropriate technical support group and its subsystems which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide radiation design limits for the DC current transformer assembly and its associated subsystems.01/08/2026ReviewedFALSE
- 6.04.06.02Physics Group shall provide the expected radiation dose in its installation location and if required the physics design of the radiation shielding for the DC Current Transformer and its associated subsystems to mitigate exposure.01/08/2026ReviewedFALSE
- 6.04.06.02Mechanical engineering group shall define the installation details of the required radiation shielding design for the DC current transformer assembly and its associated subsystems to mitigate exposure which satisfies the Physics Group design.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to install DC current transformer required radiation shielding design by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide thermal design limits and required mitigation design for the DC current transformer assembly and its associated subsystems exposure01/08/2026ReviewedFALSE
- 6.04.06.02Mechanical engineering group shall define the installation details for DC current transformer temperature control design.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to install DC current transformer temperature control design by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.04.06.02Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Infustructure mechanical cooling shall provide the piping design and procurement of materials from the cooling water distribution to connections to the DC current transformer cooling loops which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the installation of piping from the cooling water distribution to each DC current transformer cooling loops by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide design details that defines the signal cable connection for DC current transformer assembly and its associated subsystems.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to connect the DC curent transformer required signal cable by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Reference related DC Current Transformer electronics interface document for signal cable routing interface details inside of tunnel. (I-ESR-INST-DCCT-ELEC.XX)01/08/2026ReviewedFALSE
- 6.05.05.03Instrumentation Group shall provide thermal design parameters for the DC current transformer assembly and its associated subsystems exposure.01/08/2026ReviewedFALSE
- 6.05.05.03Vacuum group shall incorporate the defined thermal design limits for the DC current transformer assembly into its bakeout procedure for the ESR beamline which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.05.05.03Vacuum group shall define the cleaniness and installation details for the mechanical connection of the DC current transformer to the beamline.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to insure cleanliness installation of the DC current transformer design by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
ESR-INST-DCCT-ELEC
- 6.04.06.02Instrumentation Group shall provide design details that defines the building, rack layouts and power utilities for DC current transformer electronics and its associated subsystems.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the DC current transformer electronics and its subcomponents which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the DC current transformer electronics and its subcomponents which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure Group shall provide reliable air conditioning and humidity control in the DC current transformer electronic buildings which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the DC current transformer electronic rack and its subsystems design, including the spacial location, thermal and weight details.09/24/2025In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the plan and funding for the procurement of the racks to house DC current transformer electronics and its sub components which satisfies the design.09/24/2025In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to move and install DC current transformer electronic racks by the appropriate technical support group and its subsystems which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the installation of DC current transformer electronics into the racks by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each DC current transformer electronic rack.09/24/2025In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the design and AC cable tray to contain the AC cable distribution from the wall mount distribution to each DC current transformer electronic rack.01/08/2026In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable and AC cable tray from the wall mount distribution to each DC current transformer electronic rack by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the design details defining the signal cable routing from the DC current transformer electronic and its associated subsystems into the tunnels.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure group shall provide tunnel penetrations for signal cables which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Reference related DC current transformer interface document for signal cable termination iterface details inside of tunnel. (I-ESR-INST-BPM-PU.XX)09/24/2025In ProcessFALSE
- 6.04.06.02Machine Protection System (MPS) Group shall define the design details including input connections and data required to monitor DC current transformer electronic status.09/24/2025In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide output from the MPS interface the data which satisfies MPS design.09/24/2025In ProcessFALSE
- 6.04.06.02Reference related controls interface document for Controls interface signal details. (I-ESR-CNTRL-XXX.XX)09/24/2025In ProcessFALSE
ESR-INST-FCM : ESR Instrumentation Fast Charge Monitor (WBS 6.02.02.05.03)
- 6.02.02.05.03The ESR fast charge monitor shall be able to measure the integrated charge of a single bunch averaged over 1000 turns with a resolution of 100 (pC).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor charge measurement shall maintain its accuracy with a thermal drift of < TBD nC/K02/09/2026In ProcessFALSE
- 6.02.02.05.03The ESR fast charge monitor charge measurement shall not vary more than +/- 1% per (mm) of beam offset.02/09/2026ApprovedFALSE
- 6.02.02.05.03A fast charge monitor shall be installed in the ESR to measure bunch charge with an accuracy 2 (%).02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor system shall operate in ultra-high vacuum.02/09/2026ApprovedFALSE
- 6.02.02.05.03The impedance of the ESR fast charge monitor shall be approved by beam physics02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor system shall, at a rate of 1 Hz, provide the charge of all bunches within a single turn.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor system shall, at a rate of 1 Hz, provide an array of the average bunch charge per turn.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor system shall, at a rate of 1 Hz, provide a turn-by-turn bunch charge measurement for a user-selected bunch.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be a radiation hardened device.02/09/2026ApprovedFALSE
- ESR-INST-FCM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches.02/09/2026ApprovedFALSE
- 6.02.02.05.03The ESR fast charge monitor shall be capable of measuring witness bunch for 1/e for a fixed gain over the beam lifetime.02/09/2026ApprovedFALSE
ESR-INST-FCM-CM
- 6.04.06.02Instrumentation Group shall provide design details that defines the location limitations and the mechanical connection for the fast charge monitor assembly and its associated subsystems.01/08/2026ReviewedFALSE
- 6.04.06.02The physics group shall indicate the location of the fast charge monitor in the lattice designated by a marker that satisfies the instrumentation groups design limits01/08/2026ReviewedFALSE
- 6.04.06.02Vacuum group shall define the installation details for the mechanical connection of fast charge monitor assembly and its associated subsystems to the vacuum beamline.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to install fast charge monitor by the appropriate technical support group and its subsystems which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide radiation design limits for the fast charge monitor assembly and its associated subsystems.01/08/2026ReviewedFALSE
- 6.04.06.02Physics Group shall provide the expected radiation dose in its installation location and if required the physics design of the radiation shielding for the fast charge monitor assembly and its associated subsystems to mitigate exposure.01/08/2026ReviewedFALSE
- 6.04.06.02Mechanical engineering group shall define the installation details of the required radiation shielding design for fast charge monitor assembly and its associated subsystems to mitigate exposure which satisfies the Physics Group design.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to install fast charge monitor required radiation shielding design by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide thermal design limits and required temperature control design for the fast charge monitor assembly and its associated subsystems exposure.01/08/2026ReviewedFALSE
- 6.04.06.02Mechanical engineering group shall define the installation details for fast charge monitor temperature control design.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to install fast charge monitor temperature control design by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.01/08/2026ReviewedFALSE
- 6.04.06.02Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Infustructure mechanical cooling shall provide the piping design and procurement of materials from the cooling water distribution to connections to the fast charge monitor cooling loops which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the installation of piping from the cooling water distribution to each fast charge monitor cooling loops by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide design details that defines the signal cable connection for fast charge monitor assembly and its associated subsystems.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to connect the fast charge monitor required signal cable by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Reference related fast charge monitor electronics interface document for signal cable routing and installation interface details inside of tunnel. (I-ESR-INST-FCM-ELEC.XX)01/08/2026ReviewedFALSE
- 6.04.06.02Instrumentation Group shall provide thermal design parameters for the fast charge monitor assembly and its associated subsystems exposure.01/08/2026ReviewedFALSE
- 6.04.06.02Vacuum group shall incorporate the defined thermal design limits for the fast charge monitor assembly into its bakeout procedure for the ESR beamline which satisfies the Instrumentation Group design.01/08/2026ReviewedFALSE
- 6.04.06.02Vacuum group shall define the cleaniness and installation details for the mechanical connection of the fast charge monitor to the beamline.01/08/2026ReviewedFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to insure cleanliness installation of the fast charge monitor design by the appropriate technical support group which satisfies the Vacuum Group design.01/08/2026ReviewedFALSE
ESR-INST-FCM-ELEC
- 6.04.06.02Instrumentation Group shall provide design details that defines the building, rack layouts and power utilities for fast charge monitor electronics and its associated subsystems.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the fast charge monitor electronics and its subcomponents which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the fast charge monitor electronics and its subcomponents which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure Group shall provide reliable air conditioning and humidity control in the fast charge monitor electronic buildings which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the fast charge monitor electronic rack and its subsystems design, including the spacial location, thermal and weight details.09/24/2025In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the plan and funding for the procurement of the racks to house fast charge monitor electronics and its sub components which satisfies the design.09/24/2025In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling to move and install fast charge monitor electronic racks by the appropriate technical support group and its subsystems which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the installation of fast charge monitor electronics into the racks by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the design and funding for the procurement of the AC cable distribution from the wall mount distribution to each fast charge monitor electronic rack.09/24/2025In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the design and AC cable tray to contain the AC cable distribution from the wall mount distribution to each fast charge monitor electronic rack.01/08/2026In ProcessFALSE
- 6.04.06.02ASR System Installation and Final Integration shall provide funding and scheduling for the installation of AC cable and AC cable tray from the wall mount distribution to each fast charge monitor electronic rack by the appropriate technical support group which satisfies the Instrumentation Group design.01/08/2026In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide the design details defining the signal cable routing from the fast charge monitor electronic and its associated subsystems into the tunnels.09/24/2025In ProcessFALSE
- 6.04.06.02Infrastructure group shall provide tunnel penetrations for signal cables which satisfies the Instrumentation Group design.09/24/2025In ProcessFALSE
- 6.04.06.02Reference related fast charge monitor interface document for signal cable termination iterface details inside of tunnel. (I-ESR-INST-BPM-PU.XX)09/24/2025In ProcessFALSE
- 6.04.06.02Machine Protection System (MPS) Group shall define the design details including input connections and data required to monitor fast charge monitor electronic status.09/24/2025In ProcessFALSE
- 6.04.06.02Instrumentation Group shall provide output from the MPS interface the data which satisfies MPS design.09/24/2025In ProcessFALSE
- 6.04.06.02Reference related controls interface document for Controls interface signal details. (I-ESR-CNTRL-XXX.XX)09/24/2025In ProcessFALSE
ESR-INST-TMK : ESR Instrumentation Horizontal and Vertical Tune Meter Kicker (WBS 6.02.02.05.03)
- ESR-INST-TMK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the horizontal kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The magnitude of the kick required for the vertical kicker shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The kicker waveform (risetime and shape) requirements shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The location of the tune meter kicker striplines in the ESR shall be TBD units02/09/2026In ProcessFALSE
- 6.02.02.05.03The impedance of the kicker beamline device shall be approved by beam Physics.02/09/2026In ProcessFALSE
ESR-INST-SLM : ESR Instrumentation Synchrotron Light Monitors (WBS 6.02.02.05.04)
- ESR-INST-SLM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
ESR-INST-TRANSSLFB : ESR Instrumentation Transverse Slow Feedback (WBS 6.02.02.05.05)
- ESR-INST-TRANSSLFB EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.05.01The transverse slow feedback system bandwidth shall bs 10 Hz02/09/2026In ProcessFALSE
ESR-INST-TRANSFB : ESR Instrumentation Transverse Bunch-by-Bunch Feedback (WBS 6.02.02.05.05.01)
- ESR-INST-TRANSFB EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.01The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
- 6.02.02.05.05.01Placeholder, Input needed TBD02/09/2026In ProcessFALSE
ESR-INST-LONGIFB : ESR Instrumentation Longitudinal Bunch-by-Bunch Feedback (WBS 6.02.02.05.05.02)
- ESR-INST-LONGIFB EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02.05.05.02The requirements for longitudinal feedback are ??? TBD02/09/2026In ProcessFALSE
- 6.02.02.05.05.02The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms02/09/2026In ProcessFALSE
ESR-VAC : ESR Vacuum (WBS 6.02.02.06)
- 6.02.02.06The stray field From any Vacuum equipment on the ESR beam shall have no adverse effect on the beam.02/09/2026ApprovedFALSE
- 6.02.02.06The magnetic permeability for vacuum equipment shall be approved by beam physics.02/09/2026ApprovedFALSE
- 6.02.02.06All hardware in vacuum shall be UHV compatible.02/09/2026ApprovedFALSE
- ESR-VAC EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02The ESR vacuum chamber shall provide sufficient horizontal and vertical aperture to accommodate; a +/-15 sigma beam, where the vertical RMS beam size is based on the emittance of a fully coupled beam, plus an additional 10 mm horizontal and 5 mm vertical margin to account for expected orbit errors.02/09/2026ApprovedFALSE
- 6.02.02.06The typical (standard) vacuum chamber aperture shall be 80 x 36 mm.02/09/2026ApprovedFALSE
- 6.02.02.06Special aperture requirements and/or aperture file shall be provided by or approved by physics.02/09/2026ApprovedFALSE
- 6.02.02The dynamic pressure around the ESR shall be consistent with a beam gas lifetime of >10[hrs] with the design currents after an integrated beam current of 1000 [A.h].02/09/2026ApprovedFALSE
- 6.02.02.06There shall be no upper pressure limit as long as the average pressure is maintained.02/09/2026ApprovedFALSE
- 6.02.02.06The average vacuum level in the ESR Arc sections after conditioning (for 1000Ahrs) shall be <5x10-9 Torr.02/09/2026ApprovedFALSE
- 6.02.02.06On 15 m on each side (or one vacuum sector) of the SRF cavities shall be processed to class ISO 5.02/09/2026ApprovedFALSE
- 6.02.02There shall be no pressure bumps in the ESR exceeding (TBD)[Torr]02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber and all its components shall be designed to withstand a total synchrotron radiation load of 10 MW, considering the uneven linear load particularly related to the super-bends.02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.02.02The ESR vacuum chamber material shall be chosen such that the SR power can be intercepted by the arc chambers and in addition good radiation shielding will be provided to prevent damage to other components.02/09/2026ApprovedFALSE
- 6.02.02.06The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.02/09/2026ApprovedFALSE
- 6.04.05The impedance of the entire ESR vacuum system, including the interaction regions in IR06 and IR08, shall allow for the bunch intensities, beam currents, and bunch numbers contained in the Master Parameter Table (MPT). [Document#:EIC-SEG-RSI-005]05/16/2025ApprovedFALSE
- 6.02.02.06The vacuum system global impedance shall be less than the impedance budget as provided by accelerator physics.02/09/2026ApprovedFALSE
- 6.02.02.06The maximum beam excursion orbit shall be TBD02/09/2026In ProcessFALSE
ESR-CONT : ESR Controls System (WBS 6.02.04.02)
- ESR-CONT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.04.02The ESR control system shall facilitate all ESR global control requirements.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.01.02.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The physical magnet length shall be <2.73 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be a single function dipole with a vertical field direction.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.01.02-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.01.03-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The physical magnet length shall be <1.13 (m).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet pole gap height and width shall be H=52 (mm), W=140(mm).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.02.01-1)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet good field aperture dAx required shall be 35.0345 mm.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet-to-magnet variability shall be < 0.1%.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.03.01-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/09/2026ApprovedFALSE
- 6.02.02.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR control system shall facilitate all network, relational database and data archiving required.02/09/2026ApprovedFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The control system shall facilitate all machine protection systems required02/09/2026ApprovedFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR control system shall facilitate all EIC machine timing required.02/09/2026ApprovedFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR control system shall facilitate fast orbit feedback integration systems as required.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR control system shall facilitate all physics application support required.02/09/2026ApprovedFALSE
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
ESR-CONT-ALGNMNT : ESR Controls Beam Based Alignment (WBS 6.02.04.02)
- ESR-CONT-ALGNMNT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.04.02The BBA quad strength command output rate shall be 1 Hz02/09/2026In ProcessFALSE
ESR-CONT-FEEDBACK : ESR Controls Slow Orbit Feedback (WBS 6.02.04.02)
- ESR-CONT-FEEDBACK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.04.02The slow orbit feedback BPM data averaging period shall be tbd -02/09/2026In ProcessFALSE
- 6.02.04.02The slow orbit feedback correction output rate shall be 10 Hz02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02Placeholder for fast orbit feedback requirements02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback measurement sample rate shall be tbd Hz02/09/2026In ProcessFALSE
- 6.02.04.02The tune feedback correction rate shall be tbd Hz02/09/2026In ProcessFALSE
ESR-CONT-SPIN : ESR Controls Spin Pattern (WBS 6.02.04.02)
- ESR-CONT-SPIN EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
ESR-CRYO_CENT_PLANT : Cryogenics Central Plant (WBS 6.02.04.04)
ESR-CRYO_IR10_SAT_PLANT : IR10 Satellite Cryogenics Plant (WBS 6.02.04.04)
ESR-MPS : ESR Machine Protection System (WBS 6.04.04.03.01)
ESR-MPS-CONTROLS : Electron Storage Ring Machine Protection Control System (WBS 6.02.04.02)
ESR-MPS-GENERAL : Electron Storage Ring General Machine Protection (WBS 6.04.04.03.01)
- 6.04.04.03.01The ESR machine protection system (MPS) shall protect the detector and accelerator from permanent beam-induced or synchrotron radiation induced damage.02/09/2026ReviewedFALSE
- 6.04.04.03.01The ESR MPS shall consist of a set of inputs (beam loss monitors, detector background signals, manual operator input, quench protection system, ...), a electronic trigger and a fast abort system.02/09/2026ReviewedFALSE
- 6.04.04.03.01The ESR MPS thresholds at the input devices shall be set such that a sufficient safety margin remains until permanent damage occurs to machine or detector component.02/09/2026ReviewedFALSE
- 6.04.04.03.01The ESR electronic triggers shall be fast enough such that for any realistic failure scenario the beam loss or detector background occurring between loss or background detection and actual beam abort does not result in permanent damage.02/09/2026ReviewedFALSE
- 6.04.04.03.01The ESR electronic triggers shall be synchronized with the beam abort gap in the ESR bunch train such that the rising edge of the fast abort kicker pulse falls into the abort gap and all ESR bunches receive a sufficiently large kick to detect them safely past the extraction septum and into the fast beam dump.02/09/2026ReviewedFALSE
ESR-MPS-COLLIMATION : Electron Storage Ring Machine Protection Collimation (WBS 6.04.04.03.01.01)
- 6.04.04.03.01.01The ESR collimation system shall consist of betatron and momentum collimators.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR collimation system shall reduce the electron-induced detector background such that the detector can operate safely and efficiently.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01All ESR collimator stations shall be double-sided if available space allows. If only a single-sided collimator can be installed due to space constraints, a second collimator shall be installed on the same side as the first, at a betatron phase of 180 degrees from the first one.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR Collimators shall be placed at accelerator locations suitable for back-ground reduction at 5, 10, and 18 GeV electron beam current.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR Collimator jaws shall be independently and remotely movable over a range of TBD millimeters.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR Collimator jaw material shall be chosen such that the collimator jaw can absorb TBD electrons at TBD energies without sustaining permanent damage.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01All ESR collimator stations shall be equipped with appropriate beam loss monitors to protect the collimator jaws from excessive beam losses by aborting the beam via the Machine Protection System (MPS).02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR collimator jaws shall be wide enough to still be effective in the presence of beam orbit errors of TBD millimeters.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR collimator jaws need to be replaceable within TBD days.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR collimation stations shall be designed such as to minimize their beam impedance.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR collimation stations shall be designed for operation in a vacuum system with pressure in the TBD nTorr range.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The total number of ESR collimators shall be minimized in order to keep their contribution to the accelerator impedance at a minimum.02/09/2026ApprovedFALSE
- ESR-MPS-COLLIMATION EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.01The ESR shall have a collimation system capable of ensuring a sufficiently low background at the detector.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be vertical.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR injection absorber has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The thermal duty cycle shall be 2 Hz. 2 Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR detector absorbers shall be placed at Sector 5.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The detector absorbers have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be tbd kW. tbd kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 21 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 560-720 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator shall be horizontal.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be range shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary vertical collimator shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary horizontal collimator shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W on the tip of the jaw. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary vertical collimators shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary horizontal collimators shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR shall have a collimation system capable protecting all machine elements in case of failure.02/09/2026ApprovedFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be vertical.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR injection absorber has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The thermal duty cycle shall be 2 Hz. 2 Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR detector absorbers shall be placed at Sector 5.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The detector absorbers have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be tbd kW. tbd kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 21 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 560-720 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator shall be horizontal.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be range shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary vertical collimator shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary horizontal collimator shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W on the tip of the jaw. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary vertical collimators shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary horizontal collimators shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
ESR-MPS-ABORT : Electron Storage Ring Machine Protection Abort System (WBS 6.04.04.03.01.02)
- 6.04.04.03.01.02The ESR fast abort system shall receive its trigger from the ESR machine protection system.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR fast abort system shall consist of a set of kickers, a septum magnet, and a beam dump02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR fast abort beam dump shall be external to the circulating beam line.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR fast abort beam dump shall be installed in the stub tunnel in IR2.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR beam dump shall be capable of absorbing the entire ESR beam (1160 bunches, 28 nC each, at 5 to 10 GeV, or 290 bunches, 11 nC each, at 18 GeV) without sustaining permanent damage.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR beam dump shall be capable of absorbing the entire ESR beam at a rate of up to once every 20 minutes at 5-10 GeV, or once every 5 minutes at 18 GeV, respectively.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02Radiation shielding shall be provided as part of the ESR beam dump assembly such that the radiation on the outer surface of the beam dump does not exceed TBD after TBD beam aborts at full intensity.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR abort kickers shall be installed in the IR2 straight section, inside the IR2 experimental hall.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The rise time of the ESR abort kicker system shall not exceed 0.8 usec.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR abort kicker pulse shall remain at or near its peak value for a duration of at least 13 usec.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR abort kicker pulse amplitude shall be sufficiently large to detect the beam safely past the abort extraction septum.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02A septum magnet shall separate the extracted ESR beam from the circulating ESR Beam.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The ESR septum magnet shall be designed such that the magnetic field in the circulating ESR beam enclosure does not compromise the circulating beam.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02Between the ESR septum magnet and the ESR beam dump, DC magnets shall be installed to dilute the beam at the entrance face of the beam dump, such that at least a 50 percent safety margin on beam intensity is guaranteed before permanent damage to the beam dump occurs.02/09/2026ApprovedFALSE
- ESR-MPS-ABORT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.02The ESR shall contain an Abort system to dump the beam.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The materials shall be C / Al / Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The energy deposited during abort shall be 320 kJ02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The frequency of thermal cycle shall be 1 hour02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window thickness shall be tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window material shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Length shall be 50 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Internal diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Temperature sensors shall be yes02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The BPMs shall be 402/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Correctors shall be 402/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Corrector PS shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Cooling / pumping shall be yes02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The deflection shall be 2 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 1.2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply shall be 1600 A02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Y-chamber aperture shall be 36 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02May need to add additional window requirements for other leg of Lambertson magnet TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The gradient shall be 17 T/m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply shall be 1600 A02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 70 cm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The aperture radius shall be 50 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The number of kickers shall be 602/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Rise time shall be 900 ns02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Fall time shall be NA sec02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flat top time shall be 13 us02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The waveshape shall be trap02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The painting shall be vertical02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum field shall be 0.12 T02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The total deflection shall be 16 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum current shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum voltage shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The inductance with cable shall be TBD (uH)02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Max rep rate shall be 100 kV/pC02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flat top repeatability shall be NA Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The beam abort kicker shall be tbd %02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flatness of flat top/pulse form shall be 1 %02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The cooling type shall be w (W,A)02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The ESR Abort system shall contain a beam dump to safely absorb the energy of the stored beam in a controlled fashion.02/09/2026ApprovedFALSE
- 6.04.04.03.01.02The diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The materials shall be C / Al / Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The energy deposited during abort shall be 320 kJ02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window thickness shall be tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The frequency of thermal cycle shall be 1 hour02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window material shall be tbd02/09/2026In ProcessFALSE
ESR-MPS-EXT_KICK : ESR Machine Protection System Single Turn Extraction Kicker (WBS 6.04.04.03.01.02)
- ESR-MPS-EXT_KICK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.02The number of kickers shall be 602/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Rise time shall be 900 ns02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Fall time shall be NA sec02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flat top time shall be 13 us02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The waveshape shall be trap02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The painting shall be vertical02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum field shall be 0.12 T02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The total deflection shall be 16 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum current shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The maximum voltage shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The inductance with cable shall be TBD (uH)02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Max rep rate shall be 100 kV/pC02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flat top repeatability shall be NA Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The flatness of flat top/pulse form shall be 1 %02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The beam abort kicker shall be tbd %02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The cooling type shall be w (W,A)02/09/2026In ProcessFALSE
ESR-MPS-EXT_KICK-DUMP_BLK : ESR Machine Protection System Dump block (WBS 6.04.04.03.01.02)
- ESR-MPS-EXT_KICK-DUMP_BLK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.02The diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The materials shall be C / Al / Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The energy deposited during abort shall be 320 kJ02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The frequency of thermal cycle shall be 1 hour02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window thickness shall be tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The window material shall be tbd02/09/2026In ProcessFALSE
ESR-MPS-EXT_KICK-INST_BP : ESR Machine Protection System Beam Pipe & Instrumentation (WBS 6.04.04.03.01.02)
- ESR-MPS-EXT_KICK-INST_BP EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.02The Length shall be 50 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Internal diameter shall be 90 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Temperature sensors shall be yes02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The BPMs shall be 402/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Correctors shall be 402/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Corrector PS shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be tbd02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Cooling / pumping shall be yes02/09/2026In ProcessFALSE
ESR-MPS-EXT_KICK-LAMBERT : ESR Machine Protection System Lamberton Magnet (WBS 6.04.04.03.01.02)
- ESR-MPS-EXT_KICK-LAMBERT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.02The deflection shall be 2 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 1.2 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply shall be 1600 A02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The Y-chamber aperture shall be 36 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02May need to add additional window requirements for other leg of Lambertson magnet TBD02/09/2026In ProcessFALSE
ESR-MPS-EXT_KICK-QUAD : ESR Machine Protection System Line Quads (WBS 6.04.04.03.01.02)
- ESR-MPS-EXT_KICK-QUAD EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.02The gradient shall be 17 T/m02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply shall be 1600 A02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The power supply accuracy shall be TBD02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The length shall be 70 cm02/09/2026In ProcessFALSE
- 6.04.04.03.01.02The aperture radius shall be 50 mm02/09/2026In ProcessFALSE
ESR-MPS-RADSHIELD : Electron Storage Ring Machine Protection Radiation Shielding (WBS 6.04.04.03.01.03)
ESR-COLL : ESR Momentum Collimator System (WBS 6.04.04.03.01.01)
ESR-COLL-ABS : ESR Momentum Collimator Injection Absorbers (WBS 6.04.04.03.01.01)
- ESR-COLL-ABS EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.01The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR Injection absorber shall be vertical.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR injection absorber has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The thermal duty cycle shall be 2 Hz. 2 Hz02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
ESR-COLL-DET : ESR Momentum Collimator Detector Absorbers (WBS 6.04.04.03.01.01)
- ESR-COLL-DET EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.01The ESR detector absorbers shall be placed at Sector 5.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be fixed.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The detector absorbers have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be tbd kW. tbd kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 21 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Cu02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 560-720 mm02/09/2026In ProcessFALSE
ESR-COLL-MOM : ESR Momentum Collimator (WBS 6.04.04.03.01.01)
- ESR-COLL-MOM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.01The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator shall be horizontal.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be tbd mm half gap. tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall be range shall be tbd.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR momentum collimator has dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 550 W. 550 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be steady-state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be negligible. ~0 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 0.89 m02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Al-Ti02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 1.1 m02/09/2026In ProcessFALSE
ESR-COLL-PRIM : ESR Momentum Collimator Primary Collimators (WBS 6.04.04.03.01.01)
- ESR-COLL-PRIM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.01The ESR primary vertical collimator shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary horizontal collimator shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR primary collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W on the tip of the jaw. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
ESR-COLL-SECDRY : ESR Momentum Collimator Secondary Collimators (WBS 6.04.04.03.01.01)
- ESR-COLL-SECDRY EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04.03.01.01The ESR secondary vertical collimators shall be placed at Sector 4.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR secondary horizontal collimators shall be placed at Sector 2.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The ESR collimators have dual jaws.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The aperture shall be centered on the beam axshall be.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam energy deposition shall be 300 W. 300 W02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The duty cycle shall be steady state.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc02/09/2026In ProcessFALSE
- 6.04.04.03.01.01Jaw angle in position relative to beam axshall be 1 mrad02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The beam failure energy deposition shall be 275 J (1 bunch). 275 J02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The failure thermal duty cycle shall be intermittent.02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The tapered jaw slope shall be 1/10. 1/1002/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw tip length shall be 180 mm02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The jaw material shall be Mo-Gr02/09/2026In ProcessFALSE
- 6.04.04.03.01.01The total length shall be 762-995 mm02/09/2026In ProcessFALSE
ESR-RF : ESR RF System (WBS 6.08.03.02)
- 6.02.02The ESR RF Systems shall meet ambient magnetic field hygiene requirements.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF cryomodule maximum design ambient magnetic field amplitude shall be 700 mG.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum magnetic shield attenuation factor at SRF cavity equator shall be 250.11/17/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall be designed to protect cryogenic surfaces utilizing passive thermal control systems to reduce radiative heat transfer on all 2K, 5K and 50K surfaces.02/09/2026ApprovedFALSE
- 6.08.04.01The cold insulating maximum vacuum shall be 5.0e-7 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The warm insulating maximum vacuum shall be 1.0e-5 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The insulating vacuum maximum leak rate shall be 1.0e-8 mbar L/s.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/m^2.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 50K surfaces shall be 5 W/m^2.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall be designed to utilize common transportation methods without degradation of performance.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be capable of withstanding a maximum allowable vertical acceleration of ±4 G.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be capable of withstanding a maximum allowable beamline axis acceleration of ±5 G.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be capable of withstanding a maximum allowable lateral acceleration of ±1.5 G.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to withstand a tilt around the beamline axis (roll) up to ± 0.03 radians.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall interface to all required accelerator systems.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cavity helium jacket shall have a minimum helium bath vapor surface area of 0.049 m^2.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed helium supply operational temperature shall be 5.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed helium supply operational pressure shall be 3 to 3.5 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return temperature shall be 20 to 100 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return pressure shall be 2.4 to 2.6 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return temperature shall be 4.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return pressure shall be 30 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cooling water subsystem shall be designed to utilize the supply characteristics as defined by the EIC Infrastructure Utility Requirements Document (Doc. No. EIC-IFD-RSI-012).05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall be able to accommodate 0.1 W/m electron beam losses.02/09/2026ApprovedFALSE
- 6.02.02The ESR RF Systems shall provide conditioning capabilities necessary for ESR operations.02/09/2026ApprovedFALSE
- 6.08.04.01Conditioning for individual components shall have a maximum average cryogenic power dissipation of 200 W.05/29/2025ApprovedFALSE
- 6.08.04.01Conditioning for individual components shall be achieved with a maximum temperature of 2.1 K.05/29/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall have an active frequency tuning system to maintain and tune the resonant frequency in order to accommodate radial beam offset and changes in the synchronous trajectory circumference.02/09/2026ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity Slow Tuner tuning range shall be 600 kHz.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner resolution shall be ± 1 Hz.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity slow tuner tuning rate shall be 800 Hz/s.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner hysteresis shall be ± 20 Hz.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall be rotatable around the beamline axis (roll) to facilitate installation without degradation of performance.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to withstand a tilt around the beamline axis (roll) up to ± 0.03 radians.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall have alignment verification designed into the system.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in X shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Y shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Z shall be ± 5 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ± 0.04 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall achieve an independent beam line vacuum level equivalent to the ESR operational vacuum level.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The cold beamline maximum vacuum shall be 1.0e-9 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The warm beamline maximum vacuum shall be 5.0e-7 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The beamline vacuum maximum leak rate shall be 5e-10 mbar L/s.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall provide all necessary controls and diagnostics necessary for system operation but not extending beyond the RF System interfaces.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall provide 0.1% field amplitude stability02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cryomodule FPC external quality factors shall all be within ± 0.1e5 of all other FPC external quality factor design values.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall provide 0.02 degree phase stability02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cryomodule FPC external quality factors shall all be within ± 0.1e5 of all other FPC external quality factor design values.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall be protected from degradation due to adjacent beam lines and hardware.02/09/2026ApprovedFALSE
- 6.08.04.01The manufactured SRF Cryomodule Cavity shall produce no field emission at 4 MV.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems design shall be protected against beam loss and RF field collapse events.02/09/2026ApprovedFALSE
- 6.02.02The ESR RF Systems shall maintain ESR operations in the event of single RF system failure.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity Slow Tuner tuning range shall be 600 kHz.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems active frequency tuning system shall be maintainable in-situ.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.08.04.01The active SRF cavity tuning mechanism components (motor/gearbox/drive mechanism) shall be replaceable and maintainable in-situ.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems air-side FPC components shall be maintainable in-situ.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.02.02The ESR Crabbing RF System shall be designed to crab electron bunches.02/09/2026ApprovedFALSE
- 6.02.02The ESR Storage RF System design shall protect against common electron beam instabilities.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1.5E10.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF pressure sensitivity maximum shall be 10 Hz/mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner resolution shall be ± 1 Hz.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF maximum lorentz force detuning shall be 5 Hz/(MV/m)^2.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner hysteresis shall be ± 20 Hz.05/16/2025ApprovedFALSE
- 6.02.02The ESR Storage RF System shall be designed as identical modular components to provide the full system functionality.02/09/2026ApprovedFALSE
- 6.02.02The ESR Storage RF System shall maintain constant electron beam energy after injection.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule loaded quality factor shall be 2.9e5 ± 0.2e5.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- 6.02.02The ESR Storage RF System shall be capable of coupling 10 MW of RF power to the beam.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- 6.02.02The ESR Storage RF System shall maintain beam energy from 5 to 18 GeV.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule loaded quality factor shall be 2.9e5 ± 0.2e5.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity Slow Tuner tuning range shall be 600 kHz.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner resolution shall be ± 1 Hz.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner hysteresis shall be ± 20 Hz.05/16/2025ApprovedFALSE
- 6.02.02The ESR Storage RF System shall be able to provide 68 MV peak voltage for an RF bucket height 10x greater than the rms energy spread of the beam.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- ESR-RF EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.02The ESR RF Systems shall be designed to fulfill all necessary parameters as set by the Master Parameter Table (MPT). [Document#: EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF System shall utilize superconductivity.02/09/2026ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum designed operational temperature shall be 2 K.05/16/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum designed operational pressure shall be 30 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath designed operational pressure stability shall be ±0.1 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cavity helium jacket shall have a minimum helium bath vapor surface area of 0.049 m^2.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed helium supply operational temperature shall be 5.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed helium supply operational pressure shall be 3 to 3.5 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return temperature shall be 20 to 100 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return pressure shall be 2.4 to 2.6 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return temperature shall be 4.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return pressure shall be 30 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity maximum Niobium temperature shall be 5 K during operation.05/16/2025ApprovedFALSE
- 6.08.04.01The warm beamline maximum vacuum shall be 5.0e-7 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The cold beamline maximum vacuum shall be 1.0e-9 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The beamline vacuum maximum leak rate shall be 5e-10 mbar L/s.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall conform to the ESR lattice.02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.08.04.01The SRF cryomodule cavity beam axis to the tunnel floor shall be vertically alignable to 1381.09 ± 20 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in X shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Y shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Z shall be ± 5 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ± 0.04 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall be installed in the straight sections of the ESR lattice within the existing RHIC tunnel in IR10.02/09/2026ApprovedFALSE
- 6.02.02.05.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor02/09/2026In ProcessFALSE
- 6.02.02.05.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.02/09/2026In ProcessFALSE
- 6.02.02.05.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM shall be able to measure the Global coupling with accuracy of TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.02/09/2026In ProcessFALSE
- 6.02.02.05.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda02/09/2026In ProcessFALSE
- 6.02.02.05.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.02/09/2026In ProcessFALSE
- 6.02.02.05.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.02/09/2026In ProcessFALSE
- 6.02.02.05.04The locations of the SLM optical lab rooms shall be TBD -02/09/2026In ProcessFALSE
- 6.02.02.05.04The double-slit interferometer method shall be used to measure transverse beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04The standard transverse resolution of an SLM using visible light shall be ~60 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The resolution using the double-slit method shall equal to 10 um02/09/2026In ProcessFALSE
- 6.02.02.05.04A streak camera shall be used to measure the bunch longitudinal profiles02/09/2026In ProcessFALSE
- 6.02.02.05.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs02/09/2026In ProcessFALSE
- 6.02.02.05.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation02/09/2026In ProcessFALSE
- 6.02.02.05.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.02/09/2026In ProcessFALSE
- 6.02.02.05.04The location of the X-ray pinhole monitoring system shall be TBD02/09/2026In ProcessFALSE
- 6.02.02.05.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/09/2026In ProcessFALSE
- 6.02.02.05.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements02/09/2026In ProcessFALSE
- 6.02.02.05.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size02/09/2026In ProcessFALSE
- 6.02.02.05.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.02/09/2026In ProcessFALSE
- 6.08.04.01The SRF Cryomodule maximum length shall be 7.2 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule maximum width shall be 2.15 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule maximum height shall be 1.7 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cryomodule cavity beam axis to the tunnel floor shall be vertically alignable to 1381.09 ± 20 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in X shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Y shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Z shall be ± 5 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ± 0.04 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum height shall be 2.1m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum length (not including vacuum jacketed lines) shall be 1.5 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum width shall be 1.0 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic valve box minimum vertical stay clear height above the cryomodule shall be 0.92 m.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall conform to the EIC Code of Record.02/09/2026ApprovedFALSE
- 6.08.04.01All cryomodule surfaces accessible to workers shall be within the temperature range of 283 to 333 K.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME B31.3.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME BPVC.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASTM C1055.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 70.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 70E.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by API 520 & API 521.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by AWS.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by CGA S1.3.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards as directed by the DOE Vacuum Vessel Consensus Standards.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to meet or exceed the maximum working pressures defined by the EIC pressure document (Document No. TBD).05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems within the tunnel shall operate within its yearly radiation exposure budget.02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF Systems shall have a minimum operating lifetime of 20 years02/09/2026ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall operate through a minimum of 200 thermal cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner 1% range tuning cycles shall be 100,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner full range tuning cycles shall be 1,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.05/16/2025ApprovedFALSE
- 6.02.02The ESR RF System shall be designed to minimize unscheduled downtime, maintenance time and repair time to achieve ESR operational availability.02/09/2026ApprovedFALSE
- 6.02.04.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection02/09/2026In ProcessFALSE
- 6.02.04.02The Spin pattern control granularity shall be 1 bunch02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 18GeV shall be 290 cnt02/09/2026In ProcessFALSE
- 6.02.04.02The number of bunches @ 10GeV and below shall be 1160 cnt02/09/2026In ProcessFALSE
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 300K to 150K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve steady state temperature with the cavity bath at 4K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve a full warm-up cycle from 4K to 295K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall operate through a minimum of 200 thermal cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner 1% range tuning cycles shall be 100,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner full range tuning cycles shall be 1,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The manufactured SRF Cryomodule Cavity shall produce no field emission at 4 MV.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.08.04.01All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.05/16/2025ApprovedFALSE
- 6.08.04.01The active SRF cavity tuning mechanism components (motor/gearbox/drive mechanism) shall be replaceable and maintainable in-situ.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity slow tuner tuning rate shall be 800 Hz/s.05/16/2025ApprovedFALSE
- 6.02.02The ESR Storage RF System shall be designed to accelerate electrons.02/09/2026ApprovedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the cryomodule shall be 30 kW for all EIC design energies and currents.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E11.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
ESR-RF-SRF : Electron Storage Ring Superconducting RF (WBS 6.02.02)
ESR-RF-SRF-591_1Cell : 591MHz 1 Cell Cryomodule (WBS 6.08.04.01)
ESR-RF-HPRF : Electron Storage Ring High Power RF Amplifier (WBS 6.02.03.08)
- 6.02.03.08The HPRF System shall be interlocked to Personel Protection System (PPS), and have a coordinated shutoff sequence to minimize AC line disturbances03/02/2026In ProcessFALSE
- 6.02.03.08The HPRF System shall include interlocks to guarantee its integrity and the integrity of other equipment.03/02/2026ReviewedFALSE
- 6.02.03.08The HPRF System shall be equipped with directional power monitoring, arc detection, termometry, flow monitoring, and water supply, bleeding functionality to monitor and control all sub-systems during operation and testing.03/02/2026ReviewedFALSE
- 6.02.03.08The HPRF System components shall be accessible for installation and maintenance, maximizing the number of components which can be maintained and replaced in-situ.03/02/2026ReviewedFALSE
- 6.02.03.08The HPRF System shall be designed to operate in the environmental conditions in the space provided by EIC Infrastructure03/02/2026ReviewedFALSE
- 6.02.03.08The HPRF System shall be designed to fit within the space envelope provided03/02/2026ReviewedFALSE
- 6.02.03.08The HPRF System shall be capable of supporting both CW and Pulsed modes of operation03/02/2026ReviewedFALSE
ESR-RF-HPRF-ACAV:591 : HSR RF Amplifier SRF 591 MHz Cavity (WBS 6.02.03.08)
- 6.02.02.08HPRF shall provide a feedback signal and cabling requirements from the DC Bias power supply to the LLRF controls system.03/02/2026In ProcessFALSE
- 6.02.02.08LLRF shall provide the control and monitoring signal for the HPRF DC Bias.03/02/2026In ProcessFALSE
- 6.02.02.08Reference related Low Level RF (LLRF) interface document for HPRF to LLRF cable routing and signal details. (I-ASR-CTRL-LLRF-XXX.XXX)03/02/2026In ProcessFALSE
- 6.02.02.08Reference related Personnel Protection Systems (PPS) interface document for Personnel Protection Systems (PPS) cable routing and signal details. (I-ASR-SIFI-CNTRL-PPS-XXX.XXX)03/02/2026In ProcessFALSE
- 6.02.02.08Reference related EIC global controls interface document for controls cable routing and signal details. (I-ASR-SIFI-CNTRL-RF-XXX.XXX)03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall provide design details that defines the building, rack layouts and power, chilled water and instrument air utility requirements for HPRF systems and its associated subsystems.03/02/2026In ProcessFALSE
- 6.02.02.08Infrastructure Group shall provide buildings, detailed drawings including rack layouts for the HPRF systems and its subcomponents which satisfies the HPRF group design03/02/2026In ProcessFALSE
- 6.02.02.08Infrastructure Group shall provide funding and scheduling for the installation of the wall terminated power utilities for distribution to the HPRF systems and its subcomponents which satisfies the HPRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08Infrastructure Group shall provide funding and scheduling for the installation of the chilled water and instrument air utilities for distribution to the HPRF systems and its subcomponents which satisfies the HPRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08Infrastructure Group shall provide indoor environmental control for the HPRF buildings which satisfies the HPRF equipment operating parameters.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall provide the HPRF equipment rack and its subsystems design, including the spatial location, thermal and weight details.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall provide the HPRF equipment list, including the subsystems design, spatial location, thermal, power and weight details.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall provide the plan and funding for the procurement of the racks to house HPRF equipment and its sub components which satisfies HPRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08ASR System Installation and Final Integration shall provide funding and scheduling to move and install HPRF racks by the appropriate technical support group and its subsystems which satisfies the HPRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08ASR System Installation and Final Integration shall provide funding and scheduling for the installation of HPRF into the racks by the appropriate technical support group which satisfies the HPRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08ASR System Installation and Final Integration shall provide design, funding and scheduling for installation of the AC cable tray and AC cable from the wall mount distribution to each HPRF rack and the freestanding equipment, including termination by the appropriate technical support group which satisfies the HPRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF group shall provide the design for water flow, pressure, supply temperature, water resistivity and heat load into water.03/02/2026In ProcessFALSE
- 6.02.02.08Infustructure mechanical cooling shall provide water flow, pressure, supply temperature, water resistivity and heat load removal which satisfies the HPRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08Infustructure mechanical cooling shall provide the CHWS piping design and procurement of materials from the wall distribution to connections to the HPRF system which satisfies the HPRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08ASR System Installation and Final Integration shall provide funding and scheduling for the installation of CHWS piping from the wall mount distribution to each HPRF system by the appropriate technical support group which satisfies the HPRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF group shall provide the design and instrument air requirements for the FPC and waveguide transmission system.03/02/2026In ProcessFALSE
- 6.02.02.08SRF Systems shall provide a suitable path for air flow via FPC into the HPRF transmission system.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall specify air flow requirements for the FPC to SRF Systems03/02/2026In ProcessFALSE
- 6.02.02.08Infustructure shall provide instrument air which satisfies the HPRF and/or SRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08Infustructure shall provide the instrument air piping design and procurement of materials from the wall distribution to connections to the HPRF system which satisfies the HPRF and/or SRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08ASR System Installation and Final Integration shall provide funding and scheduling for the installation of instrument air piping from the wall mount distribution to each HPRF system by the appropriate technical support group which satisfies the HPRF and/or SRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF Group shall provide the design details defining the transmission line routing from the HPRF subsystems into the tunnels.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall provide HPRF components and specific fixtures for installing HPRF equipment03/02/2026In ProcessFALSE
- 6.02.02.08Infrastructure group shall provide tunnel penetrations for HPRF subsystems which satisfies the HPRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08ASR System Installation and Final Integration shall provide funding and scheduling for the installation of HPRF transmission line from each HPRF amplifier to its recipical cavity by the appropriate technical support group which satisfies the HPRF and/or SRF/SRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08Reference related EIC SRF cryomodule interface document for cryomodule transmission line and connection details. (I-RF-SRF-XXX.XXX)03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall provide the DC Bias power supply for biasing the FPC of the SRF Cryomodules.03/02/2026In ProcessFALSE
- 6.02.02.08HPRF shall provide a feedback signal and cabling requirments from the DC Bias power supply to the LLRF controls system.03/02/2026In ProcessFALSE
- 6.02.02.08LLRF shall provide the control and monitoring signal for the HPRF DC Bias.03/02/2026In ProcessFALSE
- 6.02.02.08Infrastructure group shall provide tunnel penetrations for HPRF DC Bias subsystems which satisfies the HPRF group design.03/02/2026In ProcessFALSE
- 6.02.02.08ASR System Installation and Final Integration shall provide funding and scheduling for the installation of HPRF DC Bias line from each HPRF amplifier to its recipical cavity by the appropriate technical support group which satisfies the HPRF and/or SRF/SRF groups design.03/02/2026In ProcessFALSE
- 6.02.02.08Reference related EIC SRF cryomodule interface document for cryomodule DC Bias line and connection details. (I-RF-SRF-XXX.XXX)03/02/2026In ProcessFALSE
ESR-RF-ACAV:591S : 591MHz 1 Cell Cryomodule (WBS 6.08.04.01)
- 6.08.04Infrastructure shall provide supply and return headers within the tunnel for Low Conductivity Water (LCW) to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04RF Pre-Installation shall provide all distribution design, materials, and installation of the piping (or hoses) for Low Conductivity Water (LCW) from the tunnel header to the Cryomodules to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide Low Conductivity Water (LCW) supply/return receptacles at the Cryomodule to facilitate installation by RF Pre-Installation.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide instrument air receptables to be utilized by TBD.07/25/2025In ProcessFALSE
- 6.08.04TBD shall provide an instrument air supply system for the required SRF Systems Cryomodule components in the tunnel.07/25/2025In ProcessFALSE
- 6.08.04TBD shall provide all routing design, installation, and control logic to the instrument air receptacles on the SRF Systems Cryomodule.07/25/2025In ProcessFALSE
- 6.08.04RF Pre-Installation shall provide the helium blowdown system design, materials (including helium gas), and installation labor to any water circuit required on the SRF cryomodule inside the tunnel.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide a helium blowdown receptacle to be utilized by Pre-Installation.07/25/2025In ProcessFALSE
- 6.08.04The 2K Cryogenics Distribution System shall provide all distribution design, materials, and installation of the helium piping for the 2K helium distribution system inside the tunnel to the Cryomodules to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04The 2K Cryogenics Distribution System shall provide Supercritical Helium to/from the supply/return header(s) to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide Helium supply and return receptacle(s) at the cryomodule to be utilized by the 2K Cryogenics Distribution System.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide calculations and analyses necessary for sizing the piping and supply by the 2K Cryogenics Distribution.07/25/2025In ProcessFALSE
- 6.08.04The 2K Cryogenics Distribution System shall provide all distribution design, materials, and installation of the helium piping for the helium relief system inside the tunnel to the cryomodule to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide a helium pressure relief valve with common relief header receptacle to be utilized by the 2K Cryogenics Distribution Relief System.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide calculations and analyses necessary for sizing the relief piping by the 2K Cryogenics Distribution relief header.07/25/2025In ProcessFALSE
- 6.08.04The 2K Cryogenics Distribution System shall provide all distribution design, materials, and installation of the helium piping for the guard vacuum inside the tunnel to the cryomodule to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide a helium pressure relief valve with common guard vacuum receptacle to be utilized by the 2K Cryogenics Distribution System.07/25/2025In ProcessFALSE
- 6.08.04Cryogenics Controls shall provide all cabling, routing design, installation, and control logic to the cryomodule control receptacles at the SRF Systems Cryomodules.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide helium liquid level monitor receptacles to be utilized by Cryogenics Controls.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide pressure monitor receptacles to be utilized by Cryogenics Controls.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide temperature monitors receptacles to be utilized by Cryogenics Controls.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide heater control receptacles to be utilized by Cryogenics Controls.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide cryogenic control valve receptacles to be utilized by Cryogenics Controls.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide Beamline All Metal Gate Valve receptacles to be utilized by ESR Vacuum System.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide Beamline Ion pump receptacles to be utilized by ESR Vacuum System.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide Vacuum pressure receptacles to be utilized by ESR Vacuum System.07/25/2025In ProcessFALSE
- 6.08.04TBD shall provide all cabling, routing design, installation, and control logic to the cryomodule vacuum control receptacles at the Cryomodules to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide RF feedback and control receptacles to be utilized by RF Controls.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide temperature monitors receptacles to be utilized by RF Controls.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide tuner control receptacles to be utilized by RF Controls.07/25/2025In ProcessFALSE
- 6.08.04RF Controls shall provide all cabling, routing design, installation, and control logic to the cryomodule RF control receptacles at the Cryomodules to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04High Power RF shall design and provide a waveguide to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide a receptacle for the High Power RF waveguide to be utilized by High Power RF.07/25/2025In ProcessFALSE
- 6.08.04Accelerator Installation shall provide the schedule and funding for the waveguide installation to the SRF Cryomodule.07/25/2025In ProcessFALSE
- 6.08.04High Power RF shall provide DC Bias to be utilized by the SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide a receptable on the cryomodule for the DC Bias connection to be used by High Power RF.07/25/2025In ProcessFALSE
- 6.08.04High Power RF shall provide all design, fabrication, and controls of the DC Bias system to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04Beamline Components shall provide a doorknob waveguide to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04Beamline Components shall provide a BLA to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04Beamline Components shall provide conditioned FPC to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide fiducialization points on the Cryomodule relating back to the electromagnetic center of the beamline to be utilized by Accelerator Installation.07/25/2025In ProcessFALSE
- 6.08.04RF Pre-Installation shall arrange the installation location/area to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04Accelerator Installation shall provide the schedule and funding for the SRF Cryomodule installation and app technician support (lag bolts/rough alignment/pedestal/etc...).07/25/2025In ProcessFALSE
- 6.08.04The Mechanical Design Group shall model the tunnel to define the required spatial locations to be utilized by SRF Systems.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide a receptacle for the beamline connection on both ends of the SRF cryomodule to be utilized by RF System07/25/2025In ProcessFALSE
- 6.08.04ESR Vacuum System shall provide installation labor of the SRF Systems Cryomodule to the beamline.07/25/2025In ProcessFALSE
- 6.08.04RF System shall ensure that during installation the SRF Systems Cryomodule cleanliness does not degrade.07/25/2025In ProcessFALSE
- 6.08.04RF System shall ensure that the beamline components surrounding the SRF Systems Cryomodule does not degrade the performance of the SRF Systems Cryomodule during its lifetime.07/25/2025In ProcessFALSE
- 6.08.04Cryomodule Verification shall ensure TJNAF has a bunker that can high power test the SRF Systems Cryomodule.07/25/2025In ProcessFALSE
- 6.08.04RF Pre-Installation shall ensure BNL has a bunker that can high power test the SRF Systems Cryomodule.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide insulating vacuum gate valve receptables to be utilized by TBD.07/25/2025In ProcessFALSE
- 6.08.04TBD shall provide a vacuum system for the insulating vacuum of the SRF Systems Cryomodule in the tunnel to be utilized by SRF Systems07/25/2025In ProcessFALSE
- 6.08.04TBD shall provide all cabling, routing design, installation, and control logic to the insulating vacuum control receptacles at the SRF Systems Cryomodule.07/25/2025In ProcessFALSE
- 6.08.04RF Pre-Installation shall ensure BNL has a bunker that can high power test the SRF Systems Cryomodule.07/25/2025In ProcessFALSE
- 6.08.04SRF Systems shall provide insulating vacuum gate valve receptables to be utilized by TBD.07/25/2025In ProcessFALSE
- 6.08.04TBD shall provide a vacuum system for the insulating vacuum of the SRF Systems Cryomodule in the tunnel to be utilized by SRF Systems07/25/2025In ProcessFALSE
- 6.08.04TBD shall provide all cabling, routing design, installation, and control logic to the insulating vacuum control receptacles at the SRF Systems Cryomodule.07/25/2025In ProcessFALSE
- ESR-RF-ACAV:591S EXTERNALSRequirements who's parents are in other sub-systems.
- 6.08.04.01The SRF Cryomodule shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing.05/16/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum designed operational temperature shall be 2 K.05/16/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum designed operational pressure shall be 30 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath designed operational pressure stability shall be ±0.1 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cavity helium jacket shall have a minimum helium bath vapor surface area of 0.049 m^2.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed helium supply operational temperature shall be 5.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed helium supply operational pressure shall be 3 to 3.5 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return temperature shall be 20 to 100 K.05/16/2025ApprovedFALSE
- 6.08.04.01The range of the designed combined helium return pressure shall be 2.4 to 2.6 bar.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return temperature shall be 4.5 K.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum designed sub-atmospheric helium return pressure shall be 30 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 300K to 150K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve steady state temperature with the cavity bath at 4K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall achieve a full warm-up cycle from 4K to 295K in a maximum of 2 days.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cooling water subsystem shall be designed to utilize the supply characteristics as defined by the EIC Infrastructure Utility Requirements Document (Doc. No. EIC-IFD-RSI-012).05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall operate through a minimum of 200 thermal cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner 1% range tuning cycles shall be 100,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cryomodule Slow Tuner full range tuning cycles shall be 1,000 cycles.05/16/2025ApprovedFALSE
- 6.08.04.01The manufactured SRF Cryomodule Cavity shall produce no field emission at 4 MV.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1 MGy.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy.05/16/2025ApprovedFALSE
- 6.08.04.01The active SRF cavity tuning mechanism components (motor/gearbox/drive mechanism) shall be replaceable and maintainable in-situ.05/16/2025ApprovedFALSE
- 6.08.04.01All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the cryomodule shall be 30 kW for all EIC design energies and currents.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to operate with a beam current up to 2.5 A.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule loaded quality factor shall be 2.9e5 ± 0.2e5.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cryomodule FPC external quality factors shall all be within ± 0.1e5 of all other FPC external quality factor design values.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1.5E10.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured voltage shall be 4 MV.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E11.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF pressure sensitivity maximum shall be 10 Hz/mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF maximum lorentz force detuning shall be 5 Hz/(MV/m)^2.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity maximum Niobium temperature shall be 5 K during operation.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cavity shall be designed to handle a minimum forward power of 800 kW.05/16/2025ApprovedFALSE
- 6.08.04.01The warm beamline maximum vacuum shall be 5.0e-7 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The cold beamline maximum vacuum shall be 1.0e-9 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The beamline vacuum maximum leak rate shall be 5e-10 mbar L/s.05/16/2025ApprovedFALSE
- 6.08.04.01The warm insulating maximum vacuum shall be 1.0e-5 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The cold insulating maximum vacuum shall be 5.0e-7 mbar.05/16/2025ApprovedFALSE
- 6.08.04.01The insulating vacuum maximum leak rate shall be 1.0e-8 mbar L/s.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity Slow Tuner tuning range shall be 600 kHz.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity slow tuner tuning rate shall be 800 Hz/s.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner resolution shall be ± 1 Hz.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner hysteresis shall be ± 20 Hz.05/16/2025ApprovedFALSE
- 6.08.04.01All cryomodule surfaces accessible to workers shall be within the temperature range of 283 to 333 K.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME B31.3.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME BPVC.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASTM C1055.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 70.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 70E.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by API 520 & API 521.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by CGA S1.3.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by AWS.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed and manufactured to meet all applicable standards as directed by the DOE Vacuum Vessel Consensus Standards.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule maximum length shall be 7.2 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule maximum width shall be 2.15 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule maximum height shall be 1.7 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cryomodule cavity beam axis to the tunnel floor shall be vertically alignable to 1381.09 ± 20 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in X shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Y shall be ± 250 μm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance in Z shall be ± 5 mm.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ± 0.04 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ± 0.01 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum height shall be 2.1m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum width shall be 1.0 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic box maximum length (not including vacuum jacketed lines) shall be 1.5 m.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule cryogenic valve box minimum vertical stay clear height above the cryomodule shall be 0.92 m.05/16/2025ApprovedFALSE
- 6.08.04.01Conditioning for individual components shall have a maximum average cryogenic power dissipation of 200 W.05/29/2025ApprovedFALSE
- 6.08.04.01Conditioning for individual components shall be achieved with a maximum temperature of 2.1 K.05/29/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be capable of withstanding a maximum allowable vertical acceleration of ±4 G.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be capable of withstanding a maximum allowable lateral acceleration of ±1.5 G.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be capable of withstanding a maximum allowable beamline axis acceleration of ±5 G.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to withstand a tilt around the beamline axis (roll) up to ± 0.03 radians.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF cryomodule maximum design ambient magnetic field amplitude shall be 700 mG.05/16/2025ApprovedFALSE
- 6.08.04.01The minimum magnetic shield attenuation factor at SRF cavity equator shall be 250.11/17/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/m^2.05/16/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 50K surfaces shall be 5 W/m^2.05/16/2025ApprovedFALSE
- 6.08.04.01The SRF Cryomodule shall be designed to meet or exceed the maximum working pressures defined by the EIC pressure document (Document No. TBD).05/16/2025ApprovedFALSE
ESR-RF-ACAV:591S-FPC
- 6.08.04.01The FPC shall be designed and fabricated to deliver an average forward RF power of 400 kW under all reflection conditions expected during SRF operation and during SRF cavity quench.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC’s window bandwidth shall be (591 ± 10) MHz with S11 < -30 dB.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC shall be capable to withstand a maximum allowable 5G acceleration in all directions.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC lowest order mode mechanical frequency of the FPC during all events shall be greater than 60 Hz.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC design shall prevent modal mechanical resonance frequencies at multiples of 60 Hz up to 240 Hz.06/05/2025ApprovedFALSE
- 6.08.04.01The maximum temperature of all surfaces on the FPC not accessible to workers shall be 100°C.06/05/2025ApprovedFALSE
- 6.08.04.01The minimum temperature for all water-cooled FPC components shall be 0°C to prevent freezing.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC shall be designed and fabricated to provide a minimum bias voltage of ± 5 kV.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC air-side components shall be maintainable in-situ.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC shall have an air side purge with a minimum flow rate of 5 SCFH nitrogen gas with less than 10 ppm water and filtered to 25 µm.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC design shall protect against multipacting.06/05/2025ApprovedFALSE
- 6.08.04.01The extension of the thermal transition from 2 K to room temperature shall not exceed 317 mm from the beam line axis.06/05/2025ApprovedFALSE
- 6.08.04.01The FPC shall include arc detectors, independent vacuum monitoring, thermosensors, and heaters.06/05/2025ApprovedFALSE
ESR-RF-400KW
- 6.08.06.01The RF Amplifier system shall utilize watercooling to dissipate a minimum of 80 % heat.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier system shall have a maximum die temperature of 130 C.11/20/2025ApprovedFALSE
- 6.08.06.01The minimum acceptable system loss of RF Amplifiers due to faults shall be 10 %.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier shutoff switch time shall be 100 ns.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier shall be designed to self-protect to a maximum overdrive power of 10 %.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier phase linearity over a 40dB dynamic range shall be 30 deg.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplified input/output impedance shall be 50 ohms.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier input VSWR shall be 1.5:1 .11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier input RF overdrive shall be 10 dBm.11/20/2025ApprovedFALSE
- 6.08.06.01All RF Amplifier harmonics shall not exceed -30 dBc.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier spurious and line harmonics shall not exceed -80 dBc.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier noise power output in the ON state shall be -90 dBm/Hz.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier noise power output in the STANDBY state shall be -154 dBm/Hz.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier group delay shall be 200 ns.11/20/2025ApprovedFALSE
- 6.08.06.01The minimum RF Amplifier AC to RF efficiency at full rated power shall be 45 %.11/20/2025ApprovedFALSE
- 6.08.06.01The operational frequency band of the RF Amplifier system shall be 591 ± 5 MHz.11/20/2025ApprovedFALSE
- 6.08.06.01The minimum RF on/off ratio (Δgain between on/standby) shall be 86 dB.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier system shall be designed to operate in an ambient temperature range of 18-30 C.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier system shall be designed to operate in a maximum relative humidity of 60-65 %.11/20/2025ApprovedFALSE
- 6.08.06.01With the load mismatch of less than 3:1, the RF Amplifier shall maintain 100% forward power. .11/20/2025ApprovedFALSE
- 6.08.06.01With the load mismatch of greater than 3:1, the RF Amplifier shall utilize 0% forward power. .11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier forward RF sample coupling across the amplifier bandwidth shall be 70±0.5 dB.11/20/2025ApprovedFALSE
- 6.08.06.01The minimum RF Amplifier forward directivity across the amplifier bandwidth shall be 30 dB.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier reflected RF sample coupling across the amplifier bandwidth shall be 70±0.5 dB.11/20/2025ApprovedFALSE
- 6.08.06.01The minimum RF Amplifier reflected directivity across the amplifier bandwidth shall be 30 dB.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier audible noise at 1m shall be 70 dB(A).11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier pulse width range shall be .001-100 ms.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier duty cycle range shall be 1-50 %.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifer rise time during pulsed operation shall be 0.75 μs.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifer fall time during pulsed operation shall be 0.75 μs.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier pulse drop at max pulse width shall be 0.5-1 dB.11/21/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier phase error over pulse at max pulse width shall be 2.5-5 deg.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier AC Voltage input shall remain between 432-528 VAC.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier controls AC Voltage input shall remain between 108-132 VAC.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier AC frequency shall be 60 Hz .11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier power factor shall be between 0.90-0.95 .11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier input power for the rated output power shall be ±1.5 dBm.11/20/2025ApprovedFALSE
- 6.08.06.01The minimum output power of the RF Amplifier for all operational modes shall be 200/400 kW.11/20/2025ApprovedFALSE
- 6.08.06.01The minimum linear power of the RF Amplifier for all operational modes shall be 200/400 kW.11/20/2025ApprovedFALSE
- 6.08.06.01The nominal power gain for the RF Amplifier system over a 40 dB dynamic range shall be 86 dB.11/20/2025ApprovedFALSE
- 6.08.06.01The maximum power gain deviation of the RF Amplifier system shall be 1.5 dB.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier maximum height shall be 3 m.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier maximum width shall be 3 m.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier maximum length shall be 3 m.11/20/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier shall be designed to have a minimum MTBF of 10,000 hours.11/20/2025ApprovedFALSE
- 6.08.06.01The RF devices shall be designed to have a minimum MTBF of 35,000 hours.11/20/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by ASTM C1055 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .11/20/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by 29 CFR 1910 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .11/20/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by NFPA 70E-2021 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .11/20/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by UL 508 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .11/20/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by Occupatioanl Radiation Protection 10 CFR 835 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .11/20/2025ApprovedFALSE
ESR-RF-CCAV:394
- 6.08.04.05The SRF CM shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing05/16/2025In ProcessFALSE
- 6.08.04.05The cavity helium bath maximum operational temperature shall be TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The cavity helium bath maximum operational pressure shall be TBD mbar05/16/2025In ProcessFALSE
- 6.08.04.05The cavity helium bath operation pressure stability shall be ±TBD mbar05/16/2025In ProcessFALSE
- 6.08.04.05The maximum helium supply operational temperature shall be TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The range of the helium supply operational pressure shall be TBD to TBD bar05/16/2025In ProcessFALSE
- 6.08.04.05The range of the combined helium return temperature shall be TBD to TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The range of the combined helium return pressure shall be TBD to TBD bar05/16/2025In ProcessFALSE
- 6.08.04.05The maximum sub-atmospheric helium return temperature shall be TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The maximum Subatmospheric helium return pressure shall be TBD mbar05/16/2025In ProcessFALSE
- 6.08.04.05The minimum cooldown rate of the SRF cavity between 300K and 4.5K shall be TBD K/hour05/16/2025In ProcessFALSE
- 6.08.04.05The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be TBD K/hour05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall achieve steady state temperature with the cavity bath at 4K in a maximum of TBD days05/16/2025In ProcessFALSE
- 6.08.04.05The chilled water and low-conductivity water operational temperature range shall be TBD to TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The chilled water and low-conductivity water operational pressure range shall be TBD to TBD bar05/16/2025In ProcessFALSE
- 6.08.04.05The minimum magnetic shield attenuation factor at SRF cavity equator shall be TBD05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall operate through a minimum of TBD thermal cycles05/16/2025In ProcessFALSE
- 6.08.04.05The minimum SRF Cavity Slow tuner minimum lifetime shall be TBD years05/16/2025In ProcessFALSE
- 6.08.04.05The minimum SRF CM Slow Tuner 1% range tuning cycles shall be TBD cycles05/16/2025In ProcessFALSE
- 6.08.04.05The minimum SRF CM Slow Tuner full range tuning cycles shall be TBD cycles05/16/2025In ProcessFALSE
- 6.08.04.05The manufactured SRF CM Cavity shall produce no field emission at TBD MV05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM components that are not replaceable in-situ shall be designed with a radiation tolerance greater than TBD MGy05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM components that are replaceable in-situ shall have a radiation tolerance greater than TBD kGy05/16/2025In ProcessFALSE
- 6.08.04.05The active SRF cavity tuning mechanism components (bearings/motor/piezo) shall be replaceable and maintainable in-situ.05/16/2025In ProcessFALSE
- 6.08.04.05All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM instrument should have maximized instruments that can be maintained and replaced in-situ05/16/2025In ProcessFALSE
- 6.08.04.05The SRF maximum (per cavity) RF longitudinal impedance shall be TBD MΩ GHz05/16/2025In ProcessFALSE
- 6.08.04.05The SRF maximum (per cavity) RF horizontal impedance shall be TBD MΩ/m05/16/2025In ProcessFALSE
- 6.08.04.05The SRF maximum (per cavity) RF vertical impedance shall be TBD MΩ/m05/16/2025In ProcessFALSE
- 6.08.04.05The minimum cavity aperture radius shall be TBD mm05/16/2025In ProcessFALSE
- 6.08.04.05The maximum broadband RF power emitted from the CM shall be TBD kW05/16/2025In ProcessFALSE
- 6.08.04.05The Maximum Quadrupole multipole content shall be TBD mT05/16/2025In ProcessFALSE
- 6.08.04.05The Maximum Sextupole multipole content shall be TBD mT/m05/16/2025In ProcessFALSE
- 6.08.04.05The Maximum Octupole multipole content shall be TBD T/m^205/16/2025In ProcessFALSE
- 6.08.04.05The Maximum Decapole multipole content shall be TBD T/m^305/16/2025In ProcessFALSE
- 6.08.04.05The SRF cavity minimum manufactured quality factor (Qo) shall be TBD05/16/2025In ProcessFALSE
- 6.08.04.05The SRF cavity minimum manufactured voltage shall be TBD MV05/16/2025In ProcessFALSE
- 6.08.04.05The SRF cavity fundamental power coupler Qext shall be TBD05/16/2025In ProcessFALSE
- 6.08.04.05The SRF cavity field probe Qext range shall be TBD05/16/2025In ProcessFALSE
- 6.08.04.05The SRF cavity nominal cold frequency shall be TBD MHz05/16/2025In ProcessFALSE
- 6.08.04.05The SRF cavity maximum Niobium temperature shall be TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The SRF Pressure sensitivity maximum shall be TBD Hz/mBar05/16/2025In ProcessFALSE
- 6.08.04.05The SRF maximum Lorentz force detuning shall be TBD Hz/(Mv/m)^205/16/2025In ProcessFALSE
- 6.08.04.05The warm beamline maximum vacuum shall be TBD mbar05/16/2025In ProcessFALSE
- 6.08.04.05The cold beamline maximum vacuum shall be TBD mbar05/16/2025In ProcessFALSE
- 6.08.04.05The beamline vacuum maximum leak rate shall be TBD mbar L/s05/16/2025In ProcessFALSE
- 6.08.04.05The warm insulating maximum vacuum shall be TBD mbar05/16/2025In ProcessFALSE
- 6.08.04.05The cold insulating maximum vacuum shall be TBD mbar05/16/2025In ProcessFALSE
- 6.08.04.05The insulating vacuum maximum leak rate shall be TBD mbar L/s05/16/2025In ProcessFALSE
- 6.08.04.05The minimum SRF Cavity Slow Tuner tuning range shall be shall be -TBD, +TBD kHz05/16/2025In ProcessFALSE
- 6.08.04.05The minimum SRF Cavity slow tuner tuning rate shall be TBD Hz/s05/16/2025In ProcessFALSE
- 6.08.04.05The maximum SRF Cavity Slow Tuner resolution shall be TBD Hz05/16/2025In ProcessFALSE
- 6.08.04.05The maximum SRF Cavity Slow Tuner hysteresis shall be ±TBD Hz05/16/2025In ProcessFALSE
- 6.08.04.05The external warm maximum allowable working pressure of the SRF cavity shall not exceed TBD bar05/16/2025In ProcessFALSE
- 6.08.04.05The external cold maximum allowable working pressure of the SRF cavity shall not exceed TBD bar05/16/2025In ProcessFALSE
- 6.08.04.05The internal maximum allowable working pressure of the SRF cavity shall not exceed TBD bar05/16/2025In ProcessFALSE
- 6.08.04.05All cryomodule surfaces accessible to workers shall be within the temperature range of TBD to TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by ASME B31.305/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by ASME BPVC05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by ASTM C105505/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by NFPA 7005/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by API 520 & API 52105/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by NFPA 70E05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by CGA S1.305/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by AWS05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM maximum length shall be TBD m05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM maximum width shall be TBD m05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM maximum height shall be TBD m05/16/2025In ProcessFALSE
- 6.08.04.05The distance from the beamline to the tunnel floor shall be TBD m05/16/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance in X shall be ±TBD μm05/16/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance in Y shall be ±TBD μm05/16/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance in Z shall be ±TBD mm05/16/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ±TBD degrees05/16/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ±TBD degrees05/16/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ±TBD degrees05/16/2025In ProcessFALSE
- 6.08.04.05Conditioning for individual cavities shall have a maximum average cryogenic power dissipation of TBD W05/16/2025In ProcessFALSE
- 6.08.04.05Conditioning for individual cavities shall be achieved with a maximum temperature of TBD K05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be capable of withstanding a maximum allowable vertical acceleration of TBD G05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be capable of withstanding a maximum allowable lateral acceleration of TBD G05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be capable of withstanding a maximum allowable beamline axis acceleration of TBD G05/16/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed to withstand a minimum tilt around the beamline axis (roll) of ±TBD radians05/16/2025In ProcessFALSE
- 6.08.04.05The SRF Cryomodule loaded quality factor shall be TBD ± TBD.05/16/2025In ProcessFALSE
- 6.08.04.05The SRF Cryomodule FPC external quality factor balance shall be TBD.05/16/2025In ProcessFALSE
ESR-PPD : EZSR Pulse Power Device System (WBS TBD)
ESR-PPD-INJ_EXT_PULSE_KICK : EIS RCS Injection/Extraction Pulsed Bump IM&HW System (WBS TBD)
ESR-PPD-INJ_EXT_PULSE_SEPTUM : EIS RCS Injection/Extraction Pulsed Septum System (WBS TBD)
ESR-PPD-INJ_EXT_PULSE_STRIPLINE_KICK : EIS RCS Injection/Extraction Pulse Power Strip-line Kicker System (WBS TBD)
Hadron Storage Ring Requirements
General, functional and performance requirements associated with the Hadron Storage Ring of the Electron Ion Collider.
- NameWBSDescriptionUpdatedStatusTBD
HSR : Hadron Storage Ring Performance Requirements (WBS 6.02.03)
- 6.02.03The range of unpolarized ion species currently produced by the Relativistic Heavy Ion Collider (RHIC) complex shall be preserved for Electron-Ion Collider (EIC) Hadron Storage Ring (HSR) operation (from deuterons to uranium) defined in the Master Parameter Table. Refer to [EIC-SEG-RSI-005].02/09/2026ApprovedFALSE
- 6.02.03The HSR shall deliver Protons bunches having at least a 70% polarization at full beam energy ready for collision.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall deliver 3He bunches having at least a 70% polarization at full beam energy ready for collision.02/09/2026ApprovedFALSE
- 6.02.03Design of the Hadron storage ring shall allow the possibility of future operation with a polarized deuteron beam.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall provide the capability to collide protons at beam energies of 41 GeV, and from 100 to 275 GeV.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall provide the capability to collide 3He at beam energies of 41 GeV/nucleon, and from 100 to 183 GeV/nucleon.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall provide the capability to collide electrons with Au ions at 41 GeV/nucleon and from 100 to 110 GeV/nucleon energies.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall provide the capability to collide other ion species at a maximum energy equivalent to a beam rigidity Bρ value of 916.67 Tm.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall provide the capability to vary the hadron revolution frequency to match it at different hadron energies (41 GeV/nucleon and 100 - 275 GeV/nucleon) with the revolution frequency of electron beam in the ESR.02/09/2026ApprovedFALSE
- 6.02.03The HSR Ion bunches shall meet the parameters specified for different species defined in MPT. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.03The HSR shall be able to provide beams of required species for collision having the beam currents as specified in the MPT. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection system transport line shall be modified to add septum magnets in the Q3-Q4 warm straight section of the HSR on 4 o’clock side of the IR4 for hadron beam transfer into the HSR beam pipe.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall utilize a vacuum system capable of operating with peak and average beam current defined in MPT. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.03The Relativistic Heavy Ion Collider (RHIC) lattice shall be preserved and where required modified for Electron Ion Collider (EIC) Hadron Storage Ring (HSR) operations defined in the Master Parameter Table (MPT). [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.03The HSR shall have a instrumentation system to operate for all beam species which need monitoring and will, where possible utilize the existing RHIC instrumentation system.02/09/2026ApprovedFALSE
- 6.02.03All HSR components and systems shall be designed and installed in line with all relevant regulatory codes and in full compliance with BNL SBMS.02/09/2026ApprovedFALSE
- 6.02.03The HSR uptime shall be consistent with the overall uptime requirements of the EIC.02/09/2026ApprovedFALSE
- 6.02.03The operational availability design target for the HSR Injection System shall be consistent with the operational availability target for the overall EIC as set forth in Electron-Ion Collider Global Requirements. Refer to [EIC-ORG-PLN-010].02/09/2026ApprovedFALSE
- 6.02.03The HSR shall meet the beam parameters specified for different species at injection defined in MPT. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall utilize the existing RHIC injector chain upstream of the RHIC-ATR D26 Dipole magnet with no modifications.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system, consisting of the transport beamline, septum magnet and injection kickers, shall be capable of transporting a maximum beam rigidity of 81.12Tm from the transport line to IR4 central area and injecting it into the HSR.02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection System design shall use a warm transport line in arc 6-4 as continuation of the Injection line to transport the hadron beam to the injection system located in IR4.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection transport beamline shall be able to transport polarized beam with less than 5% polarization loss.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to inject all beam species with less than 5% beam emittance increase.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to fill the HSR with 290 consecutive bunches without interruption.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to fill the HSR with one(1) bunch per AGS cycle for polarized proton, two(2) bunches per AGS cycle for ion beams.02/09/2026ApprovedFALSE
- 6.02.03The operational availability design target for the HSR Injection System shall be consistent with the operational availability target for the overall EIC as set forth in Electron-Ion Collider Global Requirements. Refer to [EIC-ORG-PLN-010].02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system transfer line shall provide the following physical aperture:02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection system transport line shall be modified to add septum magnets in the Q3-Q4 warm straight section of the HSR on 4 o’clock side of the IR4 for hadron beam transfer into the HSR beam pipe.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall utilize an Injection system to provide the ability for single bunch transport and Injection at IR4.02/09/2026ApprovedFALSE
- 6.02.03Any reused existing RHIC-ATR transfer line magnets shall meet the requirements of the new approved HSR Injection line lattice.02/09/2026ApprovedFALSE
- 6.02.03New magnets shall only be used where any available existing magnets do not meet the requirements of the new approved HSR Injection line lattice.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kickers shall provide a half aperture greater than 10σ for the stored beam at Collison energies.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kickers shall provide a half aperture greater than 7σ for the stored beam at injection energies.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kickers shall provide a half aperture greater than 6σ for the injected beam.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system shall be able to deflect the injected beam to be on axis02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system shall be installed in the straight section of the IR4 area.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system shall be capable of single-bunch on-axis injection to fill the ring with 290 bunches.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system rise time shall be short enough so that it does not step on the previous bunch.02/09/2026ApprovedFALSE
- 6.02.03The present RHIC injection kicker system including the Lambertson magnet and current injection kicker magnets at the 5 o’clock area shall be removed.02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection System magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets02/09/2026ApprovedFALSE
- 6.02.03The vacuum level in the HSR transport line shall be kept at the same level as in the current RHIC-ATR line.02/09/2026ApprovedFALSE
- 6.02.03A ~20m section of the warm injection beamline near the HSR including the injection septum shall have a vacuum pressure of ~1E-10 torr or better, after baking .02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall utilize the existing RHIC injector chain upstream of the RHIC-ATR D26 Dipole magnet with no modifications.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system, consisting of the transport beamline, septum magnet and injection kickers, shall be capable of transporting a maximum beam rigidity of 81.12Tm from the transport line to IR4 central area and injecting it into the HSR.02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection System design shall use a warm transport line in arc 6-4 as continuation of the Injection line to transport the hadron beam to the injection system located in IR4.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection transport beamline shall be able to transport polarized beam with less than 5% polarization loss.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to inject all beam species with less than 5% beam emittance increase.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to fill the HSR with 290 consecutive bunches without interruption.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to fill the HSR with one(1) bunch per AGS cycle for polarized proton, two(2) bunches per AGS cycle for ion beams.02/09/2026ApprovedFALSE
- 6.02.03The operational availability design target for the HSR Injection System shall be consistent with the operational availability target for the overall EIC as set forth in Electron-Ion Collider Global Requirements. Refer to [EIC-ORG-PLN-010].02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system transfer line shall provide the following physical aperture:02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection system transport line shall be modified to add septum magnets in the Q3-Q4 warm straight section of the HSR on 4 o’clock side of the IR4 for hadron beam transfer into the HSR beam pipe.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall utilize an RF system capable of operating over the parameters defined in MPT. [Document#:EIC-SEG-RSI-005]02/09/2026ApprovedFALSE
- 6.02.04.02The HSR shall have a control system which can operate the HSR consistent will the overall control of the other EIC system's and to ensure the HSR meets all the Physics requirements needed to deliver the physics goals of the EIC.02/09/2026ApprovedFALSE
- 6.02.04.04The HSR shall have a cryogenic system to cool and operate all elements which need cryogenic cooling and will, where possible utilize the existing RHIC cooling system.02/09/2026ApprovedFALSE
- 6.02.03The HSR proton beam shall be ramped from injection energy to a maximum collision energy of 275 GeV.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall be designed for changing beam optics between the injection configuration to collision configuration with beam in the machine.02/09/2026ApprovedFALSE
- 6.02.03The HSR beam at collision energies shall be synchronized to the revolution frequency of the electron beam.02/09/2026ApprovedFALSE
- 6.02.03The HSR systems shall provide the capability to operate with a radial shift, having a full range of +/-21 (mm) beam orbit in all arcs.02/09/2026ApprovedFALSE
- 6.02.03The HSR systems shall operate with a vertical orbit excursion having a full range of +/-2 (mm) beam orbit in all arcs.02/09/2026ApprovedFALSE
- 6.02.03The HSR decoupling system shall provide the capability to maintain a flat beam with the required beam size ratios.02/09/2026ApprovedFALSE
- 6.02.03The HSR orbit tune chromaticity correction, nonlinear correction and gamma-T jump systems, shall be provided with the same capability as in the present RHIC machine.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall provide a dynamic aperture greater than 6σ under colliding beam conditions.02/09/2026ApprovedFALSE
- 6.02.03The physical aperture for the circulating hadron beam at the store energies shall be greater than 10σ in the horizontal and vertical planes.02/09/2026ApprovedFALSE
- 6.02.03The physical aperture for the circulating hadron beam at the injection energy shall be greater than 7σ in the horizontal and 6σ in the vertical plane in all locations, for normalized beam emittance of 2.5 um.02/09/2026ApprovedFALSE
- 6.02.03The physical aperture for the circulating hadron beam at the injection energy in the beam dump beam pipe shall be greater than 6σ in the horizontal and vertical planes.02/09/2026In ProcessFALSE
- 6.02.03The apertures of the downstream, near-IR magnets, within the IR hadron lattice, shall be large enough to transport a 4 mrad cone of neutral particles from the IP without obstruction.02/09/2026ApprovedFALSE
- 6.02.03The apertures of the forward side near-IR magnets, within the IR hadron lattice, shall be large enough to transport particles having a transverse momentum of up to 1.3 GeV/c with a 275GeV proton beam without obstruction.02/09/2026ApprovedFALSE
- 6.02.03The apertures of the forward side, near-IR magnets, within the IR hadron lattice, shall accommodate off beam-axis detectors which can detect forward scattered protons with a transverse momentum of 0.2GeV to 1.3GeV at a proton beam energy of 275GeV.02/09/2026ApprovedFALSE
- 6.02.03At the store energies the vacuum chamber shall provide sufficient horizontal and vertical aperture to accommodate, a +/-10 sigma beam, where the vertical RMS beam size is based on the emittance of a fully coupled beam. In the arcs an additional 20 mm horizontally and 2 mm vertically shall be included to account for the radially shifted beam and orbit errors.02/09/2026ApprovedFALSE
- 6.02.03The HSR alignment requirements are established by dynamic aperture and polarization tracking. The HSR RMS alignment tolerances shall be such that all the beam parameters listed in the MPT can be satisfied. Refer to [EIC-SEG-RSI-005].02/09/2026ApprovedFALSE
- 6.02.03The operational availability design target for the HSR shall be consistent with the operational availability target for the overall EIC as set forth in Electron-Ion Collider Global Requirements Document. Refer to [EIC-ORG-PLN-010].02/09/2026ApprovedFALSE
- 6.02.03The HSR shall deliver spin polarized ion beams with store-averaged polarization of at least 70 percent at collision.02/09/2026ApprovedFALSE
- 6.02.03The HSR lattice shall have features to preserve the polarization from injection to collision energies.02/09/2026ApprovedFALSE
- 6.02.03The HSR lattice will utlise RHIC snakes and spin rotators to control the hadron spin.02/09/2026ApprovedFALSE
- 6.02.03The HSR shall be capable of delivering bunches with longitudinal spins to the IP.02/09/2026ApprovedFALSE
- 6.02.03The HSR Lattice shall contain provisions for correctors such as horizontal and vertical dipole correctors, skew quadrupoles, octupoles, etc. as needed.02/09/2026ApprovedFALSE
HSR-ARC : Hadron Storage Ring ARC Sector (WBS 6.02.03)
- 6.02.03The HSR lattice will utlise the existing RHIC arc sections02/09/2026ApprovedFALSE
- 6.02.03The changes required to the existing RHIC arc sections shall be kept to a minimum02/09/2026ApprovedFALSE
- 6.02.03For operation in the energy range 100-275 GeV the HSR shall use 6 Yellow sextants.02/09/2026ApprovedFALSE
- 6.02.03The inner arc 12-2 shall be used instead of 12-2 outer arc for maintaining synchronization of the hadron beam at 41 GeV/nucleon beam energy with the electron beam.02/09/2026ApprovedFALSE
- 6.02.03Switchyards on each side of the 12-2 arc, in IR12 and in IR2, shall be in place to redirect beam at different energies to the respective arc.02/09/2026ApprovedFALSE
HSR-CRYO : Hadron Storage Ring Cryogenic System (WBS 6.02.03)
- 6.02.04.04The cryogenic system shall provide enough cooling power to the superconducting magnets in HSR for them to operate safely.02/09/2026ApprovedFALSE
- 6.02.04.04The cryogenic system shall provide enough cooling power to the superconducting RF cavities in HSR for them to operate safely.02/09/2026ApprovedFALSE
- 6.02.04.04All cryogenic components shall meet the relevant Cryogenic pressure design codes ASME B31.3 etc.02/09/2026ApprovedFALSE
HSR-INJ : Hadron Storage Ring Injection System (WBS 6.02.03)
- 6.02.03The HSR injection line magnets excluding the induction septum shall provide a half physical aperture greater than 6σ for the injected beam.02/09/2026ApprovedFALSE
- 6.02.03The HSR induction septum shall provide a half physical aperture greater than 5σ for the injected beam.02/09/2026ApprovedFALSE
- 6.02.03The HSR induction septum shall provide a half physical aperture greater than 6σ for the circulating beam.02/09/2026ApprovedFALSE
- HSR-INJ EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.03The HSR injection system shall utilize the existing RHIC injector chain upstream of the RHIC-ATR D26 Dipole magnet with no modifications.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system, consisting of the transport beamline, septum magnet and injection kickers, shall be capable of transporting a maximum beam rigidity of 81.12Tm from the transport line to IR4 central area and injecting it into the HSR.02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection System design shall use a warm transport line in arc 6-4 as continuation of the Injection line to transport the hadron beam to the injection system located in IR4.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection transport beamline shall be able to transport polarized beam with less than 5% polarization loss.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to inject all beam species with less than 5% beam emittance increase.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to fill the HSR with 290 consecutive bunches without interruption.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system shall be able to fill the HSR with one(1) bunch per AGS cycle for polarized proton, two(2) bunches per AGS cycle for ion beams.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection system transfer line shall provide the following physical aperture:02/09/2026ApprovedFALSE
- 6.02.03The operational availability design target for the HSR Injection System shall be consistent with the operational availability target for the overall EIC as set forth in Electron-Ion Collider Global Requirements. Refer to [EIC-ORG-PLN-010].02/09/2026ApprovedFALSE
- 6.02.03The HSR Injection system transport line shall be modified to add septum magnets in the Q3-Q4 warm straight section of the HSR on 4 o’clock side of the IR4 for hadron beam transfer into the HSR beam pipe.02/09/2026ApprovedFALSE
HSR-INJ-PP : Hadron Storage Ring Injection System Pulsed Power (WBS 6.02.03)
- HSR-INJ-PP EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.03The HSR injection kickers shall provide a half aperture greater than 10σ for the stored beam at Collison energies.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kickers shall provide a half aperture greater than 7σ for the stored beam at injection energies.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kickers shall provide a half aperture greater than 6σ for the injected beam.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system shall be able to deflect the injected beam to be on axis02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system shall be installed in the straight section of the IR4 area.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system shall be capable of single-bunch on-axis injection to fill the ring with 290 bunches.02/09/2026ApprovedFALSE
- 6.02.03The HSR injection kicker system rise time shall be short enough so that it does not step on the previous bunch.02/09/2026ApprovedFALSE
- 6.02.03The present RHIC injection kicker system including the Lambertson magnet and current injection kicker magnets at the 5 o’clock area shall be removed.02/09/2026ApprovedFALSE
HSR-INJ-MAG : HSR Hadron Ring Injection Magnets (WBS 6.02.03.03)
- 6.02.03The HSR injection system shall have two septa, one DC septum and one induction septum.02/09/2026ApprovedFALSE
- 6.02.03The septa of HSR injection system shall provide a total bending angle of 69 (mrad).02/09/2026ApprovedFALSE
- HSR-INJ-MAG EXTERNALSRequirements who's parents are in other sub-systems.
- 6.02.03Any reused existing RHIC-ATR transfer line magnets shall meet the requirements of the new approved HSR Injection line lattice.02/09/2026ApprovedFALSE
- 6.02.03New magnets shall only be used where any available existing magnets do not meet the requirements of the new approved HSR Injection line lattice.02/09/2026ApprovedFALSE
HSR-INJ-MAG-DV1 : HSR Injector Magnet D1 (WBS 6.02.03.03.01)
- 6.02.03.03.01The magnet shall use am existing refurbished ATR DV1 dipole to provide a a single function vertical bending dipole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.01The physical length of the magnet shall be less than or equal to <1(m)02/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet shall have a gap capable of accommodating a beampipe with an OD of 36(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.01The magnet shall be able to fit within the following volume constraints:02/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.02/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet field axis displacement and rotational alignment shall utilize the existing fiducials to determine the field center and rotational alignment.03/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.01The Integrated Dipole Field B Shall be = 0.14(Tm)02/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-340(um) dy=+/-170(um) dz= na02/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.085 (mRad) About Y=+/-0.17 (mRad) About Z=na02/09/2026In ProcessFALSE
- 6.02.03.03.01This is an existing ATR magnet the magnet bore field should have the same multipole content as per the existing combined function ATR magnets. If the magnet needs to be repositioned it shall be confirmed the field quality of the bore remains unchanged.03/02/2026In ProcessFALSE
- 6.02.03.03.01The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.01The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0103/02/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.0102/09/2026In ProcessFALSE
- 6.02.03.03.01The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.02/09/2026In ProcessFALSE
- 6.02.03.03.01Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team02/09/2026In ProcessFALSE
HSR-INJ-MAG-DW0 : HSR Injector Magnet D2 (WBS 6.02.03.03.01)
- 6.02.03.03.01The magnet shall use an existingnew HSR DW0 dipole to porvide a single function horizontal bending dipole field centered on the injected beam axis which directs the injected beam to merge with the circulating beam .03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-340(um) dy=+/-170(um) dz= na03/02/2026In ProcessFALSE
- The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.042 (mRad) About Y=+/-0.085 (mRad) About Z=na03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
HSR-INJ-MAG-CHV1 : HSR Injector Magnet Ver Corr (WBS 6.02.03.03.02)
- 6.02.03.03.02The magnet shall use the existing ATR vertical and horizontal correctors to create a dual plane corrector by placing the horizontal corrector and vertically correctors, consecutively with both correctors centered on the injected beam axis.03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The physical length of the magnet shall be less than or equal to <0.41(m) per corrector total length being less than 0.82(m) ( 2x0.41(m))03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet shall have a gap capable of accommodating a beampipe with an OD of 36(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet shall be able to fit within the following volume constraints:02/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.02/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet field axis displacement and rotational alignment shall utilize the existing fiducials to determine the field center and rotational alignment.03/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.02The Integrated Dipole Field B Shall be = 0.058(Tm)03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-340(um) dy=+/-170(um) dz= na02/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.207 (mRad) About Y=+/-0.415 (mRad) About Z=na03/02/2026In ProcessFALSE
- 6.02.03.03.02This is an existing ATR magnet the magnet bore field is assumed to have the same multipole content as per the existing magnet. If the magnet needs to be repositioned it shall be confirmed the field quality of the bore remains unchanged.03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.03/02/2026In ProcessFALSE
- 6.02.03.03.02Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team03/02/2026In ProcessFALSE
HSR-INJ-MAG-CHV2 : HSR Injector Magnet DVERT_1 (WBS 6.02.03.03.02)
- 6.02.03.03.02The magnet shall have a dual plane corrector with a combined dual function horizontal bending dipole corrector and a vertically bending dipole field both centered on the injected beam axis.03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The physical length of the magnet shall be less than or equal to <0.4(m)03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet shall have a gap capable of accommodating a beampipe with an OD of 36(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet shall be able to fit within the following volume constraints:02/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.02/09/2026In ProcessFALSE
- 6.02.03.03.02nan03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnetic field axis displacement tolerances: The field center shall be identified to within dx=+/-500(um) dy=+/-500(um) dz=+/-500(um)03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnetic field rotational alignment tolerances: The field axes shall be identified to within the following rotational tolerances About X=+/-0.5(mrad) About Y=+/-0.5(mrad) About Z=+/-0.5(mrad)03/02/2026In ProcessFALSE
- 6.02.03.03.02The Integrated Dipole Field B Shall be = 0.025(Tm)03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-340(um) dy=+/-170(um) dz= na02/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.212 (mRad) About Y=+/-0.425 (mRad) About Z=na03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet field homogeneity shall be measured within the following constraints:02/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- 6.02.03.03.02The magnet bore field SHALL have the following multipole content as the existing ATR Dual plane corrector magnet design. Notes: The units are specified in parts of 10-4 of the main components.02/09/2026In ProcessFALSE
- 6.02.03.03.02b1=10000 , a1+/- 1000002/09/2026In ProcessFALSE
- 6.02.03.03.02b2 +/- TBD , a2=TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b3 < +/- TBD , a3 < +/-TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b4 < +/- TBD , a4 < +/-TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b5 < +/- TBD , a5 < +/- TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b6 < +/- TBD , a6 < +/- TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b7 < +/- TBD , a7 < +/- TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b8 < +/- TBD , a8 < +/- TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b9 < +/- TBD , a9 < +/- TBD02/09/2026In ProcessFALSE
- 6.02.03.03.02b10 < +/- TBD , a10 < +/- TBD02/09/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0203/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.0202/09/2026In ProcessFALSE
- 6.02.03.03.02The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.02/09/2026In ProcessFALSE
- 6.02.03.03.02Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team02/09/2026In ProcessFALSE
HSR-INJ-MAG-Q:100 : HSR Injector Magnet Q2 (WBS 6.02.03.03.03)
- The magnet shall provide a single function normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The physical length of the magnet shall be less than or equal to <1(m)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall have a pole tip radius capable of accommodating a beampipe with an OD of 36(mm)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall be able to fit within the following volume constraints:03/02/2026In ProcessFALSE
- The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.03/02/2026In ProcessFALSE
- The magnet field axis displacement and rotational alignment shall be identified by applying fiducials to locate the field center and rotational alignment of the magnet, within the following limits.03/02/2026In ProcessFALSE
- The magnetic field axis displacement tolerances: The field center shall be identified to within dx=+/-500(um) dy=+/-500(um) dz=+/-500(um)03/02/2026In ProcessFALSE
- The magnetic field rotational alignment tolerances: The field axes shall be identified to within the following rotational tolerances About X=+/-0.5(mrad) About Y=+/-0.5(mrad) About Z=+/-0.5(mrad)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The Integrated Gradient Field G Shall be = 20.47(T)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)03/02/2026In ProcessFALSE
- The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-450(um) dy=+/-230(um) dz= na03/02/2026In ProcessFALSE
- The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.115 (mRad) About Y=+/-0.225 (mRad) About Z=TBD03/02/2026In ProcessFALSE
- The field homogenity required by the magnet needs to be better than 1e-3 units.03/02/2026In ProcessFALSE
- The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- The magnet bore field Shall have the following multipole content Notes: The units are specified in parts of 10-4 of the main components.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- b2 = 10000 , a2 = +/-1003/02/2026In ProcessFALSE
- b3 < +/- 10 , a3 < +/- 1003/02/2026In ProcessFALSE
- b4 < +/- 10 , a4 < +/-1003/02/2026In ProcessFALSE
- b5 < +/- 10 , a5 < +/- 1003/02/2026In ProcessFALSE
- b6 < +/- 10 , a6 < +/- 1003/02/2026In ProcessFALSE
- b7 < +/- 10 , a7 < +/- 1003/02/2026In ProcessFALSE
- b8 < +/- 10 , a8 < +/- 1003/02/2026In ProcessFALSE
- b9 < +/- 10 , a9 < +/- 1003/02/2026In ProcessFALSE
- b10 < +/- 10 , a10 < +/- 1003/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.03/02/2026In ProcessFALSE
- Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team03/02/2026In ProcessFALSE
HSR-INJ-MAG-QD1 : HSR Injector Magnet QD1 (WBS 6.02.03.03.03)
- 6.02.03.03.03The magnet shall use an existing ATR combined function magnet to provide a combined function horizontal bending dipole field and a normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The physical length of the magnet shall be less than or equal to <3.66(m)02/09/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall have a gap capable of accommodating a beampipe with an OD of 36(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall be able to fit within the following volume constraints:02/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet field axis displacement and rotational alignment shall utilize the existing fiducials to determine the field center and rotational alignment.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The Integrated Dipole Field B Shall be = 3.96(Tm)03/02/2026In ProcessFALSE
- 6.02.03.03.03The Integrated Gradient Field G Shall be = 41.89(T)03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-340(um) dy=+/-170(um) dz= na03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.023 (mRad) About Y=+/-0.046 (mRad) About Z=na03/02/2026In ProcessFALSE
- 6.02.03.03.03This is an existing magnet the magnet bore field should have the same multipole content as per the existing combined function ATR magnets. If the magnet needs to be repositioned it shall be confirmed the field quality of the bore remains unchanged.03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.02/09/2026In ProcessFALSE
- 6.02.03.03.03Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team03/02/2026In ProcessFALSE
HSR-INJ-MAG-QD2 : (WBS 6.02.03.03.03)
- 6.02.03.03.03The magnet shall use an existing ATR combined function magnet to provide a combined function horizontal bending dipole field and a normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The physical length of the magnet shall be less than or equal to <2.95(m)02/09/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall have a gap capable of accommodating a beampipe with an OD of 36(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall be able to fit within the following volume constraints:02/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.02/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet field axis displacement and rotational alignment shall utilize the existing fiducials to determine the field center and rotational alignment.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The Integrated Dipole Field B Shall be = 3.19(Tm)03/02/2026In ProcessFALSE
- 6.02.03.03.03The Integrated Gradient Field G Shall be = 33.74(T)03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-340(um) dy=+/-170(um) dz= na03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.029 (mRad) About Y=+/-0.058 (mRad) About Z=TBD03/02/2026In ProcessFALSE
- 6.02.03.03.03This is an existing magnet the magnet bore field should have the same multipole content as per the existing combined function ATR magnets. If the magnet needs to be repositioned it shall be confirmed the field quality of the bore remains unchanged.03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.02/09/2026In ProcessFALSE
- 6.02.03.03.03Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team03/02/2026In ProcessFALSE
HSR-INJ-MAG-QLA:100 : HSR Injector Magnet Q3 (WBS 6.02.03.03.03)
- The magnet shall provide a single function normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The physical length of the magnet shall be less than or equal to <1(m)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall have a pole tip radius capable of accommodating a beampipe with associated bake out heaters having a total OD of 57(mm)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall be able to fit within the following volume constraints:03/02/2026In ProcessFALSE
- The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.03/02/2026In ProcessFALSE
- The magnet field axis displacement and rotational alignment shall be identified by applying fiducials to locate the field center and rotational alignment of the magnet, within the following limits.03/02/2026In ProcessFALSE
- The magnetic field axis displacement tolerances: The field center shall be identified to within dx=+/-500(um) dy=+/-500(um) dz=+/-500(um)03/02/2026In ProcessFALSE
- The magnetic field rotational alignment tolerances: The field axes shall be identified to within the following rotational tolerances About X=+/-0.5(mrad) About Y=+/-0.5(mrad) About Z=+/-0.5(mrad)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The Integrated Gradient Field G Shall be = 13.48(T)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)03/02/2026In ProcessFALSE
- The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-450(um) dy=+/-230(um) dz= na03/02/2026In ProcessFALSE
- The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.115 (mRad) About Y=+/-0.225 (mRad) About Z=TBD03/02/2026In ProcessFALSE
- The field homogenity required by the magnet needs to be better than 1e-3 units.03/02/2026In ProcessFALSE
- The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- The magnet bore field Shall have the following multipole content Notes: The units are specified in parts of 10-4 of the main components.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- b2 = 10000 , a2 = +/-1003/02/2026In ProcessFALSE
- b3 < +/- 10 , a3 < +/- 1003/02/2026In ProcessFALSE
- b4 < +/- 10 , a4 < +/-1003/02/2026In ProcessFALSE
- b5 < +/- 10 , a5 < +/- 1003/02/2026In ProcessFALSE
- b6 < +/- 10 , a6 < +/- 1003/02/2026In ProcessFALSE
- b7 < +/- 10 , a7 < +/- 1003/02/2026In ProcessFALSE
- b8 < +/- 10 , a8 < +/- 1003/02/2026In ProcessFALSE
- b9 < +/- 10 , a9 < +/- 1003/02/2026In ProcessFALSE
- b10 < +/- 10 , a10 < +/- 1003/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.03/02/2026In ProcessFALSE
- Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team03/02/2026In ProcessFALSE
HSR-INJ-MAG-YQ1 : (WBS 6.02.03.03.03)
- 6.02.03.03.03The magnet shall use an existing refurbished ATR YQ1 quadrupole to provide a single function normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The physical length of the magnet shall be less than or equal to <0.73(m)02/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall have a pole tip radius capable of accommodating a beampipe with an OD of 36(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall be able to fit within the following volume constraints:02/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.02/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet field axis displacement and rotational alignment shall utilize the existing fiducials to determine the field center and rotational alignment.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The Integrated Gradient Field G Shall be = 12.11(T)02/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)02/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-450(um) dy=+/-230(um) dz= na03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.158 (mRad) About Y=+/-0.308 (mRad) About Z=TBD03/02/2026In ProcessFALSE
- 6.02.03.03.03This is an existing ATR magnet the magnet bore field should have the same multipole content as per the existing combined function ATR magnets. If the magnet needs to be repositioned it shall be confirmed the field quality of the bore remains unchanged.03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet field homogeneity shall be measured at a reference radius of 23(mm)03/02/2026In ProcessFALSE
- 6.02.03.03.03The magnet field homogeneity shall be measured at a reference field comparable to the nominal operating field of the magnet.03/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0303/02/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet cooling system shall be capable of maintaning an operational temperature range of +25 (C) to +35 (C).03/02/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet coils shall pass a Hi-Pot test to confirm the magnet can meet the maximum Voltage Vmax seen in operation +500 Volts, i.e. Vmax+500(V)03/02/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.0302/09/2026In ProcessFALSE
- 6.02.03.03.03The magnet shall be able to sustain 30 years of EIC operation under nominal conditions. During this time the magnet is expected to survive 30000 power cycles.02/09/2026In ProcessFALSE
- 6.02.03.03.03Over its planned life of 30(yrs), the magnet Shall be able to survive a total integrated absorbed radiation dose from 1(MGy) to 20(MGy) without damage. The upper limit should be taken as a guide for the design process. The actual upper limit the magnet will see in operation will need further analysis and will need to be confirmed by the EIC radiation physics team02/09/2026In ProcessFALSE
HSR-INJ-MAG-Q:50 : HSR Injector Magnet Q2 (WBS 6.02.03.03.04)
- The magnet shall use am existing refurbished APS Q50 quadrupole magnet to provide a single function normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall be able to fit within the following volume constraints:03/02/2026In ProcessFALSE
- The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)03/02/2026In ProcessFALSE
- The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-450(um) dy=+/-230(um) dz= na03/02/2026In ProcessFALSE
- The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.23 (mRad) About Y=+/-0.45 (mRad) About Z=TBD03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
HSR-INJ-MAG-Q:60 : (WBS 6.02.03.03.04)
- The magnet shall use am existing refurbished APS Q60 quadrupole magnet to provide a single function normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall be able to fit within the following volume constraints:03/02/2026In ProcessFALSE
- The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)03/02/2026In ProcessFALSE
- The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-450(um) dy=+/-230(um) dz= na03/02/2026In ProcessFALSE
- The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.192 (mRad) About Y=+/-0.375 (mRad) About Z=TBD03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
HSR-INJ-MAG-Q:80 : (WBS 6.02.03.03.04)
- The magnet shall use am existing refurbished APS Q80 quadrupole magnet to provide a single function normal quadrupole field centered on the injected beam axis.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet shall be able to fit within the following volume constraints:03/02/2026In ProcessFALSE
- The magnet volume occupied shall be approved by the EIC engineering team to ensure the design does not impede any other EIC components or block egress.03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- The magnet cryostat installation position and alignment with respect to the nominal beam position defined in the lattice file and axis shall be within the following limits: (Note: Z is along the beam axis)03/02/2026In ProcessFALSE
- The magnet install center displacement shall be aligned with respect to the specified lattice field center position dx=+/-450(um) dy=+/-230(um) dz= na03/02/2026In ProcessFALSE
- The magnet rotational alignment shall be aligned with respect to the specified lattice beam axis About X=+/-0.144 (mRad) About Y=+/-0.281 (mRad) About Z=TBD03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
- 03/02/2026In ProcessFALSE
HSR-INJ-MAG-INDSEPT : HSR Injector Magnet Induction Septum (WBS 6.02.03.11.02.01)
- 6.02.03.11.02.01.01The magnet shall provide a single function horizontal bending dipole field centered on the injected beam axis which directs the injected beam to merge with the circulating beam .03/02/2026In ProcessFALSE
- 6.02.03.11.02.01.0102/09/2026In ProcessFALSE
- 6.02.03.11.02.01.0102/09/2026In ProcessFALSE
- 6.02.03.11.02.01.0102/09/2026In ProcessFALSE
- 6.02.03.11.02.01.0102/09/2026In ProcessFALSE
- 6.02.03.11.02.01.01The physical length of the magnet shall be less than or equal to <1.5(m)02/09/2026In Process