Record Date: 10/10/2024 00:00 | |||
Identifier: | F-DET-ANC-B0.3 | WBS: | 6.10.11 |
Date Modified: | TBD: | FALSE | |
Status Date: | Status: | APPROVED | |
Description: | B0-system must fit and be integrated into the warm area of the B0-dipole. The B0-systems shall require appropriate power and cabling to support operation of the detector elements.Should have survey marks to determine their physical location | ||
Comments: |
No archive versions
Parents | |
G-DET-ANC.4 | The EIC ancillary detectors will be functionally integrated with the other detectors, with the interaction region components, and with the facility infrastructure. |
G-DET-ANC.5 | The EIC ancillary detectors must fit within the available space in the experimental hall, and be consistent with the available infrastructure and resources. |
Children | |
P-DET-ANC-B0.5 | B0 tracker dimensions shall be 24 [cm] in X and 30 [cm] in Y |
P-DET-ANC-B0.6 | B0 tracker shall provide momentum resolution < 6%. |
P-DET-ANC-B0.7 | B0 calorimeter dimensions shall be 24 [cm] in X and 30 [cm] in Y |
P-DET-ANC-B0.8 | B0 calorimeter thickness shall be 7 [cm] 6.4( X/X0) |
P-DET-ANC-B0.9 | B0 calorimeter shall provide high granularity which is capable of distinguishing two showers ( calorimeter transverse cell size shall be 2 [cm] in X and 2 [cm] in Y |
P-DET-ANC-B0.11 | detectors and readout electronics must tolerate the magnetic field in the subsystem location. |