Electron Storage Ring Requirements
Electron Ion Collider
Electron Storage Ring Requirements
General, functional and performance requirements associated with the Electron Storage Ring of the Electron Ion Collider.
- NameWBSDescriptionUpdatedStatusTBD
ESR : Electron Storage Ring Interface Requirements
- 6.04The ESR shall be able to accept single bunches of spin-polarized electrons as injected from the RCS at the energies specified in the MPT. [Document#:EIC-SEG-RSI-005]01/27/2025ApprovedFALSE
- 6.04All quadrupoles shall be designed to facilitate beam based alignment.01/27/2025ReviewedFALSE
- 6.04The main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.01/27/2025ReviewedFALSE
- 6.04The ESR alignment requirements are established by dynamic aperture and polarization tracking. The ESR RMS alignment tolerances shall be such that all the beam parameter listed in [5.9] can be satisfied.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice shall provide a minimum dynamic aperture of 10 sigma w.r.t Gaussian electron beam distribution in all three dimensions, horizontal, vertical, and longitudinal. With the vertical emittance being half the horizontal design emittance.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.05The maximum beam excursion orbit shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04The ESR Lattice shall contain provisions for correctors such as skew quadrupoles, Dipole correctors etc. as needed.01/27/2025ApprovedFALSE
- 6.04The ESR shall have a depolarization time long enough to ensure an average polarization of 70%, assuming an injected polarization level of 80% and a bunch replacement rate of two bunches per second.01/27/2025ApprovedFALSE
- 6.03.03The ESR extraction system shall have another three bump magnets, its own septum magnet and use the strip line injection kicker to extract and replace the circulating bunches.01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall provide sufficient horizontal and vertical aperture to accommodate; a +/-15 sigma beam, where the vertical RMS beam size is based on the emittance of a fully coupled beam, plus an additional 10 mm horizontally and 5 mm vertically margin to account for expected orbit errors.01/27/2025In ProcessFALSE
- 6.04.05The typical (standard) vacuum chamber aperture shall be 80 x 36 mm.01/27/2025ApprovedFALSE
- 6.04.05Special aperture requirements and/or aperture file shall be provided by or approved by physics.01/27/2025ApprovedFALSE
- 6.04.05The dynamic pressure around the ring shall be consistent with a beam gas lifetime of >10[hrs] with the design currents after an integrated beam current of 1000 [A.h].01/27/2025In ProcessFALSE
- 6.04.05There shall be no upper pressure limit as long as the average pressure is maintained.01/27/2025ApprovedFALSE
- 6.04.05The average vacuum level in the ESR Arc sections after conditioning (for 1000Ahrs) shall be <5x10-9 Torr.01/27/2025ApprovedFALSE
- 6.04.05On 15 m on each side (or one vacuum sector) of the SRF cavities shall be processed to class ISO 5.01/27/2025ApprovedFALSE
- 6.04.05There shall be no pressure bumps in the ring exceeding (TBD)[Torr]01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber and all its components shall be designed to withstand a total synchrotron radiation load of 10 MW, considering the uneven linear load particularly related to the super-bends.01/27/2025In ProcessFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04.05The vacuum chamber material shall be chosen such that the SR power can be intercepted by the arc chambers and in addition good radiation shielding will be provided.01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04.05The impedance of the entire ESR vacuum system, including the interaction regions in IR 6 and IR 8, shall allow for the bunch intensities, beam currents, and bunch numbers contained in [5.9].01/27/2025In ProcessFALSE
- 6.04.05The vacuum system global impedance shall be less than the impedance budget as provided by accelerator physics.01/27/2025ApprovedFALSE
- 6.04The ESR shall deliver a spin polarized electron beam with time-averaged polarization of at least 70 percent.01/27/2025ApprovedFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
- 6.04The polarization lifetime of the beam in the ESR shall be maximized to maximize time-averaged polarization based on the given replacement frequency and polarization degree of the bunches provided by the injector.01/27/2025ApprovedFALSE
- 6.04The polarization lifetime maximum optimization in the ESR shall be accomplished by proper spin matching which minimizes the spin diffusion.01/27/2025ApprovedFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
- 6.04The ESR shall include a system to absorb the energy of bunches which are periodically ejected from the ESR to accommodate freshly polarized bunches.02/13/2025ApprovedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04The ESR shall be capable of continuously replacing all polarized electron bunches in the ring while the electron beam is in collision with a hadron beam allowing it to provide arbitrary spin patterns for the collider experiments.01/27/2025ApprovedFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
- 6.04The ESR shall be capable of delivering bunches with longitudnal spins to the IP.01/27/2025ApprovedFALSE
- 6.04The ESR shall provide electron bunches having the bunch parameters specified in in MPT. [Document#:EIC-SEG-RSI-005]01/27/2025ApprovedFALSE
- 6.04The ESR lattice arc magnet structure shall contain an array of regular FODO cells01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall contain drift spaces between the half cells all of which may beslightly different in each individual arc. To account for small differences in the required average bending radii at the different arc locations.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR super-bends shall generate additional synchrotron radiation damping to support a large beam-beam parameter of 0.1 and to create the required horizontal design emittance in the MPT when the ESR is operated at energies below 10 GeV. [Document: EIC-SEG-RSI-005]01/27/2025ReviewedFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The polarity of the center bending magnet shall be capable of being wired in reverse to control the beam emittance and to damp the beam. The polarity will be dictated by the beam energy.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The FODO cell shall operate with a horizontal and vertical betatron phase advance of 60 degrees per arc section at beam energies of 10 GeV and below.01/27/2025ReviewedFALSE
- 6.04The sextupole wiring scheme shall accomadate the required sextupole families needed per arc to create the 60 degree FODO cell phase advance at < 10 GeV.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The FODO cell shall operate with a horizontal and vertical betatron phase advance of 90 degrees per arc section to provide the required horizontal beam emittance [5.9] at 18 GeV.01/27/2025ReviewedFALSE
- 6.04The vertical emittance in the arc section shall be controlled by appropriate beam orbit manipulations and horizontal-vertical cross coupling.01/27/2025ReviewedFALSE
- 6.04.06The instrumentation system shall include dual-plane beam position monitors (BPMs) adjacent to specified quadrupoles.01/27/2025ReviewedFALSE
- 6.04.06.01There shall be Dual plane BPM pick-ups shall be located at all vertically focusing quads,and at all quadrupoles within about +/- 100 meters around IP601/27/2025In ProcessFALSE
- 6.04.06.01There shall be NO BPMs at the horizontally focusing quadrupoles.01/27/2025In ProcessFALSE
- 6.04.06.01No single plane BPMs are required for the ESR01/27/2025In ProcessFALSE
- 6.04.06.01Special BPM pick-ups shall be located between the RF cavities in sector 10, for position resolution01/27/2025In ProcessFALSE
- 6.04.06.01The total number of ESR BPM pick-ups shall be 220 BPMs01/27/2025In ProcessFALSE
- 6.04.06.01The ESR BPM pick-ups shall be designed to provide the required measurement resolution over the range of H=+/- 10 V=+/- 5 mm02/13/2025In ProcessFALSE
- 6.04.06.01No ESR BPM pick-ups shall be required to have an unusual range of position requirements.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up design shall ensure the simulated impedance is within the accepted overall ESR impedance budget.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up shall be designed to ensure the maximim temperatures of the components (due to heating by the beam) are acceptable for reliability and operations, using thermal simulation software.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-ups shall be designed to be baked to 250 °C for a sufficiently large number of thermal cycles.01/27/2025In ProcessFALSE
- 6.04.06.01The relative alignment of the BPMs and neighboring quads, after BBA shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up installation mechanical misalignment tolerance shall be same as for the quadrupoles. 100 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up operational misalignement tolerance variations due to thermal cycling of the ESR shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06The BPMs shall have turn-by-turn orbit measurement capability based on a single, remotely selectable bunch out of the fully filled bunch train to enable injection optimization.01/27/2025ReviewedFALSE
- 6.04.06.01Beam position monitor shall have electronics with turn-by-turn capability01/27/2025In ProcessFALSE
- 6.04.06.01The first 20 BPMs after injection shall be capable of measuring individual bunch positions to minimize betatron oscillations of the newly injected bunches.01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM measurement resolution shall be 100 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM maximum allowable drift shall be 50 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM measurement resolution shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM maximum allowable drift shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the measurement resolution shall be H=50 V=10 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the maximum allowable drift shall be 10 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the measurement resolution shall be H=10 V=5 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the maximum allowable drift shall be 5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches),the measurement resolution shall be H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches), the maximum allowable drift H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the measurement resolution shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the maximum allowable drift shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the measurement resolution shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the maximum allowable drift shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge(7-28 nC bunch range), the measurement resolution shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge, 7-28 nC bunch range, the maximum allowable drift shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01The BPM sum signals shall be used to measure bunch lifetime (1/e) during the 2.5 minute bunch duration01/27/2025In ProcessFALSE
- 6.04.06.01There shall be Dual plane BPM pick-ups shall be located at all vertically focusing quads,and at all quadrupoles within about +/- 100 meters around IP601/27/2025In ProcessFALSE
- 6.04.06.01There shall be NO BPMs at the horizontally focusing quadrupoles.01/27/2025In ProcessFALSE
- 6.04.06.01No single plane BPMs are required for the ESR01/27/2025In ProcessFALSE
- 6.04.06.01Special BPM pick-ups shall be located between the RF cavities in sector 10, for position resolution01/27/2025In ProcessFALSE
- 6.04.06.01The total number of ESR BPM pick-ups shall be 220 BPMs01/27/2025In ProcessFALSE
- 6.04.06.01The ESR BPM pick-ups shall be designed to provide the required measurement resolution over the range of H=+/- 10 V=+/- 5 mm02/13/2025In ProcessFALSE
- 6.04.06.01No ESR BPM pick-ups shall be required to have an unusual range of position requirements.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up design shall ensure the simulated impedance is within the accepted overall ESR impedance budget.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up shall be designed to ensure the maximim temperatures of the components (due to heating by the beam) are acceptable for reliability and operations, using thermal simulation software.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-ups shall be designed to be baked to 250 °C for a sufficiently large number of thermal cycles.01/27/2025In ProcessFALSE
- 6.04.06.01The relative alignment of the BPMs and neighboring quads, after BBA shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up installation mechanical misalignment tolerance shall be same as for the quadrupoles. 100 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up operational misalignement tolerance variations due to thermal cycling of the ESR shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a beam current monitor to measure average beam current01/27/2025ReviewedFALSE
- 6.04.06.02A DCCT shall measure the average beam current in the ESR.01/27/2025In ProcessFALSE
- 6.04.06.02The measurement resolution averaged over 1 sec shall be < 5 uA01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT beamline device impedance shall be approved by beam physics.01/27/2025In ProcessFALSE
- 6.04.06.02Measurement drift tolerance (thermal effects) shall be ≤ 1 uA/K01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor in the ring shall operate in the temperature range of 15 to 35 degrees C01/27/2025In ProcessFALSE
- 6.04.06.02The range of average beam current to be measured shall be 0.156x10-3 to 2.5 A01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall have a remote controlled self calibration system01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall provide measurements with absolute accuracy of better than 0.2 %01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be archived at a rate of 10 Hz Hz01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be provided to users at a rate of 10 Hz01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor assembly in the ring shall be radiation resistant01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure individual bunch charges and bunch pattern01/27/2025ReviewedFALSE
- 6.04.06.02A bunch pattern monitor shall be installed in the ESR to measure indiviual bunch charge with an accuracy 1 %01/27/2025In ProcessFALSE
- 6.04.06.02The Bunch pattern monitor shall be capable of measuring Bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure transverse beam profiles01/27/2025ReviewedFALSE
- 6.04.06.03The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure longitudinal beam profiles01/27/2025ReviewedFALSE
- 6.04.06.03The requirements for longitudinal feedback are ??? TBD01/27/2025In ProcessFALSE
- 6.04.06.03The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06The longitudinal bunch profile monitor needs turn-by-turn capability based on a single bunch in the fully filled bunch train to allow timing and energy adjustment for injection optimization.01/27/2025ReviewedFALSE
- 6.04.06.03The requirements for longitudinal feedback are ??? TBD01/27/2025In ProcessFALSE
- 6.04.06.03The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include system to measure H & V betatron tunes.02/13/2025ReviewedFALSE
- 6.04.06.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/13/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the horizontal kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the vertical kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The location of the tune meter kicker striplines in the ESR shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The kicker waveform (risetime and shape) requirements shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The impedance of the kicker beamline device shall be approved by beam Physics.01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall facilitate all required feedback systems (slow transverse, longitudinal and transverse bunch-by-bunch)01/27/2025ReviewedFALSE
- 6.07.02The slow orbit feedback correction output rate shall be 10 Hz01/27/2025In ProcessFALSE
- 6.07.02The slow orbit feedback BPM data averaging period shall be tbd -01/27/2025In ProcessFALSE
- 6.04.06.03The transverse slow feedback system bandwidth shall bs 10 Hz01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include beam loss monitor system with detectors located only at select regions of the ESR.01/27/2025ReviewedFALSE
- 6.04.06.05BLM shall be needed to needed to protect sensitive equipment.01/27/2025In ProcessFALSE
- 6.04.06.05The number of BLM installed in the ESR shall be TBD ea01/27/2025In ProcessFALSE
- 6.04.06.05BLM shall be installed at the following locations in the ESR TBD01/27/2025In ProcessFALSE
- 6.04.06.05The sensitivity of the BLM detectors shall be TBD units?01/27/2025In ProcessFALSE
- 6.04.06.05Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/13/2025In ProcessFALSE
- 6.04.06.05The response time from loss detection to abort shall be TBD us01/27/2025In ProcessFALSE
- 6.04The magnets shall meet the requirements defined by the physics lattice.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The magnets shall have the required field quality to meet the operational needs.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The “super-bends” in the ARC sections shall consist of two long dipoles on either end of a short dipole.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The good field region of the dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends. Confirm the 4cm with Daniel and Scott01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04All quadrupoles shall be designed to facilitate beam based alignment.01/27/2025ReviewedFALSE
- 6.04The maximum integrated field strength of the FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.04The aperture of all magnets shall be large enough to accommodate the ESR vacuum chamber. Add a link to details of vacuum.01/27/2025ReviewedFALSE
- 6.04The ESR Shall have a conventional orbit corrector scheme, with single-plane correctors located at the respective quadrupoles.The strength of these correctors needs to be chosen to correct for any source of orbit distortion and should have enough margin for beam based diagnostic purposes, Harmonic Spin bumps,emittance generating bumps.01/27/2025ReviewedFALSE
- 6.04The two types of dipole magnets in the regular arc cells, the long dipole and the short one, shall be organized into two logical power circuits capable of being optimized subject to the physical power scheme used.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.01/27/2025ReviewedFALSE
- 6.04The quadrupoles in the straight sections IR10, IR12, IR2, and IR4 and in the transition from the arc to the straight section structure shall be wired to provide the optimized betatron phase advance across each straight section, as required for dynamic aperture optimization.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04Provisions shall be made to vary individual qudrupole strengths by approximately 1% (of their maximum strength?) for beam-based alignment purposes.01/27/2025ReviewedFALSE
- 6.04The sextupole power supply scheme shall be laid out such that the sextupole family structure can be configured for both the 60 and the 90 degree lattice along with a small number of individually powered sextupoles in the transition regions between arcs and straight sections with minimal effort, cost and minimizing any risk of error.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets01/27/2025ReviewedFALSE
- 6.04The ESR magnet power supply’s shall be capable of providing the stability the ESR needs to operate02/13/2025ReviewedFALSE
- 6.08The Storage RF System shall be designed to fulfill all necessary paramaters as set by the Master Parameter Table Doc. No. EIC-SEG-RSI-005.01/27/2025ReviewedFALSE
- 6.08.04.01Conditioning for individual cavities shall be achieved with a maximum temperature of 2.1 K02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable vertical acceleration of ±4 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable beamline axis acceleration of ±5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable lateral acceleration of ±1.5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to withstand a tilt around the beamline axis (roll) up to ±1.4 radians02/18/2025ApprovedFALSE
- 6.08The Storage ring RF System shall be installed in the straight sections of the ESR lattice within the existing RHIC tunnel in IR10.01/27/2025ReviewedFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.08.04.01Conditioning for individual cavities shall be achieved with a maximum temperature of 2.1 K02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable lateral acceleration of ±1.5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable vertical acceleration of ±4 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable beamline axis acceleration of ±5 G02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall conform to the ESR lattice.01/27/2025ReviewedFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall be designed to accelerate electrons.01/27/2025ReviewedFALSE
- 6.08The Storage RF System shall utilize superconductivity.01/27/2025ReviewedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall conform to the EIC Code of Record.01/27/2025ReviewedFALSE
- 6.08The Storage RF System shall have a minimum operating lifetime of 20 years01/27/2025ReviewedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/cm^202/18/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The chilled water and low-conductivity water operational pressure range shall be 7.5 to 8.5 bar02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall have an operational availability of (TBD 95%) or better.01/27/2025ReviewedFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/cm^202/18/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The chilled water and low-conductivity water operational pressure range shall be 7.5 to 8.5 bar02/18/2025ApprovedFALSE
- 6.08The Storage RF Systems within the tunnel shall operate within its yearly radiation exposure budget.01/27/2025ReviewedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
- 6.04The phase advance of each straight section shall be tunable in order to optimize the dynamic aperture of the ESR.01/27/2025In ProcessFALSE
- 6.04The straight sections IR2,4,10 and 12 shall be based on FODO cells.01/27/2025In ProcessFALSE
- 6.04There shall be matching sections at the ends of each of the straight sections to compensate for the different FODO cell lengths wrt the arc FODO cells imposed by geometric constraints.01/27/2025In ProcessFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall provide sufficient horizontal and vertical aperture to accommodate; a +/-15 sigma beam, where the vertical RMS beam size is based on the emittance of a fully coupled beam, plus an additional 10 mm horizontally and 5 mm vertically margin to account for expected orbit errors.01/27/2025In ProcessFALSE
- 6.04.05The typical (standard) vacuum chamber aperture shall be 80 x 36 mm.01/27/2025ApprovedFALSE
- 6.04.05Special aperture requirements and/or aperture file shall be provided by or approved by physics.01/27/2025ApprovedFALSE
- 6.04.05The dynamic pressure around the ring shall be consistent with a beam gas lifetime of >10[hrs] with the design currents after an integrated beam current of 1000 [A.h].01/27/2025In ProcessFALSE
- 6.04.05There shall be no upper pressure limit as long as the average pressure is maintained.01/27/2025ApprovedFALSE
- 6.04.05The average vacuum level in the ESR Arc sections after conditioning (for 1000Ahrs) shall be <5x10-9 Torr.01/27/2025ApprovedFALSE
- 6.04.05On 15 m on each side (or one vacuum sector) of the SRF cavities shall be processed to class ISO 5.01/27/2025ApprovedFALSE
- 6.04.05There shall be no pressure bumps in the ring exceeding (TBD)[Torr]01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber and all its components shall be designed to withstand a total synchrotron radiation load of 10 MW, considering the uneven linear load particularly related to the super-bends.01/27/2025In ProcessFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04.05The vacuum chamber material shall be chosen such that the SR power can be intercepted by the arc chambers and in addition good radiation shielding will be provided.01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04.05The impedance of the entire ESR vacuum system, including the interaction regions in IR 6 and IR 8, shall allow for the bunch intensities, beam currents, and bunch numbers contained in [5.9].01/27/2025In ProcessFALSE
- 6.04.05The vacuum system global impedance shall be less than the impedance budget as provided by accelerator physics.01/27/2025ApprovedFALSE
- 6.04The ESR lattice shall provide a minimum dynamic aperture of 10 sigma w.r.t Gaussian electron beam distribution in all three dimensions, horizontal, vertical, and longitudinal. With the vertical emittance being half the horizontal design emittance.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.05The maximum beam excursion orbit shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04The ESR alignment requirements are established by dynamic aperture and polarization tracking. The ESR RMS alignment tolerances shall be such that all the beam parameter listed in [5.9] can be satisfied.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The minimum dynamic aperture shall be achieved in two optics configurations (60 and 90 degrees betatron phase advance per FODO cell) at all operational beam energies in the Master Parameter Table (MPT), and with one and with two low-beta insertions. [Document#:EIC-SEG-RSI-005]01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.05The maximum beam excursion orbit shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04The ESR Lattice shall contain provisions for correctors such as skew quadrupoles, Dipole correctors etc. as needed.01/27/2025ApprovedFALSE
- 6.04The ESR shall have an average arc bending radius of approximately 380 meters01/27/2025ApprovedFALSE
- 6.04The ESR lattice arc magnet structure shall contain an array of regular FODO cells01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall contain drift spaces between the half cells all of which may beslightly different in each individual arc. To account for small differences in the required average bending radii at the different arc locations.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The magnets shall meet the requirements defined by the physics lattice.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The magnets shall have the required field quality to meet the operational needs.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The “super-bends” in the ARC sections shall consist of two long dipoles on either end of a short dipole.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The good field region of the dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends. Confirm the 4cm with Daniel and Scott01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The aperture of all magnets shall be large enough to accommodate the ESR vacuum chamber. Add a link to details of vacuum.01/27/2025ReviewedFALSE
- 6.04The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets01/27/2025ReviewedFALSE
- 6.04The two types of dipole magnets in the regular arc cells, the long dipole and the short one, shall be organized into two logical power circuits capable of being optimized subject to the physical power scheme used.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR Lattice shall contain provisions for correctors such as skew quadrupoles, Dipole correctors etc. as needed.01/27/2025ApprovedFALSE
- 6.04The ESR electron beam shall orbit in the clockwise direction as seen from above.01/27/2025ApprovedFALSE
- 6.04The ESR lattice arc magnet structure shall contain an array of regular FODO cells01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall contain drift spaces between the half cells all of which may beslightly different in each individual arc. To account for small differences in the required average bending radii at the different arc locations.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The magnets shall meet the requirements defined by the physics lattice.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The magnets shall have the required field quality to meet the operational needs.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The “super-bends” in the ARC sections shall consist of two long dipoles on either end of a short dipole.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The good field region of the dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends. Confirm the 4cm with Daniel and Scott01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The two types of dipole magnets in the regular arc cells, the long dipole and the short one, shall be organized into two logical power circuits capable of being optimized subject to the physical power scheme used.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The aperture of all magnets shall be large enough to accommodate the ESR vacuum chamber. Add a link to details of vacuum.01/27/2025ReviewedFALSE
- 6.04The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets01/27/2025ReviewedFALSE
- 6.04The ESR Lattice shall be tilted with respect to the plane of the HSR along the chord between IP6 and IP8 by 200 microradians to avoid interference with other components and such that the ESR is above the HSR at IP12.01/27/2025ApprovedFALSE
- 6.04The ESR shall have a circumference such that the revolution frequency of the stored electron beam matches the revolution frequency of 133 GeV protons stored in the HSR, with the proton beam orbit centered in the HSR beampipe.01/27/2025ApprovedFALSE
- 6.04The ESR lattice arc magnet structure shall contain an array of regular FODO cells01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall contain drift spaces between the half cells all of which may beslightly different in each individual arc. To account for small differences in the required average bending radii at the different arc locations.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The magnets shall meet the requirements defined by the physics lattice.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The magnets shall have the required field quality to meet the operational needs.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The “super-bends” in the ARC sections shall consist of two long dipoles on either end of a short dipole.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The good field region of the dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends. Confirm the 4cm with Daniel and Scott01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The two types of dipole magnets in the regular arc cells, the long dipole and the short one, shall be organized into two logical power circuits capable of being optimized subject to the physical power scheme used.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The aperture of all magnets shall be large enough to accommodate the ESR vacuum chamber. Add a link to details of vacuum.01/27/2025ReviewedFALSE
- 6.04The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets01/27/2025ReviewedFALSE
- 6.04The ESR shall use the “inner” aisle (closest to the inner tunnel wall) of the tunnel from IR4 to IR6 and from IR8 to IR12, and the “outer” aisle (closest to the outer tunnel wall) from IR12 to IR4 and IR6 to IR802/13/2025ApprovedFALSE
- 6.04The ESR lattice arc magnet structure shall contain an array of regular FODO cells01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall contain drift spaces between the half cells all of which may beslightly different in each individual arc. To account for small differences in the required average bending radii at the different arc locations.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The sextupole wiring scheme shall accomadate the required sextupole families needed per arc to create the 90 degree FODO cell phase advance at 18 GeV.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The magnets shall meet the requirements defined by the physics lattice.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The magnets shall have the required field quality to meet the operational needs.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The “super-bends” in the ARC sections shall consist of two long dipoles on either end of a short dipole.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The good field region of the dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends. Confirm the 4cm with Daniel and Scott01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The two types of dipole magnets in the regular arc cells, the long dipole and the short one, shall be organized into two logical power circuits capable of being optimized subject to the physical power scheme used.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The aperture of all magnets shall be large enough to accommodate the ESR vacuum chamber. Add a link to details of vacuum.01/27/2025ReviewedFALSE
- 6.04The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets01/27/2025ReviewedFALSE
- 6.04The ESR shall be equipped with spin-transparent cross-overs where the ESR and the HSR transition between the inner and outer aisles at IR4 and IR12.02/13/2025ApprovedFALSE
- 6.04The ESR shall support collisions with the HSR in IR6 and IR8, equipped with colliding beam interaction regions and detectors for nuclear physics experiments. These areas include crossovers where the ESR and the HSR transition between the inner and outer aisles.01/27/2025ApprovedFALSE
- 6.04The main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.01/27/2025ReviewedFALSE
- 6.04The maximum integrated field strength of the FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.04The ESR shall support two low-beta insertions (colliding beam interaction regions) at IRs 6 and 8.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.05The maximum beam excursion orbit shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04The ESR shall have provisions made at IR8 to accommodate a future 2nd colliding beam interaction region with low-beta section, spin rotators and crab cavities.01/27/2025ApprovedFALSE
- 6.04The main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.01/27/2025ReviewedFALSE
- 6.04The maximum integrated field strength of the FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.04The ESR shall support two low-beta insertions (colliding beam interaction regions) at IRs 6 and 8.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.05The maximum beam excursion orbit shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04The ESR shall have provisions made at IR12 to accommodate beam polarimetry measurements, damper systems, special instrumentation, electron injection into and extraction out of the ESR.02/13/2025ApprovedFALSE
- 6.04The ESR shall be designed to minimize unscheduled downtime, maintenance time and repair time to achieve EIC operational availability.01/27/2025ApprovedFALSE
- 6.06.03The EIS shall contain an Abort system to dump the beam and protect the system when required01/27/2025In ProcessFALSE
- 6.03.03.03Fast valve protection shall be provided at TBD01/27/2025On HoldFALSE
- 6.03.03.03Fast valve protection shall be provided at TBD01/27/2025On HoldFALSE
- 6.06.03The ESR shall have a collimation system capable of ensuring a sufficiently low background at the detector.01/27/2025ReviewedFALSE
- 6.06.03.02The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR Injection absorber shall be vertical.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR injection absorber has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The thermal duty cycle shall be 2 Hz. 2 Hz01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR detector absorbers shall be placed at Sector 5.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The detector absorbers have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be tbd kW. tbd kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 21 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Cu01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 560-720 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator shall be horizontal.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be range shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary vertical collimator shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary horizontal collimator shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The ESR primary collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W on the tip of the jaw. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary vertical collimators shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary horizontal collimators shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).01/27/2025In ProcessFALSE
- 6.06.03.02The ESR collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
- 6.06.03The ESR shall have a collimation system capable protecting all machine elements in case of failure.01/27/2025ReviewedFALSE
- 6.06.03.02The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR Injection absorber shall be vertical.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR injection absorber has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The thermal duty cycle shall be 2 Hz. 2 Hz01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR detector absorbers shall be placed at Sector 5.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The detector absorbers have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be tbd kW. tbd kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 21 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Cu01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 560-720 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator shall be horizontal.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be range shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary vertical collimator shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary horizontal collimator shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The ESR primary collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W on the tip of the jaw. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary vertical collimators shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary horizontal collimators shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).01/27/2025In ProcessFALSE
- 6.06.03.02The ESR collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
- 6.06.03The ESR Abort system shall contain a beam to dump to safely absorb the energy of the dumped beam in a controlled fashion.01/27/2025ReviewedFALSE
- 6.06.03.01The diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 2 m01/27/2025In ProcessFALSE
- 6.06.03.01The materials shall be C / Al / Cu01/27/2025In ProcessFALSE
- 6.06.03.01The energy deposited during abort shall be 320 kJ01/27/2025In ProcessFALSE
- 6.06.03.01The window thickness shall be tbd mm01/27/2025In ProcessFALSE
- 6.06.03.01The frequency of thermal cycle shall be 1 hour01/27/2025In ProcessFALSE
- 6.06.03.01The window material shall be tbd01/27/2025In ProcessFALSE
- 6.06.03The ESR shall contain an Abort system to dump the beam01/27/2025ReviewedFALSE
- 6.06.03.01The diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 2 m01/27/2025In ProcessFALSE
- 6.06.03.01The materials shall be C / Al / Cu01/27/2025In ProcessFALSE
- 6.06.03.01The energy deposited during abort shall be 320 kJ01/27/2025In ProcessFALSE
- 6.06.03.01The frequency of thermal cycle shall be 1 hour01/27/2025In ProcessFALSE
- 6.06.03.01The window thickness shall be tbd mm01/27/2025In ProcessFALSE
- 6.06.03.01The window material shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The Length shall be 50 m01/27/2025In ProcessFALSE
- 6.06.03.01The Internal diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The Temperature sensors shall be yes01/27/2025In ProcessFALSE
- 6.06.03.01The BPMs shall be 401/27/2025In ProcessFALSE
- 6.06.03.01The Correctors shall be 401/27/2025In ProcessFALSE
- 6.06.03.01The Corrector PS shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The Cooling / pumping shall be yes01/27/2025In ProcessFALSE
- 6.06.03.01The deflection shall be 2 mrad01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 1.2 m01/27/2025In ProcessFALSE
- 6.06.03.01The power supply shall be 1600 A01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The Y-chamber aperture shall be 36 mm01/27/2025In ProcessFALSE
- 6.06.03.01May need to add additional window requirements for other leg of Lambertson magnet TBD01/27/2025In ProcessFALSE
- 6.06.03.01The gradient shall be 17 T/m01/27/2025In ProcessFALSE
- 6.06.03.01The power supply shall be 1600 A01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 70 cm01/27/2025In ProcessFALSE
- 6.06.03.01The aperture radius shall be 50 mm01/27/2025In ProcessFALSE
- 6.06.03.01The number of kickers shall be 601/27/2025In ProcessFALSE
- 6.06.03.01The Rise time shall be 900 ns01/27/2025In ProcessFALSE
- 6.06.03.01The Fall time shall be NA sec01/27/2025In ProcessFALSE
- 6.06.03.01The flat top time shall be 13 us01/27/2025In ProcessFALSE
- 6.06.03.01The waveshape shall be trap01/27/2025In ProcessFALSE
- 6.06.03.01The painting shall be vertical01/27/2025In ProcessFALSE
- 6.06.03.01The maximum field shall be 0.12 T01/27/2025In ProcessFALSE
- 6.06.03.01The total deflection shall be 16 mrad01/27/2025In ProcessFALSE
- 6.06.03.01The maximum current shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The maximum voltage shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The inductance with cable shall be TBD (uH)01/27/2025In ProcessFALSE
- 6.06.03.01The Max rep rate shall be 100 kV/pC01/27/2025In ProcessFALSE
- 6.06.03.01The flat top repeatability shall be NA Hz01/27/2025In ProcessFALSE
- 6.06.03.01The beam abort kicker shall be tbd %01/27/2025In ProcessFALSE
- 6.06.03.01The flatness of flat top/pulse form shall be 1 %01/27/2025In ProcessFALSE
- 6.06.03.01The cooling type shall be w (W,A)01/27/2025In ProcessFALSE
- 6.04The ESR shall reach an availability consistent with the overall availability of the entire EIC as specified in the MPT. [Document#:EIC-SEG-RSI-005]01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04All ESR components and systems shall be designed and installed in line with all relevant regulatory codes and in full compliance with BNL SBMS.01/27/2025ApprovedFALSE
- 6.07.02The ESR shall have a control system which can operate the ESR consistent will the overall control of the other EIC system's and to ensure the ESR meets all the Physics requirements needed to deliver the physics goals of the EIC.01/27/2025ApprovedFALSE
- 6.07.02The ESR control system shall facilitate all ESR global control requirements.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate all network, relational database and data archiving required.01/27/2025ReviewedFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The control system shall facilitate all machine protection systems required01/27/2025ReviewedFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate all EIC machine timing required.01/27/2025ReviewedFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate fast feedback integration systems required. The child for this are in the IR need to identify the magnets . These will be low inductance correctors .02/13/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate all physics application support required.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include dual-plane beam position monitors (BPMs) adjacent to specified quadrupoles.01/27/2025ReviewedFALSE
- 6.04.06.01There shall be Dual plane BPM pick-ups shall be located at all vertically focusing quads,and at all quadrupoles within about +/- 100 meters around IP601/27/2025In ProcessFALSE
- 6.04.06.01There shall be NO BPMs at the horizontally focusing quadrupoles.01/27/2025In ProcessFALSE
- 6.04.06.01No single plane BPMs are required for the ESR01/27/2025In ProcessFALSE
- 6.04.06.01Special BPM pick-ups shall be located between the RF cavities in sector 10, for position resolution01/27/2025In ProcessFALSE
- 6.04.06.01The total number of ESR BPM pick-ups shall be 220 BPMs01/27/2025In ProcessFALSE
- 6.04.06.01The ESR BPM pick-ups shall be designed to provide the required measurement resolution over the range of H=+/- 10 V=+/- 5 mm02/13/2025In ProcessFALSE
- 6.04.06.01No ESR BPM pick-ups shall be required to have an unusual range of position requirements.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up design shall ensure the simulated impedance is within the accepted overall ESR impedance budget.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up shall be designed to ensure the maximim temperatures of the components (due to heating by the beam) are acceptable for reliability and operations, using thermal simulation software.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-ups shall be designed to be baked to 250 °C for a sufficiently large number of thermal cycles.01/27/2025In ProcessFALSE
- 6.04.06.01The relative alignment of the BPMs and neighboring quads, after BBA shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up installation mechanical misalignment tolerance shall be same as for the quadrupoles. 100 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up operational misalignement tolerance variations due to thermal cycling of the ESR shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06The BPMs shall have turn-by-turn orbit measurement capability based on a single, remotely selectable bunch out of the fully filled bunch train to enable injection optimization.01/27/2025ReviewedFALSE
- 6.04.06.01Beam position monitor shall have electronics with turn-by-turn capability01/27/2025In ProcessFALSE
- 6.04.06.01The first 20 BPMs after injection shall be capable of measuring individual bunch positions to minimize betatron oscillations of the newly injected bunches.01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM measurement resolution shall be 100 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM maximum allowable drift shall be 50 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM measurement resolution shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM maximum allowable drift shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the measurement resolution shall be H=50 V=10 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the maximum allowable drift shall be 10 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the measurement resolution shall be H=10 V=5 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the maximum allowable drift shall be 5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches),the measurement resolution shall be H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches), the maximum allowable drift H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the measurement resolution shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the maximum allowable drift shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the measurement resolution shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the maximum allowable drift shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge(7-28 nC bunch range), the measurement resolution shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge, 7-28 nC bunch range, the maximum allowable drift shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01The BPM sum signals shall be used to measure bunch lifetime (1/e) during the 2.5 minute bunch duration01/27/2025In ProcessFALSE
- 6.04.06.01There shall be Dual plane BPM pick-ups shall be located at all vertically focusing quads,and at all quadrupoles within about +/- 100 meters around IP601/27/2025In ProcessFALSE
- 6.04.06.01There shall be NO BPMs at the horizontally focusing quadrupoles.01/27/2025In ProcessFALSE
- 6.04.06.01No single plane BPMs are required for the ESR01/27/2025In ProcessFALSE
- 6.04.06.01Special BPM pick-ups shall be located between the RF cavities in sector 10, for position resolution01/27/2025In ProcessFALSE
- 6.04.06.01The total number of ESR BPM pick-ups shall be 220 BPMs01/27/2025In ProcessFALSE
- 6.04.06.01The ESR BPM pick-ups shall be designed to provide the required measurement resolution over the range of H=+/- 10 V=+/- 5 mm02/13/2025In ProcessFALSE
- 6.04.06.01No ESR BPM pick-ups shall be required to have an unusual range of position requirements.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up design shall ensure the simulated impedance is within the accepted overall ESR impedance budget.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up shall be designed to ensure the maximim temperatures of the components (due to heating by the beam) are acceptable for reliability and operations, using thermal simulation software.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-ups shall be designed to be baked to 250 °C for a sufficiently large number of thermal cycles.01/27/2025In ProcessFALSE
- 6.04.06.01The relative alignment of the BPMs and neighboring quads, after BBA shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up installation mechanical misalignment tolerance shall be same as for the quadrupoles. 100 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up operational misalignement tolerance variations due to thermal cycling of the ESR shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a beam current monitor to measure average beam current01/27/2025ReviewedFALSE
- 6.04.06.02A DCCT shall measure the average beam current in the ESR.01/27/2025In ProcessFALSE
- 6.04.06.02The measurement resolution averaged over 1 sec shall be < 5 uA01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT beamline device impedance shall be approved by beam physics.01/27/2025In ProcessFALSE
- 6.04.06.02Measurement drift tolerance (thermal effects) shall be ≤ 1 uA/K01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor in the ring shall operate in the temperature range of 15 to 35 degrees C01/27/2025In ProcessFALSE
- 6.04.06.02The range of average beam current to be measured shall be 0.156x10-3 to 2.5 A01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall have a remote controlled self calibration system01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall provide measurements with absolute accuracy of better than 0.2 %01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be archived at a rate of 10 Hz Hz01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be provided to users at a rate of 10 Hz01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor assembly in the ring shall be radiation resistant01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure individual bunch charges and bunch pattern01/27/2025ReviewedFALSE
- 6.04.06.02A bunch pattern monitor shall be installed in the ESR to measure indiviual bunch charge with an accuracy 1 %01/27/2025In ProcessFALSE
- 6.04.06.02The Bunch pattern monitor shall be capable of measuring Bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure transverse beam profiles01/27/2025ReviewedFALSE
- 6.04.06.03The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure longitudinal beam profiles01/27/2025ReviewedFALSE
- 6.04.06.03The requirements for longitudinal feedback are ??? TBD01/27/2025In ProcessFALSE
- 6.04.06.03The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06The longitudinal bunch profile monitor needs turn-by-turn capability based on a single bunch in the fully filled bunch train to allow timing and energy adjustment for injection optimization.01/27/2025ReviewedFALSE
- 6.04.06.03The requirements for longitudinal feedback are ??? TBD01/27/2025In ProcessFALSE
- 6.04.06.03The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include system to measure H & V betatron tunes.02/13/2025ReviewedFALSE
- 6.04.06.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/13/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the horizontal kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the vertical kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The location of the tune meter kicker striplines in the ESR shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The kicker waveform (risetime and shape) requirements shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The impedance of the kicker beamline device shall be approved by beam Physics.01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall facilitate all required feedback systems (slow transverse, longitudinal and transverse bunch-by-bunch)01/27/2025ReviewedFALSE
- 6.07.02The slow orbit feedback correction output rate shall be 10 Hz01/27/2025In ProcessFALSE
- 6.07.02The slow orbit feedback BPM data averaging period shall be tbd -01/27/2025In ProcessFALSE
- 6.04.06.03The transverse slow feedback system bandwidth shall bs 10 Hz01/27/2025In ProcessFALSE
- 6.08The Storage RF System shall have an operational availability of (TBD 95%) or better.01/27/2025ReviewedFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/cm^202/18/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The chilled water and low-conductivity water operational pressure range shall be 7.5 to 8.5 bar02/18/2025ApprovedFALSE
- 6.04.06The instrumentation system shall include beam loss monitor system with detectors located only at select regions of the ESR.01/27/2025ReviewedFALSE
- 6.04.06.05BLM shall be needed to needed to protect sensitive equipment.01/27/2025In ProcessFALSE
- 6.04.06.05The number of BLM installed in the ESR shall be TBD ea01/27/2025In ProcessFALSE
- 6.04.06.05BLM shall be installed at the following locations in the ESR TBD01/27/2025In ProcessFALSE
- 6.04.06.05The sensitivity of the BLM detectors shall be TBD units?01/27/2025In ProcessFALSE
- 6.04.06.05Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/13/2025In ProcessFALSE
- 6.04.06.05The response time from loss detection to abort shall be TBD us01/27/2025In ProcessFALSE
- 6.08The Storage RF Systems within the tunnel shall operate within its yearly radiation exposure budget.01/27/2025ReviewedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
- 6.09The ESR shall have a cryogenic system to cool and operate all elements which need cryogenic cooling .01/27/2025ApprovedFALSE
- 6.08The ESR shall utilize an RF system capable of operating at parameters defined in MPT. [Document#:EIC-SEG-RSI-005]01/27/2025In ProcessFALSE
ESR-ARC : Arc Sections
- ESR-ARC EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04The ESR lattice arc magnet structure shall contain an array of regular FODO cells01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall consists of a quadrupole, a sextupole, a bending section, and a dipole corrector in each arc half-cell.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The ESR lattice arc magnet structure shall contain drift spaces between the half cells all of which may beslightly different in each individual arc. To account for small differences in the required average bending radii at the different arc locations.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR beamline bending sections shall contain three individual dipole magnets, referred to as “super-bends”.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The ESR super-bends shall generate additional synchrotron radiation damping to support a large beam-beam parameter of 0.1 and to create the required horizontal design emittance in the MPT when the ESR is operated at energies below 10 GeV. [Document: EIC-SEG-RSI-005]01/27/2025ReviewedFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The polarity of the center bending magnet shall be capable of being wired in reverse to control the beam emittance and to damp the beam. The polarity will be dictated by the beam energy.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The FODO cell shall operate with a horizontal and vertical betatron phase advance of 60 degrees per arc section at beam energies of 10 GeV and below.01/27/2025ReviewedFALSE
- 6.04The sextupole wiring scheme shall accomadate the required sextupole families needed per arc to create the 60 degree FODO cell phase advance at < 10 GeV.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The FODO cell shall operate with a horizontal and vertical betatron phase advance of 90 degrees per arc section to provide the required horizontal beam emittance [5.9] at 18 GeV.01/27/2025ReviewedFALSE
- 6.04The sextupole wiring scheme shall accomadate the required sextupole families needed per arc to create the 90 degree FODO cell phase advance at 18 GeV.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The vertical emittance in the arc section shall be controlled by appropriate beam orbit manipulations and horizontal-vertical cross coupling.01/27/2025ReviewedFALSE
ESR-COLL : ESR Momentum Collimator System
ESR-COLL-ABS : ESR Momentum Collimator Injection Absorbers
- ESR-COLL-ABS EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.02The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR Injection absorber shall be vertical.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR injection absorber has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The thermal duty cycle shall be 2 Hz. 2 Hz01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
ESR-COLL-DET : ESR Momentum Collimator Detector Absorbers
- ESR-COLL-DET EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.02The ESR detector absorbers shall be placed at Sector 5.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The detector absorbers have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be tbd kW. tbd kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 21 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Cu01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 560-720 mm01/27/2025In ProcessFALSE
ESR-COLL-MOM : ESR Momentum Collimator
- ESR-COLL-MOM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.02The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator shall be horizontal.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be range shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
ESR-COLL-PRIM : ESR Momentum Collimator Primary Collimators
- ESR-COLL-PRIM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.02The ESR primary vertical collimator shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary horizontal collimator shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The ESR primary collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W on the tip of the jaw. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
ESR-COLL-SECDRY : ESR Momentum Collimator Secondary Collimators
- ESR-COLL-SECDRY EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.02The ESR secondary vertical collimators shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary horizontal collimators shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).01/27/2025In ProcessFALSE
- 6.06.03.02The ESR collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
ESR-CONT : ESR Controls System
- ESR-CONT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.07.02The ESR control system shall facilitate all ESR global control requirements.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate all network, relational database and data archiving required.01/27/2025ReviewedFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The control system shall facilitate all machine protection systems required01/27/2025ReviewedFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate all EIC machine timing required.01/27/2025ReviewedFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate fast feedback integration systems required. The child for this are in the IR need to identify the magnets . These will be low inductance correctors .02/13/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR control system shall facilitate all physics application support required.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
ESR-CONT-ALGNMNT : ESR Controls Beam Based Alignment
- ESR-CONT-ALGNMNT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
ESR-CONT-FEEDBACK : ESR Controls Slow Orbit Feedback
- ESR-CONT-FEEDBACK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.07.02The slow orbit feedback BPM data averaging period shall be tbd -01/27/2025In ProcessFALSE
- 6.07.02The slow orbit feedback correction output rate shall be 10 Hz01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02Placeholder for fast orbit feedback requirements01/27/2025In ProcessFALSE
- 6.07.02The tune feedback measurement sample rate shall be tbd Hz01/27/2025In ProcessFALSE
- 6.07.02The tune feedback correction rate shall be tbd Hz01/27/2025In ProcessFALSE
ESR-CONT-SPIN : ESR Controls Spin Pattern
- ESR-CONT-SPIN EXTERNALSRequirements who's parents are in other sub-systems.
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
ESR-CRYO_CENT_PLANT : Cryogenics Central Plant
ESR-CRYO_IR10_SAT_PLANT : IR10 Satellite Cryogenics Plant
ESR-INST : ESR Instrumentation System
- ESR-INST EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06The instrumentation system shall include dual-plane beam position monitors (BPMs) adjacent to specified quadrupoles.01/27/2025ReviewedFALSE
- 6.04.06.01There shall be Dual plane BPM pick-ups shall be located at all vertically focusing quads,and at all quadrupoles within about +/- 100 meters around IP601/27/2025In ProcessFALSE
- 6.04.06.01There shall be NO BPMs at the horizontally focusing quadrupoles.01/27/2025In ProcessFALSE
- 6.04.06.01No single plane BPMs are required for the ESR01/27/2025In ProcessFALSE
- 6.04.06.01Special BPM pick-ups shall be located between the RF cavities in sector 10, for position resolution01/27/2025In ProcessFALSE
- 6.04.06.01The total number of ESR BPM pick-ups shall be 220 BPMs01/27/2025In ProcessFALSE
- 6.04.06.01The ESR BPM pick-ups shall be designed to provide the required measurement resolution over the range of H=+/- 10 V=+/- 5 mm02/13/2025In ProcessFALSE
- 6.04.06.01No ESR BPM pick-ups shall be required to have an unusual range of position requirements.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up design shall ensure the simulated impedance is within the accepted overall ESR impedance budget.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up shall be designed to ensure the maximim temperatures of the components (due to heating by the beam) are acceptable for reliability and operations, using thermal simulation software.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-ups shall be designed to be baked to 250 °C for a sufficiently large number of thermal cycles.01/27/2025In ProcessFALSE
- 6.04.06.01The relative alignment of the BPMs and neighboring quads, after BBA shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up installation mechanical misalignment tolerance shall be same as for the quadrupoles. 100 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up operational misalignement tolerance variations due to thermal cycling of the ESR shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06The BPMs shall have turn-by-turn orbit measurement capability based on a single, remotely selectable bunch out of the fully filled bunch train to enable injection optimization.01/27/2025ReviewedFALSE
- 6.04.06.01Beam position monitor shall have electronics with turn-by-turn capability01/27/2025In ProcessFALSE
- 6.04.06.01The first 20 BPMs after injection shall be capable of measuring individual bunch positions to minimize betatron oscillations of the newly injected bunches.01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM measurement resolution shall be 100 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM maximum allowable drift shall be 50 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM measurement resolution shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM maximum allowable drift shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the measurement resolution shall be H=50 V=10 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the maximum allowable drift shall be 10 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the measurement resolution shall be H=10 V=5 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the maximum allowable drift shall be 5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches),the measurement resolution shall be H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches), the maximum allowable drift H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the measurement resolution shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the maximum allowable drift shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the measurement resolution shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the maximum allowable drift shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge(7-28 nC bunch range), the measurement resolution shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge, 7-28 nC bunch range, the maximum allowable drift shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01The BPM sum signals shall be used to measure bunch lifetime (1/e) during the 2.5 minute bunch duration01/27/2025In ProcessFALSE
- 6.04.06.01There shall be Dual plane BPM pick-ups shall be located at all vertically focusing quads,and at all quadrupoles within about +/- 100 meters around IP601/27/2025In ProcessFALSE
- 6.04.06.01There shall be NO BPMs at the horizontally focusing quadrupoles.01/27/2025In ProcessFALSE
- 6.04.06.01No single plane BPMs are required for the ESR01/27/2025In ProcessFALSE
- 6.04.06.01Special BPM pick-ups shall be located between the RF cavities in sector 10, for position resolution01/27/2025In ProcessFALSE
- 6.04.06.01The total number of ESR BPM pick-ups shall be 220 BPMs01/27/2025In ProcessFALSE
- 6.04.06.01The ESR BPM pick-ups shall be designed to provide the required measurement resolution over the range of H=+/- 10 V=+/- 5 mm02/13/2025In ProcessFALSE
- 6.04.06.01No ESR BPM pick-ups shall be required to have an unusual range of position requirements.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up design shall ensure the simulated impedance is within the accepted overall ESR impedance budget.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up shall be designed to ensure the maximim temperatures of the components (due to heating by the beam) are acceptable for reliability and operations, using thermal simulation software.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-ups shall be designed to be baked to 250 °C for a sufficiently large number of thermal cycles.01/27/2025In ProcessFALSE
- 6.04.06.01The relative alignment of the BPMs and neighboring quads, after BBA shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up installation mechanical misalignment tolerance shall be same as for the quadrupoles. 100 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up operational misalignement tolerance variations due to thermal cycling of the ESR shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a beam current monitor to measure average beam current01/27/2025ReviewedFALSE
- 6.04.06.02A DCCT shall measure the average beam current in the ESR.01/27/2025In ProcessFALSE
- 6.04.06.02The measurement resolution averaged over 1 sec shall be < 5 uA01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT beamline device impedance shall be approved by beam physics.01/27/2025In ProcessFALSE
- 6.04.06.02Measurement drift tolerance (thermal effects) shall be ≤ 1 uA/K01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor in the ring shall operate in the temperature range of 15 to 35 degrees C01/27/2025In ProcessFALSE
- 6.04.06.02The range of average beam current to be measured shall be 0.156x10-3 to 2.5 A01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall have a remote controlled self calibration system01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall provide measurements with absolute accuracy of better than 0.2 %01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be archived at a rate of 10 Hz Hz01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be provided to users at a rate of 10 Hz01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor assembly in the ring shall be radiation resistant01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure individual bunch charges and bunch pattern01/27/2025ReviewedFALSE
- 6.04.06.02A bunch pattern monitor shall be installed in the ESR to measure indiviual bunch charge with an accuracy 1 %01/27/2025In ProcessFALSE
- 6.04.06.02The Bunch pattern monitor shall be capable of measuring Bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure transverse beam profiles01/27/2025ReviewedFALSE
- 6.04.06.03The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include a system to measure longitudinal beam profiles01/27/2025ReviewedFALSE
- 6.04.06.03The requirements for longitudinal feedback are ??? TBD01/27/2025In ProcessFALSE
- 6.04.06.03The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06The longitudinal bunch profile monitor needs turn-by-turn capability based on a single bunch in the fully filled bunch train to allow timing and energy adjustment for injection optimization.01/27/2025ReviewedFALSE
- 6.04.06.03The requirements for longitudinal feedback are ??? TBD01/27/2025In ProcessFALSE
- 6.04.06.03The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include system to measure H & V betatron tunes.02/13/2025ReviewedFALSE
- 6.04.06.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/13/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the horizontal kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the vertical kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The location of the tune meter kicker striplines in the ESR shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The kicker waveform (risetime and shape) requirements shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The impedance of the kicker beamline device shall be approved by beam Physics.01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall facilitate all required feedback systems (slow transverse, longitudinal and transverse bunch-by-bunch)01/27/2025ReviewedFALSE
- 6.07.02The slow orbit feedback correction output rate shall be 10 Hz01/27/2025In ProcessFALSE
- 6.07.02The slow orbit feedback BPM data averaging period shall be tbd -01/27/2025In ProcessFALSE
- 6.04.06.03The transverse slow feedback system bandwidth shall bs 10 Hz01/27/2025In ProcessFALSE
- 6.04.06The instrumentation system shall include beam loss monitor system with detectors located only at select regions of the ESR.01/27/2025ReviewedFALSE
- 6.04.06.05BLM shall be needed to needed to protect sensitive equipment.01/27/2025In ProcessFALSE
- 6.04.06.05The number of BLM installed in the ESR shall be TBD ea01/27/2025In ProcessFALSE
- 6.04.06.05BLM shall be installed at the following locations in the ESR TBD01/27/2025In ProcessFALSE
- 6.04.06.05The sensitivity of the BLM detectors shall be TBD units?01/27/2025In ProcessFALSE
- 6.04.06.05Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/13/2025In ProcessFALSE
- 6.04.06.05The response time from loss detection to abort shall be TBD us01/27/2025In ProcessFALSE
ESR-INST-BCM : ESR Instrumentation Bunch Charge and Bunch Pattern Monitor
- ESR-INST-BCM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.02A bunch pattern monitor shall be installed in the ESR to measure indiviual bunch charge with an accuracy 1 %01/27/2025In ProcessFALSE
- 6.04.06.02The Bunch pattern monitor shall be capable of measuring Bunch patterns ranging from a single bunch, to a filled ring with 1,160 bunches01/27/2025In ProcessFALSE
ESR-INST-BLM : ESR Instrumentation Beam Loss Monitor
- ESR-INST-BLM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.05BLM shall be needed to needed to protect sensitive equipment.01/27/2025In ProcessFALSE
- 6.04.06.05The number of BLM installed in the ESR shall be TBD ea01/27/2025In ProcessFALSE
- 6.04.06.05BLM shall be installed at the following locations in the ESR TBD01/27/2025In ProcessFALSE
- 6.04.06.05The sensitivity of the BLM detectors shall be TBD units?01/27/2025In ProcessFALSE
- 6.04.06.05Where possible existing RHIC BLM's can be relocated to identify ESR & HSR losses02/13/2025In ProcessFALSE
- 6.04.06.05The response time from loss detection to abort shall be TBD us01/27/2025In ProcessFALSE
ESR-INST-BPM : ESR Instrumentation Beam Postion Monitor System
ESR-INST-BPM-ELEC : ESR Instrumentation Beam Postion Monitor
- ESR-INST-BPM-ELEC EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.01Beam position monitor shall have electronics with turn-by-turn capability01/27/2025In ProcessFALSE
- 6.04.06.01The first 20 BPMs after injection shall be capable of measuring individual bunch positions to minimize betatron oscillations of the newly injected bunches.01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM measurement resolution shall be 100 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, single turn, the BPM maximum allowable drift shall be 50 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM measurement resolution shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For 2 nC pilot bunches, 1,000 turns average, the BPM maximum allowable drift shall be 30 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the measurement resolution shall be H=50 V=10 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunches , with low charge (2nC bunches), in presence of stored beam, the maximum allowable drift shall be 10 um01/27/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the measurement resolution shall be H=10 V=5 um02/13/2025In ProcessFALSE
- 6.04.06.01For newly injected refill bunch - high charge (7-28 nC bunch range), in presence of stored beam, the maximum allowable drift shall be 5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches),the measurement resolution shall be H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - low charge(2 nC bunches), the maximum allowable drift H=V=30 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the measurement resolution shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, turn-by-turn - high charge(7-28 nC bunch range), the maximum allowable drift shall be H=V=10 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the measurement resolution shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1,000 turns average, high charge(7-28 nC bunch range),the maximum allowable drift shall be H=V=5 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge(7-28 nC bunch range), the measurement resolution shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01For stored beam, 1 second average, high charge, 7-28 nC bunch range, the maximum allowable drift shall be H=V=1 um01/27/2025In ProcessFALSE
- 6.04.06.01The BPM sum signals shall be used to measure bunch lifetime (1/e) during the 2.5 minute bunch duration01/27/2025In ProcessFALSE
ESR-INST-BPM-PU : ESR Instrumentation Beam Postion Monitor Pick-up
- ESR-INST-BPM-PU EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.01There shall be Dual plane BPM pick-ups shall be located at all vertically focusing quads,and at all quadrupoles within about +/- 100 meters around IP601/27/2025In ProcessFALSE
- 6.04.06.01There shall be NO BPMs at the horizontally focusing quadrupoles.01/27/2025In ProcessFALSE
- 6.04.06.01No single plane BPMs are required for the ESR01/27/2025In ProcessFALSE
- 6.04.06.01Special BPM pick-ups shall be located between the RF cavities in sector 10, for position resolution01/27/2025In ProcessFALSE
- 6.04.06.01The total number of ESR BPM pick-ups shall be 220 BPMs01/27/2025In ProcessFALSE
- 6.04.06.01The ESR BPM pick-ups shall be designed to provide the required measurement resolution over the range of H=+/- 10 V=+/- 5 mm02/13/2025In ProcessFALSE
- 6.04.06.01No ESR BPM pick-ups shall be required to have an unusual range of position requirements.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up design shall ensure the simulated impedance is within the accepted overall ESR impedance budget.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up shall be designed to ensure the maximim temperatures of the components (due to heating by the beam) are acceptable for reliability and operations, using thermal simulation software.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-ups shall be designed to be baked to 250 °C for a sufficiently large number of thermal cycles.01/27/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up installation mechanical misalignment tolerance shall be same as for the quadrupoles. 100 µm02/13/2025In ProcessFALSE
- 6.04.06.01The relative alignment of the BPMs and neighboring quads, after BBA shall be <20 µm02/13/2025In ProcessFALSE
- 6.04.06.01The BPM pick-up operational misalignement tolerance variations due to thermal cycling of the ESR shall be <20 µm02/13/2025In ProcessFALSE
ESR-INST-DCCT : ESR Instrumentation Current and Charge Monitor
- ESR-INST-DCCT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.02A DCCT shall measure the average beam current in the ESR.01/27/2025In ProcessFALSE
- 6.04.06.02The measurement resolution averaged over 1 sec shall be < 5 uA01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT beamline device impedance shall be approved by beam physics.01/27/2025In ProcessFALSE
- 6.04.06.02Measurement drift tolerance (thermal effects) shall be ≤ 1 uA/K01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor in the ring shall operate in the temperature range of 15 to 35 degrees C01/27/2025In ProcessFALSE
- 6.04.06.02The range of average beam current to be measured shall be 0.156x10-3 to 2.5 A01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall have a remote controlled self calibration system01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT system shall provide measurements with absolute accuracy of better than 0.2 %01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be provided to users at a rate of 10 Hz01/27/2025In ProcessFALSE
- 6.04.06.02The measured average current shall be archived at a rate of 10 Hz Hz01/27/2025In ProcessFALSE
- 6.04.06.02The DCCT sensor assembly in the ring shall be radiation resistant01/27/2025In ProcessFALSE
ESR-INST-LONGIFB : ESR Instrumentation Longitudinal Bunch-by-Bunch Feedback
- ESR-INST-LONGIFB EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.03The requirements for longitudinal feedback are ??? TBD01/27/2025In ProcessFALSE
- 6.04.06.03The Longitudinal feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
ESR-INST-SLM : ESR Instrumentation Synchrotron Light Monitors
- ESR-INST-SLM EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
ESR-INST-TMK : ESR Instrumentation Horizontal and Vertical Tune Meter Kicker
- ESR-INST-TMK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.03Stripline kickers (H & V) shall be used to excite the beam so tunes can be measured using turn-by-turn BPM data.02/13/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the horizontal kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The magnitude of the kick required for the vertical kicker shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The kicker waveform (risetime and shape) requirements shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The location of the tune meter kicker striplines in the ESR shall be TBD units01/27/2025In ProcessFALSE
- 6.04.06.03The impedance of the kicker beamline device shall be approved by beam Physics.01/27/2025In ProcessFALSE
ESR-INST-TRANSFB : ESR Instrumentation Transverse Bunch-by-Bunch Feedback
- ESR-INST-TRANSFB EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
- 6.04.06.03The transverse feedback systems shall be capable of counteracting single-bunch rise times of 1 ms01/27/2025In ProcessFALSE
- 6.04.06.03Placeholder, Input needed TBD01/27/2025In ProcessFALSE
ESR-INST-TRANSSLFB : ESR Instrumentation Transverse Slow Feedback
- ESR-INST-TRANSSLFB EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.06.03The transverse slow feedback system bandwidth shall bs 10 Hz01/27/2025In ProcessFALSE
ESR-MAG : ESR Magnet
- ESR-MAG EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04The magnets shall meet the requirements defined by the physics lattice.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The magnets shall have the required field quality to meet the operational needs.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The “super-bends” in the ARC sections shall consist of two long dipoles on either end of a short dipole.02/13/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04The good field region of the dipoles shall extend over a horizontal range of at least 4 centimeters in the radial direction, for all operational beam energies from 5 to 18 GeV. This will take into account the orbit changes due to the reverse bends. Confirm the 4cm with Daniel and Scott01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04All quadrupoles shall be designed to facilitate beam based alignment.01/27/2025ReviewedFALSE
- 6.04The maximum integrated field strength of the FODO sextupoles needs to be sufficient to provide chromatic correction at all energies from 5 to 18 GeV with two low-beta interaction regions.01/27/2025ReviewedFALSE
- 6.07.02The BBA quad strength command output rate shall be 1 Hz01/27/2025In ProcessFALSE
- 6.04The aperture of all magnets shall be large enough to accommodate the ESR vacuum chamber. Add a link to details of vacuum.01/27/2025ReviewedFALSE
- 6.04The ESR Shall have a conventional orbit corrector scheme, with single-plane correctors located at the respective quadrupoles.The strength of these correctors needs to be chosen to correct for any source of orbit distortion and should have enough margin for beam based diagnostic purposes, Harmonic Spin bumps,emittance generating bumps.01/27/2025ReviewedFALSE
- 6.04The two types of dipole magnets in the regular arc cells, the long dipole and the short one, shall be organized into two logical power circuits capable of being optimized subject to the physical power scheme used.01/27/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04The main arc quadrupoles shall be powered to accommodate the Lattice requirements having the appropriate number of circuits to power the focusing and defocusing quadrupoles in each sextant of the ESR.01/27/2025ReviewedFALSE
- 6.04The quadrupoles in the straight sections IR10, IR12, IR2, and IR4 and in the transition from the arc to the straight section structure shall be wired to provide the optimized betatron phase advance across each straight section, as required for dynamic aperture optimization.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
- 6.04Provisions shall be made to vary individual qudrupole strengths by approximately 1% (of their maximum strength?) for beam-based alignment purposes.01/27/2025ReviewedFALSE
- 6.04The sextupole power supply scheme shall be laid out such that the sextupole family structure can be configured for both the 60 and the 90 degree lattice along with a small number of individually powered sextupoles in the transition regions between arcs and straight sections with minimal effort, cost and minimizing any risk of error.01/27/2025ReviewedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
ESR-MAG-B2ER
- 6.04.02.01The magnet shall be a single function magnet with a vertical dipole field along the beam axis.02/05/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/07/2025On HoldFALSE
- 6.04.02.01The physical magnet length shall be <TBD m.02/07/2025On HoldFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01The magnet bore gap and Width shall be 48 mm.02/07/2025On HoldFALSE
- 6.04.02.01The magnet shall be designed to fit within the following envelope:02/07/2025On HoldFALSE
- 6.04.02.01Magnet installation volume tolerances TBD02/07/2025On HoldFALSE
- 6.04.02.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/07/2025On HoldFALSE
- 6.04.02.01The magnet integrated dipole field (B) shall be TBD T.m.02/07/2025On HoldFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be TBD mm.02/07/2025On HoldFALSE
- 6.04.02.01The magnet good field aperture dAy required shall be TBD mm.02/07/2025On HoldFALSE
- 6.04.02.01The magnet to magnet field variability between magnets shall be 5 %.02/07/2025On HoldFALSE
- 6.04.02.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/07/2025On HoldFALSE
- 6.04.02.01The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/07/2025On HoldFALSE
- 6.04.02.01The Field at the reference radius at the design energy of 18 GeV shall be TBD (T) .02/07/2025On HoldFALSE
- 6.04.02.01The magnet bore field shall require the following multipole content:02/07/2025On HoldFALSE
- 6.04.02.01b1 = 10000 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b2 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b3 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b4 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b5 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b6 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b7 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b8 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b9 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b10 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b11 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b12 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b13 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b14 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b15 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01b16 < 100 (10^-4)02/07/2025On HoldFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.02/07/2025On HoldFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/07/2025On HoldFALSE
- 6.04.02.01The magnet shall be designed to meet the following fringe field requirements TBD02/07/2025On HoldFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-CORR_FFB_H
- 6.04.03.02The magnet shall be a single function dipole with a vertical field.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.02The physical magnet length shall be <TBD m.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet bore gap and Width shall be 48 mm.02/05/2025ReviewedFALSE
- 6.04.03.02The magnet shall be designed to fit within the following envelope:02/05/2025In ProcessFALSE
- 6.04.03.02Magnet installation volume tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The magnet integrated dipole field (B) shall be TBD T.m.02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet good field aperture dAx required shall be TBD mm.02/05/2025In ProcessFALSE
- 6.04.03.02The magnet good field aperture dAy required shall be TBD mm.02/05/2025In ProcessFALSE
- 6.04.03.02The magnet to magnet field variability between magnets shall be 5 %.02/05/2025ApprovedFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/05/2025In ProcessFALSE
- 6.04.03.02The Field at the reference radius at the design energy of 18 GeV shall be TBD (T) .02/05/2025In ProcessFALSE
- 6.04.03.02The magnet bore field shall require the following multipole content:02/05/2025In ProcessFALSE
- 6.04.03.02b1 = 10000 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b2 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b3 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b4 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b5 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b6 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b7 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b8 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b9 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b10 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b11 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b12 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b13 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b14 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b15 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b16 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.02The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-CORR_H : ESR Magnet HCOR
- 6.04.03.02The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The maximum physical magnet length shall be 0.2 m.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The minimal magnet bore gap and width shall be 52 and TBD mm.02/13/2025In ProcessFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The magnet integrated dipole field (B) shall be 12 milli T-m.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet good field aperture radius, dAx, shall be 24 mm.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet good field aperture radius, dAy, shall be 15 mm.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet to magnet variability shall be < 5e-3.02/06/2025ApprovedFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/06/2025ApprovedFALSE
- 6.04.03.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet field quality shall be measured within dAx and dAy radii utilizing a hall probe.02/06/2025ApprovedFALSE
- 6.04.03.0202/13/2025Not ApplicableFALSE
- 6.04.03.02The magnet bore field shall have a total field homogeneity of better than 0.003 dB/B and a btotal of <±100.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/06/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet axis.02/06/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet shall be designed with components capable to withstand a radiation dose of TBD MGy, or be approved by EIC for use in a specific location as shown in the “BNL Materials” List Doc. No. TBD.02/13/2025ApprovedFALSE
ESR-MAG-CORR_LA_H
- 6.04.03.02The magnet shall be a single function dipole with a vertical field.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.02The physical magnet length shall be <0.2 m.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet bore gap and Width shall be 48 mm.02/05/2025ReviewedFALSE
- 6.04.03.02The magnet shall be designed to fit within the following envelope:02/05/2025In ProcessFALSE
- 6.04.03.02Magnet installation volume tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The magnet integrated dipole field (B) shall be 60 mT.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet good field aperture dAx required shall be TBD mm.02/05/2025In ProcessFALSE
- 6.04.03.02The magnet good field aperture dAy required shall be TBD mm.02/05/2025In ProcessFALSE
- 6.04.03.02The magnet to magnet field variability between magnets shall be 5 %.02/05/2025ApprovedFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/05/2025In ProcessFALSE
- 6.04.03.02The Field at the reference radius at the design energy of 18 GeV shall be TBD (T) .02/05/2025In ProcessFALSE
- 6.04.03.02The magnet bore field shall require the following multipole content:02/05/2025In ProcessFALSE
- 6.04.03.02b1 = 10000 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b2 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b3 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b4 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b5 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b6 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b7 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b8 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b9 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b10 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b11 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b12 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b13 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b14 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b15 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b16 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.02The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-CORR_LA_V
- 6.04.03.02The magnet shall be a single function dipole with a horizontal field.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.02The physical magnet length shall be <0.2 m.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet bore gap and Width shall be 130.1 mm.02/05/2025ReviewedFALSE
- 6.04.03.02The magnet shall be designed to fit within the following envelope:02/05/2025In ProcessFALSE
- 6.04.03.02Magnet installation volume tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The magnet integrated dipole field (B) shall be 60 mT.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet good field aperture dAx required shall be TBD mm.02/05/2025In ProcessFALSE
- 6.04.03.02The magnet good field aperture dAy required shall be TBD mm.02/05/2025In ProcessFALSE
- 6.04.03.02The magnet to magnet field variability between magnets shall be 5 %.02/05/2025ApprovedFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/05/2025In ProcessFALSE
- 6.04.03.02The Field at the reference radius at the design energy of 18 GeV shall be TBD (T) .02/05/2025In ProcessFALSE
- 6.04.03.02The magnet bore field shall require the following multipole content:02/05/2025In ProcessFALSE
- 6.04.03.02b1 = 10000 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b2 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b3 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b4 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b5 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b6 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b7 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b8 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b9 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b10 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b11 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b12 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b13 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b14 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b15 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02b16 < 100 (10^-4)02/05/2025In ProcessFALSE
- 6.04.03.02The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.02The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-CORR_SPNQ_H : ESR Horizontal Corrector (SPNQ_H) Magnet
- 6.04.03.02The magnet shall be a single function dipole with a vertical field.02/05/2025On HoldFALSE
- 6.04.03.02The magnet shall require trim coils capable of trimming the field within +/- TBD (%) of the Peak field.02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall require current taps for operation TBD ( Y or N)02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/04/2025On HoldFALSE
- 6.04.03.02The physical magnet length shall be <TBD m.02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet bore gap and Width shall be 48 mm.02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall be designed to fit within the following envelope. TBD02/04/2025On HoldFALSE
- 6.04.03.02Magnet installation tolerances TBD02/04/2025On HoldFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025On HoldFALSE
- 6.04.03.02The magnet integrated dipole field (B) shall be TBD.02/12/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet good field aperture dAx required shall be TBD mm.02/04/2025On HoldFALSE
- 6.04.03.02The magnet good field aperture dAy required shall be TBD mm.02/04/2025On HoldFALSE
- 6.04.03.02The magnet to magnet field variability between magnets shall be 0.5 %.02/04/2025On HoldFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025On HoldFALSE
- 6.04.03.02The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/04/2025On HoldFALSE
- 6.04.03.02The Field at the reference radius at the design energy of 18 GeV shall be 29e-3 (T) .02/04/2025On HoldFALSE
- 6.04.03.02The magnet bore field shall require the following multipole content:02/04/2025On HoldFALSE
- 6.04.03.02b1 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b2 = 10000 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b3 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b4 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b5 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b6 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b7 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b8 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b9 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b10 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b11 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b12 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b13 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b14 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b15 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b16 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall not be designed to limit CrossTalk requirements.02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall be designed to meet the following fringe field requirements TBD02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/04/2025On HoldFALSE
ESR-MAG-CORR_SPNQ_V : ESR Vertical Corrector (SPNQ_V) Magnet
- 6.04.03.02The magnet shall be a single function dipole with a horizontal field.02/05/2025On HoldFALSE
- 6.04.03.02The magnet shall require trim coils capable of trimming the field within +/- TBD (%) of the Peak field.02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall require current taps for operation TBD ( Y or N)02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/04/2025On HoldFALSE
- 6.04.03.02The physical magnet length shall be <TBD m.02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet bore gap and Width shall be 130.1 mm.02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall be designed to fit within the following envelope. TBD02/04/2025On HoldFALSE
- 6.04.03.02Magnet installation tolerances TBD02/04/2025On HoldFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025On HoldFALSE
- 6.04.03.02The magnet integrated dipole field (B) shall be TBD.02/12/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet good field aperture dAx required shall be TBD mm.02/04/2025On HoldFALSE
- 6.04.03.02The magnet good field aperture dAy required shall be TBD mm.02/04/2025On HoldFALSE
- 6.04.03.02The magnet to magnet field variability between magnets shall be 0.5 %.02/04/2025On HoldFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025On HoldFALSE
- 6.04.03.02The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/04/2025On HoldFALSE
- 6.04.03.02The Field at the reference radius at the design energy of 18 GeV shall be 29e-3 (T) .02/04/2025On HoldFALSE
- 6.04.03.02The magnet bore field shall require the following multipole content:02/04/2025On HoldFALSE
- 6.04.03.02b1 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b2 = 10000 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b3 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b4 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b5 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b6 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b7 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b8 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b9 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b10 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b11 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b12 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b13 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b14 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b15 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02b16 < 1 (10^-4)02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall not be designed to limit CrossTalk requirements.02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall be designed to meet the following fringe field requirements TBD02/04/2025On HoldFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/04/2025On HoldFALSE
- 6.04.03.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/04/2025On HoldFALSE
ESR-MAG-CORR_V : ESR Vertical Corrector (CORR_V) Magnet
- 6.04.03.02The magnet shall operate with a single functions with a Dipole field type and Vertical field direction.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The maximum physical magnet length shall be 0.2 m.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The minimal magnet bore gap and width shall be TBD and 52 mm.02/13/2025In ProcessFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The magnet integrated dipole field (B) shall be 12 milli T-m.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet good field aperture radius, dAx, shall be 24 mm.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet good field aperture radius, dAy, shall be 15 mm.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet to magnet variability shall be < 5e-3.02/05/2025In ProcessFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The magnet shall be measured using the field homogeneity measurement methodology defined to satisfy its operational harmonic multipole content constraints.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet field quality shall be measured within dAx and dAy radii utilizing a hall probe.02/06/2025ApprovedFALSE
- 6.04.03.0202/13/2025Not ApplicableFALSE
- 6.04.03.02The magnet bore field shall have a total field homogeneity of better than 0.003 dB/B and a btotal of <±100.02/06/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet axis.02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/06/2025ApprovedFALSE
- 6.04.03.02The magnet shall be designed with components capable to withstand a radiation dose of TBD MGy, or be approved by EIC for use in a specific location as shown in the “BNL Materials” List Doc. No. TBD.02/13/2025In ProcessFALSE
ESR-MAG-D13 : ESR Dipole (D1/D3) Magnet
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/27/2025ApprovedFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).01/27/2025ApprovedFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the tunnel header(s) to the girder for Low Conductivity Water LCW) to the Accelerator Installation WBS CAM.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.01/27/2025ApprovedFALSE
- The Power Supply Group shall purchase the cables that go from the power supply to the magnets.01/27/2025ApprovedFALSE
- The Power Supply Group shall provide a power supply to be utilized by the magnets.01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the main winding magnet power cables at the magnet and power supply.01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the trim coils to the chassis cables at the magnet and the power supply.01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables.01/27/2025ApprovedFALSE
- The Magnet Group shall provide trim coils.01/27/2025ApprovedFALSE
- The Magnet Group shall provide several current taps to facilitate running at different operational modes.01/27/2025ApprovedFALSE
- The Accelerator Installation Group WBS CAM shall fund and schedule installion of cabling from the power supply to the magnets.01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s).01/27/2025ApprovedFALSE
- The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the power supply.01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the thermal switch cables at the magnet and power supply.01/27/2025ApprovedFALSE
- The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe01/27/2025ApprovedFALSE
- The Magnet Group shall procure the girders for the Accelerator Installation Group to install.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the magnets on the girders.01/27/2025ApprovedFALSE
- The Vacuum Group shall design the beampipe integration to the magnets.01/27/2025ApprovedFALSE
- The Vacuum Group will provide temporary thermal couples to the magnets for bakeout. The magnet and coil temperature shall not exceed TBD degrees F during bakeout01/27/2025ApprovedFALSE
- The Magnet Group will provide supports for beam pipes.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups. .01/27/2025ApprovedFALSE
- The Survey Group shall fiducialize the magnet based on magnetic center.01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders.01/27/2025ApprovedFALSE
- The Infrastructure Group will provide the arrangement location for the magnets and girders01/27/2025ApprovedFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- ESR-MAG-D13 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
ESR-MAG-D2 : ESR Dipole (D2) Magnet
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/27/2025In ProcessFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the magnet to the girder for Low Conductivity Water (LCW).01/27/2025In ProcessFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections from the header(s) to the girder for Low Conductivity Water LCW) to the Accelerator Installation WBS CAM.01/27/2025In ProcessFALSE
- The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/27/2025In ProcessFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit.01/27/2025In ProcessFALSE
- The Power Supply Group shall purchase the cables that go from the power supply to the magnets.01/27/2025In ProcessFALSE
- The Power Supply Group shall provide a power supply to be utilized by the magnets.01/27/2025In ProcessFALSE
- The Power Supply Group shall make connections of the main winding magnet power cables at the magnet and power supply.01/27/2025In ProcessFALSE
- The Power Supply Group shall make connections of the trim coils to the chassis cables at the magnet and the power supply.01/27/2025In ProcessFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables.01/27/2025In ProcessFALSE
- The Magnet Group shall provide trim coils.01/27/2025In ProcessFALSE
- The Magnet Group shall provide several current taps to facilitate running at different operational modes.01/27/2025In ProcessFALSE
- The Accelerator Installation Group WBS CAM shall fund and schedule installion of cabling from the power supply to the magnets.01/27/2025In ProcessFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s).01/27/2025In ProcessFALSE
- The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the power supply.01/27/2025In ProcessFALSE
- The Power Supply Group shall make connections of the thermal switch cables at the magnet and power supply.01/27/2025In ProcessFALSE
- The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection.01/27/2025In ProcessFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.01/27/2025In ProcessFALSE
- The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe01/27/2025In ProcessFALSE
- The Magnet Group shall procure the girders for the Accelerator Installation Group to install.01/27/2025In ProcessFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the magnets on the girders.01/27/2025In ProcessFALSE
- The Vacuum Group shall design the beampipe integration to the magnets.01/27/2025In ProcessFALSE
- The Vacuum Group will provide temporary thermal couples to the magnets for bakeout. The magnet and coil temperature shall not exceed TBD degrees F during bakeout01/27/2025In ProcessFALSE
- The Magnet Group will provide supports for beam pipes.01/27/2025In ProcessFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups. .01/27/2025In ProcessFALSE
- The Survey Group shall fiducialize the magnet based on magnetic center.01/27/2025In ProcessFALSE
- The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders.01/27/2025In ProcessFALSE
- The Infrastructure Group will provide the arrangement location for the magnets and girders01/27/2025In ProcessFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01nan02/13/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- ESR-MAG-D2 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
ESR-MAG-D_SR : ESR Dipole (D_SR) Magnet
- 6.04.02.01The magnet shall be a single function dipole with a vertical field.02/05/2025ApprovedFALSE
- 6.04.02.01The magnet shall require trim coils capable of trimming the field within +/- TBD (%) of the Peak field.02/04/2025In ProcessFALSE
- 6.04.02.01The magnet shall require current taps for operation TBD ( Y or N)02/04/2025In ProcessFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/04/2025In ProcessFALSE
- 6.04.02.01The physical magnet length shall be <3.8 m.02/04/2025ApprovedFALSE
- 6.04.02.01The magnet model length shall be 3.8 m.02/04/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01The magnet bore gap and Width shall be 52 mm.02/04/2025ReviewedFALSE
- 6.04.02.01The magnet shall be designed to fit within the following envelope. TBD02/04/2025In ProcessFALSE
- 6.04.02.01Magnet installation tolerances TBD02/05/2025Not ApplicableFALSE
- 6.04.02.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.02.01The magnet integrated dipole field (B) shall be 1.2 T.m.02/04/2025ReviewedFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 32.59 mm.02/04/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAy required shall be 11.46 mm.02/04/2025ApprovedFALSE
- 6.04.02.01The magnet to magnet field variability between magnets shall be 1 %.02/04/2025ApprovedFALSE
- 6.04.02.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.02.01The field shall be measured at two locations as follows: Harmonic Measurements region 1; Rref1=TBD mm centered at (-TBD,TBD) mm, Harmonic Measurements region 2; Rref2=TBD mm centered at (TBD,TBD) mm.02/13/2025In ProcessFALSE
- 6.04.02.01The reference field for the different measurements shall be: Measurement 1; Bref1=TBD (T) in Region TBD, Region TBD , Measurement 2; Bref2=TBD (T) in Region TBD, Region TBD.02/13/2025In ProcessFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region TBD, of better than dB/B <10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1 & 2.02/13/2025ApprovedFALSE
- 6.04.02.01Region 1: b1 = 10000, Region 2: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01Region1: -4<b2<4, Region2: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.02/04/2025In ProcessFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/04/2025In ProcessFALSE
- 6.04.02.01The magnet shall be designed to meet the following fringe field requirements TBD02/04/2025In ProcessFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/04/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/04/2025ApprovedFALSE
ESR-MAG-Q120
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation.02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <1.2 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 40 mm.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances, Max X = 42 cm May Y = 42 cm02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 18.9 T/m.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet center.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-Q50 : ESR Quadrupole (Q50) Magnet
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections for Low Conductivity Water LCW) to the Accelerator Installation WBS CAM.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule the installation of cabling from the power supply to the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the main winding magnet power cables at the magnet and PS ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit01/27/2025ApprovedFALSE
- The Power Supply Group shall provide a power supply to be utilized by the magnets01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)01/27/2025ApprovedFALSE
- The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe01/27/2025ApprovedFALSE
- The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/27/2025ApprovedFALSE
- The Vacuum Group shall design the beampipe integration to the magnets01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups. .01/27/2025ApprovedFALSE
- The Vacuum Group will provide temporary thermal couples to the magnets for bakeout. The magnet and coil temperature shall not exceed TBD degrees F during bakeout01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders01/27/2025ApprovedFALSE
- The Infrastructure Group will provide the arrangement location for the magnets and girders01/27/2025ApprovedFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- ESR-MAG-Q50 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
ESR-MAG-Q60 : ESR Quadrupole (Q60) Magnet
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections for Low Conductivity Water LCW) to the Accelerator Installation WBS CAM.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule the installation of cabling from the power supply to the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the main winding magnet power cables at the magnet and PS ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit01/27/2025ApprovedFALSE
- The Power Supply Group shall provide a power supply to be utilized by the magnets01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)01/27/2025ApprovedFALSE
- The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe01/27/2025ApprovedFALSE
- The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/27/2025ApprovedFALSE
- The Vacuum Group shall design the beampipe integration to the magnets01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Vacuum Group will provide temporary thermal couples to the magnets for bakeout. The magnet and coil temperature shall not exceed TBD degrees F during bakeout01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders01/27/2025ApprovedFALSE
- The Infrastructure Group will provide the arrangement location for the magnets and girders01/27/2025ApprovedFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- ESR-MAG-Q60 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
ESR-MAG-Q60x2 : ESR Quadrupole (Q60x2) Magnet
ESR-MAG-Q80 : ESR Quadrupole (Q80) Magnet
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections for Low Conductivity Water LCW) to the Accelerator Installation WBS CAM.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule the installation of cabling from the power supply to the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the main winding magnet power cables at the magnet and PS ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit01/27/2025ApprovedFALSE
- The Power Supply Group shall provide a power supply to be utilized by the magnets01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the Beam Based Alignment shunt resistor/switch power cables and connections from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Magnet Group shall provide a way to mount the Beam Based Alignment shunt resistor/switch chasssis from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the Beam Based Alignment shunt resistor/switch chasssis cables at the magnet and PS ends01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)01/27/2025ApprovedFALSE
- The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe01/27/2025ApprovedFALSE
- The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/27/2025ApprovedFALSE
- The Vacuum Group shall design the beampipe integration to the magnets01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Vacuum Group will provide temporary thermal couples to the magnets for bakeout. The magnet and coil temperature shall not exceed TBD degrees F during bakeout01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders01/27/2025ApprovedFALSE
- The Infrastructure Group will provide the arrangement location for the magnets and girders01/27/2025In ProcessFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- ESR-MAG-Q80 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
ESR-MAG-Q_LA : ESR Quadrupole (Q_LA) Magnet
- 6.04.03.01.02The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/04/2025In ProcessFALSE
- 6.04.03.01.02The physical magnet length shall be <1.2 m.02/04/2025ApprovedFALSE
- 6.04.03.01.02The magnet model length shall be 1.20 m.02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet pole tip radius shall be 80 mm.02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall be designed to fit within the following envelope:02/05/2025In ProcessFALSE
- 6.04.03.01.02Magnet installation tolerances TBD02/04/2025In ProcessFALSE
- 6.04.03.01.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet integrated grad field G shall be 9.7 T/m.02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet to magnet field variability between magnets shall be less than 0.1%.02/04/2025ApprovedFALSE
- 6.04.03.01.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01.02The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01.02The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01.02The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01.02-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01.02-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01.02-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01.02-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/04/2025In ProcessFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/04/2025In ProcessFALSE
- 6.04.03.01.02The magnet shall be designed to meet the following fringe field requirements TBD02/04/2025In ProcessFALSE
- 6.04.03.01.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/04/2025ApprovedFALSE
- 6.04.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/04/2025ApprovedFALSE
ESR-MAG-Q_NARROW : ESR Quadrupole (Q_NARROW) Magnet
- 6.04.03.01.02The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation.02/04/2025In ProcessFALSE
- 6.04.03.01.02The physical magnet length shall be <0.6 m.02/04/2025ApprovedFALSE
- 6.04.03.01.02The magnet model length shall be 0.60 m.02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet pole tip radius shall be 40 mm.02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall be designed to fit within the following envelope:02/05/2025In ProcessFALSE
- 6.04.03.01.02Magnet installation tolerances, Max X = 42 cm May Y = 42 cm02/13/2025In ProcessFALSE
- 6.04.03.01.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet integrated grad field G shall be 18.9 T/m.02/04/2025ApprovedFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet to magnet field variability between magnets shall be less than 0.1%.02/04/2025ApprovedFALSE
- 6.04.03.01.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01.02The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01.02The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01.02The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01.02-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01.02-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01.02-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01.02-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall not be designed to limit CrossTalk requirements.02/04/2025In ProcessFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet shall be designed to specifically constrain the external fringe field Y (Yes or No)02/04/2025In ProcessFALSE
- 6.04.03.01.02The magnet shall be designed to meet the following fringe field requirements The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at TBD mm from the magnet center.02/04/2025In ProcessFALSE
- 6.04.03.01.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.01.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/04/2025ApprovedFALSE
- 6.04.03.01.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/04/2025ApprovedFALSE
ESR-MAG-Q_SKEW : ESR Quadrupole (Q_SKEW) Magnet
- 6.04.03.02The magnet shall be a single function quadrupole with a skew field rotation.02/05/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/04/2025In ProcessFALSE
- 6.04.03.02The physical magnet length shall be <0.25 m.02/04/2025ApprovedFALSE
- 6.04.03.02The magnet model length shall be 0.25 m.02/04/2025ApprovedFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet pole tip radius shall be 67.9 mm.02/04/2025ReviewedFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to fit within the following envelope. TBD02/04/2025In ProcessFALSE
- 6.04.03.02Magnet installation tolerances TBD02/04/2025In ProcessFALSE
- 6.04.03.02The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet integrated grad field G shall be 1.5 T/m.02/04/2025ReviewedFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet to magnet field variability between magnets shall be less than 0.1%.02/04/2025ApprovedFALSE
- 6.04.03.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.02The harmonic reference radius at the design energy of 18 GeV shall be TBD (mm) .02/04/2025In ProcessFALSE
- 6.04.03.02The field at the reference radius at the design energy of 18 GeV shall be TBD (T/m) .02/04/2025In ProcessFALSE
- 6.04.03.02The magnet bore field shall require the following multipole content:02/04/2025In ProcessFALSE
- 6.04.03.02b1 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b2 = 10000 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b3 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b4 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b5 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b6 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b7 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b8 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b9 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b10 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b11 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b12 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b13 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b14 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b15 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02b16 < 1 (10^-4)02/04/2025In ProcessFALSE
- 6.04.03.02The magnet shall not be designed to limit CrossTalk requirements.02/04/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/04/2025In ProcessFALSE
- 6.04.03.02The magnet shall be designed to meet the following fringe field requirements TBD02/04/2025In ProcessFALSE
- 6.04.03.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.03.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/04/2025ApprovedFALSE
- 6.04.03.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/04/2025ApprovedFALSE
ESR-MAG-QLS1 : ESR Magnet QLS1
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <1.22 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 1.22 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope:02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-QLS2 : ESR Magnet QLS2
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <1.97 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 1.97 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-QLS3 : ESR Magnet QLS3
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <1.2 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 1.22 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-QLS4 : ESR Magnet QLS4
ESR-MAG-QSS1 : ESR Magnet QSS1
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <0.87 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 0.87 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-QSS2 : ESR Magnet QSS2
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <1.22 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 1.22 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-QSS3 : ESR Magnet QSS3
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <1.97 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 1.97 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-QSS4 : ESR Magnet QSS4
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <1.22 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 1.22 m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-QSS5 : ESR Magnet QSS5
- 6.04.03.01The magnet shall be single function quadrupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/05/2025In ProcessFALSE
- 6.04.03.01The physical magnet length shall be <0.87 m.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet model length shall be 0.87m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet pole tip radius shall be 67.9 mm.02/05/2025ReviewedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to fit within the following envelope. TBD02/05/2025In ProcessFALSE
- 6.04.03.01Magnet installation tolerances TBD02/05/2025In ProcessFALSE
- 6.04.03.01The magnet install center and install alignment must be within a translational value of +/-TBD (um) and a rotational alignment value of +/- TBD (mrad).02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet integrated grad field G shall be 24 T/m.02/05/2025ApprovedFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet to magnet field variability between magnets shall be less than 0.1%.02/05/2025ApprovedFALSE
- 6.04.03.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/- TBD (um) and a rotational alignment value of +/- TBD(mrad).02/05/2025In ProcessFALSE
- 6.04.03.01The field harmonic measurements shall be measured at Rref=25mm.02/07/2025ApprovedFALSE
- 6.04.03.01The reference field for the measurement shall be TBD (T/m).02/07/2025In ProcessFALSE
- 6.04.03.01The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.owing multipole content:02/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01b2=1000002/07/2025ApprovedFALSE
- 6.04.03.01-2.71<b3<0.72, -2.13<a3<2.1102/07/2025ApprovedFALSE
- 6.04.03.01-3.58<b4<1.2, -0.47<a4<0.4602/07/2025ApprovedFALSE
- 6.04.03.01-0.53<b5<0.23, -0.29<a5<0.3502/07/2025ApprovedFALSE
- 6.04.03.01-1.75<b6<-1.12, -0.07<a6<0.0802/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-1.06<b10<-0.66, 0<a10<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-0.12<b14<-0.08, 0<a14<002/07/2025ApprovedFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01-, -02/07/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall not be designed to limit CrossTalk requirements.02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet shall be designed to specifically constrain the external fringe field TBD (Yes or No)02/05/2025In ProcessFALSE
- 6.04.03.01The magnet shall be designed to meet the following fringe field requirements TBD02/05/2025In ProcessFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.03.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.03.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-SXT : ESR Sextupole (SXT) Magnet
- Infrastructure shall provide Low Conductivity Water (LCW) to/from the common supply/return header(s) with isolation valves in the tunnel to be utilized by the magnets.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule piping installation (or hosing) between the Low Conductivity Water (LCW) infrastructure and the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Magnet Group shall provide a design and materials to facilitate installation of the piping (or hose) and connections for Low Conductivity Water LCW) to the Accelerator Installation WBS CAM.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule the installation of cabling from the power supply to the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall purchase the cables that go from the Power Supply to the magnets.01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the main winding magnet power cables at the magnet and PS ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet power circuit01/27/2025ApprovedFALSE
- The Power Supply Group shall provide a power supply to be utilized by the magnets01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to facilitate installation of the main winding magnet power cables from the Accelerator Installation WBS CAM01/27/2025ApprovedFALSE
- The Magnet Group shall provide a terminal block to support shorting out of the main SXT winding if required.01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing cabling from the thermal switch lugs on the magnet to the Power Supply Group by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Power Supply Group shall procure the cables that go from the thermal switch on the magnet to the Power Supply01/27/2025ApprovedFALSE
- The Power Supply Group shall make connections of the thermal switch cables at the magnet and Power Supply ends01/27/2025ApprovedFALSE
- The Power Supply Group shall create schematics with the number of cables and connectors made to the magnet protection circuit(s)01/27/2025ApprovedFALSE
- The Magnet Group shall provide the terminal block to facilitate installation of the connection for thermal protection01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide a girder design and CAD model to be utilized by the magnet group. They will ensure the design does not interfere with magnet or cable installation around the beampipe01/27/2025ApprovedFALSE
- The BNL Magnet Group shall procure the girders for the Accelerator Installation WBS CAM to install.01/27/2025ApprovedFALSE
- The Vacuum Group shall design the beampipe integration to the magnets01/27/2025ApprovedFALSE
- The Accelerator Installation WBS CAM shall fund and schedule installing the beampipe through the magnets and installing the magnets by the appropriate technical support groups.01/27/2025ApprovedFALSE
- The Vacuum Group will provide temporary thermal couples to the magnets for bakeout. The magnet and coil temperature shall not exceed TBD degrees F during bakeout01/27/2025ApprovedFALSE
- The Mechanical Design Group shall provide the volume within the tunnel required by the magnets and girders01/27/2025ApprovedFALSE
- 6.04.02.02The magnet shall have a single function02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The magnet shall have a Sextupole field02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The magnet shall have a normal field rotation02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The physical magnet length shall be <0.32 m.02/05/2025ApprovedFALSE
- 6.04.02.02The effective magnet length shall be 0.24 m02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The magnet integrated grad field G shall be 97.2 T/m.02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The harmonic reference radius and current shall be 25 (mm) and 161 (A).02/05/2025ApprovedFALSE
- 6.04.02.02The Field at the reference radius and current shall be 405 (T/m^2).02/05/2025ApprovedFALSE
- 6.04.02.02The magnet bore field shall require the following multipole content:02/05/2025ApprovedFALSE
- 6.04.02.02b1 = N/A a1 = N/A02/05/2025ApprovedFALSE
- 6.04.02.02b2 = N/A a2 = N/A02/05/2025ApprovedFALSE
- 6.04.02.023 = 10000 a3 = N/A02/05/2025ApprovedFALSE
- 6.04.02.02b4 = HV +/- 5.3 a4 = HV +/- 7.102/05/2025ApprovedFALSE
- 6.04.02.025 = HV +/- 1.6 a5 = HV +/- 1.902/05/2025ApprovedFALSE
- 6.04.02.02b6 = HV +/- 1.0 a6 = +/- 2.002/05/2025ApprovedFALSE
- 6.04.02.02b7 = HV +/- 1.0 a7 = +/- 0.802/05/2025ApprovedFALSE
- 6.04.02.02b8 = HV +/- 1.0 a8 = +/- 0.502/05/2025ApprovedFALSE
- 6.04.02.02b9 = HV +/- 1.0 a9 = +/- 0.502/05/2025ApprovedFALSE
- 6.04.02.02b10 = HV +/- 1.0 a10 = +/- 0.302/05/2025ApprovedFALSE
- 6.04.02.02b11 = HV +/- 1.0 a11 = +/- 0.2802/05/2025ApprovedFALSE
- 6.04.02.02b12 = HV +/- 1.0 a12 = +/- 0.2602/05/2025ApprovedFALSE
- 6.04.02.02b13 = HV +/- 1.0 a13 = +/- 0.2402/05/2025ApprovedFALSE
- 6.04.02.02b14 = HV +/- 1.0 a14 = +/- 0.2202/05/2025ApprovedFALSE
- 6.04.02.02b15 = HV +/- 1.0 a15 = +/- 0.2102/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The magnetic field, center and alignment, within the magnet mustbe known to within a translational value of +/-50 (um) and arotational alignment value of +/-0.5 (mrad).02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The magnet install center and install alignment must be within atranslational value of +/-150(um) and a rotational alignment valueof +/-0.5 (mrad).02/05/2025ApprovedFALSE
- 6.04.02.02The magnet shall be designed to have a splitable pole toaccommodate the Vacuum beam pipe installation.02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/05/2025Not ApplicableFALSE
- 6.04.02.02The magnet design and verification process shall ensure the finalmagnet will meet the reliability needs of the EIC over it plannedoperational life of >20 Years.02/05/2025ApprovedFALSE
- 6.04.02.02The magnet shall be designed to operate reliably given thecumulative radiation dose of TBD Rads it will experience over thelifetime of the EIC of >20 Years.02/05/2025ApprovedFALSE
ESR-MAG-SXT_LONG : ESR Sextupole (Long_SXT) Magnet
- 6.04.02.02The magnet shall be a single function sextupole with a normal field rotation.02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02The magnet shall be designed to have a splitable pole to accommodate the Vacuum beam pipe installation TBD02/04/2025In ProcessFALSE
- 6.04.02.02The physical magnet length shall be <0.57 m.02/04/2025ApprovedFALSE
- 6.04.02.02The magnet model length shall be 0.57 m.02/04/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02The magnet pole tip radius shall be 68 mm.02/04/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02The magnet shall be designed to fit within the following envelope. TBD02/04/2025In ProcessFALSE
- 6.04.02.02Magnet installation tolerances TBD02/04/2025In ProcessFALSE
- 6.04.02.02The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).02/05/2025ApprovedFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02The magnet integrated grad field G shall be 112.3 T/m2.02/04/2025ReviewedFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02The magnet to magnet field variability between magnets shall be less than 0.1%.02/04/2025ApprovedFALSE
- 6.04.02.02The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).02/05/2025In ProcessFALSE
- 6.04.02.02The field harmonic measurements shall be measured at Rref=27mm.02/07/2025ApprovedFALSE
- 6.04.02.02The reference field for the measurement shall be TBD (T/m^2)02/07/2025In ProcessFALSE
- 6.04.02.02The magnet bore field shall have a field homogeneity in the region of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in the region.02/07/2025ApprovedFALSE
- 6.04.02.02-83.71<b1<-21.31, -44.04<a1<26.8202/07/2025ApprovedFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02b3=1000002/07/2025ApprovedFALSE
- 6.04.02.02-4.15<b4<5.66, -4.83<a4<8.3902/07/2025ApprovedFALSE
- 6.04.02.02-1.97<b5<0.82, -2.36<a5<0.9902/07/2025ApprovedFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02-0.12<b7<0.54, -0.35<a7<0.5202/07/2025ApprovedFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02-12.67<b9<-12.52, -0.19<a9<-0.0402/07/2025ApprovedFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02-0.47<b15<-0.46, 0<a15<002/07/2025ApprovedFALSE
- 6.04.02.02-, -02/07/2025Not ApplicableFALSE
- 6.04.02.02The magnet shall not be designed to limit CrossTalk requirements.02/04/2025In ProcessFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02The magnet shall be designed to specifically constrain the external fringe field.02/04/2025In ProcessFALSE
- 6.04.02.02The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS Beamline.02/04/2025In ProcessFALSE
- 6.04.02.02< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02< Requirement Not Applicable >02/04/2025Not ApplicableFALSE
- 6.04.02.02The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.02/04/2025ApprovedFALSE
- 6.04.02.02The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.02/04/2025ApprovedFALSE
ESR-MPS : ESR Machine Protection System
ESR-MPS-EXT_KICK : ESR Machine Protection System Single Turn Extraction Kicker
- ESR-MPS-EXT_KICK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.01The number of kickers shall be 601/27/2025In ProcessFALSE
- 6.06.03.01The Rise time shall be 900 ns01/27/2025In ProcessFALSE
- 6.06.03.01The Fall time shall be NA sec01/27/2025In ProcessFALSE
- 6.06.03.01The flat top time shall be 13 us01/27/2025In ProcessFALSE
- 6.06.03.01The waveshape shall be trap01/27/2025In ProcessFALSE
- 6.06.03.01The painting shall be vertical01/27/2025In ProcessFALSE
- 6.06.03.01The maximum field shall be 0.12 T01/27/2025In ProcessFALSE
- 6.06.03.01The total deflection shall be 16 mrad01/27/2025In ProcessFALSE
- 6.06.03.01The maximum current shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The maximum voltage shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The inductance with cable shall be TBD (uH)01/27/2025In ProcessFALSE
- 6.06.03.01The Max rep rate shall be 100 kV/pC01/27/2025In ProcessFALSE
- 6.06.03.01The flat top repeatability shall be NA Hz01/27/2025In ProcessFALSE
- 6.06.03.01The flatness of flat top/pulse form shall be 1 %01/27/2025In ProcessFALSE
- 6.06.03.01The beam abort kicker shall be tbd %01/27/2025In ProcessFALSE
- 6.06.03.01The cooling type shall be w (W,A)01/27/2025In ProcessFALSE
ESR-MPS-EXT_KICK-DUMP_BLK : ESR Machine Protection System Dump block
- ESR-MPS-EXT_KICK-DUMP_BLK EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.01The diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 2 m01/27/2025In ProcessFALSE
- 6.06.03.01The materials shall be C / Al / Cu01/27/2025In ProcessFALSE
- 6.06.03.01The energy deposited during abort shall be 320 kJ01/27/2025In ProcessFALSE
- 6.06.03.01The frequency of thermal cycle shall be 1 hour01/27/2025In ProcessFALSE
- 6.06.03.01The window thickness shall be tbd mm01/27/2025In ProcessFALSE
- 6.06.03.01The window material shall be tbd01/27/2025In ProcessFALSE
ESR-MPS-EXT_KICK-INST_BP : ESR Machine Protection System Beam Pipe & Instrumentation
- ESR-MPS-EXT_KICK-INST_BP EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.01The Length shall be 50 m01/27/2025In ProcessFALSE
- 6.06.03.01The Internal diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The Temperature sensors shall be yes01/27/2025In ProcessFALSE
- 6.06.03.01The BPMs shall be 401/27/2025In ProcessFALSE
- 6.06.03.01The Correctors shall be 401/27/2025In ProcessFALSE
- 6.06.03.01The Corrector PS shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The Cooling / pumping shall be yes01/27/2025In ProcessFALSE
ESR-MPS-EXT_KICK-LAMBERT : ESR Machine Protection System Lamberton Magnet
- ESR-MPS-EXT_KICK-LAMBERT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.01The deflection shall be 2 mrad01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 1.2 m01/27/2025In ProcessFALSE
- 6.06.03.01The power supply shall be 1600 A01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The Y-chamber aperture shall be 36 mm01/27/2025In ProcessFALSE
- 6.06.03.01May need to add additional window requirements for other leg of Lambertson magnet TBD01/27/2025In ProcessFALSE
ESR-MPS-EXT_KICK-QUAD : ESR Machine Protection System Line Quads
- ESR-MPS-EXT_KICK-QUAD EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03.01The gradient shall be 17 T/m01/27/2025In ProcessFALSE
- 6.06.03.01The power supply shall be 1600 A01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 70 cm01/27/2025In ProcessFALSE
- 6.06.03.01The aperture radius shall be 50 mm01/27/2025In ProcessFALSE
ESR-PPD : EZSR Pulse Power Device System
ESR-PPD-INJ_EXT_PULSE_KICK : EIS RCS Injection/Extraction Pulsed Bump IM&HW System
- 6.03.03.05.07The location (Section) shall be 12 o'clock01/27/2025In ProcessFALSE
- 6.03.03.05.07The dimension in W shall be 4 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.07The dimension in L shall be 10 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.07The dimension in H shall be 7 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.07The num magnets shall be 301/27/2025In ProcessFALSE
- 6.03.03.05.07The mag gap shall be tbd (cm)01/27/2025In ProcessFALSE
- 6.03.03.05.07The rise time shall be 0.001 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.07The fall time shall be 0.002 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.07The flat top time shall be 0.0015 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.07The waveshape shall be trap01/27/2025In ProcessFALSE
- 6.03.03.05.07The flat top repeatability shall be tbd %01/27/2025In ProcessFALSE
- 6.03.03.05.07The uniformity of the flattop shall be tbd (V)01/27/2025In ProcessFALSE
- 6.03.03.05.07The deflecting Angle shall be 1 (mRad)01/27/2025In ProcessFALSE
- 6.03.03.05.07The rep rate spec shall be 1 (Hz)01/27/2025In ProcessFALSE
- 6.03.03.05.07The output voltage Spec shall be 60 (Volts)01/27/2025In ProcessFALSE
- 6.03.03.05.07The output current spec shall be 1000 (Amps)01/27/2025In ProcessFALSE
- 6.03.03.05.07The inductance with cable shall be 11 (uH)01/27/2025In ProcessFALSE
- 6.03.03.05.07The cooling type shall be water01/27/2025In ProcessFALSE
ESR-PPD-INJ_EXT_PULSE_SEPTUM : EIS RCS Injection/Extraction Pulsed Septum System
- 6.03.03.05.08The location (Section) shall be 12 o'clock01/27/2025In ProcessFALSE
- 6.03.03.05.08The dimension in W shall be 4 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.08The dimension in L shall be 10 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.08The dimension in H shall be 7 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.08The num magnets shall be 2 ea01/27/2025In ProcessFALSE
- 6.03.03.05.08The mag gap shall be tbd01/27/2025In ProcessFALSE
- 6.03.03.05.08The rise time shall be 0.001 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.08The fall time shall be 0.0015 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.08The flat top time shall be 0.002 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.08The waveshape shall be Trap01/27/2025In ProcessFALSE
- 6.03.03.05.08The flat top repeatability shall be tbd %01/27/2025In ProcessFALSE
- 6.03.03.05.08The uniformity of the flattop shall be tbd (V)01/27/2025In ProcessFALSE
- 6.03.03.05.08The deflecting Angle shall be 15 (mRad)01/27/2025In ProcessFALSE
- 6.03.03.05.08The rep rate spec shall be 1 (Hz)01/27/2025In ProcessFALSE
- 6.03.03.05.08The output voltage Spec shall be 80 (Volts)01/27/2025In ProcessFALSE
- 6.03.03.05.08The output current spec shall be 1000 (Amps)01/27/2025In ProcessFALSE
- 6.03.03.05.08The inductance with cable shall be 25 (uH)01/27/2025In ProcessFALSE
- 6.03.03.05.08The cooling type shall be water01/27/2025In ProcessFALSE
ESR-PPD-INJ_EXT_PULSE_STRIPLINE_KICK : EIS RCS Injection/Extraction Pulse Power Strip-line Kicker System
- 6.03.03.05.06The location (Section) type shall be 12 o'clock01/27/2025In ProcessFALSE
- 6.03.03.05.06The dimension in W type shall be 5 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.06The dimension in L type shall be 5 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.06The dimension in H type shall be 7 (ft)01/27/2025In ProcessFALSE
- 6.03.03.05.06The num magnets type shall be 1001/27/2025In ProcessFALSE
- 6.03.03.05.06The mag gap type shall be 3 (cm)01/27/2025In ProcessFALSE
- 6.03.03.05.06The rise time type shall be 0.000000002 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.06The fall time type shall be 0.0000000025 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.06The flat top time type shall be 0.000000013 (sec)01/27/2025In ProcessFALSE
- 6.03.03.05.06The waveshape type shall be Trap01/27/2025In ProcessFALSE
- 6.03.03.05.06The flat top repeatability type shall be tbd %01/27/2025In ProcessFALSE
- 6.03.03.05.06The uniformity of the flattop type shall be tbd (V)01/27/2025In ProcessFALSE
- 6.03.03.05.06The deflecting Angle type shall be 1.3 (mRad)01/27/2025In ProcessFALSE
- 6.03.03.05.06The rep rate spec type shall be 1 (Hz)01/27/2025In ProcessFALSE
- 6.03.03.05.06The output voltage Spec type shall be 20000 (Volts)01/27/2025In ProcessFALSE
- 6.03.03.05.06The output current spec type shall be 400 (Amps)01/27/2025In ProcessFALSE
- 6.03.03.05.06The inductance with cable type shall be 50 ohm (TBD) (uH)01/27/2025In ProcessFALSE
- 6.03.03.05.06The cooling type shall be water01/27/2025In ProcessFALSE
ESR-PROT : Machine Protection
- ESR-PROT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03The ESR shall contain an Abort system to dump the beam01/27/2025ReviewedFALSE
- 6.06.03.01The diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 2 m01/27/2025In ProcessFALSE
- 6.06.03.01The materials shall be C / Al / Cu01/27/2025In ProcessFALSE
- 6.06.03.01The energy deposited during abort shall be 320 kJ01/27/2025In ProcessFALSE
- 6.06.03.01The frequency of thermal cycle shall be 1 hour01/27/2025In ProcessFALSE
- 6.06.03.01The window thickness shall be tbd mm01/27/2025In ProcessFALSE
- 6.06.03.01The window material shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The Length shall be 50 m01/27/2025In ProcessFALSE
- 6.06.03.01The Internal diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The Temperature sensors shall be yes01/27/2025In ProcessFALSE
- 6.06.03.01The BPMs shall be 401/27/2025In ProcessFALSE
- 6.06.03.01The Correctors shall be 401/27/2025In ProcessFALSE
- 6.06.03.01The Corrector PS shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be tbd01/27/2025In ProcessFALSE
- 6.06.03.01The Cooling / pumping shall be yes01/27/2025In ProcessFALSE
- 6.06.03.01The deflection shall be 2 mrad01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 1.2 m01/27/2025In ProcessFALSE
- 6.06.03.01The power supply shall be 1600 A01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The Y-chamber aperture shall be 36 mm01/27/2025In ProcessFALSE
- 6.06.03.01May need to add additional window requirements for other leg of Lambertson magnet TBD01/27/2025In ProcessFALSE
- 6.06.03.01The gradient shall be 17 T/m01/27/2025In ProcessFALSE
- 6.06.03.01The power supply shall be 1600 A01/27/2025In ProcessFALSE
- 6.06.03.01The power supply accuracy shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 70 cm01/27/2025In ProcessFALSE
- 6.06.03.01The aperture radius shall be 50 mm01/27/2025In ProcessFALSE
- 6.06.03.01The number of kickers shall be 601/27/2025In ProcessFALSE
- 6.06.03.01The Rise time shall be 900 ns01/27/2025In ProcessFALSE
- 6.06.03.01The Fall time shall be NA sec01/27/2025In ProcessFALSE
- 6.06.03.01The flat top time shall be 13 us01/27/2025In ProcessFALSE
- 6.06.03.01The waveshape shall be trap01/27/2025In ProcessFALSE
- 6.06.03.01The painting shall be vertical01/27/2025In ProcessFALSE
- 6.06.03.01The maximum field shall be 0.12 T01/27/2025In ProcessFALSE
- 6.06.03.01The total deflection shall be 16 mrad01/27/2025In ProcessFALSE
- 6.06.03.01The maximum current shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The maximum voltage shall be TBD01/27/2025In ProcessFALSE
- 6.06.03.01The inductance with cable shall be TBD (uH)01/27/2025In ProcessFALSE
- 6.06.03.01The Max rep rate shall be 100 kV/pC01/27/2025In ProcessFALSE
- 6.06.03.01The flat top repeatability shall be NA Hz01/27/2025In ProcessFALSE
- 6.06.03.01The beam abort kicker shall be tbd %01/27/2025In ProcessFALSE
- 6.06.03.01The flatness of flat top/pulse form shall be 1 %01/27/2025In ProcessFALSE
- 6.06.03.01The cooling type shall be w (W,A)01/27/2025In ProcessFALSE
- 6.06.03The ESR Abort system shall contain a beam to dump to safely absorb the energy of the dumped beam in a controlled fashion.01/27/2025ReviewedFALSE
- 6.06.03.01The diameter shall be 90 mm01/27/2025In ProcessFALSE
- 6.06.03.01The length shall be 2 m01/27/2025In ProcessFALSE
- 6.06.03.01The materials shall be C / Al / Cu01/27/2025In ProcessFALSE
- 6.06.03.01The energy deposited during abort shall be 320 kJ01/27/2025In ProcessFALSE
- 6.06.03.01The window thickness shall be tbd mm01/27/2025In ProcessFALSE
- 6.06.03.01The frequency of thermal cycle shall be 1 hour01/27/2025In ProcessFALSE
- 6.06.03.01The window material shall be tbd01/27/2025In ProcessFALSE
ESR-PROT-COLL : Collimation System
- ESR-PROT-COLL EXTERNALSRequirements who's parents are in other sub-systems.
- 6.06.03The ESR shall have a collimation system capable of ensuring a sufficiently low background at the detector.01/27/2025ReviewedFALSE
- 6.06.03.02The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR Injection absorber shall be vertical.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR injection absorber has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The thermal duty cycle shall be 2 Hz. 2 Hz01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR detector absorbers shall be placed at Sector 5.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The detector absorbers have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be tbd kW. tbd kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 21 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Cu01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 560-720 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator shall be horizontal.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be range shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary vertical collimator shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary horizontal collimator shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The ESR primary collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W on the tip of the jaw. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary vertical collimators shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary horizontal collimators shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).01/27/2025In ProcessFALSE
- 6.06.03.02The ESR collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
- 6.06.03The ESR shall have a collimation system capable protecting all machine elements in case of failure.01/27/2025ReviewedFALSE
- 6.06.03.02The ESR Injection absorber shall be placed in Sector 12 adjacent to the Momentum collimator.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR Injection absorber shall be vertical.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be 17.5 mm half gap. +/- 17.5 mm01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR injection absorber has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The thermal duty cycle shall be 2 Hz. 2 Hz01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR detector absorbers shall be placed at Sector 5.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be fixed.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The detector absorbers have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be tbd kW. tbd kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 21 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Cu01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 560-720 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimators shall be placed in Sector 12 adjacent to the Injection absorber.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator shall be horizontal.01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be tbd mm half gap. tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall be range shall be tbd.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR momentum collimator has dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 550 W. 550 W01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be steady-state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be negligible. ~0 W01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 0.89 m01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Al-Ti01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 1.1 m01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary vertical collimator shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR primary horizontal collimator shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 5 to 10 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 8 to 23 mm (half gap, +/- 10 µm).02/13/2025In ProcessFALSE
- 6.06.03.02The ESR primary collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W on the tip of the jaw. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch) on the tip of the jaw. 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary vertical collimators shall be placed at Sector 4.01/27/2025In ProcessFALSE
- 6.06.03.02The ESR secondary horizontal collimators shall be placed at Sector 2.01/27/2025In ProcessFALSE
- 6.06.03.02The vertical aperture shall range from 6 to 11 mm (half gap, +/- tbd). tbd mm01/27/2025In ProcessFALSE
- 6.06.03.02The horizontal aperture shall range from 9 to 25 mm (half gap, +/- tbd).01/27/2025In ProcessFALSE
- 6.06.03.02The ESR collimators have dual jaws.01/27/2025In ProcessFALSE
- 6.06.03.02The aperture shall be centered on the beam axshall be.01/27/2025In ProcessFALSE
- 6.06.03.02The beam energy deposition shall be 300 W. 300 W01/27/2025In ProcessFALSE
- 6.06.03.02The duty cycle shall be steady state.01/27/2025In ProcessFALSE
- 6.06.03.02The energy deposition from synchrotron radiation shall be 2.5 kW (horizontal only). 2.5 kW01/27/2025In ProcessFALSE
- 6.06.03.02The impedance shall be less than the Impedance budget (100 kV/pc). 100 kV/pc01/27/2025In ProcessFALSE
- 6.06.03.02Jaw angle in position relative to beam axshall be 1 mrad01/27/2025In ProcessFALSE
- 6.06.03.02The beam failure energy deposition shall be 275 J (1 bunch). 275 J01/27/2025In ProcessFALSE
- 6.06.03.02The failure thermal duty cycle shall be intermittent.01/27/2025In ProcessFALSE
- 6.06.03.02The tapered jaw slope shall be 1/10. 1/1001/27/2025In ProcessFALSE
- 6.06.03.02The jaw material shall be Mo-Gr01/27/2025In ProcessFALSE
- 6.06.03.02The jaw tip length shall be 180 mm01/27/2025In ProcessFALSE
- 6.06.03.02The total length shall be 762-995 mm01/27/2025In ProcessFALSE
ESR-PS : ESR Magnet Power Supply
- ESR-PS EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04The ESR magnets shall be fed by a system of power supplies matched in voltage and maximum current to the specifications and requirements of the respective magnets01/27/2025ReviewedFALSE
- 6.04The ESR magnet power supply’s shall be capable of providing the stability the ESR needs to operate02/13/2025ReviewedFALSE
ESR-PS-CORR_H : ESR Magnet Power Supply HCOR
- 6.04.04The magnet type being powered is KICKER01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is CORR_H01/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 101/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 22001/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 3 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.78 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 10 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 1000 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 100 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 1000 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-CORR_SPNQ_H : ESR Horizontal Corrector (SPNQ_H) Power Supply
ESR-PS-CORR_SPNQ_V : ESR Vertical Corrector (SPNQ_V) Power Supply
ESR-PS-CORR_V : ESR Vertical Corrector (CORR_V) Power Supply
- 6.04.04The magnet type being powered is KICKER01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is CORR_V01/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 101/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 22901/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 2.2 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 18 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 1000 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 100 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 1000 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-D13 : ESR Magnet Power Supply D13
- 6.04.04The magnet type being powered is DIPOLE TRIM01/28/2025ApprovedFALSE
- 6.04.04The magnet model being powered is D1301/28/2025ApprovedFALSE
- 6.04.04Number of unique magnet strings 1, 144, 2, 352, 4, 5, 801/28/2025ApprovedFALSE
- 6.04.04Quantity of magnets per unique string 1, 1, 2, 1, 1, 1, 401/28/2025ApprovedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ApprovedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.439 Ohm01/28/2025ApprovedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0 H01/28/2025ApprovedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 10.39 A01/28/2025ApprovedFALSE
- 6.04.04The current tuning margin shall be 1 %01/28/2025ApprovedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ApprovedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ApprovedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025ApprovedFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5-10 ppm02/13/2025ApprovedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Ripple 1kHz – 8kHz 200 ppm02/13/2025ApprovedFALSE
- 6.04.04Ripple 8kHz-40kHz See chart ppm01/28/2025ApprovedFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-D2 : ESR Magnet Power Supply D2
- 6.04.04The magnet type being powered is DIPOLE TRIM01/28/2025ApprovedFALSE
- 6.04.04The magnet model being powered is D201/28/2025ApprovedFALSE
- 6.04.04Number of unique magnet strings 1, 176, 401/28/2025ApprovedFALSE
- 6.04.04Quantity of magnets per unique string 2, 1, 401/28/2025ApprovedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ApprovedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.26 Ohm01/28/2025ApprovedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0 H01/28/2025ApprovedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 10.13 A01/28/2025ApprovedFALSE
- 6.04.04The current tuning margin shall be 1 %01/28/2025ApprovedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ApprovedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ApprovedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025ApprovedFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5-10 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Ripple 1kHz – 8kHz 200 ppm02/13/2025In ProcessFALSE
- 6.04.04Ripple 8kHz-40kHz See chart ppm01/28/2025In ProcessFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-D_SR : ESR Dipole (D_SR) Power Supply
- 6.04.04The magnet type being powered is DIPOLE TRIM01/28/2025ApprovedFALSE
- 6.04.04The magnet model being powered is D_SR01/28/2025ApprovedFALSE
- 6.04.04Number of unique magnet strings 1001/28/2025ApprovedFALSE
- 6.04.04Quantity of magnets per unique string 101/28/2025ApprovedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ApprovedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.879 Ohm01/28/2025ApprovedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.0046 H01/28/2025ApprovedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 7.45 A01/28/2025ApprovedFALSE
- 6.04.04The current tuning margin shall be 1 %01/28/2025ApprovedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ApprovedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ApprovedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5-10 ppm02/13/2025ApprovedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Ripple 1kHz – 8kHz 200 ppm02/13/2025ApprovedFALSE
- 6.04.04Ripple 8kHz-40kHz See chart ppm01/28/2025ApprovedFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-ESRMainDipole
- 6.04.04The magnet type being powered is DIPOLE01/28/2025ApprovedFALSE
- 6.04.04The magnet model being powered is ESRMainDipole01/28/2025ApprovedFALSE
- 6.04.04Number of unique magnet strings 12501/28/2025ApprovedFALSE
- 6.04.04Quantity of magnets per unique string 601/28/2025ApprovedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ApprovedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.019 Ohm01/28/2025ApprovedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.000148192771084337 H01/28/2025ApprovedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be 1600 V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 391 A01/28/2025ApprovedFALSE
- 6.04.04The current tuning margin shall be 1 %01/28/2025ApprovedFALSE
- 6.04.04The current ps reliability margin shall be TBD %01/28/2025ApprovedFALSE
- 6.04.04The PS current type shall be TBD (DC or Ramped )01/28/2025ApprovedFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04The PS waveshape required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The peak waveshape di/dt during ramping shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025ApprovedFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5-10 ppm02/13/2025ApprovedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Ripple 1kHz – 8kHz 200 ppm02/13/2025ApprovedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-FAST_CORR
- 6.04.04The magnet type being powered is KICKER01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is FAST_CORR01/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings TBD01/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string TBD01/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be TBD Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be TBD H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be TBD %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be TBD %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be TBD (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04Small SIgnal Bandwidth 3kHz01/28/2025In ProcessFALSE
- 6.04.04The PS waveshape required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The peak waveshape di/dt during ramping shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 1000 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 100 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 1000 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-Q50 : ESR Magnet Power Supply Q50
- 6.04.04The magnet type being powered is QUADRUPOLE01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is Q5001/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 1, 11, 12, 15, 18, 19, 201/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 22, 1, 1, 4, 3, 3, 1201/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.03 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.017 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 420 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) 100 A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 50 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-Q60 : ESR Magnet Power Supply Q60
- 6.04.04The magnet type being powered is QUADRUPOLE01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is Q6001/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 101/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 901/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.035 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.02 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 420 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) 100 A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 50 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-Q60x2 : ESR Quadrupole (Q60x2) Power Supply
- 6.04.04The magnet type being powered is QUADRUPOLE01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is Q60x201/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 101/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 2201/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.066 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.0405 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 420 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) 100 A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 50 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-Q80 : ESR Magnet Power Supply Q80
- 6.04.04nan02/13/2025ReviewedFALSE
- 6.04.04The magnet model being powered is Q8001/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 1, 4, 701/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 147, 2, 201/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.044 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.027 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 420 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) 100 A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 50 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-Q_LA : ESR Quadrupole (Q_LA) Power Supply
- 6.04.04The magnet type being powered is QUADRUPOLE01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is Q_LA01/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 501/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 201/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.066 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.0405 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 176.49 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) 100 A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 50 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-Q_NARROW : ESR Quadrupole (Q_NARROW) Power Supply
ESR-PS-Q_SKEW : ESR Quadrupole (Q_SKEW) Power Supply
- 6.04.04The magnet type being powered is QUADRUPOLE01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is Q_SKEW01/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 101/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 1901/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.015 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.0085 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 420 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) 100 A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 5 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 50 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-PS-QLS1 : ESR Magnet Power Supply QLS1
ESR-PS-QLS2 : ESR Magnet Power Supply QLS2
- ESR-PS-QLS2 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
ESR-PS-QLS3 : ESR Magnet Power Supply QLS3
- ESR-PS-QLS3 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
ESR-PS-QSS1 : ESR Magnet Power Supply QSS1
- ESR-PS-QSS1 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
ESR-PS-QSS2 : ESR Magnet Power Supply QSS2
- ESR-PS-QSS2 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
ESR-PS-QSS3 : ESR Magnet Power Supply QSS3
- ESR-PS-QSS3 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
ESR-PS-QSS4 : ESR Magnet Power Supply QSS4
- ESR-PS-QSS4 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
ESR-PS-QSS5 : ESR Magnet Power Supply QSS5
- ESR-PS-QSS5 EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The nominal current of the magnets being powered shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/28/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/28/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/28/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/28/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/28/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/28/2025In ProcessFALSE
ESR-PS-SXT : ESR Magnet Power Supply SXT
- ESR-PS-SXT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/27/2025ReviewedFALSE
- 6.04.04The maximum magnet string resistance to be powered shall be TBD ohm01/27/2025In ProcessFALSE
- 6.04.04The maximum magnet string inductance to be powered shall be TBD H01/27/2025In ProcessFALSE
- 6.04.04The magnets being powered shall be saturated TBD Y/N01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The voltage to ground of the magnet being powered shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The minimum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be TBD A01/27/2025In ProcessFALSE
- 6.04.04The PS current type shall be NC (DC or AC)01/27/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The full power bandwidth required shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The ppm of full scale current (peak to peak) shall be TBD %01/27/2025In ProcessFALSE
- 6.04.04The time period for specified stability shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The long term stability shall be TBD A/s01/27/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/27/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/27/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (peak to peak) TBD A01/27/2025In ProcessFALSE
- 6.04.04The max current ripple frequency range (Hz) TBD Hz01/27/2025In ProcessFALSE
- 6.04.04WRT the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04The max voltage ripple (peak to peak) shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04An NMR shall be required to measure the field TBD A/s01/27/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.04The current required to be shunted through the magnet shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The magnet turns ratio shall be TBD01/27/2025In ProcessFALSE
- 6.04.04The terminal voltage shall be TBD V01/27/2025In ProcessFALSE
- 6.04.04The design shall have thermal switches TBD01/27/2025In ProcessFALSE
ESR-PS-SXT_LONG : ESR Sextupole (Long_SXT) Power Supply
- 6.04.04The magnet type being powered is SEXTUPOLE01/28/2025ReviewedFALSE
- 6.04.04The magnet model being powered is SXT_LONG01/28/2025ReviewedFALSE
- 6.04.04Number of unique magnet strings 1, 2, 5, 801/28/2025ReviewedFALSE
- 6.04.04Quantity of magnets per unique string 1, 1, 3, 601/28/2025ReviewedFALSE
- 6.04.04The number of Independent functions on the magnets being powered shall be 101/28/2025ReviewedFALSE
- 6.04.04The minimum magnet resistance to be powered shall be TBD Ohm01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet resistance to be powered shall be 0.22 Ohm01/28/2025ReviewedFALSE
- 6.04.04The minimum magnet inductance to be powered shall be TBD H01/28/2025In ProcessFALSE
- 6.04.04The maximum magnet inductance to be powered shall be 0.056 H01/28/2025ReviewedFALSE
- 6.04.04The max voltage to ground of the magnet being powered shall be TBD V01/28/2025In ProcessFALSE
- 6.04.04The minimum opperating current the PS must operate at shall be TBD A01/28/2025In ProcessFALSE
- 6.04.04The maximum current the PS must operate at shall be 161 A01/28/2025ReviewedFALSE
- 6.04.04The current tuning margin shall be 5 %01/28/2025ReviewedFALSE
- 6.04.04The current ps reliability margin shall be 10 %01/28/2025ReviewedFALSE
- 6.04.04The PS current type shall be DC (DC or Ramped )01/28/2025ReviewedFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04Small SIgnal Bandwidth TBD01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04The stability ppm of full scale current (rms) over (many hours) shall be 100 ppm01/28/2025In ProcessFALSE
- 6.04.04The short term stability shall be TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The long term stability shall be (1 sec to 12 hrs?) TBD A/s01/28/2025In ProcessFALSE
- 6.04.04The current setpoint resolution (min size in bits) shall be TBD bits01/28/2025In ProcessFALSE
- 6.04.04The synchronization required between PS's shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The synchronization timing of synchronization shall be TBD s01/28/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (0-1kHz) 20 ppm02/13/2025In ProcessFALSE
- 6.04.04The max allowable current ripple (rms) of full scale current (> 1kHz) 1000 ppm02/13/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04WRT to the ripple frequency the following resonant frequencies shall be avoided TBD Hz01/27/2025In ProcessFALSE
- 6.04.04An NMR in a reference magnet shall be required to measure the field No Y/N01/28/2025In ProcessFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
- 6.04.04< Requirement Not Applicable >01/28/2025Not ApplicableFALSE
ESR-RF : ESR RF System
- 6.08The Storage RF System shall be protected from degradation due to adjacent beam lines and hardware.01/27/2025ApprovedFALSE
- 6.08The Storage RF System design shall be protected against beam loss and RF field collapse events.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall maintain ESR operations in the event of single RF system failure.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall be designed as identical modular components to provide the full system functionality.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall interface to all required accelerator systems.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall provide all necessary controls and diagnostics necessary for system operation but not extending beyond the RF System interfaces.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall achieve an independent beam line vacuum level equivalent to the ESR operational vacuum level.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall meet ambient magnetic field hygiene requirements.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall have alignment verification designed into the system.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall provide conditioning capabilities necessary for ESR operations.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall be rotatable around the beamline axis (roll) to facilitate installation without degredation of performance.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall have an active frequency tuning system to maintain and tune the resonant frequency in order to accommodate radial beam offset and changes in the synchronous trajectory circumference.01/27/2025ApprovedFALSE
- 6.08The Storage RF System active frequency tuning system shall be maintainable in-situ.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall be designed to utilize common transportation methods without degredation of performance.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall maintain constant electron beam energy after injection.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall be capable of coupling 10 MW of RF power to the beam.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall maintain beam energy from 5 to 18 GeV.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall be able to provide 68 MV peak voltage for an RF bucket height 10x greater than the rms energy spread of the beam.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall be able to accommodate 0.1 W/m electron beam losses.01/27/2025ApprovedFALSE
- 6.08The Storage RF System shall provide (TBD 0.1%) field amplitude stability01/27/2025In ProcessFALSE
- 6.08The Storage RF System shall provide (TBD 0.5) degree phase stability01/27/2025In ProcessFALSE
- ESR-RF EXTERNALSRequirements who's parents are in other sub-systems.
- 6.08The Storage RF System shall be designed to fulfill all necessary paramaters as set by the Master Parameter Table Doc. No. EIC-SEG-RSI-005.01/27/2025ReviewedFALSE
- 6.08.04.01Conditioning for individual cavities shall be achieved with a maximum temperature of 2.1 K02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable vertical acceleration of ±4 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable beamline axis acceleration of ±5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable lateral acceleration of ±1.5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to withstand a tilt around the beamline axis (roll) up to ±1.4 radians02/18/2025ApprovedFALSE
- 6.08The Storage ring RF System shall be installed in the straight sections of the ESR lattice within the existing RHIC tunnel in IR10.01/27/2025ReviewedFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.08.04.01Conditioning for individual cavities shall be achieved with a maximum temperature of 2.1 K02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable lateral acceleration of ±1.5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable vertical acceleration of ±4 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable beamline axis acceleration of ±5 G02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall conform to the ESR lattice.01/27/2025ReviewedFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall be designed to accelerate electrons.01/27/2025ReviewedFALSE
- 6.08The Storage RF System shall utilize superconductivity.01/27/2025ReviewedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall conform to the EIC Code of Record.01/27/2025ReviewedFALSE
- 6.08The Storage RF System shall have a minimum operating lifetime of 20 years01/27/2025ReviewedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/cm^202/18/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The chilled water and low-conductivity water operational pressure range shall be 7.5 to 8.5 bar02/18/2025ApprovedFALSE
- 6.08The Storage RF System shall have an operational availability of (TBD 95%) or better.01/27/2025ReviewedFALSE
- 6.07.02The ESR controls system shall be capable of producing arbitrary spin pattern at injection01/27/2025In ProcessFALSE
- 6.07.02The Spin pattern control granularity shall be 1 bunch01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 18GeV shall be 290 cnt01/27/2025In ProcessFALSE
- 6.07.02The number of bunches @ 10GeV and below shall be 1160 cnt01/27/2025In ProcessFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/cm^202/18/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The chilled water and low-conductivity water operational pressure range shall be 7.5 to 8.5 bar02/18/2025ApprovedFALSE
- 6.08The Storage RF Systems within the tunnel shall operate within its yearly radiation exposure budget.01/27/2025ReviewedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
ESR-RF-NCRF : ESR Main RF System
ESR-RF-NCRF-TBD : ESR Main RF Fundamental Mode
ESR-RF-SRF : Superconducting RF
ESR-RF-SRF-394_Crab
- 6.08.04.05The SRF CM shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing02/20/2025In ProcessFALSE
- 6.08.04.05The cavity helium bath maximum operational temperature shall be TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The cavity helium bath maximum operational pressure shall be TBD mbar02/20/2025In ProcessFALSE
- 6.08.04.05The cavity helium bath operation pressure stability shall be ±TBD mbar02/20/2025In ProcessFALSE
- 6.08.04.05The maximum helium supply operational temperature shall be TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The range of the helium supply operational pressure shall be TBD to TBD bar02/20/2025In ProcessFALSE
- 6.08.04.05The range of the combined helium return temperature shall be TBD to TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The range of the combined helium return pressure shall be TBD to TBD bar02/20/2025In ProcessFALSE
- 6.08.04.05The maximum sub-atmospheric helium return temperature shall be TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The maximum Subatmospheric helium return pressure shall be TBD mbar02/20/2025In ProcessFALSE
- 6.08.04.05The minimum cooldown rate of the SRF cavity between 300K and 4.5K shall be TBD K/hour02/20/2025In ProcessFALSE
- 6.08.04.05The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be TBD K/hour02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall achieve steady state temperature with the cavity bath at 4K in a maximum of TBD days02/20/2025In ProcessFALSE
- 6.08.04.05The chilled water and low-conductivity water operational temperature range shall be TBD to TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The chilled water and low-conductivity water operational pressure range shall be TBD to TBD bar02/20/2025In ProcessFALSE
- 6.08.04.05The minimum magnetic shield attenuation factor at SRF cavity equator shall be TBD02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall operate through a minimum of TBD thermal cycles02/20/2025In ProcessFALSE
- 6.08.04.05The minimum SRF Cavity Slow tuner minimum lifetime shall be TBD years02/20/2025In ProcessFALSE
- 6.08.04.05The minimum SRF CM Slow Tuner 1% range tuning cycles shall be TBD cycles02/20/2025In ProcessFALSE
- 6.08.04.05The minimum SRF CM Slow Tuner full range tuning cycles shall be TBD cycles02/20/2025In ProcessFALSE
- 6.08.04.05The manufactured SRF CM Cavity shall produce no field emission at TBD MV02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM components that are not replaceable in-situ shall be designed with a radiation tolerance greater than TBD MGy02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM components that are replaceable in-situ shall have a radiation tolerance greater than TBD kGy02/20/2025In ProcessFALSE
- 6.08.04.05The active SRF cavity tuning mechanism components (bearings/motor/piezo) shall be replaceable and maintainable in-situ.02/20/2025In ProcessFALSE
- 6.08.04.05All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM instrument should have maximized instruments that can be maintained and replaced in-situ02/20/2025In ProcessFALSE
- 6.08.04.05The SRF maximum (per cavity) RF longitudinal impedance shall be TBD MΩ GHz02/20/2025In ProcessFALSE
- 6.08.04.05The SRF maximum (per cavity) RF horizontal impedance shall be TBD MΩ/m02/20/2025In ProcessFALSE
- 6.08.04.05The SRF maximum (per cavity) RF vertical impedance shall be TBD MΩ/m02/20/2025In ProcessFALSE
- 6.08.04.05The minimum cavity aperture radius shall be TBD mm02/20/2025In ProcessFALSE
- 6.08.04.05The maximum broadband RF power emitted from the CM shall be TBD kW02/20/2025In ProcessFALSE
- 6.08.04.05The Maximum Quadrupole multipole content shall be TBD mT02/20/2025In ProcessFALSE
- 6.08.04.05The Maximum Sextupole multipole content shall be TBD mT/m02/20/2025In ProcessFALSE
- 6.08.04.05The Maximum Octupole multipole content shall be TBD T/m^202/20/2025In ProcessFALSE
- 6.08.04.05The Maximum Decapole multipole content shall be TBD T/m^302/20/2025In ProcessFALSE
- 6.08.04.05The SRF cavity minimum manufactured quality factor (Qo) shall be TBD02/20/2025In ProcessFALSE
- 6.08.04.05The SRF cavity minimum manufactured voltage shall be TBD MV02/20/2025In ProcessFALSE
- 6.08.04.05The SRF cavity fundamental power coupler Qext shall be TBD02/20/2025In ProcessFALSE
- 6.08.04.05The SRF cavity field probe Qext range shall be TBD02/20/2025In ProcessFALSE
- 6.08.04.05The SRF cavity nominal cold frequency shall be TBD MHz02/20/2025In ProcessFALSE
- 6.08.04.05The SRF cavity maximum Niobium temperature shall be TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The SRF Pressure sensitivity maximum shall be TBD Hz/mBar02/20/2025In ProcessFALSE
- 6.08.04.05The SRF maximum Lorentz force detuning shall be TBD Hz/(Mv/m)^202/20/2025In ProcessFALSE
- 6.08.04.05The warm beamline maximum vacuum shall be TBD mbar02/20/2025In ProcessFALSE
- 6.08.04.05The cold beamline maximum vacuum shall be TBD mbar02/20/2025In ProcessFALSE
- 6.08.04.05The beamline vacuum maximum leak rate shall be TBD mbar L/s02/20/2025In ProcessFALSE
- 6.08.04.05The warm insulating maximum vacuum shall be TBD mbar02/20/2025In ProcessFALSE
- 6.08.04.05The cold insulating maximum vacuum shall be TBD mbar02/20/2025In ProcessFALSE
- 6.08.04.05The insulating vacuum maximum leak rate shall be TBD mbar L/s02/20/2025In ProcessFALSE
- 6.08.04.05The minimum SRF Cavity Slow Tuner tuning range shall be shall be -TBD, +TBD kHz02/20/2025In ProcessFALSE
- 6.08.04.05The minimum SRF Cavity slow tuner tuning rate shall be TBD Hz/s02/20/2025In ProcessFALSE
- 6.08.04.05The maximum SRF Cavity Slow Tuner resolution shall be TBD Hz02/20/2025In ProcessFALSE
- 6.08.04.05The maximum SRF Cavity Slow Tuner hysteresis shall be ±TBD Hz02/20/2025In ProcessFALSE
- 6.08.04.05The external warm maximum allowable working pressure of the SRF cavity shall not exceed TBD bar02/20/2025In ProcessFALSE
- 6.08.04.05The external cold maximum allowable working pressure of the SRF cavity shall not exceed TBD bar02/20/2025In ProcessFALSE
- 6.08.04.05The internal maximum allowable working pressure of the SRF cavity shall not exceed TBD bar02/20/2025In ProcessFALSE
- 6.08.04.05All cryomodule surfaces accessible to workers shall be within the temperature range of TBD to TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by ASME B31.302/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by ASME BPVC02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by ASTM C105502/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by NFPA 7002/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by API 520 & API 52102/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the EIC code of records to meet all applicable safety standards as defined by NFPA 70E02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by CGA S1.302/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by AWS02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM maximum length shall be TBD m02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM maximum width shall be TBD m02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM maximum height shall be TBD m02/20/2025In ProcessFALSE
- 6.08.04.05The distance from the beamline to the tunnel floor shall be TBD m02/20/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance in X shall be ±TBD μm02/20/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance in Y shall be ±TBD μm02/20/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance in Z shall be ±TBD mm02/20/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ±TBD degrees02/20/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ±TBD degrees02/20/2025In ProcessFALSE
- 6.08.04.05The Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ±TBD degrees02/20/2025In ProcessFALSE
- 6.08.04.05Conditioning for individual cavities shall have a maximum average cryogenic power dissipation of TBD W02/20/2025In ProcessFALSE
- 6.08.04.05Conditioning for individual cavities shall be achieved with a maximum temperature of TBD K02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be capable of withstanding a maximum allowable vertical acceleration of TBD G02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be capable of withstanding a maximum allowable lateral acceleration of TBD G02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be capable of withstanding a maximum allowable beamline axis acceleration of TBD G02/20/2025In ProcessFALSE
- 6.08.04.05The SRF CM shall be designed to withstand a minimum tilt around the beamline axis (roll) of ±TBD radians02/20/2025In ProcessFALSE
ESR-RF-SRF-400KW
- 6.08.06.01The RF Amplifier system shall utilize watercooling to dissipate a minimum of 80 % heat.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier system shall have a maximum die temperature of 130 C.01/27/2025ApprovedFALSE
- 6.08.06.01The minimum acceptable system loss of RF Amplifiers due to faults shall be 10 %.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier shutoff switch time shall be 100 ns.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier shall be designed to self-protect to a maximum overdrive power of 10 %.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier phase linearity over a 40dB dynamic range shall be 30 deg.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplified input/output impedance shall be 50 ohms.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier input VSWR shall be 1.5:1 .01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier input RF overdrive shall be 10 dBm.01/27/2025ApprovedFALSE
- 6.08.06.01All RF Amplifier harmonics shall not exceed -30 dBc.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier spurious and line harmonics shall not exceed -80 dBc.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier noise power output in the ON state shall be -90 dBm/Hz.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier noise power output in the STANDBY state shall be -154 dBm/Hz.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier group delay shall be 200 ns.01/27/2025ApprovedFALSE
- 6.08.06.01The minimum RF Amplifier AC to RF efficiency at full rated power shall be 45 %.01/27/2025ApprovedFALSE
- 6.08.06.01The operational frequency band of the RF Amplifier system shall be 591 ± 5 MHz.01/28/2025ApprovedFALSE
- 6.08.06.01The minimum RF on/off ratio (Δgain between on/standby) shall be 86 dB.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier system shall be designed to operate in an ambient temperature range of 18-30 C.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier system shall be designed to operate in a maximum relative humidity of 60-65 %.01/27/2025ApprovedFALSE
- 6.08.06.01With the load mismatch of less than 3:1, the RF Amplifier shall maintain 100% forward power. .01/27/2025ApprovedFALSE
- 6.08.06.01With the load mismatch of greater than 3:1, the RF Amplifier shall utilize 0% forward power. .01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier forward RF sample coupling across the amplifier bandwidth shall be 70±0.5 dB.01/28/2025ApprovedFALSE
- 6.08.06.01The minimum RF Amplifier forward directivity across the amplifier bandwidth shall be 30 dB.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier reflected RF sample coupling across the amplifier bandwidth shall be 70±0.5 dB.01/28/2025ApprovedFALSE
- 6.08.06.01The minimum RF Amplifier reflected directivity across the amplifier bandwidth shall be 30 dB.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier audible noise at 1m shall be 70 dB(A).01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier pulse width range shall be .001-100 ms.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier duty cycle range shall be 1-50 %.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifer rise time during pulsed operation shall be 0.75 μs.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifer fall time during pulsed operation shall be 0.75 μs.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier pulse droop at max pulse width shall be 0.5-1 dB.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum RF Amplifier phase error over pulse at max pulse width shall be 2.5-5 deg.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier AC Voltage input shall remain between 432-528 VAC.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier controls AC Voltage input shall remain between 108-132 VAC.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier AC frequency shall be 60 Hz .01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier power factor shall be between 0.90-0.95 .01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier input power for the rated output power shall be ±1.5 dBm.01/28/2025ApprovedFALSE
- 6.08.06.01The minimum output power of the RF Amplifier for all operational modes shall be 200/400 kW.01/27/2025ApprovedFALSE
- 6.08.06.01The minimum linear power of the RF Amplifier for all operational modes shall be 200/400 kW.01/27/2025ApprovedFALSE
- 6.08.06.01The nominal power gain for the RF Amplifier system over a 40 dB dynamic range shall be 86 dB.01/27/2025ApprovedFALSE
- 6.08.06.01The maximum power gain deviation of the RF Amplifier system shall be 1.5 dB.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier maximum height shall be 3 m.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier maximum width shall be 3 m.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier maximum length shall be 3 m.01/27/2025ApprovedFALSE
- 6.08.06.01The RF Amplifier shall be designed to have a minimum MTBF of 10,000 hours.01/27/2025ApprovedFALSE
- 6.08.06.01The RF devices shall be designed to have a minimum MTBF of 35,000 hours.01/27/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by ASTM C1055 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .01/27/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by 29 CFR 1910 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .01/27/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by NFPA 70E-2021 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .01/27/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by UL 508 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .01/27/2025ApprovedFALSE
- 6.08.06.01The RF amplifier shall be designed to meet all applicable standards as defined by Occupatioanl Radiation Protection 10 CFR 835 as directed by the EIC Code of Records Doc No. EIC-ORG-RSI-026 and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA) .01/27/2025ApprovedFALSE
ESR-RF-SRF-591_1Cell : 591MHz 1 Cell Cryomodule
- 6.08.04.01The SRF CM shall be outfitted with flow control, thermometry, pressure, and RF instrumentation as to monitor and control all sub-systems during the cooldown, warm-up, operation and testing02/18/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum operational temperature shall be 2 K02/18/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath maximum operational pressure shall be 30 mbar02/18/2025ApprovedFALSE
- 6.08.04.01The cavity helium bath operation pressure stability shall be ±0.1 mbar02/18/2025ApprovedFALSE
- 6.08.04.01The maximum helium supply operational temperature shall be 5.5 K02/18/2025ApprovedFALSE
- 6.08.04.01The range of the helium supply operational pressure shall be 3 to 3.5 bar02/18/2025ApprovedFALSE
- 6.08.04.01The range of the combined helium return temperature shall be 64 to 66 K02/18/2025ApprovedFALSE
- 6.08.04.01The range of the combined helium return pressure shall be 2.4 to 2.6 bar02/18/2025ApprovedFALSE
- 6.08.04.01The maximum sub-atmospheric helium return temperature shall be 4.5 K02/18/2025ApprovedFALSE
- 6.08.04.01The maximum Subatmospheric helium return pressure shall be 30 mbar02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 300K to 150K shall be 10 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 50K surfaces shall be 2 W/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall achieve steady state temperature with the cavity bath at 4K in a maximum of 2 days02/18/2025ApprovedFALSE
- 6.08.04.01The chilled water and low-conductivity water operational temperature range shall be 295 to 315 K02/18/2025ApprovedFALSE
- 6.08.04.01The minimum magnetic shield attenuation factor at SRF cavity equator shall be 5002/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall operate through a minimum of 200 thermal cycles02/18/2025ApprovedFALSE
- 6.08.04.01The minimum SRF CM Slow Tuner 1% range tuning cycles shall be 100000 cycles02/18/2025ApprovedFALSE
- 6.08.04.01The minimum SRF CM Slow Tuner full range tuning cycles shall be 1000 cycles02/18/2025ApprovedFALSE
- 6.08.04.01The manufactured SRF CM Cavity shall produce no field emission at 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM components that cannot be maintained in-situ shall be designed with a minimum lifetime radiation tolerance of 1000 kGy02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM components that can be maintained in-situ shall have an annual minimum radiation tolerance of 1 kGy02/18/2025ApprovedFALSE
- 6.08.04.01The active SRF cavity tuning mechanism components (bearings/motor/piezo) shall be replaceable and maintainable in-situ.02/18/2025ApprovedFALSE
- 6.08.04.01The SRF Cavity Fast Tuner shall have a minimum lifetime of 1 year between maintenance cycles .02/18/2025ApprovedFALSE
- 6.08.04.01All critical monitoring and control instruments that cannot be maintained in-situ shall utilize a back-up instrument.02/18/2025ApprovedFALSE
- 6.08.04.01The SRF pressure sensitivity maximum shall be 10 Hz/mbar02/18/2025ApprovedFALSE
- 6.08.04.01The SRF maximum lorentz force detuning shall be 5 Hz/(MV/m)^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity maximum Niobium temperature shall be 5 K02/18/2025ApprovedFALSE
- 6.08.04.01The warm beamline maximum vacuum shall be 5.0e-7 mbar02/18/2025ApprovedFALSE
- 6.08.04.01The cold beamline maximum vacuum shall be 1.0e-9 mbar02/18/2025ApprovedFALSE
- 6.08.04.01The beamline vacuum maximum leak rate shall be 5e-10 mbar L/s02/18/2025ApprovedFALSE
- 6.08.04.01The warm insulating maximum vacuum shall be 1.0e-5 mbar02/18/2025ApprovedFALSE
- 6.08.04.01The cold insulating maximum vacuum shall be 5.0e-7 mbar02/18/2025ApprovedFALSE
- 6.08.04.01The insulating vacuum maximum leak rate shall be 1.0e-8 mbar L/s02/18/2025ApprovedFALSE
- 6.08.04.01The The minimum SRF Cavity Slow Tuner tuning range shall be shall be 600 kHz02/18/2025ApprovedFALSE
- 6.08.04.01The minimum SRF Cavity slow tuner tuning rate shall be 800 Hz/s02/18/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner resolution shall be ±1 Hz02/18/2025ApprovedFALSE
- 6.08.04.01The maximum SRF Cavity Slow Tuner hysteresis shall be ±10 Hz02/18/2025ApprovedFALSE
- 6.08.04.01The external warm maximum allowable working pressure of the SRF cavity shall not exceed 2.2 bar02/18/2025ApprovedFALSE
- 6.08.04.01The external cold maximum allowable working pressure of the SRF cavity shall not exceed 5.2 bar02/18/2025ApprovedFALSE
- 6.08.04.01The internal maximum allowable working pressure of the SRF cavity shall not exceed 1.8 bar02/18/2025ApprovedFALSE
- 6.08.04.01All cryomodule surfaces accessible to workers shall be within the temperature range of 283 to 333 K02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME B31.302/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASME BPVC02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by ASTM C105502/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 7002/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to meet all applicable standards,as directed by the EIC Code of Record and/or all applicable excluded items governed by the EIC Memorandum of Agreements (MOA), and defined by NFPA 70E02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by API 520 & API 52102/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by CGA S1.302/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed and manufactured as directed by the JLAB ES&H Manual to meet all applicable safety standards as defined by AWS02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to meet all applicable standards as directed by the DOE Vacuum Vessel Consensus Standards02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM maximum length shall be 7.2 m02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM maximum width shall be 2.15 m02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM maximum height shall be 1.7 m02/18/2025ApprovedFALSE
- 6.08.04.01The distance from the beamline to the tunnel floor shall be 1381.09 mm02/18/2025ApprovedFALSE
- 6.08.04.01The Combined Cavity Electromagnetic Center Alignment Tolerance in X shall be ±250 μm02/18/2025ApprovedFALSE
- 6.08.04.01The Combined Cavity Electromagnetic Center Alignment Tolerance in Y shall be ±250 μm02/18/2025ApprovedFALSE
- 6.08.04.01The Combined Cavity Electromagnetic Center Alignment Tolerance in Z shall be ±5 mm02/18/2025ApprovedFALSE
- 6.08.04.01The Combined Cavity Electromagnetic Center Alignment Tolerance for the roll shall be ±0.04 radians02/18/2025ApprovedFALSE
- 6.08.04.01The Combined Cavity Electromagnetic Center Alignment Tolerance for the pitch shall be ±0.01 radians02/18/2025ApprovedFALSE
- 6.08.04.01The Combined Cavity Electromagnetic Center Alignment Tolerance for the yaw shall be ±0.01 radians02/18/2025ApprovedFALSE
- 6.08.04.01Conditioning for individual cavities shall have a maximum average cryogenic power dissipation of 200 W02/18/2025ApprovedFALSE
- ESR-RF-SRF-591_1Cell EXTERNALSRequirements who's parents are in other sub-systems.
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 150K to 50K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 50K to 4.5K shall be 10 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cooldown rate of the SRF cavity between 4.5K to 2K shall be 0.5 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The minimum warmup rate of the SRF cavity between 50K to 150K shall be 30 K/hour02/18/2025ApprovedFALSE
- 6.08.04.01The maximum thermal radiative heat transfer to all 2K and 5K surfaces shall be 2 W/cm^202/18/2025ApprovedFALSE
- 6.08.04.01The chilled water and low-conductivity water operational pressure range shall be 7.5 to 8.5 bar02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF longitudinal impedance (accelerator definition) shall be 52 MΩ Ghz02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF horizontal impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The total SRF maximum RF vertical impedance (accelerator definition) shall be 24 MΩ/m02/18/2025ApprovedFALSE
- 6.08.04.01The minimum cavity aperture radius shall be 30 mm02/18/2025ApprovedFALSE
- 6.08.04.01The maximum broadband RF power emitted from the CM shall be 30 kW02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to operate with a beam current up to 2.5 A02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Dipole multipole content shall be 50 mT-m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Quadrupole multipole content shall be 4 mT02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Sextuple multipole content shall be 2 T/m02/18/2025ApprovedFALSE
- 6.08.04.01The Maximum Octupole multipole content shall be 50 mT/m^202/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured quality factor (Qo) shall be 1500000000002/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity minimum manufactured gradient shall be 4 MV02/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity field probe Qext range shall be 1.00E11 to 2.00E1102/18/2025ApprovedFALSE
- 6.08.04.01The SRF cavity nominal cold frequency shall be 591.149 MHz02/18/2025ApprovedFALSE
- 6.08.04.01Conditioning for individual cavities shall be achieved with a maximum temperature of 2.1 K02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable vertical acceleration of ±4 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable lateral acceleration of ±1.5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be capable of withstanding a maximum allowable beamline axis acceleration of ±5 G02/18/2025ApprovedFALSE
- 6.08.04.01The SRF CM shall be designed to withstand a tilt around the beamline axis (roll) up to ±1.4 radians02/18/2025ApprovedFALSE
ESR-STRAIGHT : Straight Sections
- ESR-STRAIGHT EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04The phase advance of each straight section shall be tunable in order to optimize the dynamic aperture of the ESR.01/27/2025In ProcessFALSE
- 6.04The straight sections IR2,4,10 and 12 shall be based on FODO cells.01/27/2025In ProcessFALSE
- 6.04There shall be matching sections at the ends of each of the straight sections to compensate for the different FODO cell lengths wrt the arc FODO cells imposed by geometric constraints.01/27/2025In ProcessFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/-3.5 (%) of the Peak main bus field. (See figure P-ESR-MSG-D13.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D13.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm)01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <2.73 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.79(T.m) (See figure P-ESR-MSG-D13.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (see figure P-ESR-MSG-D13-19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different design energies shall be Measurement 1; Bref1=0.07 (T) at R1 Measurement 2; Bref2=0.13 (T) at R2 Measurement 3; Bref3=0.284 (T) at R2 Measurement 4; Bref4=0.284(T) at R=002/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: b1 = 10000, Region 2: b1 = 10000, *Region 3: b1 = 10000,02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -4<b2<4, Region2: -4<b2<4, *Region3: -4<b2<402/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b3<0.6, Region 2: -0.5<b3<0.6, *Region 3: -0.5<b3<0.6,02/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -1<b4< 0.5, Region 2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01< Requirement Not Applicable >01/27/2025Not ApplicableFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be equipped with trim coils which are capable of trimming the field within +/- 2.8(%) of the Peak main bus field. (See figure P-ESR-MSG-D2.2-1)01/27/2025ApprovedFALSE
- 6.04.02.01To operate at a fixed bus current for 5GeV,10GeV and 18GeV the magnet shall be designed to accommodate turns. (See figure P-ESR-MSG-D2.3-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall have a Dipole field.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet field shall have a vertical field direction.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet pole gap height and width shall be H=52 (mm), W=140(mm).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet good field aperture dAx required shall be 35.0345 mm.01/27/2025ApprovedFALSE
- 6.04.02.01The physical magnet length shall be <1.13 (m).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be able to deliver an absolute nominal integrated dipole field ranging from Bmin=0(T.m) to Bmax=0.33(T.m). (See figure P-ESR-MSG-D2.15-1)01/27/2025ApprovedFALSE
- 6.04.02.01The magnet-to-magnet variability shall be < 0.1%.01/27/2025ApprovedFALSE
- 6.04.02.01The field shall be measured at 4 locations (See figure P-ESR-MSG-D2.19-1) as follows Harmonic Measurements region 1; Rref1=13mm centered at (-14,0) mm, Harmonic Measurements region 2; Rref2=13mm centered at (0,0) mm, Harmonic Measurements region 3; Rref3=13mm centered at (14,0), Relative field Measurements region 4; Relative to the central field at B(0,0), sampled in an Annulus 25mm>dRvol3>31mm02/13/2025ApprovedFALSE
- 6.04.02.01The reference field for the different measurements shall be Measurement 1; Bref1=-0.375 (T) in Region1, Region2 and Region3 Measurement 2; Bref2=0.12 (T) in Region1, Region2 and Region3 Measurement 3; Bref2=0.23 (T) in Region1, Region2 and Region3 Measurement 3; Bref3=0.23 (T) in Region402/13/2025ApprovedFALSE
- 6.04.02.01The magnet bore field shall have a field homogeneity in region 4, of better than dB/B<10-3 with respect to the central field at R(0,0) and shall meet the following harmonic multipole content in regions 1, 2 and 3. (Note: The following calculated multipoles values are given for reference radius of 17mm and centered on axis at x=0, y=0. *The multipole values need to be scaled accordingly in regions 1 and 3 with appropriate off axis values.)02/13/2025ApprovedFALSE
- 6.04.02.01*Region1: b1 = 10000, Region2: b1 = 10000, *Region3: b1 = 1000002/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -6<b2<6, Region 2: -6<b2<6, *Region 3: -6<b2<602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -0.5<b3<0.6, Region2: -0.5<b3<0.6, *Region3: -0.5<b3<0.602/13/2025ApprovedFALSE
- 6.04.02.01*Region1: -1<b4< 0.5, Region2: -1<b4< 0.5, *Region 3: -1<b4< 0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b5 <0.5, Region 2: -0.5<b5 <0.5, *Region 3: -0.5<b5 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b6 <0.5, Region 2: -0.5<b6 <0.5, *Region 3: -0.5<b6 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b7 <0.5, Region 2: -0.5<b7 <0.5, *Region 3: -0.5<b7 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b8 <0.5, Region 2: -0.5<b8 <0.5, *Region 3: -0.5<b8 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b9 <0.5, Region 2: -0.5<b9 <0.5, *Region 3: -0.5<b9 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b10 <0.5, Region 2: -0.5<b10 <0.5, *Region 3: -0.5<b10 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b11 <0.5, Region 2: -0.5<b11 <0.5, *Region 3: -0.5<b11 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b12 <0.5, Region 2: -0.5<b12 <0.5, *Region 3: -0.5<b12 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b13 <0.5, Region 2: -0.5<b13 <0.5, *Region 3: -0.5<b13 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b14 <0.5, Region 2: -0.5<b14 <0.5, *Region 3: -0.5<b14 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b15 <0.5, Region 2: -0.5<b15 <0.5, *Region 3: -0.5<b15 <0.502/13/2025ApprovedFALSE
- 6.04.02.01*Region 1: -0.5<b16 <0.5, Region 2: -0.5<b16 <0.5, *Region 3: -0.5<b16 <0.502/13/2025ApprovedFALSE
- 6.04.02.01The magnet shall not be designed to limit CrossTalk requirements.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to constrain the external fringe field01/27/2025ApprovedFALSE
- 6.04.02.01The magnet fringe field shall not exceed 10 Gauss at a radial distance greater than 900mm from the magnet centerline01/27/2025ApprovedFALSE
- 6.04.02.01Magnetic field position and alignment within the magnet, the magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01Magnet installation tolerances, the magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.02.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.02.01The magnet shall be designed to operate reliably given the cumulative radiation dose it will experience over the lifetime of the EIC of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.5 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 9.4 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.6 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 11.0 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
- 6.04.03.01.01The magnet shall have a single function.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall require shunt resistors for beam-based alignment, 5A at 5 GeV.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a Quadrupole field.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a normal field rotation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The pole tip radius of the magnet shall be 40 mm.01/27/2025ApprovedFALSE
- 6.04.03.01.01The physical magnet length shall be <0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The effective magnet length shall be 0.8 m.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet integrated grad field G shall be 15.1 T.01/27/2025ApprovedFALSE
- 6.04.03.01.01The harmonic reference radius and current at 18 GeV shall be 25 (mm) and 412 (A).01/27/2025ApprovedFALSE
- 6.04.03.01.01The Field at the reference radius and current at 18 GeV shall be 18.9 (T/m).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet bore field shall require the following multipole content:01/27/2025ApprovedFALSE
- 6.04.03.01.01b2 = 10000 , a2 = N/A01/27/2025ApprovedFALSE
- 6.04.03.01.01b3 = HV +/- 2.2 , a3 = +/- 201/27/2025ApprovedFALSE
- 6.04.03.01.01b4 = HV +/- 2.4 , a4 = +/- 0.701/27/2025ApprovedFALSE
- 6.04.03.01.01b5 = HV +/- 1.0 , a5 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b6 = HV +/- 1.0 , a6 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b7 = HV +/- 1.0 , a7 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b8 = HV +/- 1.0 , a8 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b9 = HV +/- 1.0 , a9 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b10 = HV +/- 1.0 , a10 = +/- 0.501/27/2025ApprovedFALSE
- 6.04.03.01.01b11 = HV +/- 1.0 , a11 = +/- 0.401/27/2025ApprovedFALSE
- 6.04.03.01.01b12 = HV +/- 1.0 , a12 = +/- 0.301/27/2025ApprovedFALSE
- 6.04.03.01.01b13 = HV +/- 1.0 , a13 = +/- 0.201/27/2025ApprovedFALSE
- 6.04.03.01.01b14 = HV +/- 1.0 , a14 = +/- 0.1501/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to specifically constrain the external fringe field to 10 Gauss at 40mm from the RCS beamline.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnetic field, center and alignment, within the magnet must be known to within a translational value of +/-50 (um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet install center and install alignment must be within a translational value of +/-150(um) and a rotational alignment value of +/-0.5(mrad).01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet design and verification process shall ensure the final magnet will meet the reliability needs of the EIC over it planned operational life of >20 Years.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall have a splitable pole to facilitate the vacuum beam pipe installation.01/27/2025ApprovedFALSE
- 6.04.03.01.01The magnet shall be designed to operate reliably given the cumulative radiation dose of TBD Rads it will experience over the lifetime of the EIC of >20 Years.01/27/2025In ProcessFALSE
ESR-VAC : ESR Vacuum
- 6.04.05The stray field From any Vacuum equipment on the ESR beam shall have no adverse effect on the beam.01/27/2025ApprovedFALSE
- 6.04.05The magnetic permeability for vacuum equipment shall be approved by beam physics.01/27/2025ApprovedFALSE
- 6.04.05All hardware in vacuum shall be UHV compatible.01/27/2025ApprovedFALSE
- ESR-VAC EXTERNALSRequirements who's parents are in other sub-systems.
- 6.04.05The vacuum chamber shall provide sufficient horizontal and vertical aperture to accommodate; a +/-15 sigma beam, where the vertical RMS beam size is based on the emittance of a fully coupled beam, plus an additional 10 mm horizontally and 5 mm vertically margin to account for expected orbit errors.01/27/2025In ProcessFALSE
- 6.04.05The typical (standard) vacuum chamber aperture shall be 80 x 36 mm.01/27/2025ApprovedFALSE
- 6.04.05Special aperture requirements and/or aperture file shall be provided by or approved by physics.01/27/2025ApprovedFALSE
- 6.04.05The dynamic pressure around the ring shall be consistent with a beam gas lifetime of >10[hrs] with the design currents after an integrated beam current of 1000 [A.h].01/27/2025In ProcessFALSE
- 6.04.05There shall be no upper pressure limit as long as the average pressure is maintained.01/27/2025ApprovedFALSE
- 6.04.05The average vacuum level in the ESR Arc sections after conditioning (for 1000Ahrs) shall be <5x10-9 Torr.01/27/2025ApprovedFALSE
- 6.04.05On 15 m on each side (or one vacuum sector) of the SRF cavities shall be processed to class ISO 5.01/27/2025ApprovedFALSE
- 6.04.05There shall be no pressure bumps in the ring exceeding (TBD)[Torr]01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber and all its components shall be designed to withstand a total synchrotron radiation load of 10 MW, considering the uneven linear load particularly related to the super-bends.01/27/2025In ProcessFALSE
- 6.04.06.04The ESR shall have two synchrotron light monitors (SLM) and one X-ray pin hole monitor01/27/2025In ProcessFALSE
- 6.04.06.04The longitudinal bunch profile monitor shall have a turn-by-turn capability based on a single bunch in the fully filled bunch train.01/27/2025In ProcessFALSE
- 6.04.06.04TThe SLM systems shall measure the crabbing angle, longitudinal bunch parameters, H & V beam size and global coupling.02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure a crabbing angle of 12.5 mrad with accuracy of 10 %01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Longitudinal bunch parameters with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the H & V beam size with accuracy of H=?? V=?? units02/13/2025In ProcessFALSE
- 6.04.06.04The SLM shall be able to measure the Global coupling with accuracy of TBD -01/27/2025In ProcessFALSE
- 6.04.06.04One SLM port shall be located downstream of a dipole in an appropriate location in the ESR, exact location not critical.01/27/2025In ProcessFALSE
- 6.04.06.04The second SLM port shall be located in a complimentary location in the lattice to ensure all the necessary SLM measurements can be made. TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be good quality, having a surface finish better than 1/10 Lambda01/27/2025In ProcessFALSE
- 6.04.06.04The SLM light extraction port mirrors shall be water cooled to avoid image distortion.01/27/2025In ProcessFALSE
- 6.04.06.04There shall be an enclosed SL transport from the light extraction port to the SLM optical lab rooms. Length to be determined by the distance to optical lab room, should be minimized to reduce vibration problems.01/27/2025In ProcessFALSE
- 6.04.06.04The locations of the SLM optical lab rooms shall be TBD -01/27/2025In ProcessFALSE
- 6.04.06.04The double-slit interferometer method shall be used to measure transverse beam size01/27/2025In ProcessFALSE
- 6.04.06.04The standard transverse resolution of an SLM using visible light shall be ~60 um01/27/2025In ProcessFALSE
- 6.04.06.04The resolution using the double-slit method shall equal to 10 um01/27/2025In ProcessFALSE
- 6.04.06.04A streak camera shall be used to measure the bunch longitudinal profiles01/27/2025In ProcessFALSE
- 6.04.06.04A position sensitive photo-diode will provide photon beam centroid information which shall supplement the orbit stability measurements by the BPMs01/27/2025In ProcessFALSE
- 6.04.06.04A GigE CCD/CMOS camera, externally triggerable with exposure times ranging from 10 nsec to 5 sec, shall be used to image the visible radiation01/27/2025In ProcessFALSE
- 6.04.06.04A commercially available gated camera with gate width of <2 nsec (compared to a minimum bunch spacing of 10 nsec) shall be used to detect injection oscillations and for beam studies.01/27/2025In ProcessFALSE
- 6.04.06.04The location of the X-ray pinhole monitoring system shall be TBD01/27/2025In ProcessFALSE
- 6.04.06.04The target resolution of the X-ray pin hole monitoring system shall be ~ 5 um (or as best that can be achieved with the machine parameters and commercial equipment) 5 um01/27/2025In ProcessFALSE
- 6.04.06.04The X-ray pin hole monitor shall provide independent measurement of the energy spread and horizontal/vertical emittance. H=V=15.4 nm02/13/2025In ProcessFALSE
- 6.04.06.04The X-ray pinhole photon beamline shall be equipped with gated cameras that will be employed to provide high resolution turn-by-turn profile measurements01/27/2025In ProcessFALSE
- 6.04.06.04Several different size pinholes sizes shall be incorporated to allow easy alignment and measurements at different beam currents and energies.01/27/2025In ProcessFALSE
- 6.04.06.04A pinhole assembly including tungsten slits shall provide sufficient resolution to precisely measure the beam size01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04.05The vacuum chamber material shall be chosen such that the SR power can be intercepted by the arc chambers and in addition good radiation shielding will be provided.01/27/2025In ProcessFALSE
- 6.04.05The vacuum chamber shall be able to absorb synchrotron radiation and carry away 10 MW of power.01/27/2025ApprovedFALSE
- 6.04.05The impedance of the entire ESR vacuum system, including the interaction regions in IR 6 and IR 8, shall allow for the bunch intensities, beam currents, and bunch numbers contained in [5.9].01/27/2025In ProcessFALSE
- 6.04.05The vacuum system global impedance shall be less than the impedance budget as provided by accelerator physics.01/27/2025ApprovedFALSE
- 6.04.05The maximum beam excursion orbit shall be TBD01/27/2025In ProcessFALSE